Focus® DAILIES® and Focus® DAILIES® Toric (neifilcon A) ONE-DAY CONTACT LENSES are indicated for daily wear for the optical correction of refractive ametropia (myopia, hyperopia and astigmatism) in not-aphakic persons with non-diseased eyes. Focus® DAILIES® Progressives (nelfilcon A) One-Day soft contact lenses are indicated for daily wear for the optical correction of presbyopia in not aphakic persons with non-diseased eyes who require a reading addition of +3.00 diopters (D) or less and who may have 2.00 diopters (D) or less of astigmatism that does not interfere with visual acuity. The lenses are to be prescribed for single-use daily disposable wear. DAILIES lenses are not intended to be cleaned or disinfected and should be discarded after a single use.
Device Story
Soft contact lenses (nelfilcon A) for daily disposable wear; spherical, toric (double thin zone), and progressive aspheric simultaneous vision designs. Lens material: 69% water, 31% nelfilcon A polymer. Supplied sterile in blister packs with phosphate-acetate buffered saline. Prescribed by eye care professionals for patients with refractive errors or presbyopia. Lenses provide optical correction; discarded after single use; no cleaning or disinfection required. Benefits include convenience of daily replacement and visual correction for distance, intermediate, and near ranges.
Clinical Evidence
One-month clinical study comparing subject lenses to concurrent predicate control. Evaluated performance in vision, ocular health, comfort, and fit. Results demonstrated similar overall performance to the predicate.
Technological Characteristics
Material: 69% water, 31% nelfilcon A (polyvinyl alcohol partially acetalized with N-formylmethyl acrylamide). FDA Group 2 (nonionic). Oxygen permeability (Dk): 26 x 10^-11. Refractive index: 1.38. Manufacturing: Full Mold Cast Lightstream Technology. Sterilization: Steam/autoclave. Packaging: Blister pack with phosphate-acetate buffered saline containing up to 0.02% Poloxamer 108.
Indications for Use
Indicated for daily wear optical correction of refractive ametropia (myopia, hyperopia, astigmatism) or presbyopia in non-aphakic persons with non-diseased eyes. Presbyopic patients require reading addition ≤ +3.00D and astigmatism ≤ 2.00D. Single-use daily disposable wear.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
Predicate Devices
Focus® DAILIES® (nelfilcon A) One-Day Contact Lens (K943487)
Focus® DAILIES® Progressives (nelfilcon A) One-Day Visitint lenses (K003826)
Submission Summary (Full Text)
{0}------------------------------------------------
DEC 1 5 2003
| CIBA Vision® Corporation | | | | Page 1 of 4 |
|------------------------------------------------------------|--|--|--|-------------|
| 11460 Johns Creek Parkway | | | | |
| Duluth, GA USA 30097 | | | | |
| Focus® DAILIES®, Focus® DAILIES® Toric and Focus® DAILIES® | | | | |
| Progressives | | | | |
| (nelfilcon A) ONE-DAY Soft Contact Lenses | | | | |
| 510(k) Summary of Safety and Effectiveness | | | | |
510(k) Summary , K
O33701
| <b>1. Submitter Information:</b> | | |
|----------------------------------|-------------------------------------------------------------------------------------|---------|
| Company: | CIBA Vision Corporation | |
| | 11460 Johns Creek Parkway | |
| | Duluth, Georgia USA 30097 | |
| Contact Person: | Alicia M. Plesnarski, RAC | |
| | Director, Global Regulatory Affairs | |
| Telephone: | 678-415-3924 | |
| FAX: | 678-415-3454 | |
| Date Prepared: | 24 November 2003 | |
| <b>2. Device Name:</b> | | |
| Common Name: | Soft Contact Lens | |
| Trade/Proprietary Name: | Focus® DAILIES®, Focus® DAILIES® Toric and | |
| | Focus® DAILIES® Progressives | |
| | (nelfilcon A) ONE-DAY CONTACT LENS | |
| Classification Name: | Daily Wear | |
| | Soft (hydrophilic) Contact Lens | |
| Device Classification: | Class II [21 CFR 886.5925 (b) (1)] | |
| <b>3. Predicate Device(s):</b> | | |
| | Lens Material: CIBA Vision's Focus® DAILIES® (nelfilcon A) One-Day Contact Lens | |
| | Clear lenses (spherical & toric): | K943487 |
| | VISITINT® lenses: | K984273 |
| | Manufacturing Change-Surfactant Additive: | K010636 |
| | Multifocal Design: CIBA Vision's Focus® DAILIES® Progressives (nelfilcon A) One-Day | |
| | Visitint lenses: | K003826 |
### 4. Description of Device:
The Focus® DAILIES®, Focus® DAILIES® Toric and Focus® DAILIES® Progressives (nelfilcon A) ONE-DAY CONTACT LENS are daily wear soft contact lenses intended for single use daily disposable wear. The Dailies lens is a spherical soft contact, Dailies toric lenses have a double thin zone design, and the Dailies Progressives iens is a progressive aspheric simultaneous vision soft contact lens. A constant near power profile is incorporated into each Progressive lens across the full range of distance powers. The near and intermediate powers are concentrated primarily in the central portion of the optical zone while the surrounding portion is weighted toward distance. The continuous changes in power across the surface of the lens allow patients
{1}------------------------------------------------
| CIBA Vision.<br>A Novartis Company | CIBA Vision® Corporation<br>11460 Johns Creek Parkway<br>Duluth, GA USA 30097 | Page 2 of 4 |
|------------------------------------|-------------------------------------------------------------------------------|-------------|
|------------------------------------|-------------------------------------------------------------------------------|-------------|
Focus® DAILIES®, Focus® DAILIES® Toric and Focus® DAILIES®
Progressives
(nelfilcon A) ONE-DAY Soft Contact Lenses
510(k) Summary of Safety and Effectiveness
requiring a reading addition of up to +3.00 diopters to see clearly at far, intermediate and near distances.
The lens material is 69% water and 31% nelfilcon A polymer (polyviny) alcohol partially acetalized with N-formylmethyl acrylamide). The lenses are clear or tinted from edge to edge for visibility purposes with the color additive copper phthalocyanine (CuP).
Lenses are supplied sterile in foil sealed blister packs containing isotonic phosphateacetate buffered saline solution. The package storage saline may contain up to 0.02% Poloxamer 108.
The physical properties of the lens are:
| Refractive Index: | 1.38 (hydrated) |
|------------------------------------------------------|----------------------------------------------------|
| Center Thickness: | 0.09 to0.17 mm<br>(0.10 at -3.00D; 0.15 at +3.00D) |
| Light Transmittance: | 96% (approx) |
| Oxygen Permeability (Dk):<br>[35° C, Fatt corrected] | 26 x 10-11 (cm²/sec) (ml O2/ml x mm Hg) |
| Water Content: | 69% by weight in normal saline |
#### 5. Indications for Use:
Focus® DAILIES® and Focus® DAILIES® Toric (nelfilcon A) One-Day soft contact lenses are indicated for daily wear for the optical correction of refractive ametropia (myopia, hyperopia and astigmatism) in not aphakic persons with non-diseased eyes.
Focus® DAILIES® Progressives (nelfilcon A) One-Day soft contact lenses are indicated for daily wear for the optical correction of presbyopia in not aphakic persons with non-diseased eyes who require a reading addition of +3.00 diopters (D) or less and who may have 2.00 diopters (D) or less of astigmatism that does not interfere with visual acuity.
The lenses are to be prescribed for single use daily disposable wear. DAILIES® lenses are not intended to be cleaned or disinfected and should be discarded after a single use.
{2}------------------------------------------------
Image /page/2/Picture/0 description: This image shows a product information label for CIBA Vision Corporation, located at 11460 Johns Creek Parkway in Duluth, GA 30097. The label indicates that it is page 3 of 4. The product is Focus DAILIES, Focus DAILIES Toric, and Focus DAILIES Progressives, which are (nelfilcon A) ONE-DAY Soft Contact Lenses, and the label is a 510(k) Summary of Safety and Effectiveness.
# 6. Description of Safety and Substantial Equivalence
## 6.1 Comparison to Predicate Device (s):
- Lens Material [Predicate Lens current Focus DAILIES (nelfilcon A)]: . Lens material, chemical composition, formulation (except for addition of the manufacturing additive), manufacturing process, packaging and the sterilization method and cycle remain unchanged from the descriptions previously provided in motion and order remarket Notifications 510(k) K963487, K992446, K003826, K010636.
- Lens Design: . No change to established spherical, toric or multi-focal lens designs.
| | Comparison to CIBA Vision's Predicate Device | |
|-------------------------------------------|-----------------------------------------------------------------------------------|---------------------------------------------------------------------|
| Table 1: | Predicate Device<br>Focus DAILIES (nelfilcon A) made without additive nelfilcon A | Focus DAILIES (nelfilcon A) made with additive nelfilcon A |
| Lens Material: | | |
| Material<br>Classification: | FDA Group 2<br>(> 50% H2O, nonionic polymer) | FDA Group 2<br>(> 50% H2O, nonionic polymer) |
| Water Content: | 69.4% | 68.4% |
| Light Transmittance<br>(clear lenses): | 100% | 100% |
| Oxygen Permeability<br>(Dk, Coulometric): | ~ 25 barrer | 25.86 barrer |
| Power Range: | +20.00 to -20.00D | +20.00 to -20.00D |
| Visibility Tint: | With or without Copper Phthalocyanine | With or without Copper<br>Phthalocyanine |
| Manufacturing<br>Method: | Full Mold Cast<br>Lightstream Technology | Full Mold Cast<br>Lightstream Technology |
| Lens Design: | Spherical and/or Multi-focal | Spherical and/or Multi-focal |
| Sterilization: | Steam sterilization, Validated<br>autoclave | Steam sterilization, Validated<br>autoclave |
| Packaging: | Blister Pack | Blister Pack |
| Package Storage<br>saline solution | Phosphate-acetate buffered saline with<br>up to 0.02% Poloxamer 108 | Phosphate-acetate buffered saline<br>with up to 0.02% Poloxamer 108 |
{3}------------------------------------------------
Image /page/3/Picture/0 description: This image shows a document from CIBA Vision Corporation located in Duluth, GA. The document is labeled as "Page 4 of 4". The document discusses Focus DAILIES Toric and Focus DAILIES Progressives (nelfilcon A) ONE-DAY Soft Contact Lenses and is a 510(k) Summary of Safety and Effectiveness.
### 6.2 Non-clinical Testing
Results from a series of physical/chemical tests confirm that DAILIES lenses made r with or without the manufacturing additive were equivalent and within established specifications for the lenses. Successful results from in-vivo and in-vitro toxicology tests confirm the lenses remain non-toxic and biocompatible with the ocular environment.
### 6.3 Clinical Testing:
A one-month clinical study demonstrated similar overall performance to the concurrent predicate control in the clinically relevant areas of vision, health, and comfort and fit when worn for daily wear.
#### 7. Substantial Equivalence
DAILIES lenses made with or without the manufacturing additive are equivalent and within established specifications for the lens. The lenses maintain clinical performance expectations, established physical/chemical characteristics, and are stable and biocompatible with the ocular environment.
Any differences which may exist between lenses made with or without the additive do not adverselv affect the established performance characteristics and safety and effectiveness profile of the device.
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract image of an eagle.
DEC 1 5 2003
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Alicia M. Plesnarski Director, Global Regulatory Affairs CIBA Vision Corporation 11460 Johns Creek Parkway Duluth, Georgia 30097-1556
Re: K033701
Trade Name: Focus® DAILIES®, Focus® DAILIES® Toric and Focus® DAILIES® Progressives (nelfilcon A) One-Day Contact Lenses for Daily Wear Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (hydrophilic) Contact Lens Regulatory Class: Class II Product Code: LPL; MVN Dated: November 24, 2003 Received: November 28, 2003
Dear Ms. Plesnarski:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
Page 2 - Alicia M. Plesnarski
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4613. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
A. Palgi Rosenthal
A. Ralph Rosenthal, M.D Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Part III. Indications for Use Statement v01, 24-Nov-03, Page 1 of 1 Premarket Notification 510(k): (nelfilcon A) Soft Contact Lens
#### INDICATIONS FOR USE STATEMENT PART III.
#### 510(k) Number:
Device Name(s):
Focus® DAILIES®, Focus® Toric, Focus® DAILIES® Progressives (nelfilcon A) One-Day Contact Lens
#### Indications For Use:
Focus® DAILIES® and Focus® DAILIES® Toric (neifilcon A) ONE-DAY CONTACT LENSES are indicated for daily wear for the optical correction of refractive ametropia (myopia, hyperopia and astigmatism) in not-aphakic persons with non-diseased eyes.
Focus® DAILIES® Progressives (nelfilcon A) One-Day soft contact lenses are indicated for daily wear for the optical correction of presbyopia in not aphakic persons with non-diseased eyes who require a reading addition of +3.00 diopters (D) or less and who may have 2.00 diopters (D) or less of astigmatism that does not interfere with visual acuity.
The lenses are to be prescribed for single-use daily disposable wear. DAILIES lenses are not intended to be cleaned or disinfected and should be discarded after a single use.
| Prescription Use | |
|-----------------------------|--|
| (Part 21 CFR 801 Subpart D) | |
AND/OR
Over-The-Counter Use ((21 CFR 807 Subpart C)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Ophthalmic Ear,
**Nose and Throat Devices**
510(k) Number K 033701
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.