K033696 · Union Medico Aps · KZH · Feb 16, 2005 · General Hospital
Device Facts
Record ID
K033696
Device Name
PERSONAL INJECTOR
Applicant
Union Medico Aps
Product Code
KZH · General Hospital
Decision Date
Feb 16, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6920
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Device is intended to allow patients or carers to self-administer inject able FDA approved drugs or biologics with a safe, simple and easy injection system. Personal Injector is intended to be used in any setting including the home. Personal Injector is intended for use by patients to self-administer or by care-givers for injection of FDA approved drugs or biologics. Personal Injector is intended to introduce the needle below the skin surface. Personal Injector is designed to accommodate two different and replaceable syringeholders. Personal Injector is intended for the use with FDA approved fixed or non-fixed needle syringes supplied with specific injection reqimens. Personal Injector is intended to be used in any setting including the home.
Device Story
Personal Injector is a spring-loaded mechanical syringe needle introducer; designed for home or clinical use by patients or caregivers. Unlike traditional auto-injectors that perform needle insertion and drug delivery in one motion, this device uses a spring-loaded mechanism solely to insert the needle to a pre-determined depth (subcutaneous or intramuscular). After insertion, the user manually controls the plunger to inject the drug, allowing for visual inspection of the site and air bubbles. Accommodates various FDA-approved syringes with fixed or non-fixed needles via replaceable syringe-holders. Benefits include reduced injection pain, bruising, and swelling; provides users with greater control over injection speed and comfort.
Clinical Evidence
A focus group study was conducted in 2001 at Rigshospitalet, Denmark, involving patients using the injector for multiple sclerosis treatment. Results were reported as positive, with high satisfaction from both patients and healthcare professionals.
Technological Characteristics
Mechanical spring-loaded syringe needle introducer. Features replaceable syringe-holders to accommodate various FDA-approved fixed or non-fixed needle syringes. Designed for manual drug delivery following automated needle insertion. No electronic components, software, or energy source other than mechanical spring tension.
Indications for Use
Indicated for patients or caregivers to self-administer FDA-approved injectable drugs or biologics. Suitable for subcutaneous or intramuscular needle introduction. Applicable for various treatment regimens including multiple sclerosis, cancer, growth hormone therapy, erectile dysfunction, and medications like heparin, adrenaline, insulin, and apomorphine.
Regulatory Classification
Identification
A syringe needle introducer is a device that uses a spring-loaded mechanism to drive a hypodermic needle into a patient to a predetermined depth below the skin surface.
{0}------------------------------------------------
### FEB 1 6 2005
## 510(K) SUMMARY
: 上一篇:
510(k) Number: KO33696 . Product:
| Submitter's Name and Address: | Union Medico<br>Emdrupvej 22<br>2100 Copenhagen<br>Denmark | |
|-------------------------------|----------------------------------------------------------------------------|--|
| Contact Persons: | Michael Perthu<br>Jesper Hulbaek<br>Tel: +45 70266010<br>Fax: +45 70266011 | |
| Injector: | Personal Injector™ | |
| Class: | Class II | |
| Classification Name: | Syringe Needle Introducers | |
| Classification Panel: | General Hospital, Panel 80<br>Regulation Number 21 C.F.R § 880.6920 | |
| Predicate Device: | Autoject 2 (K993385)<br>Autoject 2 (K013362) | |
| Manufacturing Facilities: | GPV Teknik A/S<br>Smedeland 22, Albertslund<br>2600 Glostrup<br>Denmark | |
{1}------------------------------------------------
# 15433696 P.Z. 22
#### Device Description
Personal Injector is designed for the user - by a user, and here for a lot of emphasis has been put on functionality and injection comfort, appearance, surface, materials, weight, balance, and sound. Personal Injector is an injection system allowing the user to have maximum control and comfort while injecting.
Personal Injector differs from commonly known Auto-injector. Until today, marketed Autoinjectors are only for subcutaneous injections, and introduce both the needle and the drug in one motion. The injections are associated with discomfort, such as pain, blue marks, drug accumulation and swelling at the injection site. Personal Injector has a spring-loaded mechanism that only serves to introduce the needle to a pre-determined depth, either subcutaneous or intramuscular. Hereafter the patient can check for air bobbles and vein penetration, and perform a safe traditional manual injection of the drug, by slowly and steady pushing the plunger down. Personal Injector minimises pain and support the patient perform the manual drug introduction, avoiding almost all discomfort.
Personal Injector two unique Syringe-Holders is designed to accommodate FDA approved syringes with fixed and non-fixed needles. Personal Injector is suitable for many injection regimes, including multiple sclerosis patients treatment with interferon beta, cancer treatment, children receiving growth hormone treatment, men injecting local vasodilators to treat erectile dysfunction, and injections of heparin, adrenaline, insulin, apomorphine, and many others medications.
#### Intended Use
The Device is intended to allow patients or carers to self-administer inject able FDA approved drugs or biologics with a safe, simple and easy injection system. Personal Injector is intended to be used in any setting including the home.
#### Operational
Personal Injector is designed to accommodate two different and replaceable syringe-holders. The injector is here for designed to use multiple FDA approved fixed or non-fixed needle syringes. Additionally, Personal Injector is intended for Prescription Use and Over-The-Counter Use.
#### Focus Group Study
Personal Injector has been tested at a focus group study in 2001 at Rigshospitalet, the largest MS hospital in Denmark, amongst patients using the Injector. The results were outstanding, and Personal Injector has received wide praise from Patients and Healthcare Professionals alike.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a stylized symbol that resembles three overlapping lines or stripes, with a wavy line at the bottom. The seal is black and white.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 1 6 2005
Mr. Michael Perthu Managing Director, CEO Union Medico APS Emdrupvej 22 2100 Copenhagen DENMARK
Re: K033696
Trade/Device Name: Personal Injector Regulation Number: 21 CFR 880.6920 Regulation Name: Syringe Needle Introducer Regulatory Class: II Product Code: KZH Dated: January 11, 2005 Received: January 13, 2005
Dear Mr. Perthu:
We have reviewed your Section 510(k) premarket notification of intent to market the device reformed above and have determined the device is substantially cquivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Lederal Register.
{3}------------------------------------------------
Page 2 - Mr. Perthu
Please be advised that FDA's issuance of a substantial equivalence determination does not mcan that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Sujata Medicine Omd.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
### Indications for Use
510(k) Number (if known): K033696
Device Name:
Personal Injector
Indications For Use:
Personal Injector is intended for use by patients to self-administer or by care-givers for injection of FDA approved drugs or biologics. Personal Injector is intended to introduce the needle below the skin surface.
Personal Injector is designed to accommodate two different and replaceable syringeholders. Personal Injector is intended for the use with FDA approved fixed or non-fixed needle syringes supplied with specific injection reqimens.
Personal Injector is intended to be used in any setting including the home.
| Prescription Use | ✓ |
|------------------|---|
| | |
AND/OR
Over-The-Counter Use
(Part 21 CFR 801 Subpart D)
(21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Chris Vance
Cirison (Sign-Off) Inision of Anesthesiology, General Hospital, Infoction Control, Dental Devices
510(K) Number: K933696
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.