SET GRI-FILL 2.0-1 WAY; 2.0-2 WAYS; LUER CONNECTION TUBE
K033682 · Laboratorios Grifols, S.A. · LHI · Dec 17, 2003 · General Hospital
Device Facts
Record ID
K033682
Device Name
SET GRI-FILL 2.0-1 WAY; 2.0-2 WAYS; LUER CONNECTION TUBE
Applicant
Laboratorios Grifols, S.A.
Product Code
LHI · General Hospital
Decision Date
Dec 17, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
SETS GRI-FILL 2.0 are disposable components of the GRI-FILL 2.0 pharmacy compounding system used to provide a fluid pathway through which 1 or more source solutions are delivered into a single final solution. The device is NOT intended to be connected directly to the patient.
Device Story
Disposable fluid transfer set for pharmacy compounding; used with GRI-FILL 2.0 system. Components include syringe, distributor, and tubing; available in 1 or 2-way configurations with optional Luer adapter. Device provides fluid pathway from source solutions to final container (Gri-bag, Gri-flex, syringe, or pump). Operated by pharmacy personnel in hospital settings. Not for direct patient connection. Closed system design prevents contact between fluids and reusable compounding device parts. Benefits include accurate, automated delivery of multi-ingredient solutions; reduces manual handling risks.
Clinical Evidence
Bench testing only. No clinical data. Functional laboratory testing verified accurate delivery of source solutions under normal and stress conditions, and confirmed fluid/air leakage integrity.
Technological Characteristics
Disposable fluid transfer set; materials include PVC with DEHP plasticizer. Components: syringe, distributor, tubing, Luer female-female adapter. Closed system architecture. Sterilization: Ethylene Oxide. Connectivity: mechanical interface with GRI-FILL 2.0 compounding system.
Indications for Use
Indicated for use in hospital pharmacies as an ancillary device with the GRI-FILL 2.0 compounding system to provide a fluid pathway for delivering one or more source solutions into a single Gri-bag, Gri-flex container, standard syringe, or pump. Contraindicated for use with lipids. Restricted to sale by or on order of a physician.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
Predicate Devices
Exacta-Mix 2400 Compounding System Administration Set (BAXA)
Automix 3+3 Compounder Transfer Set (BAXTER HEALTHCARE)
Submission Summary (Full Text)
{0}------------------------------------------------
K033682
# DEC 1 7 2003
Attachment la Page 2 of 6 Revised: 7/1/96
# 12 2017 10:30 10 10 10 10 1 REELL 2.0: 510(K)
#### DATE OF SUBMISSION:
SUBMITTER NAME: SUBMITTER ADDRESS: 2003-07-30
Laboratorios Grifols, S.A. C/ Can Guasch, 2 08150 PARETS DEL VALLES BARCELONA SPAIN
TELEPHONE: FAX: e-mail:
sebastian.gascon@grifols.com Sebastian Gascón
+ 34 93 571 01 00
+ 34 93 573 09 12
Technical Director
SETS GRI-FILL 2.0
I.V. FLUID TRANSFER SETS
CONTACT:
DEVICE TRADE NAME: COMMON NAME: CLASSIFICATION NAME:
PREDICATE DEVICE:
DEVICE DESCRIPTION:
SETS GRI-FILL 2.0 are fluid transfer sets for use with the GRI-FILL 2.0 pharmacy compounding device in order tc compound or mix different multi-ingredient solutions and to channel them into a final suitable container. The set is a disposable component of the compounding device made up of a syringe, a distributor and tubing to channel the fluid. Sets are available for 1 or 2 source solutions. Also a Luer female - female adapter is available as an accessory to the 1 or 2 way transfer sets. The model Luer Connection Tube can be joined to the 1 way or 2 ways sets as prolongation for connection to source solution containers with luer terminals.
I.V. FLUID TRANSFER SETS (21 CFR 880.5440)
Exacta-Mix 2400 Compounding System Administration Set (BAXA) Automix 3+3 Compounder Transfer Set (BAXTER HEALTHCARE)
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows a document with the title "TECHNICAL EVALUATION DOCUMENTATION" and the logo "GRIFOLS" on the left side. On the right side, it says "Document: TED-SETS GRI-FILL 2.0-01". The bottom of the document says "SECTION 1 - SETS GRI-FILL 2.0: 510(k) SUMMARY".
#### INTENDED USE:
SETS GRI-FILL 2.0 are disposable components of the GRI-FILL 2.0 pharmacy compounding system used to provide a fluid pathway through which 1 or more source solutions are delivered into a single final solution. The device is NOT intended to be connected directly to the patient.
## SUMMARY OF COMPARISON WITH PREDICATE DEVICE:
In the establishment of substantial equivalence, SETS GRI-FILL 2.0 are compared with other transfer sets used in pharmacy compounding.
The following table summarizes the similanties of the principal technological characteristics and features of both predicate and new devices.
| # | Characteristic /<br>Feature | SETS GRI-FILL | PREDICATES | |
|----|----------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | | BAXA<br>Exacta-Mix Set | BAXTER<br>Automix 3+3<br>(set presented<br>together with<br>compounder) |
| 1. | Intended use / Claims | SETS GR-FILL 2.0 are<br>disposable components of<br>the GRI-FILL 2.0 pharmacy<br>compounding system used<br>to provice a fluid pathway<br>through which 1 or more<br>source solutions are<br>delivered into a single final<br>solution. The device is NOT<br>Intended to be connected<br>directly to the patient | Exacta-Mix 2400<br>Compounding System<br>Administration set is a<br>disposable component of a<br>compounding device used in<br>the pharmacy to compound<br>multiple source ingrediens<br>into one final solution. This<br>device is not intended for<br>direct patient hook-up. | Positive displacement<br>fluid compounding<br>system employing the<br>3+3 compounder,<br>TRANSFER SETS,<br>empty Viafex or<br>Travamulsion<br>containers and multitask<br>computer software to<br>provide compounding of<br>a wide var ety of fuids |
| 2. | Technological features:<br>- Sterilization<br>- Direct patient<br>hook-up<br>- Source<br>solutions | Ethylene Oxide<br>NO<br>2 | Ethylene Oxide<br>NO<br>Multiple | Radiation<br>NO<br>6 |
| 3. | Main Transfer<br>Materials | Set PVC with DEHP<br>plasticizer | PVC | PVC with DEHP<br>plasticizer |
| 4. | Physical, Mechanical<br>and Biological<br>Specifications | Sterile / Non pyrogenic | Sterile / Non pyrogenic | Sterile / Non<br>pyrogenic |
| 5. | Closed system (fluid not<br>in contact with any<br>resusable part of the<br>compounding device). | YES | YES | YES |
{2}------------------------------------------------
Image /page/2/Figure/1 description: The image shows a document with the title "TECHNICAL EVALUATION DOCUMENTATION" and the logo "GRIFOLS" on the left side. On the right side, it says "# Document: TED-SETS GRI-FILL 2.0-01". The bottom of the image says "SECTION 1 - SETS GRI-FILL 2.0: 510(k) SUMMARY".
From the above table, it can be established that the new device and the predicate devices are very similar
### SUMMARY DISCUSSION OF NON-CLINICAL DATA:
All materials used in the construction of SETS GRI-FILL 2.0 have been subject to chemical and biological testing in accordance with the applicable requirements taking account of its intended use as a parenteral drug solution transfer set.
Functional laboratory testing performed in foreseeable operating conditions showed correct operation of the device as per its intended use, specifically including:
- accurate delivery of specified source solutions under normal conditions and stress conditions
- -fluid / air leakage checking
#### CONCLUSIONS:
We believe the intended use, the indications for use, the functionality and the operation of both SETS GRI-FILL 2.0 and the predicate devices for fluid transfer in pharmacy compounding are essentially the same. Hence, substantial equivalence of SETS GRI-FILL 2.0 with the legally marketed devices may be established.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of a human figure in profile, composed of three curved lines that suggest a head, body, and legs. The figure is positioned to the right of a circular border containing the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in capital letters. The text is arranged around the circumference of the circle, with the department's name at the top and "USA" at the bottom.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 17 2003
Laboratories Grifols, S.A. Ms. Susan Gill Responsible Third Party Official Underwriters Laboratories, Incorporated 12 Laboratory Drive Research Triangle Park, North Carolina 27709-3995
Re: K033682
Trade/Device Name: Set GRI-FILL 2.0-1 way 2.0-2 ways . Luer Connection Tube Regulation Number: 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: LHI, NEP Dated: December 11, 2003 Received: December 12, 2003
Dear Ms. Gill:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 -Ms. Gill
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-14618. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Cloris
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
GRIFOLS
TECHNICAL EVALUATION
DOCUMENTATION
# Document:
TED-SETS GRI-FILL 2.0-09
SECTION 09 – SETS GRI-FILL 2.0: INDICATIONS FOR USE
STATEMENT
# PREMARKET NOTIFICATION INDICATIONS FOR USE STATEMENT (as required by ODE for all 510(k) received after Jan. 1, 1996)
510(k) Number:
Device Name:
SETS GRI-FILL 2.0
Indications for Use:
All models of SETS GRI-FILL 2.0 fluid transfer sets are ancillary devices used in conjunction with the GRI-FILL 2.0 pharmacy compounding device in hospites in hospites for a macy to provide a fluid pathway through which one or more source solutions are delivered into a single Gri-bag or Gri-flex container or into a standard syringe or pump. The device should not be used with lipids.
It is restricted to sale by or on order of a physician.
(Do not write below this line. Continue on another page in needed)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Patrizia Cucorete
(Division Sign-Off) Division of Anesthesiology, General nospilal Infection Control, Dental Devices
510(k) Number: N033682
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter Use
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.