K033343 · Varian Medical Systems, Inc. · IYE · Dec 23, 2003 · Radiology
Device Facts
Record ID
K033343
Device Name
TRILOGY RADIOTHERAPY DELIVERY SYSTEM
Applicant
Varian Medical Systems, Inc.
Product Code
IYE · Radiology
Decision Date
Dec 23, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Trilogy™ Radiotherapy Delivery System is a radiation therapy accelerator intended deliver megavoltage x-ray treatments for conventional radiotherapy (three dimensional conformal radiotherapy and intensity modulated radiotherapy) and stereotactic radiosurgery and radiotherapy. Stereotactic treatments are intended for therapy of lesions, e.g., arteriovenous malformations, primary tumors and metastases. Stereotactic treatments may be intracranial or extracranial and consist of single-session or fractionated delivery.
Device Story
Trilogy Radiotherapy Delivery System is an image-guided, dual-energy, high-dose medical linear accelerator; optimized for 3D conformal radiation therapy, intensity-modulated radiation therapy (IMRT), and stereotactic applications. Inputs include patient anatomical data for treatment planning; device delivers megavoltage x-ray beams. Features include Millennium multileaf collimator (MLC) with Dynamic MLC software, AS1000 electronic portal imaging device, asymmetric jaws, enhanced dynamic wedge, and remote couch motion. System provides 0.75mm radius isocenter accuracy across gantry, collimator, and table axes. Operated by clinical staff in radiation oncology settings; output used by physicians to target lesions (e.g., AVMs, tumors, metastases). Stereotactic mode supports 6MV, 1000 MU/min, max field 15cm x 15cm. Benefits include precise, high-dose delivery for complex intracranial/extracranial lesions.
Clinical Evidence
No clinical data provided; substantial equivalence based on technological characteristics and design modifications.
Technological Characteristics
Dual-energy medical linear accelerator; megavoltage x-ray source. Includes Millennium MLC, AS1000 electronic portal imaging device, asymmetric jaws, and enhanced dynamic wedge. Features remote couch motion and 0.75mm isocenter accuracy. Stereotactic mode: 6MV, 1000 MU/min, 15cm x 15cm max field. Software includes Dynamic MLC control. System is a modification of the Clinac 2300 C/D platform.
Indications for Use
Indicated for patients requiring conventional radiotherapy (3D conformal or intensity-modulated) or stereotactic radiosurgery/radiotherapy for lesions, including arteriovenous malformations, primary tumors, and metastases. Applicable for intracranial or extracranial sites; delivery may be single-session or fractionated.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
Brainlab Novalis Shaped Beam Surgery System (K002509)
Submission Summary (Full Text)
{0}------------------------------------------------
K03343
DEC 2 3 2003
Image /page/0/Picture/12 description: The image shows the logo for Varian Medical Systems. The word "VARIAN" is written in all caps, with a stylized "I" that resembles a person. Below the word "VARIAN" is the phrase "medical systems" in a smaller font.
### 510(k) Summary
rian Medical Systems, Inc 100 flansen Way Palo Alto, CA 94304-1038 +1 650 493 4000
The following information is provided following the format of 21 CFR 807.92 for the Trilogy Radiotherapy Delivery System
- Varian Medical Systems 1. Submitter: 3100 Hansen Way M/S H055 Palo Alto, CA 94304-1129 Contact Name: Vy Tran Phone: (650) 424-5731 (650) 842-5040 Fax: Email: vy.tran@varian.com Date summary was prepared: October 16, 2003
| 2. Name of the Device: | Trilogy Radiotherapy Delivery System |
|-------------------------|-----------------------------------------------------------------------------------------------------------|
| Trade/Proprietary Name: | Trilogy Radiotherapy Delivery System |
| Common or Usual Name: | Trilogy system |
| Classification Name: | Medical Charged Particle Radiation Therapy System<br>21 CFR §892.5050<br>Class II<br>Product Code: 90 IYE |
- 3. Predicate Devices to claim substantial equivalence:
- a. Varian Medical Systems' Clinac 2300 C/D, K913119
- b. Brainlab Novalis Shaped Beam Surgery System, K002509
- 4. Description of the Device: The modified 2300C/D, which will now be called the Trilogy™ Radiotherapy Delivery System, will include stereotactic functionality and remote couch motions as new features. The Trilogy™ Radiotherapy Delivery System is an image-guided, dual-energy, high-dose medical linear accelerator optimized for 3D conformal radiation therapy, intensity-modulated radiation therapy and stereotactic applications. The stereotactic applications include singlesession radiosurgery, fractionated stereotactic radiation therapy and intensity modulated radiosurgery. The Trilogy™ system delivers megavoltage x-ray mountated fucilobargery. The treatment of cancer. The Treatment of cancer. The Trilogy™ system includes a dynamic multileaf collimator (Millennium™ MLC with Dynamic MLC software), electronic portal imaging device (AS1000™), asymmetric jaws, enhanced dynamic wedge, a stereotactic treatment delivery mode (6MV, 1000 MU/min, maximum field size of 15cm x 15cm) and remote couch motion. There will be a 0.75mm radius isocenter for all three rotational axes, which include the gantry, collimator and table axes. The Trilogy™ system will allow for stereotactic treatments that may be intracranial or extracranial and
{1}------------------------------------------------
consist of single-session or fractionated delivery. Stereotactic treatments are intended for therapy of lesions, e.g., arteriovenous malformations, primary turnors and metastases.
- 5, Intended Use Statement: The Trilogy™ Radiotherapy Delivery System is a radiation therapy accelerator intended deliver megavoltage x-ray treatments for conventional radiotherapy (three dimensional conformal radiotherapy and intensity modulated radiotherapy) and stereotactic radiosurgery and radiotherapy. Stereotactic treatments are intended for therapy of lesions, e.g., arteriovenous malformations, primary tumors and metastases. Stereotactic treatments may be intracranial or extracranial and consist of single-session or fractionated delivery.
- Summary of the Technological Characteristics: The 2300 C/D, K913119 has 6. been modified to modifications made to include radiotherapy stereotactic functionality and remote couch motions as new features. These modifications are substantially equivalent to the features of the Brainlab Novalis Shaped Beam Surgery System, K002509. The addition of these new features also result in a name change from the Varian Clinac to the Varian Trilogy™ Radiotherapy Delivery System. The Substantial Equivalence Comparison Chart provides a comparison of the technological characteristics to those of the predicate devices. This chart is located in Tab 9 of the submission.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circle around the eagle. The eagle is black, and the text is also black. The logo is simple and recognizable.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# DEC 2 3 2003
Ms. Vy Tran Corporate Director, Regulatory Affairs Varian Medical Systems, Inc. 3100 Hansen Way PALO ALTO CA 94304-1038
Re: K033343 Trade/Device Name: Trilogy Radiotherapy Delivery System Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: 90 IYE Dated: October 16, 2003 Received: October 17, 2003
Dear Ms. Tran:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brydon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### PAGE 1 of 1
## 510(k) Number (if known): Device Name: Trilogy Radiotherapy Delivery System
Indications For Use:
The Trilogy™ Radiotherapy Delivery System is a radiation therapy accelerator intended deliver megavoltage x-ray treatments for conventional radiotherapy (three dimensional conformal radiotherapy and intensity modulated radiotherapy) and stereotactic radiosurgery and radiotherapy. Stereotactive and are intended for therapy of lesions, e.g., arteriorenous malformations, primary tumors and metastases. Stereotactic treatments may be intracranial or extracranial and consist of single-session or fractionated delivery.
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Optional Format 3-10-98)
*Prescription Use*
David Ch. Hageman.
(Division Sign-Off Division of Reproductive, a and Radiological Devices 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.