HOLOGIC QDR EXPLORER X-RAY BONE DENSITOMETER, MODEL EXPLORER
K033224 · Hologic, Inc. · KGI · Nov 5, 2003 · Radiology
Device Facts
Record ID
K033224
Device Name
HOLOGIC QDR EXPLORER X-RAY BONE DENSITOMETER, MODEL EXPLORER
Applicant
Hologic, Inc.
Product Code
KGI · Radiology
Decision Date
Nov 5, 2003
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1170
Device Class
Class 2
Indications for Use
The Hologic QDR Explorer is a fan beam X-Ray Bone Densitometer indicated for (1) the measurement of bone mineral content (BMC) and the estimation of bone mineral density (BMD), (2) comparison of measurements to reference databases, (3) the estimation of fracture risk, (4) body composition analysis, and (5) measurement of periproshtetic BMD.
Device Story
Hologic QDR Explorer is a fan beam X-Ray Bone Densitometer; utilizes switched pulse dual-energy x-ray tube (100/140 kV) and 54-channel detector array (CdWO4 scintillators/silicon diodes) to measure bone mineral content and density. Operated by clinicians in clinical settings; uses laser positioning for patient alignment. System processes scan data to produce BMD/BMC measurements, fracture risk estimates, and body composition analysis. Output allows healthcare providers to assess bone health and fracture risk, aiding clinical decision-making for osteoporosis management. Device is a modification of the QDR-3000 platform, featuring updated Windows XP-based software and hardware components.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on bench testing and comparison of technological characteristics to the predicate device.
Technological Characteristics
Fan beam X-ray bone densitometer; switched pulse dual-energy X-ray tube (100/140 kV, tungsten target); 54-channel detector array (CdWO4 scintillators/silicon diodes); laser positioning diode (<1mW); Windows XP-based operating system; PC hardware (>1.5 GHz, 256 MB RAM). Complies with 21 CFR 1020.30(k) for leakage radiation.
Indications for Use
Indicated for patients requiring bone mineral content (BMC) and bone mineral density (BMD) assessment, fracture risk estimation, body composition analysis, and periprosthetic BMD measurement.
Regulatory Classification
Identification
A bone densitometer is a device intended for medical purposes to measure bone density and mineral content by x-ray or gamma ray transmission measurements through the bone and adjacent tissues. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
Hologic QDR-3000 X-Ray Bone Densitometer (K943505)
Hologic Discovery Package for QDR X-Ray Bone Densitometers (K023398)
Submission Summary (Full Text)
{0}------------------------------------------------
K033224
Hologic, Inc. October 3, 2003 Hologic QDR Explorer X-Ray Bone Densitometer SPECIAL 510(k) Premarket Notification
Section E
510(k) Summary
000108
{1}------------------------------------------------
## E. 510(k) Summary
## E.1 Company Identification
Hologic, Inc. 35 Crosby Drive Bedford. MA 01730 781-999-7300
#### E.2 Contact Information
Daniel F. Phelan Senior Regulatory Affairs Specialist
## E.3 Date of Submission
October 3, 2003
### E.4 Device Identification
Proprietary Name Classification Name: Common/Usual Name Requlation Number: Product Code: Classification Classification Panel
Hologic QDR Explorer X-Ray Bone Densitometer Bone Densitometer Bone Densitometer 21 CFR 892.1170 90 KGI 11 Radiology
### E.5 Predicate Device Information
- K943505 Hologic QDR-3000 X-Ray Bone Densitometer
- Hologic Discovery Package for QDR X-Ray Bone Densitometers K023398
#### E.6 Device Description and Intended Use
The Hologic QDR Explorer is a fan beam X-Ray Bone Densitometer indicated for (1) the measurement of bone mineral content (BMC) and the estimation of bone mineral density (BMD), (2) comparison of measurements to reference databases, (3) the estimation of fracture risk, (4) body composition analysis, and (5) measurement of periproshtetic BMD.
000109
{2}------------------------------------------------
Hologic, Inc.
October 3, 2003
# Hologic QDR Explorer X-Ray Bone Densitometer SPECIAL 510(k) Premarket Notification
KT333222
## E.7 Substantial Equivalence
| | QDR-3000 | QDR Explorer |
|---------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K943505 | |
| System Footprint | 3.02m (119 in) L x 1.50m (59 in) W x<br>1.42m (56 in) H +/- 25mm (1 in). | Same |
| System Weight | 320 kg (720 lb) | 327 kg (729 lb) |
| System Power Requirements | 100V (23A), 120V (18A), and 230V<br>(9A) models available | Same |
| X-Ray Tube | Switched pulse dual-energy x-ray tube<br>operating at 100 kV and 140 kV. 2.5<br>mA average at 25% duty cycle.<br>Tungsten target | Same. |
| Detector Array | 64 multichannel detector consisting of<br>CdWO4 scintillators coupled to silicon<br>diodes | 54 multichannel detector consisting of<br>CdWO4 scintillators coupled to silicon<br>diodes |
| X-Ray Source-to-Image-Detector<br>Distance | 1070 mm (42 in) +/- 8 mm (0.315 in) | 883.4 mm (34.8 in) +/- 8 mm (0.315 in) |
| X-Ray Source-to-Patient Distance | 424 mm (16.7 in) +/- 6 mm (0.236 in) | Same |
| Collimation | Aperture with 1.0 mm slit. | Same |
| Laser Positioning Device | Laser diode < 1mW cross hair with<br>emergency mechanical shutter. | Same |
| Leakage Radiation | The QDR-3000 meets the<br>requirements of 21 CFR 1020.30(k) for<br>leakage from the x-ray source. | The QDR Explorer meets the<br>requirements of 21 CFR 1020.30(k) for<br>leakage from the x-ray source |
| Scatter Radiation | 10μGy/hr at 1m from the examination<br>table (Nominal) | Same |
| Scan Time | 15 seconds - 407 seconds, depending<br>on scan mode | 62 seconds - 403 seconds, depending<br>on scan mode |
| Scan Sites | Lumbar spine (L1, L2, L3, L4),<br>Proximal femur (hip), Forearm (radius<br>and ulna), Whole body | Same |
| Software Operating System | Hologic Eagle Software Windows 98-<br>Based Operating System (K992677) | Hologic Discovery Software Package<br>Windows XP-Based Operating System<br>(K023398) |
| PC Hardware Requirements | > 1.0 GHz processor<br>> 256 MB RAM<br>> 20 GB hard disk<br>3.5" 1.44 MB diskette drive<br>17" monitor (CRT)<br>120 MB SuperDisk drive<br>CD-ROM drive<br>16 MB video card<br>Standard keyboard and mouse | > 1.5 GHz processor<br>Same<br>Same<br>Same<br>17" monitor (CRT or LCD)<br>CD/RW drive<br>N/A<br>Same<br>Same |
| BMC Measurement | Standard | Same |
| BMD Measurement | Standard | Same |
| Fracture Risk Estimation (K983028) | Standard | Same |
| Reference Databases (K963363) | Standard | Same |
| Body Composition Analysis (K961787) | Optional | Same |
| Periprosthetic BMD Measurement<br>(K002711) | Optional | Same |
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo features a stylized image of an eagle with three lines forming its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV - 5 2003
Mr. Daniel F. Phelan Senior Regulatory Affairs Specialist Hologic, Inc. 35 Crosby Drive BEDFORD MA 01730-1401
Re: K033224 Trade/Device Name: Hologic ODR X-Ray Bone Densitometer Regulation Number: 21 CFR 892.1170 Regulation Name: Bone densitometer Regulatory Class: II Product Code: 90 KGI Dated: October 3, 2003 Received: October 6, 2003
Dear Mr. Phelan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050,
{4}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely vours.
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
Page 2
{5}------------------------------------------------
## A.2 Indications for Use Statement
510(k) Number (if known): K033224
Device Name: Hologic QDR Explorer X-Ray Bone Densitometer
Indications for Use:
The Hologic QDR Explorer is a fan beam X-Ray Bone Densitometer indicated for (1) the measurement of bone mineral content (BMC) and the estimation of bone mineral density (BMD), (2) comparison of measurements to reference databases, (3) the estimation of fracture risk, (4) body composition analysis, and (5) measurement of periprosthetic BMD
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
V Prescription Use______________________________________________________________________________________________________________________________________________________________
OR
Over-The-Counter-Use
(Per 21 CFR 801.109)
(Optional Format 1)
Nancy C Brogdon
(Division Sign-Division of Reproductive and Radiological Devic 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.