K033421 · Erbe USA, Inc. · GEI · Oct 14, 2004 · General, Plastic Surgery
Device Facts
Record ID
K033421
Device Name
ERBE BICLAMP OPEN AND LAPAROSCOPIC INSTRUMENTS
Applicant
Erbe USA, Inc.
Product Code
GEI · General, Plastic Surgery
Decision Date
Oct 14, 2004
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ERBE BiClamp™ Open and Laparoscopic Instruments are intended for use in general surgery, laparoscopic, gynecologic, urological, and thoracic procedures where fusion of vessels or tissues is desired. The devices can be used on vessels up to 7 mm and bundles as large as will fit in the jaws of the instrument. A vessel fusion is created by the application of bipolar electrosurgical RF energy (coagulation) to the vessels placed between the jaws of the instrument. The ERBE BiClamp™ Open and Laparoscopic Instruments are designed for use with an ERBE VIO Electrosurgical Generator (ESU) System having an Optional Bipolar Mode/ BiClamp™ upgrade and the multi-function receptacle. Not recommended for use with other manufacturer's generators.
Device Story
Bipolar electrosurgical instruments for vessel/tissue fusion; used with ERBE VIO Electrosurgical Generator. Input: HF electrical energy from generator; Output: thermal coagulation/desiccation of tissue held between instrument jaws. Used in OR by surgeons for general, laparoscopic, gynecologic, urological, and thoracic procedures. Instruments are reusable, non-sterile, stainless steel/plastic construction. Physician manually operates jaws to grasp tissue; generator delivers RF energy to create seal. Physician visually inspects seal integrity post-fusion. Benefits: effective vessel/tissue bundle fusion up to 7mm.
Clinical Evidence
No clinical data provided; substantial equivalence based on design verification and validation by manufacturer.
Technological Characteristics
Reusable bipolar electrosurgical instruments; stainless steel and plastic construction. Energy source: ERBE VIO Electrosurgical Generator (RF energy). Form factor: Open (200-270mm) and Laparoscopic (5mm O.D., 340mm length). Connectivity: proprietary multi-function receptacle for ERBE VIO system. Sterilization: validated cleaning and sterilization processes required.
Indications for Use
Indicated for patients undergoing general, laparoscopic, gynecologic, urological, or thoracic surgery requiring vessel or tissue bundle fusion (up to 7mm). Procedures include bowel resections, hysterectomies, Nissen fundoplication, adhesiolysis, oophorectomy, and laparoscopic cholecystectomies. Contraindicated for tubal sterilization/coagulation.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
Valleylab Inc. LigaSure™ Open and Laparoscopic Instruments (K981916)
Submission Summary (Full Text)
{0}------------------------------------------------
# KO33421
## 510K SUMMARY - Revised August 6, 2004
| Submitted By: | OCT 1 4 2004<br>ERBE USA, Inc.<br>2225 Northwest Parkway<br>Marietta, GA 30067<br>Tel: 770-955-4400 Fax: 770-955-2577 |
|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Julie Stephens, President/Consultant<br>Regulatory Resources Group, Inc. |
| 510(k) Number: | K033421 |
| Date Prepared: | August 6, 2004 |
| Common Name: | Bipolar Electrosurgical Open and Laparoscopic Instruments |
| Trade/Proprietary Name: | ERBE BiClamp™ Open and Laparoscopic Instruments |
| Classification Name: | Electrosurgical cutting and coagulation device and accessories<br>(21CFR878.4400) and Gynecologic electrocautery and<br>accessories (21 CFR Part 884.4120) |
| Product Code: | GEI and HGI |
| Legally Marketed<br>Predicate Device: | Valleylab Inc. LigaSure™ Open and Laparoscopic<br>Instruments 510(k) Number: K981916 |
#### Device Description:
The ERBE BiClamp™ Open and Laparoscopic Instruments are used with an ERBE VIO Electrosurgical Generator (ESU) System having the Optional Bipolar Mode, BiClamp. High Frequency (HF) energy from the ESU is delivered through the jaws of the ERBE BiClamp" Instruments to coagulate/desiccate tissue. The ERBE BiClamp™ Open Instruments are made of stainless steel with plastic insulation except at the jaw surfaces (which isolates the energy to only the jaw surfaces). The ERBE BiClamp'™ Laparoscopic Instruments are made of metals and plastics with the electrical energy isolated to the jaws. They have various jaw types, which are standard in the industry. The ERBE BiClamp™ Open Instruments range in size from 200 mm (7.9 inches) to 270 mm (10.6 inches) in length with bent jaws that have a smooth surface. The ERBE BiClamp™ Laparoscopic Instruments have a 5 mm outside diameter (O.D.) and a 340 mm (13.4 inches) working length. The Instruments are provided non-sterile and are reusable (Note: The cleaning and sterilization processes have been validated and are provided in the Notes on Use to the customer.).
#### Intended Use:
The ERBE BiClamp™ Open and Laparoscopic Instruments are intended for use in general surgery, laparoscopic, gynecologic, urological, and thoracic procedures where fusion of vessels or tissues is desired. The devices can be used on vessels up to 7 mm and bundles as large as will fit in the jaws of the instrument. A vessel fusion is created by the application of bipolar electrosurgical RF energy (coagulation) to the vessels placed between the jaws of the instrument.
The ERBE BiClamp™ Open and Laparoscopic Instruments are designed for use with an ERBE VIO Electrosurgical Generator (ESU) System having an Optional Bipolar Mode/ BiClamp™ upgrade and the multi-function receptacle. Not recommended for use with other manufacturer's generators.
{1}------------------------------------------------
K033421 (pg 2 of 2)
### 510K SUMMARY - Revised August 6, 2004
The indications for use with Open Instruments include: general surgery, gynecologic, urological. and thoracic procedures where fusion of vessels and tissue bundles is performed including such procedures as bowel resections, hysterectomies (both vaginal and abdominal), Nissen fundoplication, adhesiolysis (lysis of adhesions), oophorectomy, etc.
The indications for use with Laparoscopic Instruments include: all laparoscopic procedures (including gynecologic, general, urological, and thoracic surgery) where fusion of vessels or tissue bundles is performed including such procedures as bowel resections, hysterectomies (both vaginal and abdominal), faparoscopic cholecystectomies (gall bladder procedures), Nissen fundoplication, adhesiolysis (lysis of adhesions), oophorectomy, etc.
CAUTION: Vessel fusion can be affected by patient factors such as age, elasticity of vessels, thickness of vessel walls, etc.; therefore, the physician should review each vessel fusion for seal integrity. This device is not effective for use in tubal sterilization/tubal coaqulation for sterilization purposes.
Similarities and Differences of the Proposed Devices to the Predicate Devices Comparison/Substantial Equivalence):
#### Similarities
The ERBE BiClamp™ Open and Laparoscopic Instruments have similar physical and dimensional characteristics as the predicate devices. They have the same basic technological characteristics and the intended use is the same.
#### Differences
The ERBE BiClamp™ Open and Laparoscopic Instruments are different in that all of the product line components or parts are provided non-sterile and reusable: where as. the electrodes of the predicate's Open Instruments and Laparoscopic Instruments are sterile, single use. The ERBE BiClamp Open and Laparoscopic Instruments do not have closure mechanisms to hold the instrument closed (or clamped), where as, the predicate devices have closure mechanisms.
All the instrument designs have been verified or validated in design control by ERBE Elektromedizin GmbH.
#### Conclusion:
The ERBE BiClamp™ Open and Laparoscopic Instruments have the same intended use, principles of operation, and technological characteristics as the predicate devices that were previously cleared for market in a 510(k).
The ERBE BiClamp™ Open and Laparoscopic Instruments differ only in that all the instruments and their parts are reusable and they do not have a closure mechanism to hold the instruments closed.
In conclusion, there are no issues with the ERBE BiClamp™ Open and Laparoscopic Instruments that would raise additional safety or efficacy issues when compared to the predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract symbol that resembles an eagle or a stylized human figure.
Public Health Service
OCT 1 4 2004
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
ERBE USA, Inc. c/o Ms. Julie Stephens Regulatory Resources Group, Inc. 111 Laurel Ridge Drive Alpharetta, Georgia 30004
Re: K033421
Trade/Device Name: ERBE BiClamp™ Open and Laparoscopic Instruments Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: II Product Code: GEI Dated: September 23, 2004 Received: September 24, 2004
Dear Ms. Stephens:
We have reviewed your Section 510(k) premarket notification of intent to market the device w o no ro ro ro row a your and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for associated in the May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). and Cosmetic fore, market the device, subject to the general controls provisions of the Act. The r ou may, are every inions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
## Page 2 - Ms. Julie Stephens
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Miriam C. Provost
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use - REVISED August 6, 2004
## 510(k) Number (if known): K033421
Device Name: ERBE BiClamp™ Open and Laparoscopic Instruments
Indications For Use:
The ERBE BiClamp™ Open and Laparoscopic Instruments are intended for use in general The ERDE Dronalip - Opecologic, urological, and thoracic procedures where fusion of vessels surgery, laparoscopies, griesers, and e used on vessels up to 7 mm and bundles as large as of this us is a donrod. The instrument. A vessel fusion is created by the application of bipolar will in in the juve of argy (coagulation) to the vessels placed between the jaws of the instrument.
The ERBE BiClamp™ Open and Laparoscopic Instruments are designed for use with an ERBE VIO Electrosurgical Generator (ESU) System having an Optional Bipolar Mode/ BiClamp™ upgrade and the multi-function receptacle. Not recommended for use with other manufacturer's generators.
The indications for use with Open Instruments include: general surgery, gynecologic, urological, and thoracic procedures where fusion of vessels and tissue bundles is performed including such procedures as bowel resections, hysterectomies (both vaginal and abdominal), Nissen fundoplication, adhesiolysis (lysis of adhesions), oophorectomy, etc.
The indications for use with Laparoscopic Instruments include: all laparoscopic procedures (including gynecologic, general, urological, and thoracic surgery) where fusion of vessels or (incrading gracess), bootegory book procedures as bowel resections, hysterectomies (both vaginal and abdominal), laparoscopic cholecystectomies (gall bladder procedures), Nissen fundoplication, adhesiolysis (lysis of adhesions), oophorectomy, etc.
CAUTION: Vessel fusion can be affected by patient factors such as age, elasticity of vessels, thickness of vessel walls, etc.; therefore, the physician should review each vessel fusion for seal integrity. This device is not effective for use in tubal sterilization/tubal coagulation for sterilization purposes.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
uriam C. Phiorrst
Page 1 of 1
(Division Sign-Off) Division of General, Restorative. and Neurological Devices
510(k) Number K63342/
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.