The SPO2.COM oximetry sensors and cables are intended for the continuous nominvasive monitoring of functional oxygen saturation of arterial hermoglobin (SpO2) and pulse rate (measured by an SpO2 sensor) for adult, pediatric, infant, and neonatal patients in hospitals, hospital-type facilities, mobile, and home environments.
Device Story
SPO2.COM Oximetry Sensors and cables are disposable or reusable sensors designed for compatibility with Nellcor pulse oximeter monitors. Sensors utilize red (658 nm) and infrared (905 nm) light emitters and detectors to perform non-invasive optical assessment of tissue oxygenation. Reusable sensors feature a molded clothespin-style housing; disposable sensors use adhesive bandages for attachment to fingers or toes. Devices are used in hospitals, mobile, and home environments by clinicians or patients. Output consists of raw optical signals processed by the host oximeter monitor to calculate SpO2 and pulse rate. This data assists healthcare providers in continuous patient monitoring, enabling timely clinical intervention for hypoxia or abnormal pulse rates.
Clinical Evidence
Clinical studies conducted on healthy adult volunteers using induced hypoxia. Arterial hemoglobin saturation measured by SPO2.COM sensors compared against CO-Oximeter blood sample analysis. Results: Accuracy <2% SpO2 for 70-100% range in adults, pediatrics, and infants; <3% SpO2 for neonates.
Technological Characteristics
Non-invasive optical sensor; red (658 nm) and infrared (905 nm) emitters; photodetector. Form factors: reusable clothespin-style and disposable adhesive bandage sensors. Materials identical to Masimo LNOP line; biocompatible, non-toxic, non-irritating. Non-sterile (disposable) and reusable configurations. Connectivity: wired interface to Nellcor-compatible pulse oximeter monitors.
Indications for Use
Indicated for continuous non-invasive monitoring of functional arterial hemoglobin oxygen saturation (SpO2) and pulse rate in adult, pediatric, infant, and neonatal patients in hospital, mobile, and home settings.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
Predicate Devices
Masimo SET Radical Pulse Oximeter with SatShare and LNOP series of Sensors and Cables (K031330)
{0}------------------------------------------------
DEC 1 2 2003
510(k) SUMMARY
Image /page/0/Picture/2 description: The image shows the logo and contact information for Masimo, a company located at 2852 Kelvin Avenue in Irvine, CA 92614. The phone number is 949-250-9688, and the fax number is 949-250-9686. The logo features the company name in a stylized font.
| Submitted by: | Masimo Corporation<br>2852 Kelvin Ave<br>Irvine, CA 92614-5826<br>(714) 250-9688<br>FAX (714) 250-9686 |
|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company Contact: | James J. Cronin, Vice President, Regulatory Affairs/Quality Assurance |
| Date Summary Prepared: | October 11, 2003 |
| Trade Name | SPO2.COM Oximetry Sensors |
| Common Name | Oximeter Sensor |
| Classification Name and Product Code: | Oximeter (74DQA) (870.2700)<br>Cable, Transducer and Electrode (74DSA) (870.2900) |
| Substantially Equivalent Devices: | Masimo SET <sup>®</sup> Radical Pulse Oximeter with SatShare <sup>™</sup> and LNOP <sup>®</sup> series of Sensors and<br>Cables<br>510(k) Number - K031330<br>Nellcor N-395 Pulse Oximeter - K991823<br>Nellcor N-200 Pulse Oximeter - K863784 |
Device Description
The SPO2.COM Oximetry Sensors are fully compatible disposable replacement sensors for use with Nellon and Nelloor compatible pulse oximeter monitors. They represent a design change to the Masimo LNOP Oximetry Sensors.
The SPO2.COM RSI reusable sensor is essentially identical to Masimo LNOP-DCI sensor except for the end of the cable which interfaces with the Neller. The emitter and detector are mounted in opposing halves of the molded clothespin style sensor with contoured pads to maintain contact with the patient's fingers. The same emitter (with Red wavelength of 658 nm and Infrared wavelength of 905 nm) and detector are used in both the SPO2.COM RSI and LNOP-DCI sensors. The patient contacting materials in the SPO2.COM RSI and LNOP-DCI sensors are the same. The SPO2.COM RSI sensors are supplied non-sterile.
The SPO2.COM disposable sensors are similar in construction to the predicate devices. The emitter and detector are connected to the cable assembly. The sensors have an adhesive bandage to allow the sensor to be attached to the patient's finger or toe. The same emitter (with Red wavelength of 658 nm and Infrared wavelength of 905 nm) is used in Masimo's LNOP series of stime that that tour sizes of disposable SPO2.COM sensors are available for use with adult, pediatric, infant and neonatal cuspount our sensors are cssentially identical except for the emitter and detector spacing and size and orientation of the bandage material. The patient contacting materials in the SPO2.COM disposable schsors are the same that is used in Masimo's LNOP sensor line. The SPO2.COM disposable sensors are supplied non-sterile for single patient use.
0083
{1}------------------------------------------------
# 510(k) SUMMARY
The SPO2.COM patient cables are similar in construction to the predicate device enabling the SPO2.COM Oximetry Sensors to be connected to Nellcor and Nellcor compatible pulse oximeter monitors.
### Predicate Devices
:
| SPO2.COM Sensor Line | Masimo Predicate LNOP Sensors<br>- in K031330 | Nellcor Predicate Sensors found in<br>K863784 and K991823 |
|------------------------------------------|-----------------------------------------------|-----------------------------------------------------------|
| SPO2.COM RSI - Reusable Adult Sensor | LNOP-DCI | DS-100A |
| SPO2.COM A – Adult Disposable Sensor | LNOP-Adt | D-25 |
| SPO2.COM P - Pediatric DisposableSensor | LNOP- Pdt | D-20 |
| SPO2.COM I - Infant Disposable Sensor | LNOP-Neo | I-25 |
| SPO2.COM N -- Neonatal Disposable Sensor | LNOP-Neo | N-25 |
| SPO2.COM 200 - Patient Cable | PC Patient Cable | N/A |
| SPO2.COM 395 - Patient Cable | PC Patient Cable | N/A |
#### Intended Use
The SPO2.COM oximetry sensors and cables are intended for the continuous nominvasive monitoring of functional oxygen saturation of arterial hermoglobin (SpO2) and pulse rate (measured by an SpO2 sensor) for adult, pediatric, infant, and neonatal patients in hospitals, hospital-type facilities, mobile, and home environments.
#### Technology Comparison
The SPO2.COM oximetry sensors are substantially equivalent in intended use, design, principles of operation, materials, and performance to predicate sensors and operate on identical principles of non-invasive optical assessment of tissue oxygenation using emitters and detectors.
The SPO2.COM oximetry sensors are designed, configured, and manufactured for full compatibility with Nellcor and Nellcor The or oz. Out oximeters. The SPO2. COM oximetry sensors are constructed of similar materials and components of equivalent specifications as used in the predicate devices.
The accuracy of the SPO2.COM oximetry sensors is equivalent to those of the predicate devices.
#### Performance Testing
#### Biocompatibility
All the patient contacting materials used in the SPO2.COM sensors are the same materials that are used in Masimo's LNOP series of sensors. Test results demonstrated that the materials were non-toxic, non-irritating, and non sensitizing.
#### Environmental Testing
Applicable environmetal testing per the Reviewers Guidance for Premarket Submissions - November 1993, i.e. electrical, mechanical and environmental were performed and all tests passed
#### Clinical Testing
Clinical studies were performed using the SPO2.COM Adult and Reusable oximetry sensors on healthy adult volunteer enhined studios nore perfective who were subjected to a progressive induced hypoxia and measuring the arterial hemoglobin saturation value with the instruments against the arterial hemoglobin oxygen determined from arterial blood samples with a CO-Oximeter. Clinical testing of the SPO2.COM Adult and Reusable schsors resulted in an accuracy of less than 2% SpO2 Aranievel of 70%-100% SaO2 for adults, pediatrics and infants and less than 3% AsMs for neonates.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with its wings spread, surrounded by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. The eagle is composed of three thick, curved lines, giving it a modern and abstract appearance.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 1 2 2003
Mr. James J. Cronin Vice President, Regulatory Affairs/Quality Assurance Masimo Corporation 2852 Kelvin Avenue Irvine, California 92614-5826
Re: K033298
Trade/Device Name: SPO2.COM A, P, I, N, RS-I Pulse Oximeter Sensors Regulation Number: 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DOA Dated: December 5, 2003 Received: December 8, 2003
Dear Mr. Cronin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. James J. Cronin
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Carls
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
#### 510(k) Number (if known): K033298
SPO2.COM Sensors and Cables Device Name:
Indications For Use:
The SPO2.COM Sensors and Cables are indicated for the continuous nonitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by an SpO2 sensor) for use with adult, pediatic, infant, and neonatal patients in hospitals, hospital-type facilities, mobile, and home environments.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Per 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (Per 21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
6. ANN Antoshina
510(k) Numbe
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.