← Product Code [JHI](/productcode/JHI) · K033297

# ZAP HCG TEST, MODEL 8001 (K033297)

_Zbx Corporation · JHI · Mar 18, 2004 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K033297

## Device Facts

- **Applicant:** Zbx Corporation
- **Product Code:** [JHI](/productcode/JHI.md)
- **Decision Date:** Mar 18, 2004
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1155
- **Device Class:** Class 2
- **Review Panel:** Clinical Chemistry

## Indications for Use

The ZAP™ hCG Test is an immunoassay used for the qualitative detection of human chorionic gonadotropin in human whole blood (capillary and heparinized venous), plasma, or serum and is indicated as an aid for health care professionals in the diagnosis of early pregnancy.

## Device Story

ZAP™ hCG Test is a point-of-care immunochromatographic assay for qualitative hCG detection. Input: human whole blood, plasma, or serum. Operation: sample flows through a plasma separator membrane containing colloidal gold-conjugated monoclonal mouse anti-hCG antibodies; hCG binds to these antibodies and migrates to an analytical nitrocellulose membrane; test band contains immobilized polyclonal goat anti-hCG antibodies; control band contains rabbit anti-mouse antibodies. Output: visual test band indicating presence of hCG. Used by healthcare professionals in clinical settings. Provides rapid qualitative results to aid in early pregnancy diagnosis.

## Clinical Evidence

Clinical evaluation involved 40 capillary blood samples from female patients tested by blinded clinical personnel. Results were compared against patient diagnosis (ultrasound or quantitative serum testing). The device correctly identified 26/26 pregnant patients (including those tested near the missed menstrual period) and 14/14 non-pregnant patients, showing 100% agreement with clinical diagnosis. Analytical performance included precision testing across three hospitals (100% agreement), analytical specificity (no cross-reactivity with hTSH, hLH, hFSH), and interference testing.

## Technological Characteristics

Immunochromatographic assay; utilizes monoclonal antibody to hCG conjugated to colloidal gold. Form factor is a single-step test device. Requires 35 µL specimen volume. Qualitative visual readout via Control and Test bands. Cut-off sensitivity is 10 mIU/mL. No external energy source required.

## Regulatory Identification

A human chorionic gonadotropin (HCG) test system is a device intended for the early detection of pregnancy is intended to measure HCG, a placental hormone, in plasma or urine. A human chorionic goadotropin (HCG) test system is a device intended for any uses other than early detection of pregnancy (such as an aid in the diagnosis, prognosis, and management of treatment of persons with certain tumors or carcinomas) is intended to measure HCG, a placental hormone, in plasma or urine.

## Predicate Devices

- ICON® II HCG ImmunoConcentration Assay (k860542)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
DEVICE ONLY TEMPLATE

A. 510(k) Number:
k033297

B. Analyte:
Human chorionic gonadotropin

C. Type of Test:
Qualitative

D. Applicant:
ZBX Corporation

E. Proprietary and Established Names:
ZAP™ hCG Test

F. Regulatory Information:
1. Regulation section:
21 CFR 862.1155
2. Classification:
Class II
3. Product Code:
JHI
4. Panel:
75

G. Intended Use:
1. Intended use(s):
The ZAP™ hCG Test is an immunoassay used for the qualitative detection of human chorionic gonadotropin in human whole blood (capillary and heparinized venous), plasma, or serum and is indicated as an aid for health care professionals in the diagnosis of early pregnancy.
2. Indication(s) for use:
The ZAP™ hCG Test is an immunoassay used for the qualitative detection of human chorionic gonadotropin in human whole blood (capillary and heparinized venous), plasma, or serum and is indicated as an aid for health care professionals in the diagnosis of early pregnancy.
3. Special condition for use statement(s):
Prescription use; Point of care use.
4. Special instrument Requirements:
Not applicable

{1}

Page 2 of 5

H. Device Description:

The ZAP™ hCG Test is composed of a sample application region; receiving channel that fills with sample before it flows to the plasma separator membrane; plasma separator membrane that traps red blood cells and contains monoclonal mouse anti-hCG antibodies conjugated to colloidal gold (detector antibodies); and an analytical nitrocellulose membrane containing polyclonal goat anti-hCG antibodies immobilized in the test band region and rabbit anti-mouse antibodies immobilized in the control band region. The tests strips are packaged in individually sealed pouches.

I. Substantial Equivalence Information:

1. Predicate device name(s):
ICON® II HCG ImmunoConcentration Assay

2. Predicate K number(s):
k860542

3. Comparison with predicate:

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Intended Use | For health care professional use for the early detection of pregnancy | For health care professional use for the early detection of pregnancy  |
|  Differences  |   |   |
|  Item | Device | Predicate  |
|  Intended Use/Samples | Qualitative detection of hCG in whole blood, plasma or serum | Qualitative detection of hCG in urine; semi-quantitative detection of hCG in serum  |
|  Detector Antibody | Monoclonal antibody to hCG linked to colloidal gold | Monoclonal antibody to hCG linked to alkaline phosphatase  |
|  Test Procedure | One-step, no reagents required | Multi-step, multiple reagent additions  |
|  Sensitivity | 10 mIU/mL (all samples) | 10 mIU/mL (serum)
20 mIU/mL (urine)  |

J. Standard/Guidance Document Referenced (if applicable):
None Referenced

K. Test Principle:
The ZAP™ hCG Test is an immunochromatographic assay.

L. Performance Characteristics (if/when applicable):
1. Analytical performance:

{2}

Page 3 of 5

a. Precision/Reproducibility:
Three samples of normal human whole blood were spiked with hCG to attain concentrations less than or above the cut-off value of 10 mIU/mL. Three different hospitals were each provided with five (5) replicates of each sample (15 blind labeled tubes in total). The sponsor also analyzed the fifteen samples. The results showed complete agreement between sites for each sample tested.

b. Linearity/assay reportable range:
Not applicable

c. Traceability (controls, calibrators, or method):
The test was calibrated with the Abbott ARCHITECT Total β-hCG assay, which is calibrated with material referenced to the WHO 3rd International Standard.

d. Detection limit:
The test band was designed to be visible when a specimen containing an hCG concentration of approximately 10 mIU/mL is analyzed by the test.

Three hCG positive patient serum pools were prepared to have concentrations near 20, 10, and 5 mIU/mL hCG. Twenty (20) replicates of each pool for a total of sixty (60) samples were tested with the ZAP™ hCG Test. The ZAP™ hCG Test correctly identified the forty (40) positive samples and the twenty (20) negative samples.

e. Analytical specificity:
The following related hormones were added to human negative serum: thyroid stimulating hormone (1000 μIU/mL hTSH), luteinizing hormone (500 mIU/mL hLH), and follicle stimulating hormone (1000 mIU/mL hFSH). Samples were tested on the ZAP™ hCG Test in duplicate. The results were all negative for the samples containing hTSH, hLH, and hFSH at the specified concentrations.

Interfering substances (human albumin, unconjugated bilirubin, free hemoglobin, and triglycerides) were added to human negative serum and human positive hCG sera (10 mIU/mL and 60 mIU/mL). Samples were tested on the ZAP™ hCG Test in duplicate. The negative and positive hCG sera gave the expected results at 14 g/dL, 15 mg/dL, 250 mg/dL, and 2000 mg/dL, respectively.

f. Assay cut-off:
See Detection limit above.

2. Comparison studies:

a. Method comparison with predicate device:
A comparison study between the ZAP™ hCG Test and the predicate device was performed using one hundred and sixteen (116) female serum samples. Samples were quantitated with another commercially available assay, and fifty-nine (59) samples ranged between 10 and

{3}

Page 4 of 5

200 mIU/mL. Fifty-seven (57) samples were below the 10 mIU/mL cut-off.

For the hCG positive samples, there was 100% agreement between the two assays. There were four discrepant samples in which the other test gave positive results and the ZAP™ hCG Test gave negative results. The discrepant samples had hCG concentrations in the range of 2-6 mIU/mL and were correctly reported as negative by the ZAP™ hCG Test based on the test’s cut-off. There was 100% agreement between the two assays for the remaining fifty-three (53) samples.

b. Matrix comparison:

A correlation between whole blood and plasma was performed. Heparinized whole blood from a non-pregnant female was spiked with hCG to attain samples with concentrations of &lt;1.2, 4, 9, and 19 mIU/mL hCG when plasma recovered from the spiked blood was quantitated. Both blood and corresponding plasma samples for a total of four (4) sets of samples were tested with the ZAP™ hCG Test. Each set was tested five (5) times.

The results were as follows: The agreement between whole blood and plasma was 100% for hCG values greater than or equal to 9 mIU/mL. Two tests gave a weak positive signal at 4 mIU/mL hCG for whole blood.

3. Clinical studies:

a. Clinical sensitivity: Not applicable
b. Clinical specificity: Not applicable
c. Other clinical supportive data (when a and b are not applicable): Clinical personnel obtained forty (40) capillary blood samples from female patients and tested the samples using the ZAP™ hCG Test. The operators were blinded to the diagnoses of the patients. Diagnoses were based on patient history and the results of either ultrasound or quantitative testing of serum.

There were no discrepancies between the ZAP™ hCG Test results and patient diagnosis. The ZAP™ hCG Test positively identified twenty-six (26) of the forty (40) patients that were pregnant. Of the twenty-six (26) pregnant patients, eight (8) were tested within -1 day and +9 days of the missed menstrual period. The other eighteen (18) patients were tested between two weeks to seventeen weeks after the missed menstrual period.

4. Clinical cut-off: Not applicable
5. Expected values/Reference range:

{4}

Page 5 of 5

The expected values were established using literature.

**M. Conclusion:**

The ZAP™ hCG Test is similar to the predicate device in that both tests detect hCG as an aid in early detection of pregnancy by health care professionals. The two tests differ in sample matrices that can be used; however, this does not raise new issues of safety or effectiveness. There are other commercially available assays that process whole blood (via a plasma separating mechanism) to determine the presence of hCG.

The ZAP™ hCG Test is similar to the predicate device in that both tests are immunoassays, although the new test uses monoclonal antibodies conjugated to colloidal gold as opposed to the predicate which uses monoclonal antibodies conjugated to alkaline phosphatase. Both reagents are well-established in the scientific community. Additionally, the plasma separating mechanism in the new device has been reviewed in other commercially available assays and does not raise any new issues of safety or effectiveness.

In addition to the descriptive characteristics, the various performance data provided demonstrated that the ZAP™ hCG Test was similar to the predicate device and other commercially available assays. Therefore, a substantial equivalence determination is recommended for the ZAP™ hCG Test.

---

**Source:** [https://fda-staging.innolitics.com/device/K033297](https://fda-staging.innolitics.com/device/K033297)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).