PAJUNK MODULAR HANDLE INSTRUMENTS

K033249 · Pajunk GmbH · GCJ · Mar 5, 2004 · Gastroenterology, Urology

Device Facts

Record IDK033249
Device NamePAJUNK MODULAR HANDLE INSTRUMENTS
ApplicantPajunk GmbH
Product CodeGCJ · Gastroenterology, Urology
Decision DateMar 5, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Instruments insulated for optional monopolar coagulation which enable a surgeon to grasp, manipulate, dissect, retrieve, biopsy, cut or coagulate internal tissue or organs while performing laparoscopic procedures. Caution: Federal (USA) law restricts this device to sale, distribution, and use by, or on the order of, a physician.

Device Story

Modular surgical instrument system comprising handles, guidance tubes, and instrument inserts (grasping forceps, scissors, biopsy punches, clamps/needle holders). Used in laparoscopic procedures; operated by surgeons. Handles feature ergonomic design for left/right-handed use, rotary wheel for positioning, and optional ratchet locking. Guidance tubes (3mm, 5mm, 10mm) are insulated for optional monopolar coagulation. Mechanical snap-on mechanism secures inserts to shafts and handles; double-security bracing minimizes torsion stress. Device enables tissue manipulation and coagulation; benefits include modularity, stability, and precise mechanical control. Used in clinical/OR settings.

Clinical Evidence

Bench testing only. Compliance with ISO 11134 (moist heat sterilization), ISO 11135 (ethylene oxide sterilization), and ISO 10993-1, -5, -12 (biological evaluation) was demonstrated to ensure safety and effectiveness.

Technological Characteristics

Modular system: handles, guidance tubes, and inserts. Materials: glass fiber-reinforced plastic jacketed components. Energy: optional monopolar coagulation. Dimensions: 3mm, 5mm, 10mm shaft diameters. Connectivity: none. Sterilization: moist heat (ISO 11134) or ethylene oxide (ISO 11135). Mechanical operation: snap-on locking mechanism, rotary wheel for positioning, rotation stop.

Indications for Use

Indicated for use by surgeons to grasp, manipulate, dissect, retrieve, biopsy, cut, or coagulate internal tissue or organs during laparoscopic procedures.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K033249 (pg i of 3) ## SAFETY & EFFECTIVENESS DATA SUMMARY | Submitters Name, Address & Phone Number: | PAJUNK GmbH | |------------------------------------------|-------------------| | | Karl-Hall-Str. 1 | | | D-78187 Geisingen | | | Germany | | Submission Correspondent: | Lyle Howard Corporation<br>203 Main Street, PMB 166<br>Flemington, NJ 08822<br>Attention: Lynette Howard | |---------------------------|----------------------------------------------------------------------------------------------------------| | Classification Name: | Reusable, Monopolar, Surgical | | Classification Name: | Reusable, Monopolar, Surgical Instruments | |----------------------|-------------------------------------------| | Common / Usual Name: | Modular Handle Instruments | | Proprietary Name: | PAJUNK Modular Handled Instruments | Establishment Registration Number: 9611612 | Classification: | Class II, Reg. # 876.1500, GCJ, Laparascope, General &<br>Plastic Surgery | |-----------------|---------------------------------------------------------------------------| |-----------------|---------------------------------------------------------------------------| No performance standards have been developed Performance Standards: for this device. Substantial Equivalence: The PAJUNK Modular Handle Instruments are substantially equivalent to the Richard Wolf Medical, Modular Forcep and Scissor System, K935270 in design, materials, methods of construction and intended use. The intended use of the devices to which we claim substantial equivalence: Modular forceps and scissors system for optional monopolar coagulation which enable a surgeon to grasp, manipulate, dissect, retrieve, biopsy, cut or coagulate internal tissue or organs while performing laparoscopic procedures. {1}------------------------------------------------ K033249 (pg. 2 of 3) Testing conducted or standards applied to assure safety and effectiveness include but is not limited to: ISO 11134 - Sterilization of health care products - Requirements for validation and routine control -- Industrial moist heat sterilization ISO 11135 - Sterilization of health care products - Requirements for validation and routine control - Ethylene oxide sterilization. Biological Evaluation of Medical Devices - ISO 10993-1, ISO 10993-5, ISO 10993-12 Description of the new device: The Paiunk GmbH Modular Handle Instruments are identical in terms of materials, modes of construction, to the Richard Wolf Medical Modular Forcep and Scissor System. PAJUNK GmbH Modular Handled Instruments consist of three basic parts: Handles, Guidance Tubes and Instrument Inserts. Three handles constitute the basis of the system. While the free-moving handle opens and closes without locking into place, the handle with ratchet is especially suitable for securing and holding in place. The combination handle incorporates the advantages of these two handles. The Guidance Tubes have insulated shafts and include diameter of 3mm (Shaft + Insert = One Unit), 5mm and 10mm. The instrument inserts consist of four basic types: Grasping Forceps, Scissors, Biopsy Punches and Clamps / Needle Holders. All handles can be combined with all insulated shafts and instrument inserts, regardless of their size and shape. The shaft screws onto the handle tightly, and it ensures the safe and secure connection of the two elements. Every module of the handle-instrument is kept in place in its bracing fixture, secured against distortion. A snap-on mechanism locks the instrument insert in the shaft (exception: 3mm tube): This whole clement can then in turn be latched onto the handle by a special locking device. This latching of the connecting rod and the shaft minimizes the torsion stress exerted upon the instrument insert; the double security against distortion transfers the torque also to the entire shaft. Jacketed with glass fiber-reinforced plastic, the precise mechanical construction of the PAJUNK handle ensures maximum stability. {2}------------------------------------------------ K033249 (pg 3 of 3) Designed for left and right handed use, the ergonomically shaped handles are simple and comfortable to operate. The rotary wheel can be adjusted with a little pressure applied by the index finger. A built-in rotation stop secures the desired position when grasping. Intended Use: Instruments insulated for optional monopolar coagulation which enable a surgeon to grasp, manipulate, dissect, retrieve, biopsy, cut or coagulate internal tissue or organs while performing laparoscopic procedures. Caution: Federal (USA) law restricts this device to sale, distribution, and use by, or on the order of, a physician. Safety and Efficacy Information: The Modular Handled Instruments are well recognized as being safe and effective for the stated intended use. The PAJUNK GmbH Modular Handle Instruments have the same operating principals and intended uses as the competitive modular handle instrument systems already in commercial distribution. {3}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes representing the department's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 MAR - 5 2004 PAJUNK GmbH c/o Ms. Lynette Howard Lyle Howard Corporation 203 Main Street, PMB 166 Flemington, New Jersey 08822 Re: K033249 R055249 Trade/Device Name: PAJUNK Modular Handle Instruments Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Codc: GCJ Dated: February 18, 2004 Received: February 23, 2004 Dear Ms. Howard: Wc have reviewed your Section 510(k) premarket notification of intent to market the indication indication we nave reviewed your Section 970(x) premained is substantially equivalent (for tinglestions for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encrosure) to regally to regally and the Medical Device American so to comments provision to May 20, 1776, the excordance with the provisions of the Federal Food, Drug, devices that have been reclassified in accordance was a proval application (PMA), and Cosment Act (Act) that do not requent appt o the general controls provisions of the Act. The You may, therefore, manxet the device, sequirements for annual registration, listing of general controls provisions of the rice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (sec above) into either class II (Spccial Controls) or class III (PMA), it If your device is classified (500 above) als. Existing major regulations affecting your device can may be subject to such additional controls. Existing major regulations of may be subject to suell additional controllations, Title 21, Parts 800 to 898. In addition, FDA may be found in the Oods overning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean Picase be advised that i Dri 3 hisdance or were complies with other requirements of the Act that IDA has made a dolorimiation administered by other Federal agencies. You must or any Federal statutes and regulations and assisted to: registration and listing (21 comply with an the 7te 31equirements, to state manufacturing practice requirements as set CFR Part 807), labeling (21 CFR Part 807), good if applicable, the electronic forth in the quality systems (QS) regulation (21 CFR Part 820); and 11 GFR 1800-1050 form in the quality systems (Sections 531-542 of the Act); 21 CFR 1000-1050. {4}------------------------------------------------ Page 2 - Ms. Lynette Howard This letter will allow you to begin marketing your device as described in your Section 510(k) I his letter will and in you to ough finding of substantial equivalence of your device to a legally premaired predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you deem's specific an arrantiance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general international and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Miriam C. Provost Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## PAJUNK 510(k) K033249 revision (page 64) February 18, 2004 – New FDA Form ## Indications for Use 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ Device Name:__ PAJUNK Modular Handle Instruments Indications For Use: Instruments insulated for optional monopolar coagulation which enable a surgeon to ination manipulate, dissect, retrieve, biopsy, cut or coagulate internal tissue or organs while performing laparoscopic procedures. Prescription Use Yes Yes (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use No (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Miriam C. Provost | Concurrence of CDRH, Office of Device Evaluation (ODE) | | |--------------------------------------------------------|-------------| | (Division Sign-Off) | Page 1 of 1 | | Division of General, Restorative, | | | and Neurological Devices | | 510(k) Number K033249 く
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...