← Product Code [JIN](/productcode/JIN) · K033114

# KETOCHECKS (K033114)

_Neo Diagnostics, Inc. · JIN · Apr 26, 2004 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K033114

## Device Facts

- **Applicant:** Neo Diagnostics, Inc.
- **Product Code:** [JIN](/productcode/JIN.md)
- **Decision Date:** Apr 26, 2004
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1435
- **Device Class:** Class 1
- **Review Panel:** Clinical Chemistry

## Indications for Use

The Neo Diagnostics KetoChecks is intended for the identification of ketones in urine and can be used in the diagnosis of acidosis (a condition characterized by abnormally high acidity of body fluids) or ketosis (a condition characterized by increased production of ketone bodies such as Acetoacetic acid) and for monitoring patients on ketogenic diets and patients with diabetes. The test is indicated for use by professionals and for Over the Counter use by lay people.

## Device Story

KetoChecks is a disposable, firm plastic reagent strip for qualitative/semi-quantitative detection of acetoacetic acid in urine. User dips strip into fresh, un-centrifuged urine; compares resulting color change on reagent pad against color blocks printed on bottle label. No instruments or calculations required. Used by professionals or lay people (OTC) in clinical or home settings. Provides immediate visual feedback on carbohydrate metabolism status; assists in monitoring diabetes or ketogenic diet adherence. Results interpreted by direct visual comparison; aids clinical decision-making regarding metabolic state.

## Clinical Evidence

Clinical correlation study compared KetoChecks Lot 1 against Bayer Ketostix. Results showed 97.2% agreement on the same color block and 99.5% agreement within one color block. A secondary study compared two lots of KetoChecks, showing 97.1% agreement on the same color block and 99.3% within one color block. No clinical prospective/retrospective trials on patient outcomes were reported; evidence is based on bench-top analytical performance and comparative testing.

## Technological Characteristics

Firm plastic strip with reagent area containing 6.0% w/w sodium nitroprusside, buffer, and nonreactive ingredients. Sensing principle: chemical colorimetric reaction. Form factor: dipstick. Storage: 10-30°C, protected from light/moisture. No electronic components, software, or energy source.

## Regulatory Identification

A ketones (nonquantitative) test system is a device intended to identify ketones in urine and other body fluids. Identification of ketones is used in the diagnosis and treatment of acidosis (a condition characterized by abnormally high acidity of body fluids) or ketosis (a condition characterized by increased production of ketone bodies such as acetone) and for monitoring patients on ketogenic diets and patients with diabetes.

## Predicate Devices

- Bayer Ketostix

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
DEVICE ONLY TEMPLATE

A. 510(k) Number:
k033114

B. Purpose for Submission:
Premarket Notification 510(k) of intention to manufacture and market
the Neo Diagnostics, Inc., KetoChecks for over the counter use.

C. Analyte:
Urinary Ketones

D. Type of Test:
Qualitative and semi-quantitative

E. Applicant:
Neo Diagnostics, Inc.

F. Proprietary and Established Names:
Neo Diagnostics, Inc. KetoChecks

G. Regulatory Information:
1. Regulation section: 21 CFR § 862.1435; 21 CFR 862.9 (c)(9)
2. Classification: Class I meets limitation of exemptions
3. Product Code: JIN
4. Panel: 75 Clinical Chemistry

H. Intended Use:
1. Intended use(s):
The Neo Diagnostics KetoChecks is intended for the identification of ketones in urine and can be used in the diagnosis of acidosis (a condition characterized by abnormally high acidity of body fluids) or ketosis (a condition characterized by increased production of ketone bodies such as Acetoacetic acid) and for monitoring patients on ketogenic diets and patients with diabetes. The test is indicated for use by professionals and for Over the Counter use by lay people.
2. Indication(s) for use:
The Neo Diagnostics KetoChecks is intended for the identification of ketones in urine and can be used in the diagnosis of acidosis (a condition characterized by abnormally high acidity of body fluids) or ketosis (a condition characterized by

{1}

Page 2 of 5

increased production of ketone bodies such as Acetoacetic acid) and for monitoring patients on ketogenic diets and patients with diabetes. The test is indicated for use by professionals and for Over the Counter use by lay people.

3. Special condition for use statement(s):
This submission is for a professional and over-the-counter version of a cleared professional device for urine ketones. Professionals and consumers can perform the test using the midstream technique or the dip and read technique.

4. Special instrument Requirements: Not Applicable

I. Device Description:
The Neo Diagnostics, Inc. KetoChecks consists of a firm plastic strip to which a regent area is affixed. Depending on the product being used, KetoChecks provides a test for Ketone (Acetoacetic acid) in urine. Test results may provide information regarding the status of carbohydrate metabolism.

KetoChecks is packaged along with a drying agent in a plastic bottle with a twist-off cap. Each strip is stable and ready to use upon removal from the bottle. The entire reagent strip is disposable. The directions must be followed exactly. Results are obtained by direct comparison to the test strip with a color blocks printed on the bottle label. Accurate timing is essential to provide optimal results. To obtain optimal results, it is necessary to use fresh, well mixed, and un-centrifuged urine. No calculations or laboratory instruments are needed.

J. Substantial Equivalence Information:

1. Predicate device name(s):
Bayer Ketostix reagent strips, Urine Dipstick or tablet analysis, non automated

2. Predicate K number(s): k801270

3. Comparison with predicate:

|  Strip Name Distributor | KetoChecks Neo Diagnostics, Inc. | Bayer Ketostix Bayer  |
| --- | --- | --- |
|  Reagents for: |  |   |
|  Ketone (Acetoacetic acid) | Sodium nitroprusside | Sodium nitroprusside Buffer  |
|  Time required to Read strips | 30 to 120 seconds | 15 to 120 seconds  |
|  Storage | Between 10-30°C (50°F - 86°F). Do Not Store in refrigerator or freezer. Do Not expose direct heat, light or Moisture. | Between 10-30°C (50°F - 86°F). Do Not Store in refrigerator or freezer. Do Not Store in Direct Sunlight  |

{2}

Page 3 of 5

# K. Standard/Guidance Document Referenced (if applicable): NCCLS GP16 T

## L. Test Principle:

This test is based on the reaction of acetoacetic acid with sodium nitroprusside in a strongly basic medium. The colors range from beige or buff-pink for a “Negative” reading to a pink and pink-purple for a “Positive” reading.

## M. Performance Characteristics (if/when applicable):

### 1. Analytical performance:

a. **Precision/Reproducibility:**

The data demonstrated adequate reproducibility between masked operators and between different strip lots

b. **Linearity/assay reportable range:**

The data demonstrated a linear relationship between spiked-in values and the semi-quantitative reading for ketones.

c. **Traceability (controls, calibrators, or method):**

None Stated

d. **Detection limit:**

The sensitivity was determined by spiking in the different concentration of the Lithium Acetoacetate to determine the point that can identify or notice the color change on the strips. Sensitivity was demonstrated to the lowest reported block of 5 mg/mL.

e. **Analytical specificity:**

Color reaction that could be interpreted as false positive may be obtained with urine specimens containing MESNA or large amounts of phenylketones or L-dopa metabolites.

f. **Assay cut-off:**

Not Applicable

### 2. Comparison studies:

a. **Method comparison with predicate device:**

In this submission the sponsor demonstrated that untrained consumers, after reading the proposed package insert could use the test correctly and achieve results comparable to trained professional users. Clinical studies were conducted at 5 different sites. (Site 1 Neo Diagnostics on site = 90 samples, Site 2 = 96 samples, Site 3 = 44 samples, Site 4 = 43 samples, and Site 5 = 43 samples). The volunteers were given the package insert to read and then requested to perform the test on specified control solutions that spanned a range from (0 to 160 mg/dl) ketone concentration, using both the Neo Diagnostics KetoChecks and the

{3}

Page 4 of 5

predicate Bayer Ketostix, comparing the reagent areas to corresponding color charts.

The qualitative comparison of % agreement for each of the 5 sites is listed below:

Site 1 = 97.8% (in-house data)
Site 2 = 97.9%
Site 3 = 97.9%
Site 4 = 95.4%
Site 5 = 97.7%

In a separate study, samples were tested with the proposed device and compared with Bayer using one lot. In addition, two lots of the proposed device were compared with each other. The following results were obtained:

**KetoChecks Lot1 versus Bayer Ketostix Lot 3A04A**

|   | Same Color Block |   |  | Within 1 Color Block  |   |
| --- | --- | --- | --- | --- | --- |
|  Evaluation One-Lot 1 | 212/218 | 97.2% |  | 217/218 | 99.5%  |
|  Evaluation Bayer | 254/254 | 100.0% |  | 254/254 | 100.00%  |

**KetoChecks Lot 1 versus KetoChecks Lot 2**

|   | Same Color Block |   |  | Within 1 Color Block  |   |
| --- | --- | --- | --- | --- | --- |
|  Evaluation One-Lot 1 vs Lot 2 | 134/138 | 97.1% |  | 137/138 | 99.3%  |

b. Matrix comparison:
Not Applicable

3. Clinical studies:
a. Clinical sensitivity:
Not Applicable
b. Clinical specificity:
Not Applicable
c. Other clinical supportive data (when a and b are not applicable):

4. Clinical cut-off:
Not Applicable

{4}

Page 5 of 5

5. Expected values/Reference range:
Normally, no ketones are present in urine. Detectable levels of ketone may occur in urine during physiological stress conditions such as fasting, pregnancy, and frequent strenuous exercise. In starvation diets, or in other abnormal carbohydrate situations, ketones appear in the urine in excessively large amounts before serum ketones are elevated.

N. Conclusion:
The submitted material in this premarket notification is complete and supports a substantial equivalence decision.

---

**Source:** [https://fda-staging.innolitics.com/device/K033114](https://fda-staging.innolitics.com/device/K033114)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).