K033474 · Zoll Medical Corp · MKJ · May 21, 2004 · Cardiovascular
Device Facts
Record ID
K033474
Device Name
ZOLL AED PLUS
Applicant
Zoll Medical Corp
Product Code
MKJ · Cardiovascular
Decision Date
May 21, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5310
Device Class
Class 3
Attributes
Therapeutic, Pediatric
Indications for Use
The ZOLL AED Plus external defibrillator is intended to be used by personnel who are qualified by training in basic life support, advanced life support, or other physician-authorized emergency medical response to defibrillate victims of cardiac arrest.
Device Story
ZOLL AED Plus is a portable, battery-powered semi-automatic external defibrillator; provides voice prompts and visual icons to guide rescuers through cardiac arrest treatment. Device inputs: patient ECG rhythm via electrodes. Operation: device analyzes ECG to detect shockable rhythms; if detected, charges capacitor and prompts user to press treatment button for defibrillation using ZOLL Rectilinear Bi-Phasic waveform. Used by trained emergency responders in various settings. Device performs automatic periodic self-tests to ensure readiness. Pediatric electrodes allow use on children <8 years. Output affects clinical decision-making by providing automated rhythm analysis and therapy delivery, potentially improving survival outcomes in cardiac arrest.
Clinical Evidence
No clinical trials in humans were required. Evidence includes bench testing for functional requirements and safety standards (EN60601-1, EN60601-1-2, IEC 601-2-4, AAMI DF39), electrode performance testing, animal studies for defibrillation energy dosing, and validation of the rhythm classification algorithm using a pediatric database.
Technological Characteristics
Lightweight, portable, battery-powered AED. Utilizes ZOLL Rectilinear Bi-Phasic defibrillation waveform. Includes pediatric electrode system for patients <8 years/55 lbs. Complies with EN60601-1, EN60601-1-2, IEC 601-2-4, and AAMI DF39 standards. Features automated periodic self-testing and voice/visual guidance.
Indications for Use
Indicated for victims of cardiac arrest exhibiting unconsciousness, absence of normal breathing, and absence of pulse/signs of circulation. Pediatric use (age <8 years or weight <55 lbs/25kg) requires ZOLL AED Plus Pediatric Electrodes. Contraindicated for patients who are conscious, breathing, or have a detectable pulse/signs of circulation.
Regulatory Classification
Identification
An automated external defibrillator (AED) system consists of an AED and those accessories necessary for the AED to detect and interpret an electrocardiogram and deliver an electrical shock (e.g., battery, pad electrode, adapter, and hardware key for pediatric use). An AED system analyzes the patient's electrocardiogram, interprets the cardiac rhythm, and automatically delivers an electrical shock (fully automated AED), or advises the user to deliver the shock (semi-automated or shock advisory AED) to treat ventricular fibrillation or pulseless ventricular tachycardia.
Predicate Devices
Medtronic Physio-Control Infant / Child Reduced Energy Electrodes used with the LIFEPAK CR Plus or biphasic LIFEPAK 500 AEDs (K022732)
Submission Summary (Full Text)
{0}------------------------------------------------
K033474
# 510(K) SUMMARY
### Applicant's Name and Address
ZOLL Medical Corporation 269 Mill Road Chelmsford, MA 01824
### Contact Person
Paul Dias (978) 421-9413
### Date Summary Prepared
October 17, 2003
#### Device
ZOLL AED Plus
### Device Classification
Automatic External Defibrillator 21 CFR 870.1025 Device Class: 111
Multifunction Electrocardiograph Electrode 21 CFR 870.2360 MLN Device Class: 111
### Device Description
The ZOLL AED Plus is a lightweight, portable, battery powered semi-automatic external defibrillator that uses voice prompts and visual icons to quide a user through a cardiac The device utilizes the ZOLL Rectilinear Bi-Phasic defibrillation arrest rescue. waveform. The device is designed to be used by trained emergency responders for the treatment of cardiac arrest.
When connected with ZOLL AED Plus defibrillation electrodes to a patient, the device will analyze the electrocardiographic (ECG) rhythm of the patient and detect whether the rhythm is shockable or not shockable. If the device detects a shockable rhythm, the device charges the capacitor, enables the treatment button and issues the audio warning "Don't Touch Patient, Press Treatment Button". The user can then press the treatment button to deliver defibrillation therapy to the patient. If the patient is not responsive to treatment, additional shocks may be advised and administered after automatically repeated analyses of the patient's heart rhythm.
The ZOLL AED Plus is lightweight. It can easily be transported to any rescue site. It is designed to hang on the wall in an area where it is easily accessible and ready to use at a moment's notice. The device automatically performs a periodic self-test and continually indicates its state of readiness to the user.
<sup>·</sup> ZOLL AED Plus Pediatric Electrode System
<sup>· 510(</sup>k) Summary
<sup>·</sup> Page 1 of 3
{1}------------------------------------------------
When the ZOLL AED Plus is used in conjunction with the ZOLL AED Plus Pediatric Electrodes the device can be used on patients less than 8 years of age. Labeling and packaging of the electrodes is designed to promote a clear visual distinction between adult and pediatric electrodes.
### Indications for Use
The ZOLL AED Plus external defibrillator is intended to be used by personnel who are qualified by training in basic life support, or advanced life support, or other physicianauthorized emergency medical response to defibrillate victims of cardiac arrest.
Use the AED when a suspected cardiac arrest victim has an apparent LACK OF CIRCULATION as indicated by:
- Unconsciousness and .
- Absence of normal breathing and .
- Absence of a pulse or signs of circulation. .
When a victim is a child or infant less than 8 years of age, or less than 55 lbs (25kg), the ZOLL AED Plus should be used with the ZOLL AED Plus Pediatric Electrodes. Therapy should not be delayed to determine the patient's exact age or weight.
## Substantial Equivalence
ZOLL AED Plus and ZOLL AED Plus Pediatric Electrodes are substantially equivalent to other legally marketed AED and pediatric defibrillation electrode combinations. Specifically, the ZOLL AED Plus with the ZOLL AED Plus Pediatric Electrodes is substantially equivalent to the Medtronic Physio-Control Infant / Child Reduced Energy Electrodes used with the LIFEPAK CR Plus or biphasic LIFEPAK 500 AEDs cleared under premarket 510(k) notification k022732.
### Performance Data
The ZOLL AED Plus external defibrillator was subject to extensive safety and performance testing prior to release. Final testing for the system included various performance tests designed to ensure that the device meets all of its functional requirements and performance specifications. Safety testing was performed to ensure the device complies with applicable sections of the following industry and safety standards.
- EN60601-1, Medical Electrical Equipment Part 1: General Requirements for . Safety.
- EN60601-1-2, Medical Electrical Equipment Part 1: General Requirements . for Safety; Electromagnetic Compatibility.
- IEC 601-2-4, Medical Electrical Equipment Part 2: Particular Requirements . for the Safety of Cardiac Defibrillators and Cardiac Defibrillator-Monitors.
- . AAAMI DF39, AED and Remote Control Defibrillators.
In addition the premaket 510(k) notification includes:
<sup>·</sup> ZOLL AED Plus Pediatric Electrode System
<sup>· 510(</sup>k) Summary
<sup>·</sup> Page 2 of 3
{2}------------------------------------------------
- Electrode performance testing, .
- . Animal testing of defibrillation energy dosing, and
- Rhythm classification algorithm performance testing on a pediatric database. .
### Conclusion
The information in this premarket notification demonstrates that the ZOLL AED Plus used in conjunction with the ZOLL AEP Plus Pediatric Electrodes is substantially equivalent to the predicate device with respect to safety, effectiveness, and performance.
<sup>·</sup> ZOLL AED Plus Pediatric Electrode System
<sup>· 510(</sup>k) Summary
<sup>·</sup> Page 3 of 3
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three horizontal lines forming its body and wings. The bird is facing to the right. Encircling the bird is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY 2 1 2004
Zoll Medical Corporation c/o Mr. Paul Dias Director, Quality Assurance and Regulatory Affairs World Wide Headquarters 269 Mill Road Chelmsford, MA 01824
Re: K033474
Trade Name: ZOLL AED Plus Regulation Number: 21 CFR 870.5310 Regulation Name: Automated External Defibrillators Regulatory Class: III (three) Product Code: MLN, MKJ Dated: April 27, 2004 Received: April 29, 2004
Dear Mr. Dias:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your becalled the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use surved in the encreases 1976, the enactment date of the Medical Device Amendments, or to conniner of the to 1725 20, 2019 11:12 provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). and Cosmeter For (110) lear the device, subject to the general controls provisions of the Act. The r ou may, mererere, manns of the Act include requirements for annual registration, listing of general controls pro receive, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 – Mr. Paul Dias
Please be advised that FDA's issuance of a substantial equivalence determination does not mean r that FDA has made a determination that your device complies with other requirements of the Act that + + + + + + + = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = or any I cocur statues and equirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set Of R Fart 877) its systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic forth in the qualisms wys would be (Sections 531-542 of the Act); 21 CFR 1000-1050. product radially w you to begin marketing your device as described in your Section 510(k) I mis letter will and in you to cuba finding of substantial equivalence of your device to a legally premaired predicated device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please if you desire spoonly arriver 9 % (301) 594-4648. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general mionmation and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Neil R. Poole
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# SECTION 6 - Indications for Use
510(k) Number (if known): __________________________________________________________________________________________________________________________________________________
ZOLL AED Plus Device Name:
### Indications for Use:
Use the AED when a suspected cardiac arrest victim has an apparent LACK OF CIRCULATION as indicated by:
- Unconsciousness and .
- Absence of normal breathing and .
- Absence of a pulse or signs of circulation. .
When a victim is less than 8 years of age, or weighs less than 55 lbs (25kg), the ZOLL AED Plus should be used with the ZOLL AED Plus Pediatric Electrodes. Therapy should not be delayed to determine the patient's exact age or weight.
### Contraindications:
Do NOT use the AED when patient is:
- Consciousness; or .
- . breathing; or
- has a detectable pulse or other signs of circulation. .
#### Intended Use:
The ZOLL AED Plus external defibrillator is intended to be used by personnel who are qualified by training in basic life support, advanced life support, or other physician-authorized emergency medical response to defibrillate victims of cardiac arrest.
Net RPOgden for BDZ
(Division Sign-Off)
Division of Cardiovascular Devices
233474 510(k) Number
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
or
Over-The-Counter-Use
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.