IMMULITE 2500 AUTOMATED IMMUNOASSAY ANALYZER

K033234 · Diagnostic Products Corp. · DHA · Dec 12, 2003 · Clinical Chemistry

Device Facts

Record IDK033234
Device NameIMMULITE 2500 AUTOMATED IMMUNOASSAY ANALYZER
ApplicantDiagnostic Products Corp.
Product CodeDHA · Clinical Chemistry
Decision DateDec 12, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1155
Device ClassClass 2

Indications for Use

The DPC IMMULITE 2500 analyzer is an automated immunoassay system intended to assay the same broad range of analytes in patient samples as does the IMMULITE 2000. The intent of the systems is to impart the same automation to the same array of immunoassays in the same hospital and commercial laboratory settings as the IMMULITE 2000. The system is intended to produce safe and effective performance when used by medical laboratory personnel as is the predicate system, IMMULITE 2000.

Device Story

Automated immunoassay analyzer; processes patient serum/urine samples using chemiluminescent technology. System utilizes conveyor to move antibody-coated polystyrene beads through processing steps: sample/reagent addition, incubation, substrate addition, photon counting. Employs competition or immunometric (sandwich) assays with alkaline phosphatase-labeled reagents. Cleavage of phosphate ester dioxetane produces light; photomultiplier tube quantifies photon emission proportional to analyte concentration. Operated by medical laboratory personnel in clinical/hospital settings. Provides quantitative results for HCG and TSH to assist in clinical decision-making. Calibration uses manufacturer-provided master curves adjusted via 2-point customer instrument correlation.

Clinical Evidence

No clinical data provided. Substantial equivalence is supported by bench testing and performance data for specific assays (hCG and TSH) demonstrating equivalent performance to the predicate IMMULITE 2000 system.

Technological Characteristics

Automated random-access chemiluminescent immunoassay analyzer. Materials: 1/4 inch polystyrene antibody-coated beads. Detection: Photomultiplier tube (350–500 nm spectral response). Energy: Electrical (37°C incubation). Connectivity: PC-based interface, barcode reader, dual port RAM. Software: Automated control of pipetting, incubation, and photon counting. Calibration: Stored master curve with 2-point adjustment.

Indications for Use

Indicated for use by medical laboratory personnel in hospital and commercial laboratory settings to perform automated immunoassays on patient samples for a broad range of analytes, including hCG and TSH.

Regulatory Classification

Identification

A human chorionic gonadotropin (HCG) test system is a device intended for the early detection of pregnancy is intended to measure HCG, a placental hormone, in plasma or urine. A human chorionic goadotropin (HCG) test system is a device intended for any uses other than early detection of pregnancy (such as an aid in the diagnosis, prognosis, and management of treatment of persons with certain tumors or carcinomas) is intended to measure HCG, a placental hormone, in plasma or urine.

Predicate Devices

Submission Summary (Full Text)

{0} 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY DEVICE AND INSTRUMENT TEMPLATE A. 510(k) Number: k033234 B. Analyte: HCG, TSH C. Type of Test: Chemiluminescent quantitative assays, utilizing competition assays and immunometric assays D. Applicant: Edward Levine, Ph.D. Diagnostics Products Corporation Director, Clinical Affairs 5700 West 96th Street Los Angeles, CA 90045 E. Proprietary and Established Names: Proprietary name: IMMULITE 2500 Automated Immunoassay Analyzer; Established name – Random access chemiluminescent immunoassay analyzer F. Regulatory Information: 1. Regulation section: 21 CFR 862.2160, 862.1155, 862.1690 2. Classification: Class II & Class I (IMMULITE 2500) 3. Product Code: DHA, JLW, JJE 4. Panel: 75 G. Intended Use: 1. Intended use(s): The DPC IMMULITE 2500 analyzer is an automated immunoassay system intended to assay the same broad range of analytes in patient samples as does the IMMULITE 2000. The intent of the systems is to impart the same automation to the same array of immunoassays in the same hospital and commercial laboratory settings as the IMMULITE 2000. The system is intended to produce safe and effective performance when used by medical laboratory personnel as is the predicate system, IMMULITE 2000. 2. Indication(s) for use: The DPC IMMULITE 2500 analyzer is an automated immunoassay system intended to assay the same broad range of analytes in patient samples as does the IMMULITE 2000. The intent of the systems is to impart the same automation to the same array of {1} Page 2 of 8 immunoassays in the same hospital and commercial laboratory settings as the IMMULITE 2000. The system is intended to produce safe and effective performance when used by medical laboratory personnel as is the predicate system, IMMULITE 2000. 3. Special condition for use statement(s): 4. Special instrument Requirements: For use with IMMULITE 2500 immunoassay reagents. H. Device Description: Both the IMMULITE 2500 Automated Immunoassay and the predicate IMMULITE 2000 employ an automated conveyor system to move beads, which are the solid support for the assay through the various processing steps of sample and reagent addition, incubation, substrate addition and photon counting. Each system uses $\frac{1}{4}$ inch polystyrene antibody coated beads and assay specific antibody or antigen labeled with alkaline Phosphatase. The chemiluminescent detection system is a phosphate ester stabilized dioxetane. Cleavage of the Phosphatase ester alkaline phosphatase results in the decomposition of the dioxetane and the emission of light. The photon emission is quantified by luminometer via photon counting and is proportional to the quantity of the analyte present. I. Substantial Equivalence Information: 1. Predicate device name(s): IMMULITE 2000 2. Predicate K number(s): k970227 3. Comparison with predicate: | Similarities and Differences | | | | --- | --- | --- | | Item | DPC IMMULITE 2500 Device | DPC IMMULITE 2000 Predicate | | Operating principle | Chemiluminescent photon generation and detection | Chemiluminescent photon generation and detection | | Intended use | IN Vitro diagnostic measurement of analytes in patient samples | IN Vitro diagnostic measurement of analytes in patient samples | | Light Source | Chemiluminescent | Chemiluminescent | | Detector | Photomultiplier tube | Photomultiplier tube | {2} Page 3 of 8 | Detector Maximum/Minimum | 20 Million CPS/ 200 CPS | 20 Million CPS/ 200 CPS | | --- | --- | --- | | Attenuation | Programmable 100 fold via neutral density filter | Programmable 100 fold via neutral density filter | | Spectral Response | 350 – 500 nm (FWHM) | 350 – 500 nm (FWHM) | | Incubation Time | Fully random access | 30, 60, 90 minutes | | Integration Time | One second | One second | | Temperature Control | Assays incubated and read at 37°C | Assays incubated and read at 37°C | | Pipetting of specimen and reagents | Automated, Programmable volume range of 5 – 20ul | Automated, Programmable volume range of 5 – 20ul | | Separation method | Washing and axial centrifugation | Washing and axial centrifugation | | Display | Flat panel Display | Flat panel Display | | Printout | External graphics printer | External graphics printer | | Interface | Dual port ram | Dual port ram | | Assay principle | ¼" Antibody coated plastic bead solid support with liquid alkaline Phosphatase conjugated reagent. | ¼" Antibody coated plastic bead solid support with liquid alkaline Phosphatase conjugated reagent. | J. Standard/Guidance Document Referenced (if applicable): FDA Guidance Deciding When to Submit a 510(k) for a Change of an Existing Device" FOD 935; NCCLS EP-5 K. Test Principle: Competition Assays and Immunometric Assays utilizing Chemiluminescence {3} Page 4 of 8 # L. Performance Characteristics (if/when applicable): ## 1. Analytical performance: a. Precision/Reproducibility: The patient or patient pool samples were evaluated by an accelerated format of the NCCLS EP-5 protocol; two replicates of each specimen run on each of four assays per day over a 10 day period. This format yields 80 data points, which will be analyzed according to the NCCLS EP-5 algorithm. The IMMULITE 2500 average CV for the total precision shall be no more than 2% higher than the corresponding IMMULITE 2000 CV. HCG – Average total interassay precision 6.7% C.V. Acceptable Criteria <= 7.22% C.V. TSH – Average total interassay precision 5.5% C.V. Acceptable criteria - <= 10.0% b. Linearity/assay reportable range: At least three specimens at various concentrations covering the assay range are used. The specimens are analyzed neat (undiluted) and diluted 1:2, 1:4, and 1:8. Slope of the linear regression comparing the observed and the expected values for each sample is between 0.90 – 1.10. Regression slopes ranging from 0.99 – 1.07, acceptable 0.90 – 1.10 (HCG) ranging from 0.95 – 1.05, acceptable 0.90 – 1.10 (TSH) c. Traceability (controls, calibrators, or method): The IMMULITE 2500 calibration method employs a stored master curve in conjunction with a 2 point adjustment procedure. Analyte concentration is determined by a stored master curve. This standard curve is generated by the manufacturer for each lot of reagents and is provided as a "master curve". Adjustors are then used to correlate counts per second (CPS) of the customer's instrument to the CPS of the Instrument used by the manufacturer to generate a master curve. Master curves for each lot of reagents are generated on a single instrument by running numerous replicates of each standard, using a set of standards spanning the range of the assay. The standard replicates are collected in multiple runs. Replicates of a low and high adjustor are included in every run. Since the relationship between the laboratory instrument and the master curve signals is a straight line, only two points are needed to determine the line. These two points are defined by Adjustors run on the master instrument and the CPS run on the customer instrument during an adjustment. The slope and intercept of this line is then calculated. The slope and intercept are then used to observe to transform the CPS for an unknown to the CPS that would be observed on the master instrument. The purpose of the first adjustment of a new kit lot is {4} Page 5 of 8 to correlate the CPS of the customer instrument to the master instrument. Adjustment is then performed at variable increments depending upon the assay. d. Detection limit: Using calibrator A (zero calibrator as a sample), each method comparison study contains 20 replicates of the zero calibrator. The data are analyzed to determine the value associated with the counts that are 2SDs away from the average counts of the zero calibrator replicates. If the analytical sensitivity calculated from the IMMULITE 2500 is no greater than 110% of the analytical sensitivity of the IMMULITE 2000, the IMMULITE 2500 analytical sensitivity will be accepted. HCG analytical sensitivity 1.0 mIU/mL, TSH 0.01 mIU/mL. Functional sensitivity for the TSH assay was not determined. e. Analytical specificity: At least three specimens at various concentrations covering the assay range are utilized. The assays are evaluated for the effects of the presence of hemoglobin up to 500 mg/dL, conjugated and unconjugated bilirubin up to 20 mg/dL, and lipids up to 3000 mg/dL. Criteria are not established but rather effects, if present, are reported in the package insert as to the type of sample that is acceptable for analysis. f. Assay cut-off: To determine the minimum detectable concentration (MDC), or analytical sensitivity, of the IMMULITE 2500 HCG and TSH assay, 20 samples of the zero dose calibrator for each analyte was assayed in a single separate run for each analyte. Mean and standard deviations were calculated from the counts per second (CPS) for the 20 replicates. The apparent concentrations were determined at increasing standard deviations from the mean. The analytical sensitivity, defined as the concentration corresponding to two standard deviations above the average signal response of a sample free of the respective hormones, TSH and HCG, was assessed on five different runs and found to be – HCG 0.851, 0.628, 0.147, 0.782, and 0.588 mIU/mL, TSH – 0.011, 0.009, 0.009, 0.006 and 0.009 uIU/mL, on the IMMULITE 2500. Based on these studies the analytical sensitivity for the IMMULITE 2500 HCG is 1.0 mIU/mL and 0.01 mIU/mL for TSH. Functional sensitivity for the TSH assay was not determined. 2. Comparison studies: a. Method comparison with predicate device: Comparisons were performed with 100 specimens covering the calibration range. Specimens on both the IMMULITE 2500 and the IMMULITE 2000 platforms. Controls and calibrators as well as patient samples are {5} Page 6 of 8 run as unknowns; data are analyzed by linear regression. QC specifications for the IMMULITE 2000 lot to lot comparison for a specific assay will form the basis of the acceptance criteria for the linear regression comparison between the device and the predicate. b. Matrix comparison: The HCG assay was compared to DPC IMMULITE 2000 on 145 patient serum samples and compared to the IMMULITE 2500 via linear regression: $(\mathrm{IML2500}) = (1.07 \times \mathrm{IML2000} + 11.6 \mathrm{mlU/mL}, r = 0.993)$. Urine samples – a total of 123 urine samples from pregnant women were processed by the IMMULITE 2000 and by DPC's Double Antibody HCG. The samples, some of which were diluted, had HCG values ranging up to approximately $300 \mathrm{mlU/mL}$ with a linear regression of IML2000 $= 1.04 \times$ Dab $-10 \mathrm{mlU/mL}, r = 0.988$. The TSH assay was compared to the IMMULITE 2000 Rapid TSH on 100 serum samples with a concentration range to approximately $60~\mathrm{uIU/mL}$ and was determined comparable with a linear regression of $(\mathrm{IML}2500) = 1.03(\mathrm{IML}2000) - 0.01~\mathrm{uIU/mL}, r = 0.997$ 3. Clinical studies: a. Clinical sensitivity: N/A b. Clinical specificity: N/A c. Other clinical supportive data (when a and b are not applicable): N/A 4. Clinical cut-off: N/A 5. Expected values/Reference range: HCG – males and non pregnant females, $95\%$ of the males measured below the assay detection limit of $1.1 \mathrm{mlU/mL}$ and all were below $2.5 \mathrm{mlU/mL}$. Non pregnant females were $83\%$ below detection limit $95\%$ were below $2.5 \mathrm{mlU/mL}$ and all were below $5.3 \mathrm{mlU/mL}$. A total of 593 serum samples from apparently healthy women were processed by the IMMULITE 2500 HCG Assay and are summarized in the product labeling. Urine – a result of greater than or equal to $30 \mathrm{mlU/mL}$ is considered positive. A result less than $30 \mathrm{mlU/mL}$ is considered negative. TSH – Based on its relationship with the IMMULITE 2000, the TSH 2500 assay can be expected to have essentially the same reference ranges. Euthyroid: $0.4 - 4.0 \mathrm{ulU/mL}$ and Hyperthyroid: $< 0.01 \mathrm{ulU/mL}$ {6} Page 7 of 8 M. Instrument Name: Diagnostics Products Corporation IMMULITE 2500 N. System Descriptions: 1. Modes of Operation: Routine random analyzer 2. Software: The IMMULITE 2500 utilizes a PC based interface to the user with a color matrix display and a barcode reader for input and display of data. 3. FDA has reviewed applicant's Hazard Analysis and software development processes for this line of product types: Yes ☐ X ☐ or No ☐ 4. Sample Identification: Bar code, computer based or user input 4. Specimen Sampling and Handling: samples vary depending upon assay. TSH - serum samples, HCG - serum samples, qualitative urine samples - urine 5. Assay Types: Competition assay or Immunometric (sandwich) assays 6. Reaction Types: Chemiluminescent 7. Calibration: The IMMULITE 2500 calibration method employs a stored master curve in conjunction with a 2 point adjustment procedure. Analyte concentration is determined by a stored master curve. This standard curve is generated by the manufacturer for each lot of reagents and is provided as a "mater curve". Adjustors are then used to correlate counts per second (CPS) of the customer's instrument to the CPS of the Instrument used by the manufacturer to generate a master curve. Master curves for each lot of reagents are generated on a single instrument by running numerous replicates of each standard, using a set of standards spanning the range of the assay. The standard replicates are collected in multiple runs. Replicates of a low and high adjustor are included in every run. Since the relationship between the laboratory instrument and the master curve signals is a straight line, only two points are needed to determine the line. These two points are defined by Adjustors run on the master instrument and the CPS run on the customer instrument during an adjustment. The slope and intercept of this line is then calculated. The slope and intercept are then used to observe to transform the CPS for an unknown to the CPS that would be observed on the master instrument. The purpose of the first adjustment of a new kit lot is to correlate the CPS of the customer instrument to the master instrument. Adjustment is then performed at variable increments depending upon the assay. {7} Page 8 of 8 8. **Quality Control**: Use two controls low and high of HCG and TSH controls O. Other Supportive Instrument Performance Characteristics Data Not Covered In The “L. Performance Characteristics” Section Of The SE Determination Decision Summary. P. **Conclusion**: Based upon the information provided, I recommend that the IMMULITE 2500 Analyzer and the assays for HCG and TSH be found substantially equivalent to the respective predicate devices.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...