K033650 · Siemens Medical Solutions USA, Inc. · OBJ · Feb 24, 2004 · Cardiovascular
Device Facts
Record ID
K033650
Device Name
ACUNAV 90/10 DIAGNOSTIC ULTRASOUND CATHETER
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
OBJ · Cardiovascular
Decision Date
Feb 24, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1200
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The AcuNav™ Diagnostic Ultrasound Catheter is intended for intra-cardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart.
Device Story
AcuNav 90/10 is a single-use, disposable ultrasonic phased-array imaging transducer integrated into a 10 Fr (3.33 mm) diameter, 90 cm long catheter. Distal tip features four-way deflection (160° range) in two orthogonal planes for precise positioning. Device connects to compatible ultrasound systems (Aspen, Cypress, Sequoia) via a reusable system cable. Operates by emitting ultrasound waves and receiving echoes to generate real-time images of cardiac/vascular anatomy and physiology, blood flow, and device placement. Used by clinicians in clinical settings (e.g., cath lab) to guide procedures and assess anatomy. Output displayed on ultrasound system monitor; assists in clinical decision-making by providing high-resolution visualization of intracardiac structures and hemodynamic data.
Clinical Evidence
Bench testing only. Device safety and performance verified through compliance with electrical safety (UL 2601-1, IEC 601-1-1/2) and biocompatibility (ISO 10993) standards. No clinical trial data presented.
Technological Characteristics
Phased-array ultrasonic transducer; 10 Fr (3.33 mm) diameter; 90 cm insertable length; 4-way deflection mechanism; reusable system cable interface. Complies with UL 2601-1, CSA C22.2 No. 601-1, CISPR 11, IEC 601-1-1/2, and ISO 10993 biocompatibility standards.
Indications for Use
Indicated for intra-cardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology, and visualization of other devices in the heart. Applicable to adult and pediatric populations requiring intracardiac or intravascular ultrasound imaging.
Regulatory Classification
Identification
An intravascular diagnostic catheter is a device used to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels. Included in this generic device are right-heart catheters, left-heart catheters, and angiographic catheters, among others.
{0}------------------------------------------------
# 510(k) SUMMARY
FEB 2-4 2004 .
# AcuNav 90/10 Diagnostic Ultrasound Catheter
This summary of safety and effectiveness is provided as part of this Fremarket Notification in compliance with the Safe Medical Device Act of 1990 revisions to 21 CFR. Part 807.92. Content and Format of a 510(k) Summary.
#### 1. Submitted by:
Siemens Medical Solutions USA, Inc., Ultrasound Division 1230 Shorebird Way Mountain View, CA 94039
### Contact Person:
Iskra Mraković Manager, Regulatory Affairs
Phone: (650) 694-5004 Fax: (650) 961-6168
#### 2. Proprietary Name:
AcuNav 90/10 Diagnostic Ultrasound Catheter
### Common/Usual Name:
Intracardiac/Intravascular Ultrasound Catheter.
### Classification Name:
Regulatory Class: II Review Category: Tier II
| | FR Number | Product Code |
|-----------------------------------|-----------|--------------|
| Diagnostic Intravascular Catheter | 870.1200 | 74-DQO |
| Diagnostic Ultrasound Transducer | 892.1570 | 90-ITX |
#### 3. Predicate Device:
ACUSON AcuNav® 90/10 Diagnostic Ultrasound Catheter
#### 4. Device Description:
As described in the previously cleared 510(k) submissions, (K992631 on December 15, 1999, and K010950 on June 27, 2001), the AcuNav catheter is comprised of a single-use, disposable ultrasonic phased-array imaging transducer and a catheter which is 10 Fr (3.33 mm) in diameter and 90 cm in insertable length. The device is capable of imaging at multiple frequencies and obtaining blood flow data in multiple ultrasound modes. The distal portion of the catheter can be deflected in four directions in two orthogonal planes: leftright (in a plane perpendicular to the image plane) and anterior-posterior (in a
{1}------------------------------------------------
plane coincident with the image plane). The range of deflection is 160° in each direction. The AcuNav catheter is comprised of three major components: (1) the catheter itself; (2) the steering mechanism; and (3) the reusable system cable.
The AcuNav 90/10 Diagnostic Ultrasound Catheter has been designed to meet the following product safety standards [as required by 21 CFR § 807.87(i)].
- UL 2601-1, Safety Requirements for Medical Equipment .
- . CSA C22.2 No. 601-1, Safety Requirements for Medical Equipment
- . CISPR 11, Class A
- IEC 601-1-1 ●
- IEC 601-1-2 .
- . ISO 10993 Biocompatibility
#### Intended Use: 5.
The AcuNay catheter is intended for use in the visualization of vascular anatomy and physiology, cardiac and great vessel anatomy and physiology, visualization of other devices in the heart or vasculature, as well as measurements of blood flow, cardiac and vascular anatomic dimensions.
#### Technological Comparison to Predicate Device: 6.
The AcuNav 90/10 Diagnostic Ultrasound Catheter is substantially equivalent to the ACUSON AcuNav® 90/10 Diagnostic Ultrasound Catheter (K992631,
11/8/2001).
Both catheters are ultrasound-tipped catheter devices used directly within the vasculature and/or the heart for intravascular or intracardiac ultrasound imaging. Devices are specifically indicated for use in visualization of vascular anatomy, cardiac and great vessel anatomy and physiology, or other devices in the heart, as well as measurement of blood flow.
### End of 510(k) Summary.
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" around the top half of the circle. Inside the circle is an abstract symbol that resembles an eagle or bird-like figure, composed of three human profiles facing to the right.
MAY 2 4 2007
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Siemens Medical Solutions USA, Inc. c/o Dr. Iskra Mrakovic Regulatory Affairs Manager 1230 Shorebird Way P.O. Box 7393 Mountain View, CA 94043
Re: K033650
Trade/Device Name: Acunav 90/10 Diagnostic Ultrasound Catheter Regulation Number: 21 CFR 870.1200 and 892.1570 Regulation Name: Diagnostic Intravascular Catheter and Diagnostic Ultrasound Transducer
Regulatory Class: II Product Code: OBJ and ITX Dated: November 19, 2003 Received: November 20, 2003
Dear Dr. Mrakovic:
This letter corrects our substantially equivalent letter of February 24, 2004.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Dr. Mrakovic
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to continue marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-4008. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
fa
compe
Bram D. Zuckerman, M.D.
Director
Division of Cardiovascular Devices
Office of Device Evaluation
Center for Devices and
Radiological Health
Enclosure
{4}------------------------------------------------
### Diagnostic Ultrasound Indications for Use Form Aspen Diagnostic Ultrasound System
| 510(k) Number: | K033650 |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | AcuNav 90/10 (IC10V5) Diagnostic Ultrasound Catheter |
| Indications for Use: | The AcuNav™ Diagnostic Ultrasound Catheter is intended for intra-<br>cardiac and intra-luminal visualization of cardiac and great vessel<br>anatomy and physiology as well as visualization of other devices in the<br>heart. |
| Ultrasound System: | Aspen |
| Transducer: | AcuNav |
| | Mode of Operation | | | | | | | | | |
|----------------------------------|-------------------|---|---|-----|-----|------------------|---------------------------------|------------------------------|-----------------------|-------------------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other:<br>Harmonic<br>Imaging |
| Ophtalmir | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative<br>(Vascular) | | | | | | | | | | |
| Intraoperative<br>(Neurological) | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ<br>(Specify)** | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | P | P | P | P | P | P | | P* | P |
| Trans-esophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intra-Luminal | | P | P | P | P | P | P | | P* | P |
| Peripheral Vessel | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | |
| Other (Intra-Cardiac) | | P | P | P | P | P | P | | P+ | P |
P=Previously cleared by the FDA under premarket notification K992631.
#### Additional Comments:
*Combinations include: B+M, B+PWD, B+CWD, B+Color Doppler, B+M+ Color Doppler, B+PWD+Color Doppler, B+CWD+Color Doppler, B+Rower Doppler, B+M+Power Doppler B+PWD+Power Doppler, B+CWD+Power Doppler
# (PLESE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
Donna R. Lochner
(Division Sign-Off) Division of Cardiovascular Devices
510(k) number k033650
{5}------------------------------------------------
## Diagnostic Ultrasound Indications for Use Form Cypress Diagnostic Ultrasound System
| 510(k) Number: | k033650 |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | AcuNav 90/10 (IC10V5) Diagnostic Ultrasound Catheter |
| Indications for Use: | The AcuNav™ Diagnostic Ultrasound Catheter is intended for intra-<br>cardiac and intra-luminal visualization of cardiac and great vessel<br>anatomy and physiology as well as visualization of other devices in the<br>heart. |
| Ultrasound System: | Cypress |
| Transducer: | AcuNav |
| | Mode of Operation | | | | | | | | | |
|----------------------------------|-------------------|---|---|-----|-----|------------------|---------------------------------|------------------------------|-----------------------|-------------------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other:<br>Harmonic<br>Imaging |
| Ophtalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative<br>(Vascular) | | | | | | | | | | |
| Intraoperative<br>(Neurological) | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ<br>(Specify) ** | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | P | P | P | P | P | | | P* | P |
| Trans-esophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intra-Luminal | | P | P | P | P | P | | | P* | P |
| Peripheral Vessel | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | |
| Other (Intra-Cardiac) | | P | P | P | P | P | | | P* | P |
P=Previously cleared by the FDA under premarket notification K992631.
#### Additional Comments:
:
| *The Cypress Ultrasound System does not provide combined modes where more than one scanning mode |
|--------------------------------------------------------------------------------------------------|
| is active simultaneously. |
# (PLESE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
DUMKA R. Vochuel
(Division Dir. - OM
(Division Sign-Off) Division of Čardiovascular Devices
510(k) Number k033650
{6}------------------------------------------------
### Diagnostic Ultrasound Indications for Use Form Sequoia Diagnostic Ultrasound System
| 510(k) Number: | K033650 | | | | | | |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|
| Device Name: | AcuNav 90/10 (IC10V5) Diagnostic Ultrasound Catheter | | | | | | |
| Indications for Use: | The AcuNav™ Diagnostic Ultrasound Catheter is intended for intra-<br>cardiac and intra-luminal visualization of cardiac and great vessel<br>anatomy and physiology as well as visualization of other devices in the<br>heart. | | | | | | |
| Ultrasound System: | Sequoia | | | | | | |
| Transducer: | AcuNav | | | | | | |
| Clinical Application | Mode of Operation | | | | | | | | | | |
|----------------------------------|-------------------|---|---|-----|-----|------------------|---------------------------------|------------------------------|-----------------------|-------------------------------|--|
| | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other:<br>Harmonic<br>Imaging | |
| Ophtalmic | | | | | | | | | | | |
| Fetal | | | | | | | | | | | |
| Abdominal | | | | | | | | | | | |
| Intraoperative<br>(Vascular) | | | | | | | | | | | |
| Intraoperative<br>(Neurological) | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | |
| Small Organ<br>(Specify)** | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac | | P | P | P | P | P | P | | P* | P | |
| Trans-esophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | |
| Intra-Luminal | | P | P | P | P | P | P | | P* | P | |
| Peripheral Vessel | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | | |
| Other (Intra-Cardiac) | | P | P | P | P | P | P | | P* | P | |
P=Previously cleared by the FDA under premarket notification K992631.
Additional Comments:
*Combinations include: B+M, B+PWD, B+Color Doppler, B+M+ Color Doppler, B+PWD+Color Doppler, B+CWD+Color Doppler, B+Power Doppler, B+Power Doppler, B+M+Power Doppler, B+PWD+Power Doppler, B+CWD+Power Doppler
# (PLESE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
#### Prescription Use (Per 21 CFR 801.109)
Donna R. Holmes
(Division Sign-Off) Division of Cardiovascular Devices
510(k) Number_Ko 33650
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.