MOBILE EXTERNAL COUNTER PULSATION SYSTEM ANGIONEW-V
K033657 · Living Data Technologies Corporation · DRN · Dec 19, 2003 · Cardiovascular
Device Facts
Record ID
K033657
Device Name
MOBILE EXTERNAL COUNTER PULSATION SYSTEM ANGIONEW-V
Applicant
Living Data Technologies Corporation
Product Code
DRN · Cardiovascular
Decision Date
Dec 19, 2003
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.5225
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Mobile External Counter Pulsation System AngioNew-V is a non-invasive external counter pulsating device intended for use in the treatment of patients with congestive heart failure, stable or unstable angina pectoris, actual myocardial infraction or cardiogenic shock.
Device Story
Mobile noninvasive external counter pulsation system; treats cardiac conditions via sequential lower extremity compression. System comprises central control unit and three pairs of inflatable cuffs (calves, thighs, buttocks). Input: ECG and pulse wave signals (via finger plethysmography or temporal artery sensor). Operation: Cuffs inflate sequentially during diastole to increase coronary perfusion/collateral circulation; deflate rapidly before systole to reduce cardiac workload (systolic unloading). Features: Automated inflation timing adjustment based on heart rate; integrated cuff connection sensors; finger oximetry for blood oxygen monitoring; built-in printer for real-time ECG/pulse wave recording. Used in clinical settings to enhance coronary blood flow and reduce heart workload.
Clinical Evidence
Bench testing only. Testing included safety and EMC compliance (IEC 60601-1, EN 60601-1, IEC 60601-1-2, EN 55011, EN 61000-3-2, EN 61000-3-3) and performance verification of R-wave detection, heart rate calculation, and inflation/deflation timing control.
Technological Characteristics
Mobile external counter pulsation system. Components: Central control unit, three pairs of inflatable cuffs. Sensors: Finger plethysmography, finger oximetry, optional temporal artery sensor. Connectivity: Built-in printer for real-time data recording. Power: AC drive with compressor motor control loop. Safety: Integrated cuff connection sequence sensors. Software: Heart-rate dependent inflation timing adjustment.
Indications for Use
Indicated for patients suffering from congestive heart failure, stable or unstable angina pectoris, acute myocardial infarction, and cardiogenic shock.
Regulatory Classification
Identification
An external counter-pulsating device is a noninvasive, prescription device used to assist the heart by applying positive or negative pressure to one or more of the body's limbs in synchrony with the heart cycle.
Special Controls
*Classification.* (1) Class II (special controls) when the device is intended for the treatment of chronic stable angina that is refractory to optimal anti-anginal medical therapy and without options for revascularization. The special controls for this device are:(i) Nonclinical performance evaluation of the device must demonstrate a reasonable assurance of safety and effectiveness for applied pressure, synchronization of therapy with the appropriate phase of the cardiac cycle, and functionality of alarms during a device malfunction or an abnormal patient condition;
(ii) Reliabilities of the mechanical and electrical systems must be established through bench testing under simulated use conditions and matched by appropriate maintenance schedules;
(iii) Software design and verification and validation must be appropriately documented;
(iv) The skin-contacting components of the device must be demonstrated to be biocompatible;
(v) Appropriate analysis and testing must be conducted to verify electrical safety and electromagnetic compatibility of the device; and
(vi) Labeling must include a detailed summary of the device-related and procedure-related complications pertinent to use of the device.
(2) Class III (premarket approval) for the following intended uses: Unstable angina pectoris; acute myocardial infarction; cardiogenic shock; congestive heart failure; postoperative treatment of patients who have undergone coronary artery bypass surgery; peripheral arterial disease associated with ischemic ulcers rest pain or claudication, threatened gangrene, insufficient blood supply at an amputation site, persisting ischemia after embolectomy or bypass surgery, and/or pre- and post-arterial reconstruction to improve runoff; diabetes complicated by peripheral arterial disease or other conditions possibly related to arterial insufficiency including nocturnal leg cramps and/or necrobiosis diabeticorum; venous diseases, including prophylaxis of deep vein thrombophlebitis, edema (e.g., chronic lymphedema) and/or induration (e.g., stasis dermatitis) associated with chronic venous stasis, venous stasis ulcers, and/or thrombophlebitis; athletic injuries, including Charley horses, pulled muscles and/or edematous muscles; necrotizing cellulitis.
(c)
*Date premarket approval application (PMA) or notice of completion of product development protocol (PDP) is required.* A PMA or notice of completion of a PDP is required to be filed with FDA on or before March 31, 2014, for any external counter-pulsating device, with an intended use described in paragraph (b)(2) of this section, that was in commercial distribution before May 28, 1976, or that has, on or before March 31, 2014, been found to be substantially equivalent to any external counter-pulsating device, with an intended use described in paragraph (b)(2) of this section, that was in commercial distribution before May 28, 1976. Any other external counter-pulsating device with an intended use described in paragraph (b)(2) of this section shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.
Predicate Devices
The Mobile External Counter Pulsation System AngioNew-IV (K023701)
Submission Summary (Full Text)
{0}------------------------------------------------
### Living Data Technology Corp.
## Special 510(k) for AngioNew-V
## EXHIBIT 1
## 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 870.92.
The assigned 510(k) number is: _ K033657
#### Date Summary Prepared 1.
November 19, 2003
### 2. Submitter's Identification
Living Data Technology Corporation 140 53rd Street Brooklyn, NY 11232
Contact: Dr. Jun Ma (718) 492-7400
### Name of the Device 3.
The Mobile External Counter Pulsation System AngioNew-V
### Predicate Device Information 4.
K023701. The Mobile External Counter Pulsation System AngioNew-IV Living Data Technology Corporation, Brooklyn, NY
### ട. Device Description
The Mobile External Counter Pulsation System AngioNew-V is a mobile noninvasive external counter pulsation device for the treatment of patients suffering from congestive heart failure, stable or unstable angina pectoris, acute myocardial infarction and cardiogenic shock.
The Living Data Technology AngioNew-V system consists of a central control unit and three pairs of inflatable cuffs that are wrapped respectively around a patient's calves, thighs and buttocks. External pressure is applied via the cuffs to the lower extremities of the patient in synchronization with the heartbeat. When the heart is in the relaxed state during diastolic period, pressure is applied sequentially from the distal to proximal extremities to force blood back to the heart, increase coronary perfusion pressure (diastolic augmentation) and, at the same time, increases coronary blood flow and enhances the development of
{1}------------------------------------------------
coronary collateral circulation. Just before the heart starts ejecting blood during systolic period, air is quickly withdrawn from all cuffs simultaneously to remove all the external applied pressure, leaving behind empty vasculature in the lower extremities to receive the output of the heart, thereby reducing systolic pressure (systolic unloading) and the workload of the heart.
### Indications for Use 6.
The Mobile External Counter Pulsation System AngioNew-V is a mobile noninvasive external counter pulsation device for the treatment of patients suffering from congestive heart failure, stable or unstable angina pectoris, acute myocardial infarction and cardiogenic shock.
### Comparison to Predicate Device 7.
The modified device AngioNew-V is essentially same in specifications as well as in technological and functional characteristics as the current device AngioNew-IV (cleared under K023701), except for the following technological upgrading:
- A control loop is added to the AC drive regulate the compressor motor for . stabilizing the air reservoir pressure.
- A finger pulse plethysmography sensor is employed and the original . temporal artery pressure sensor is optional now.
- The central control unit CPU is upgraded. The display is also upgraded . from 9" CRT to 8.4" LCD.
- The quick connectors between cuffs and valves are improved with . integrated senor to identify cuff connection sequence so that the machine adjusts inflation sequence accordingly, therefore adding an additional safety feature to the device.
- A new inflation timing adjustment method for better counter pulsation . effect is used in AngioNew-V. Using a predetermined, heart rate dependent inflation timing as the initial value, the software updates the inflation timing to maximize counterpulsation effect.
- A finger oximetry sensor is added to provide measurement and display of . blood oxygen saturation level.
- A built-in printer is added to the device to provide real time recording of . ECG and pulse wave, as well as other information such as heart rate, D/S ratios, etc., during ECP treatment.
### 8. Discussion of Non-clinical Tests Performed for Determination of Substantial Equivalence
In addition to safety tests in accordance with IEC 60601-1 and EN 60601-1; EMC tests in accordance with IEC 60601-1-2, EN 55011, EN 61000-3-2, and EN 61000-3-3; other testing was also conducted, such as:
{2}------------------------------------------------
### Living Data Technology Corp.
- R-wave Detection and Heart Rate Calculation; .
- Inflation/deflation Timing Control and Counterpulsation Effect; .
etc.
It is concluded that AngioNew-V meets all safety and EMC requirement, as well as all design specifications. Therefore it is substantially equivalent to AngioNew-IV in safety and effect.
#### ਹੈ. Conclusions
The subject device, AngioNew-V, has the same intended use and similar characteristics as the predicate device. Moreover, bench testing contained in our submission and non-clinical testing supplied demonstrate that there are no differences in their technological characteristics, thereby not raising any new Thus, the AngioNew-V device is questions of safety or effectiveness. substantially equivalent to the predicate device.
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle-like design with three curved lines emanating from its head, representing the department's mission to promote health and well-being. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the emblem.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 1 9 2003
Living Data Technology Corporation c/o Ms. Susan D. Goldstein-Falk MDI Consultants, Inc. 55 Northern Boulevard, Suite 200 Great Neck, NY 11021
Re: K033657
Mobile External Counter Pulsation System AngioNew-V Regulation Number: 21 CFR 870.5225 Regulation Name: External Counter-Pulsating Device Regulatory Class: Class III (three) Product Code: DRN Dated: November 19, 2003 Received: November 21, 2003
Dear Ms. Goldstein-Falk:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Ms. Susan D. Goldstein-Falk
Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that FDA's issuance of a substanter equires with other requirements of the Act
that FDA has made a determination that your device adder Edgeral Pourmust that FDA has made a determilation administered by other Federal agencies. You must
or any Federal statutes and regulations administered by other Federation and listing or any Federal statutes and regulations and including, but not limited to: registration and listing (21)
comply with all the Act's requirements, including, and circus provine comply with all the Act s requirements, menanting practice requirements as set
CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as settroni CFR Part 807); labeling (21 CFK Patl on (21 CFR Patt 820); and if applicable, the electronic
forth in the quality systems (QS) regulation (21 CFR Path 21 CFR 1000-1050 forth in the quality systems (QS) regulation (Sections 531-542 of the Act); 21 CFF 1000-1050.
product radiation control provisions (Sections 531-542 of the Action Section S product radiation control provisions (Scetions 35 - 3 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 - 2 -This letter will allow you to begin marketing your active as walence of your device to a legally
premarket notification. The FDA finding of substantial equivalence of the , premarket notification. The FDA inding of subsantial equivalence of 1.
marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please
rs of the results of the most (2011/64), also are the regulation entitled If you desire specific advice for your ac tree of 666. Also, please note the regulation entitled, and the may obtain contact the Office of Compliance at (301) 37 - 10 - (21CFR Patt 807.97). You may obtain "Misbranding by reference to prematics houltiedable in the Act from the Division of Small other general information on your responsion.com and with toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address.
Sincerely yours,
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Special 510(k) for AngioNew-V
## EXHIBIT B
of Page 1
K033657 510(k) Number (if known): __
Device Name: The Mobile External Counter Pulsation System AngioNew-V
Indications for Use:
The Mobile External Counter Pulsation System AngioNew-V is a non-invasive external counter pulsating device intended for use in the treatment of patients with congestive heart failure, stable or unstable angina pectoris, actual myocardial infraction or cardiogenic shock.
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
O. M. 22
(Division Sign-Off)
Division of Cardiovascular Devices
510/k) Numbe
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.