BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM- CEFUROXIME GRAM NEGATIVE

K033362 · Becton, Dickinson & CO · LON · Apr 9, 2004 · Microbiology

Device Facts

Record IDK033362
Device NameBD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM- CEFUROXIME GRAM NEGATIVE
ApplicantBecton, Dickinson & CO
Product CodeLON · Microbiology
Decision DateApr 9, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.1645
Device ClassClass 2

Indications for Use

The BD Phoenix™ Automated Microbiology System is intended for the rapid identification and in vitro antimicrobial susceptibility testing of isolates from pure culture of most aerobic and facultative anaerobic Gram-negative and Gram-positive bacteria of human origin.

Device Story

The BD Phoenix System is an automated microbiology instrument for rapid identification and antimicrobial susceptibility testing (AST) of bacterial isolates. The system uses sealed, self-inoculating polystyrene panels containing 136 micro-wells with dried biochemicals and antimicrobial agents. Input consists of a pure culture bacterial isolate prepared in Phoenix ID or AST broth (0.5 McFarland standard). The instrument continuously incubates panels at 35°C, taking optical readings every 20 minutes. It utilizes a redox indicator and turbidity measurements to detect bacterial growth in the presence of antimicrobial agents. The system processes these inputs to determine minimum inhibitory concentration (MIC) values and categorical interpretations (S, I, or R). Results are provided to clinicians to guide antimicrobial therapy decisions, potentially improving patient outcomes by enabling targeted treatment. The system is intended for use in clinical laboratory settings.

Clinical Evidence

Performance evaluated via multi-site clinical study (6 sites) comparing BD Phoenix™ to NCCLS broth dilution reference method. Combined dataset (n=1068) showed 96.3% Essential Agreement (EA) and 93.3% Category Agreement (CA). Very major error rate was 1.3% (3/224). Reproducibility study (15 isolates) demonstrated >95% reproducibility. QC testing with E. coli ATCC 25922 confirmed results within expected ranges.

Technological Characteristics

Automated microbiology system using sealed, self-inoculating molded polystyrene panels (136 micro-wells). Sensing principle: redox indicator and turbidity measurement via optical readings every 20 minutes. Incubation: 35°C. Connectivity: integrated instrument and software. Software-based interpretation of MIC values and categorical results.

Indications for Use

Indicated for in vitro quantitative determination of antimicrobial susceptibility by MIC of Gram-negative aerobic/facultative anaerobic bacteria (Enterobacteriaceae and Non-Enterobacteriaceae) and Gram-positive bacteria (Staphylococcus and Enterococcus) from pure culture. Specifically for use with Cefuroxime (1-64 µg/mL) in Gram-negative ID/AST or AST only panels.

Regulatory Classification

Identification

A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.

Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”

Predicate Devices

Submission Summary (Full Text)

{0} 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY DEVICE ONLY TEMPLATE A. 510(k) Number: k033362 B. Purpose of Submission: Addition of cefuroxime to the BD Phoenix™ Automated Microbiology System C. Analyte: Cefuroxime 1 – 64 µg/mL Gram-Negative AST D. Type of Test: Antimicrobial Susceptibility Test (Quantitative) colorimetric oxidation-reduction, growth-based E. Applicant: Becton, Dickinson &amp; Company F. Proprietary and Established Names: BD Phoenix™ Automated Microbiology System – Cefuroxime Gram-Negative G. Regulatory Information: 1. Regulation section: 21 CFR 866.1645 Fully Automated Short-Term Incubation Cycle Antimicrobial 2. Classification: Class II 3. Product Code: LON 4. Panel: 83 H. Intended Use: 1. Intended use(s): BD Phoenix™ Automated Microbiology System: The BD Phoenix™ Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration of gram-negative aerobic and facultative anaerobic bacteria belonging to the family Enterobacteriaceae and non-Enterobacteriaceae and gram-positive bacteria belonging to the genera Staphylococcus and Enterococcus. The BD Phoenix™ GN Panel: {1} Page 2 of 5 The BD Phoenix™ Automated Microbiology System is intended for the in vitro rapid identification (ID) and quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of gram-negative aerobic and facultative anaerobic bacteria belonging to the family Enterobacteriaceae and non-Enterobacteriaceae. 2. Indication(s) for use: This submission is for the addition of the antibiotic cefuroxime at concentrations of 1 – 64 µg/mL to the gram-negative susceptibility panel. 3. Special condition for use statement(s): Results for Klebsiella pneumoniae and Providencia rettgeri have been excluded in the BD Phoenix™ therefore no results will be reported. An alternate method should be performed when these combinations are identified. Prescription use only 4. Special instrument Requirements: Not applicable I. Device Description: The BD Phoenix™ Automated Microbiology System includes instrumentation and software, sealed and self-inoculating molded polystyrene trays with 136 micro-wells containing dried reagents, and specific inoculum broth formulations for ID and AST Indicator. The organism to be tested must be a pure culture and be preliminarily identified as gram-positive or gram-negative. Colonies are then suspended in broth, and equated to a 0.5 McFarland with the recommendation to use the BD CrystalSpec™ Nephelometer. A further dilution is made into an AST broth, which contains an AST indicator, prior to inoculating the panel. The AST broth is a cation-adjusted formulation of Mueller-Hinton broth containing 0.01% Tween 80. After adding the indicator solution to the AST inoculum the color is blue and after inoculation and incubation goes to pink to colorless as reduction in the panel well proceeds. Inoculated panels are barcode scanned and loaded into the BD Phoenix™ Automated Microbiology System instrument where the panels are continuously incubated at 35°C. The AST has a final inoculum of 5 × 10⁵ CFU/ml. The instrument incubates, reads and records the results of the biochemical substrates and antimicrobial agents and interprets the reactions to give an ID of the isolate and MIC value and category interpretation of the antimicrobial agents. Organisms growing in the presence of a given antimicrobic agent reduce the indicator, signaling organism growth and resistance to the antimicrobic agent. Organisms killed or inhibited by a given antimicrobic do not cause reduction of the indicator and therefore do not produce a color change. Additional interpretation is done using software driven "EXPERT" System using rules derived from the NCCLS standards. Readings are taken every 20 minutes with an ID result available between 2-12 hours and an AST result available between 4-16 hours. This is only an autoread result; there are no manual readings possible. {2} Page 3 of 5 # J. Substantial Equivalence Information: 1. Predicate device name(s): VITEK® System 2. Predicate K number(s): N50510 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | 1. | Isolated colonies from culture used | Isolated colonies from culture used | | 2. | Report results as minimum inhibitory concentration (MIC) and categorical interpretation (SIR) | Report results as minimum inhibitory concentration (MIC) and categorical interpretation (SIR) | | 3. | <16 hours | <16 hours | | Differences | | | | Item | Device | Predicate | | 1. | Results are determined from serial twofold dilutions of antimicrobial agents | Results are determined from extrapolation of doubling dilutions | | 2. | Inoculum density equated to 0.5 McFarland standard | Inoculum density equated to 1.0 McFarland standard | | 3. | Automated growth based enhanced by use of a redox indicator (colorimetric oxidation-reduction) to detect organism growth. | Automated growth based with detection using an attenuation of light measured by an optical scanner. | # K. Standard/Guidance Document Referenced (if applicable): "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA"; NCCLS M7 (M100-S14) "Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria That Grow Aerobically; Approved Standard." # L. Test Principle: The system employs conventional, colorimetric, fluorogenic and chromogenic substrates to identify the genus and species of the isolate. The AST portion of the BD Phoenix™ Automated Microbiology System is a broth based microdilution method that utilizes a redox indicator (colorimetric oxidation-reduction) to enhance detection of organism growth. The MIC is determined by comparing growth in wells containing serial two-fold dilutions of an antibiotic to the growth in "growth control wells" which contain no antibiotic. {3} Page 4 of 5 # M. Performance Characteristics (if/when applicable): ## 1. Analytical performance: a. Precision/Reproducibility: Fifteen isolates were evaluated for site to site reproducibility and inter site reproducibility demonstrating &gt;95% reproducibility. The ten isolate study described in the guidance document was used (10 organisms tested 3 times on 3 days at 3 sites). b. Linearity/assay reportable range: Not applicable c. Traceability (controls, calibrators, or method): The recommended QC isolate was tested a sufficient number of times with acceptable results with the reference method. The Phoenix results demonstrated that the system can produce QC results in the recommended range. | ORGANISM | conc. | Reference | | Phoenix | | | --- | --- | --- | --- | --- | --- | | | | | | | | | E. coli ATCC 25922 Exp. Range: 2-8 μg/mL | 4 | | 202 | | 361 | | | 8 | | 178 | | 22 | | | 16 | | 3 | | 1 | | | 32 | | | | 1 | | | ≥64 | | 1 | | 1 | | | | | | | | All results are in the expected range. The BD Phoenix™ and the reference device had the same mode for the QC organism. Inoculum density control: The organism suspension density of the ID broth was equivalent to a 0.5 McFarland standard using the BBL™ CrystalSpec™ Nephelometer which was verified each day of testing. Internal data was used to demonstrate that the use of the BBL™ CrystalSpec™ Nephelometer would produce reproducible results. Five different instruments were used. d. Detection limit: Not applicable e. Analytical specificity: Not applicable f. Assay cut-off: Not applicable {4} Page 5 of 5 2. Comparison studies: a. Method comparison with predicate device: The NCCLS recommended broth dilution reference panel was prepared according to the NCCLS recommendation. Clinical testing was performed at six sites. The testing included both fresh clinical isolates and stock isolates along with a challenge set with known results. The test device had a growth rate of 99.9%. A comparison was provided to the reference method with the following agreement. | | EA Tot | EA N | EA % | Eval EA Tot | Eval EA N | Eval EA % | CA N | CA % | #R | min | maj | vmj | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Clinical | 1026 | 990 | 96.5 | 736 | 715 | 97.1 | 958 | 93.4 | 208 | 59 | 6 | 3 | | Challenge | 42 | 39 | 92.9 | 27 | 25 | 92.6 | 38 | 90.5 | 16 | 3 | 1 | 0 | | Combined | 1068 | 1029 | 96.3 | 763 | 740 | 97.0 | 996 | 93.3 | 224 | 62 | 7 | 3 | EA-Essential Agreement CA-Category Agreement R-resistant isolates maj-major discrepancies vmj-very major discrepancies min- minor discrepancies Essential agreement (EA) is when the BD Phoenix™ panels agree with the reference test panel results exactly or within one doubling dilution of the reference method. Category agreement (CA) is when the BD Phoenix™ panel result interpretation agrees exactly with the reference panel result interpretation. There were 3 vmj errors (3/224) noted in this study for an acceptable rate of 1.3%. b. Matrix comparison: Not applicable 3. Clinical studies: a. Clinical sensitivity: Not applicable b. Clinical specificity: Not applicable c. Other clinical supportive data (when a and b are not applicable): Not applicable 4. Clinical cut-off: Not applicable 5. Expected values/Reference range: ≤8 (S), 16 (I), ≥32 (R) The expected value range, interpretative criteria and QC are the same as recommended by NCCLS. All values will be included in the package insert. N. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...