OLYMPUS VIDEO URETEROSCOPE, NTSC

K033651 · Olympus Winter & Ibe GmbH · FGB · Jan 30, 2004 · Gastroenterology, Urology

Device Facts

Record IDK033651
Device NameOLYMPUS VIDEO URETEROSCOPE, NTSC
ApplicantOlympus Winter & Ibe GmbH
Product CodeFGB · Gastroenterology, Urology
Decision DateJan 30, 2004
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesTherapeutic

Indications for Use

This instrument has been designed to be used with an Olympus video system, light source, display monitor, endo-therapy accessories, and other ancillary equipment for endoscopic diagnosis and treatment within the urethra, bladder, and ureter.

Device Story

The Olympus Video Ureteroscope, NTSC, is a flexible endoscope used for visualization and therapeutic intervention in the urinary tract. It functions as an optical system that captures images of the urethra, bladder, and ureter. The device represents a modification of the predicate A2560 ureteroscope, specifically replacing the traditional image guide with a Charge-Coupled Device (CCD) sensor to improve image resolution. The device is used in clinical settings by physicians to perform endoscopic procedures. The physician views the output on an external display monitor connected to an Olympus video system. This visualization allows for real-time diagnosis and guidance during endo-therapy procedures, potentially benefiting patients by enabling minimally invasive access to the upper urinary tract.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Flexible endoscope featuring a CCD sensor for image acquisition. Designed for use with Olympus video systems, light sources, and display monitors. The device is a modification of a fiber-optic based predicate, transitioning to digital imaging via CCD. Sterilization and material specifications are not detailed in the provided text.

Indications for Use

Indicated for endoscopic diagnosis and treatment within the urethra, bladder, and ureter. Intended for use by clinicians in a professional healthcare setting.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K033651 # SMDA 5120(k) SUMMERY # OLYMPUS VIDEO URETEROSCOPE, NTSC ### Submitter's Name, Address and Phone and Fax Numbers A. | Name & Address of manufacturer: | Olympus Winter & Ibe, GmbH | |---------------------------------|----------------------------| | | Kuehnstr. 61 | | | Hamburg 22045, Germany | | | Phone: +49 40 669 66-0 | | | Fax: +49 40 668 15 91 | | Registration Number: | 8010313 | #### Name of Contact Person B. | Address of initial importer and contact | Olympus America 2 Corporate Center Drive Melville, N.Y. 11747 | |-----------------------------------------|---------------------------------------------------------------| | Registration Number: | 2429304 | | Contact: | Tina Steffanie-Oak, Associate R.A. Manager | | Phone: | 631-844-5477 | | Fax: | 631-844-5554 | ## Device name, Common Name, Classification Name and Predicate Devices C. | Trade NAME: | OLYMPUS VIDEO URETEROSCOPE, NTSC | |-----------------------------|----------------------------------------------------------------------| | Common Name: | Ureteroscope and Accessories | | Classification:<br>Class II | 21 CFR 876.1500 Endoscope and accessories. -<br>FDA Product Code FGB | | Predicate Device: | A2560 510(k)# K951855 | #### Description of the Device D. The subject device is used for endoscopic diagnosis and treatment within the urethra, I ho storeet de roo is as optical system is modified from image guide to CCD and the resolution is improved. #### Intended Use of the Device E. This instrument has been designed to be used with an Olympus video system, light I mis instrament has ovequipment, display monitor, endo-therapy accessories, and other ancillary equipment for endoscopic diagnosis and treatment within the urethra, bladder, and ureter. {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its body and wings. The eagle is facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JAN 3 0 2004 Olympus Winter & Ibe, GmbH c/o Ms. Tina Steffanie-Oak Associate Manager, Regulatory Affairs Olympus America, Inc. Two Corporate Center Drive MELVILLE NY 11747 Re: K033651 Trade/Device Name: OLYMPUS VIDEO URETEROSCOPE, NTSC (S-1546/2) Modification to A2560 Ureteroscope (K951855) Regulation Number: 21 CFR §876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: 78 FGB Dated: January 21, 2004 Received: January 22, 2004 Dear Ms. Steffanie-Oak: We have reviewed your Scction 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for association to May 28, 1976, the enactment date of the Medical Device Amendments, or to conimered process that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). and Cosmotion Fee (110) that the device, subject to the general controls provisions of the Act. The r ou may, aterest, mains of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean I that FDA has made a determination that your device complies with other requirements of the Act that I Dr Has Intatutes and regulations administered by other Federal agencies. You must of uny I outhates the Act s requirements, including, but not limited to: registration and listing comply with an the rice broquiant Part 801); good manufacturing practice requirements as set (21 OF It in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050. {2}------------------------------------------------ This letter will allow you to begin marketing your device as described in your Section 510(k) This letter will allow you to ocglif hill kemine of substantial equivalence of your device to a legally premarket notification. The FDA finding of substantial equivaleries a premarket notification. The FDA inding of substantial or your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice for your device on our line of good number at the top of the letter: | 8xx. 1xxx | (301) 594-4591 | |----------------------------------|----------------| | 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 | | 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 | | 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 | | Other | (301) 594-4692 | Additionally, for questions on the promotion and advertising of your device, please contact the Additionally, for questions on the pronocol an and the regulation entitled, "Misbranding Office of Compliance at (301) 594-4639. Area note the regulation Other segment Office of Compliance at (501) 594-4057. This, production of the general (1) 10 may obtain. Other general by reference to premarket nollication (21GI IT at out of Jaired from the Division of Strail information on your responsional and Consumer Assistance at its toll-free number (800) 638-2041 or Manufacturers, International and Consumer Assistance in the del (4cmarkers) Manufacturers, Thiernational and Collsumior Passworldrh/dsma/dsmamain.html. (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmamain.html. Sincerely yours, Nancy C. Brogdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure Page 2 {3}------------------------------------------------ K033651 | 510(k) Number (if known): | Not assigned yet K033651 | |---------------------------|--------------------------| |---------------------------|--------------------------| Olympus video ureteroscope, NTSC Device Name:_ > Indications for Use: This instrument has been designed to be used with an Olympus Indications for Ose. - Fins meatation equipment, display monitor, endo-therapy video system, ingli source, accumsion for endoscopic diagnosis and treatment within the urethra, bladder, and ureter. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) David A. Symm (Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...