SELECT 3D ARTERIAL CANNULA WITH CARMEDA BIOACTIVE SURFACE AND SELECT CAP ARTERIAL CANNULA WITH CARMEDA BIOACTIVE SURFACE

K033416 · Medtronic Perfusion Systems · DWF · Nov 5, 2003 · Cardiovascular

Device Facts

Record IDK033416
Device NameSELECT 3D ARTERIAL CANNULA WITH CARMEDA BIOACTIVE SURFACE AND SELECT CAP ARTERIAL CANNULA WITH CARMEDA BIOACTIVE SURFACE
ApplicantMedtronic Perfusion Systems
Product CodeDWF · Cardiovascular
Decision DateNov 5, 2003
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.4210
Device ClassClass 2
AttributesTherapeutic

Indications for Use

These cannulae are intended for use in perfusion of the ascending aorta during short term (6 hours or less) cardiopulmonary bypass.

Device Story

Select 3D™ and Select CAP™ Arterial Cannulae are wire-wound PVC vascular catheters used for cardiopulmonary bypass. Select 3D™ features three integrated flutes for blood flow diffusion; Select CAP™ includes an integrated pressure monitoring port for real-time central arterial pressure measurement. Both devices feature angled, beveled tips and optional Carmeda® BioActive Surface coating. Used by perfusionists or surgeons in clinical settings to facilitate extracorporeal circulation during cardiac procedures. The device connects to perfusion lines; vented connectors allow air removal prior to connection. The Carmeda® coating is intended to improve hemocompatibility during short-term use.

Clinical Evidence

Bench testing only. In vitro visual and functional testing performed to evaluate performance characteristics of materials post-Carmeda® coating. Additional testing included coverage, bioactivity, and leach testing of the bioactive surface.

Technological Characteristics

Wire-wound PVC body; angled, beveled tips. Select 3D™ has three integrated flutes; Select CAP™ has integrated pressure monitoring port. Available in 18-24 Fr sizes. Optional Carmeda® BioActive Surface coating. Non-powered, mechanical device.

Indications for Use

Indicated for patients requiring perfusion of the ascending aorta during short-term (≤6 hours) cardiopulmonary bypass.

Regulatory Classification

Identification

A cardiopulmonary bypass vascular catheter, cannula, or tubing is a device used in cardiopulmonary surgery to cannulate the vessels, perfuse the coronary arteries, and to interconnect the catheters and cannulas with an oxygenator. The device includes accessory bypass equipment.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ NOV = 5 2003 K033416 f1/2 : # 510(k) Summary | Date Prepared: | October 24, 2003 | |-----------------|------------------------------------------------------------------------------------| | Submitter: | Medtronic Perfusion Systems<br>7611 Northland Boulevard<br>Brooklyn Park, MN 55428 | | Contact Person: | Ronald W. Bennett<br>Principal Regulatory Affairs Specialist | | | Phone: (763)-391-9086<br>Fax: (763) 391-9603 | # Device Name and Classification: | Trade Name: | Select 3D™ Arterial Cannula<br>22, 24 Fr. with Carmeda® BioActive Surface<br>Select CAP™ Arterial Cannula<br>18, 20, 22, 24 Fr. with Carmeda® BioActive Surface | |-------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------| |-------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------| | Common Name: | Cardiopulmonary bypass vascular catheter, cannula or<br>tubing | |-----------------|------------------------------------------------------------------------------| | Classification: | Class II | | | Predicate Devices:<br>Arterial Cannula<br>K840002, K000776, K013013, K010737 | | | Extracorporeal Circuit with BioActive Surface | Extracorporeal Cir K891687 {1}------------------------------------------------ ### Device Description: The Select 3D™ and Select CAP™ Arterial Cannulae both have clear flexible, thin w.ill wire-wound PVC bodies with angled, beveled tips. The proximal end of each cannula includes a 3/8" (0.95 cm) vented or non-vented connector with a peel cap. The vented connector allows air to be vented from the cannula before connection to the perfusion line. The Select 3D™ Arterial Cannula tip has three integrated flutes which help diffuse and disperse blood flow. The cannula body features a tip orientation line indicate direction of the cannula tip during cannulation. Overall cannula length is 11.5" (29.2 cm). It is available in 22 and 24 Fr sizes. The Select CAP™ Arterial Cannula tips have an integrated pressure monitoring port. The pressure monitoring port on the tip of the cannulae provides the real time capability of accurately measuring central arterial pressure within the aorta. Overall cannula length is 12" (30.5 cm). It is available in 18, 20, 22, and 24 Fr. sizes. The devices may include a Carmeda® BioActive Surface. #### Indication for Use These cannulae are intended for use in perfusion of the ascending aorta during short term (6 hours or less) cardiopulmonary bypass. ### Comparison to Predicate Device The predicate devices are Select 3DTM and Select CAP™ Arterial Cannulae with the same design characteristics. The predicate cannulae were uncoated. The predicate cannulae have the same indications for use, with the addition of the clarification "(6 hours or less)" for the Select CAPTM. #### Summary of Performance Data In vitro visual and functional testing was used to establish the performance characteristic of the materials of these devices after Carmeda® coating. In addition coverage, bioactivity, and leach testing was performed on Carmeda® coated devices. ### Conclusion Medtronic Perfusion Systems has demonstrated that the modified Select 3D™ and Select CAP™ Arterial Cannulae are substantially equivalent to the predicate devices based upon design, test results, and indications for use. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a circular design with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or bird with three wing-like shapes. Public Health Service NOV - 5 2003 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Medtronic Perfusion Systems c/o Mr. Ronald W. Bennett Director, Regulatory/Clinical Affairs 7611 Northland Drive N Brooklyn Park, MN 55428-1088 Re: K033416 Select 3D™ Arterial Cannula and Select CAP™ Arterial Cannula with Carmeda® BioActive Surface Regulation Number: 870.4210 Regulation Name: Cardiopulmonary Bypass Vascular Catheter, Cannula, and Tubing Regulatory Class: Class II (two) Product Code: 74 DWF Dated: October 24, 2003 Received: October 27, 2003 Dear Mr. Bennett: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 - Mr. Ronald W. Bennett Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Page _1_ of _1__ 510(k) Number (if known): KD3341 b Device Name: # Select 3D™ Arterial Cannula and Select CAP™ Arterial Cannula with Carmeda® BioActive Surface Indications for Use: These cannulae are intended for use in perfusion of the ascending aorta during short term (6 hours or less) cardiopulmonary bypass. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Q. M. C. of Cardiovascular Devices 510(k) Number k033416 (Optional Format 3-10-98)
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...