FUJI MEDICAL DRY LASER IMAGERS, MODELS DRYPIX 7000, DRYPIX 5000 AND FM-DP L
K033377 · Fujifilm Medical System U.S.A., Inc. · LMC · Jan 9, 2004 · Radiology
Device Facts
Record ID
K033377
Device Name
FUJI MEDICAL DRY LASER IMAGERS, MODELS DRYPIX 7000, DRYPIX 5000 AND FM-DP L
Applicant
Fujifilm Medical System U.S.A., Inc.
Product Code
LMC · Radiology
Decision Date
Jan 9, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2040
Device Class
Class 2
Indications for Use
Fuji Medical Dry Laser Imagers, models DRYPIX 7000, DRYPIX 5000 and FM-DP L are indicated for use in providing diagnostic quality medical images on film for aid in physician diagnosis, including the printing of images and associated identification information from various digital imaging source modalities, including but not limited to, Computed Tomography (CT), Magnetic Resonance Imaging (MRI), Ultrasound, Computed Radiography, Digital Radiography, Digital Mammography and Nuclear Medicine.
Device Story
Fuji Medical Dry Laser Imagers (DRYPIX 7000, 5000, FM-DP L) function as medical image hardcopy devices. Input: digital image data from diagnostic modalities (CT, MRI, Ultrasound, etc.). Operation: device receives digital data, performs image processing, and utilizes laser exposure with thermal development to print images onto medical dry imaging film. Output: hardcopy diagnostic quality film with identification information. Used in clinical settings; operated by healthcare professionals. Output used by physicians to aid in diagnosis. Benefit: provides permanent, high-quality physical records of digital diagnostic images for clinical review.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on bench testing, hazard analysis, and compliance with recognized safety and performance standards for medical electrical equipment and laser products.
Technological Characteristics
Laser exposure and thermal development printing system. Connectivity via DICOM. Complies with UL 60601-1, IEC 60601-1-2, and 21 CFR 1040.10. Software level of concern: minor.
Indications for Use
Indicated for use in providing diagnostic quality medical images on film to aid physician diagnosis. Applicable to patients undergoing digital imaging procedures including CT, MRI, Ultrasound, Computed Radiography, Digital Radiography, Digital Mammography, and Nuclear Medicine.
Regulatory Classification
Identification
A medical image hardcopy device is a device that produces a visible printed record of a medical image and associated identification information. Examples include multiformat cameras and laser printers.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the Fujifilm logo. The logo consists of the Fujifilm symbol on the left and the word "FUJIFILM" in bold, black letters on the right. Below the word "FUJIFILM" is the text "I&I-Imaging & Information".
K033377
Page 1 of 2
# FUJIFILM MEDICAL SYSTEM
419 WEST AVENUE STAMFORD, CT 06902 Telephone: 203/324-2000 Fax: 203/353-0926
# 510(K) Summary
in accordance with the requirements of the Safe Medical Device Act, FUJIFILM Medical Systems, USA, Inc. herewith submits a 510(K) summary of safety and effectiveness for the following device.
| SUBMITTER NAME / ADDRESS: | FUJIFILM Medical Systems, USA, Inc.<br>419 West Avenue<br>Stamford, CT 06902 |
|---------------------------|------------------------------------------------------------------------------|
| CONTACT PERSON / TEL NO: | Frank Gianelli<br>Regulatory Coordinator<br>Tel No: (203) 602-3774 |
| DATE SUMMARY PREPARED: | October 17, 2003 |
| ESTABLISHMENT NO.: | 2443168 |
| TRADE/PROPRIETARY NAME: | Fuji Medical Dry Laser Imagers, models<br>DRYPIX 7000, DRYPIX 5000, FM-DP L |
| COMMON/USUAL NAME: | Medical Dry Laser Imager |
| CLASSIFICATION NAME: | Medical Image Hardcopy Device |
| CLASS/PANEL: | Class II, 90-LMC, 21CFR 892.2040 |
#### DEVICE DESCRIPTION:
Fuji Medical Dry Laser Imagers models DRYPIX 7000, DRYPIX 5000 and FM-DP L are image recording devices that use laser exposure and thermal development to print on medical dry imaging films the digital image data that is sent from medical diagnostic imaging modalities. Image processing is performed on the obtained digital images, and via laser exposure, and thermal development the le perfermou on tilm. The main difference between the FM-DP L device and the DRYPIX devices are image throughput. The only difference in the DRYPIX devices is that the DRYPIX 7000 is a 220 Vac version while the DRYPIX 5000 is a 110 Vac version.
## INTENDED USE:
Fuji Medical Dry Laser Imagers, models DRYPIX 7000, DRYPIX 5000 and FM-DP L are indicated for use in providing diagnostic quality medical images on film for aid in physician diagnosis, including the printing of images and associated identification information from various digital imaging source modalities, including but not limited to, Computed Tomography (CT), Magnetic Resonance Imaging (MRI), Ultrasound, Computed Radiography, Digital Radiography, Digital Mammography and Nuclear Medicine.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows a handwritten note with the identifier "K033377" at the top. Below the identifier, the text "page 2 of 2" is written, indicating that this is the second page of a two-page document. The handwriting appears to be somewhat cursive and slightly messy.
Image /page/1/Picture/1 description: The image shows the Fujifilm logo. The logo consists of the word "FUJIFILM" in bold, black letters, with a smaller symbol to the left. Below the main logo, the text "I&I-Imaging & Information" is printed in a smaller, lighter font, indicating the company's focus on imaging and information technologies.
FUJIFILM MEDICAL SYSTEMS US
419 WEST AVENUE STAMFORD, CT 06902 Telephone: 203/324-2000 Fax: 203/353-0926
### PREDICATE DEVICE AND SUBSTANTIAL EQUIVALENCE INFORMATION:
Fuji Medical Dry Laser Imagers models DRYPIX 7000, DRYPIX 5000 and FM-DP L are considered comparable and substantially equivalent to the Agfa LR 5200 Laser Film Recorder (K012010) manufactured by Agfa Corporation and the Kodak DRYVIEW 8610 Laser Imager (K002146) manufactured by the Eastman Kodak Company.
#### SAFETY INFORMATION:
Fuji Medical Dry Laser Imagers, models DRYPIX 7000, DRYPIX 5000 and FM-DP L introduce no new safety and efficacy issues other than those already identified with the predicate devices. The results of a hazard analysis, combined with the appropriate preventive measures taken indicate that the device is of minor level of concern as per the May 29, 1998 issue of the "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices".
Fuji Medical Dry Laser Imagers, models DRYPIX 7000, DRYPIX 5000 and FM-DP L comply with the following mandatory and voluntary standards:
- 21 CFR 1040.10 Performance Standards for Light Emitting Products (Laser Products) .
- . Medical Electrical Equipment Part 1: General Requirements for Safety UL Standard 60601-1 (IEC 60601-1-1 included)
- Medical Electrical Equipment Part 1: General Requirements for Safety; Electromagnetic . Compatibility - Requirements and Tests: IEC 60601-1-2
- DICOM (Digital Imaging and Communications in Medicine) Developed by the American . College of Radiology and the National Electrical Manufacturers Association.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three lines forming its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle. The logo is black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
'JAN - 9 2004
Mr. Frank Gianelli Regulatory Coordinator FUJIFILM Medical Systems, USA, Inc. 419 West Avenue STAMFORD CT 06902
Re: K033377
Trade/Device Name: Fuji Medical Dry Laser Imagers Models; DRYPIX 7000, DRYPIX 5000, and FM-DP L Regulation Number: 21 CFR 892.2040 Regulation Name: Medical image hardcopy device Regulatory Class: II Product Code: 90 LMC Dated: October 20, 2003 Received: October 22, 2003
Dear Mr. Gianelli:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k)
premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
Page 2
{4}------------------------------------------------
# 510(k) Number (if known): _
#### Fuji Medical Dry Laser Imagers, models DRYPIX 7000, DRYPIX Device Name : 5000 and FM-DP L
Indications For Use :
Fuji Medical Dry Laser Imagers, models DRYPIX 7000, DRYPIX 5000 and FM-DP L are indicated for use in providing diagnostic quality medical images on film for aid in physician diagnosis, including the printing of images and associated identification information from various digital imaging source modalities, including but not limited to, Computed Tomography (CT), Magnetic Resonance Imaging (MRI), Ultrasound, Computed Radiography, Digital Radiography, Digital Mammography and Nuclear Medicine.
(Please Do Not WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use ( Per 21 CFR 801.109
OR
Over-The-Counter Use
Nancy C. Brogdon
(Division Sign-Off)
(Optional Format 1-2-96)
Radiological Devices
510(k) Number K033377
Division of Reproductive, Abdominal.
Page A3-2
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.