PHERON PTA CATHETER

K033217 · Biotronik, Inc. · DQY · Oct 31, 2003 · Cardiovascular

Device Facts

Record IDK033217
Device NamePHERON PTA CATHETER
ApplicantBiotronik, Inc.
Product CodeDQY · Cardiovascular
Decision DateOct 31, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1250
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Pheron PTA balloon catheter is indicated to dilate stenosis in the renal, iliac, femoral, popliteal and infrapopliteal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae.

Device Story

Pheron PTA balloon catheter; dual-lumen design; used for percutaneous transluminal angioplasty. Input: guide wire (max 0.035"). Operation: inflation device connected to proximal Luer port inflates balloon to specific diameter/pressure; radiopaque markers facilitate fluoroscopic visualization/positioning across lesion. Features: tapered soft tip for advancement; BIOC (silicone) coating for trackability. Used in clinical setting by physicians to dilate peripheral vessel/fistula stenoses, improving blood flow. Output: mechanical dilation of vessel/fistula.

Clinical Evidence

Bench testing only.

Technological Characteristics

Dual-lumen balloon catheter; BIOC (silicone) coating; radiopaque markers; compatible with 0.035" guide wires; Luer port inflation system. Mechanical device; no software or electronic components.

Indications for Use

Indicated for patients requiring dilation of stenosis in renal, iliac, femoral, popliteal, and infrapopliteal arteries, and treatment of obstructive lesions in native or synthetic arteriovenous dialysis fistulae.

Regulatory Classification

Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ \$\ast\$ OCT 3 1 2003 Ko33217 BIOTRONIK, Inc., Pheron Peripheral PTA Catheter, 510(k) October 2, 2003 # Pheron Peripheral Percutaneous Transluminal Angioplasty (PTA) Catheter 510(k) Premarket Notification #### 510(K) SUMMARY 1. Name and Address of Sponsor: BIOTRONIK, Inc. 6024 Jean Road Lake Oswego, OR 97035 Establishment Registration Number: 1028232 Device Name: Proprietary Name: Classification: Classification Name: Product Code: October 2, 2003 Date Prepared: Pheron PTA Catheter Class II (21 CFR 870.1250) Percutaneous Catheter LIT; DQY ### General Description: The Pheron Percutaneous Transluminal Angioplasty (PTA) balloon catheter is indication of stendlic seqments in peripheral vessels. The dilatation balloon is designed to inflate to a known diameter at a specific inflation pressure consistent with the compliance chart on the label. Two radiopaque markers are located at either end of the ballitate fluoroscopic visualization and positioning of the balloon catheter towards and across the lesion. The balloon catheter includes a tapered soft tip to facilitate advancement of the catheter. The balloon catheter shaft has two Luer ports at the proximal end. One port (inflation port) serves for connecting an inflation device to inflate/deflate the balloon. The other port enables insertion of the guide wire lumen. The balloon catheter is a dual lumen design, with both lumens contained within one tube. The smaller lumen is the balloon inflation lumen, as described above. The larger lumen permits the use of guide wires with a maximum diameter of 0.035" to facilitate advancement of the balloon catheter towards and through the lesion(s) to be dilated. The balloon catheter is compatible with introducer sheath (introducer) sizes according to the recommendations on the label. The balloon catheter has a BIOC (silicone) coating to improve the trackability characteristics. BIOTRONIK proposes the following PTA catheter cleared through 510(k) notifications as a predicate device for the Pheron Peripheral PTA Catheter: - Jomed AG's Fox PTA catheter (K010838, cleared 06-21-01 and K020854, cleared 04-11-02) #### Indication for Use: The Pheron PTA balloon catheter is indicated to dilate stenosis in the renal, iliac, femoral, popliteal and infrapopliteal arteries and for the treatment of obstructive or synthetic arteriovenous dialysis fistulae. #### Name and Address of Manufacturing Site: BIOTRONIK GmbH & Co. (reg. no. 9610139) Woermannkehre 1, 12359 Berlin, Germany 011-49-30-689-05-304 #### Name and Address of Contract Manufacturing Site: BIOTRONIK AG (reg. no. 8043892) Ackerstrasse 6 8180 Bülach, Switzerland 011-41-1-864-5247 ## Contact Person and Phone Number: Jon Brumbaugh Director, Regulatory Affairs Fax (503) 635-9936 Phone (888) 345-0374 {1}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Image /page/1/Picture/2 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized emblem featuring three curved shapes that resemble a bird or abstract human figures. The emblem is black, and the text is also in black. OCT 3 1 2003 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Biotronik, Inc. c/o Mr. Jon Brumbaugh Director, Regulatory Affairs 6024 Jean Road Oswego. OR 97035 Re: K033217 Pheron PTA Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: October 2, 2003 Received: October 3, 2003 Dear Mr. Brumbaugh: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {2}------------------------------------------------ Page 2 - Mr. Jon Brumbaugh Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4648. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Ashley B. Braun Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ Page_ __of_1 510(k) Number (if known):__ Device Name: Pheron Peripheral Percutaneous Transluminal Angioplasty (PTA) Catheter Indications For Use: : : : : : > The Pheron PTA balloon catheter is indicated to dilate stenosis in the renal, iliac, femoral, popliteal and infrapopliteal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. (Please Do NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) | Prescription Use | |----------------------| | (Per 21 CFR 801.109) | (Optional Format 3-10-98) Division of Cardiovascular & Respiratory Devices 510(K) Number _______________________________________________________________________________________________________________ Adelle B. Evans laboz Division of Cardiovascular Devices 510(k) Number_1(0532(7
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...