Retrospective clinical samples were used to assess the clinical sensitivity and specificity of the HSV 1 IgG ELISA by comparing results against Western Blot and culture reference methods.
Expectant mothers; Sample Size: 210; Number of Sites: 1
HSV 1 Western Blot (WB)
Percent agreement positive and negative
Retrospective analysis of banked/masked sera
Sexually active adults over the age of 14; Sample Size: 198; Number of Sites: 1
HSV 1 Western Blot (WB)
Percent agreement positive and negative
Retrospective analysis of banked/masked sera
Low prevalence population; Sample Size: 184; Number of Sites: 1
HSV 1 Western Blot (WB)
Percent agreement positive and negative
Retrospective analysis of banked/masked sera
Patients 6 weeks to 1 year post clinical presentation and culture HSV 1 positive; Sample Size: 53; Number of Sites: 1
Culture and HSV 1 Western Blot (WB)
Percent agreement positive
Indications for Use
The Trinity Biotech Captia™ Herpes Simplex Virus (HSV) 1 Type Specific IgG kit is an Enzyme-linked Immunosorbent Assay (ELISA) intended for qualitatively detecting the presence or absence of human IgG class antibodies to HSV-1 in human serum. In conjunction with the Trinity Biotech Captia™ Herpes Simplex Virus (HSV) 2 Type Specific IgG kit, the test is indicated for testing sexually active adults or expectant mothers for aiding in the presumptive diagnosis of HSV infection. Due to the implications of positive results, it is recommended they be confirmed in a low prevalence population with Western blot. The performance of this assay has not been established for use in a pediatric population, for neonatal screening, for testing of immunocompromised patients, or for use with automated equipment. The user is responsible for establishing assay performanace in these populations and with automated equipment
Device Story
The Captia™ HSV 1 Type Specific IgG is an ELISA kit for qualitative detection of human IgG antibodies to HSV-1 in serum. The device uses recombinant HSV-1 glycoprotein g antigen bound to polystyrene microplate wells. Patient serum is added; if specific IgG is present, it binds to the antigen. After washing, goat anti-human IgG conjugated with horseradish peroxidase is added, followed by TMB chromogen/substrate. The enzymatic reaction produces a blue color, turning yellow upon addition of 1N H2SO4 stop solution. The intensity, proportional to antibody concentration, is measured via spectrophotometer at 450 nm (with optional 650 nm reference). Results are calculated as an index value relative to a cutoff calibrator. The test is performed in a laboratory setting by trained personnel. It aids clinicians in the presumptive diagnosis of HSV infection in specific adult populations. It is not for use with automated equipment.
Clinical Evidence
Clinical performance was evaluated using banked/masked sera from expectant mothers (n=210), sexually active adults (n=198), and a low-prevalence population (n=184), compared against Western Blot (WB). Positive agreement with WB ranged from 84.0% to 88.3%; negative agreement ranged from 98.2% to 100.0%. In culture-positive patients (n=53), positive agreement was 75.5% against culture and 85.7% against WB. Analytical precision (intra/inter-assay) was reported with CVs generally <15% for positive samples. CDC panel testing showed 96.0% total agreement.
Technological Characteristics
Enzyme-Linked Immunosorbent Assay (ELISA) using purified recombinant HSV gG1 antigen immobilized on solid-phase microtiter wells. Detection via enzyme-labeled anti-human IgG conjugate and substrate-induced colorimetric reaction measured photometrically. Manual assay format; not established for automated equipment.
Indications for Use
Indicated for sexually active adults and expectant mothers to aid in the presumptive diagnosis of HSV-1 infection. Not established for pediatric populations, neonatal screening, immunocompromised patients, or use with automated equipment.
Regulatory Classification
Identification
Herpes simplex virus serological assays are devices that consist of antigens and antisera used in various serological tests to identify antibodies to herpes simplex virus in serum. Additionally, some of the assays consist of herpes simplex virus antisera conjugated with a fluorescent dye (immunofluorescent assays) used to identify herpes simplex virus directly from clinical specimens or tissue culture isolates derived from clinical specimens. The identification aids in the diagnosis of diseases caused by herpes simplex viruses and provides epidemiological information on these diseases. Herpes simplex viral infections range from common and mild lesions of the skin and mucous membranes to a severe form of encephalitis (inflammation of the brain). Neonatal herpes virus infections range from a mild infection to a severe generalized disease with a fatal outcome.
Special Controls
*Classification.* Class II (special controls). The device is classified as class II (special controls). The special control for the device is FDA's revised guidance document entitled “Class II Special Controls Guidance Document: Herpes Simplex Virus Types 1 and 2 Serological Assays.” For availability of the guidance revised document, see § 866.1(e).
Predicate Devices
Western Blot
Submission Summary (Full Text)
{0}
13
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY DEVICE ONLY TEMPLATE
A. 510(k) Number: K033105
B. Purpose for Submission: This submission follows a Pre-IDE protocol with a premarket notification submitted and given a final review on July 9, 2004 after response to a desk hold on June 23, 2004.
C. Analyte: Captia™ HSV 1 Type Specific IgG
D. Type of Test: Enzyme-linked Immunosorbent Assay (ELISA)
E. Applicant: Trinity Biotech USA
F. Proprietary and Established Names: Trinity Biotech Captia™ Herpes Simplex Virus (HSV) 1 Type Specific IgG
G. Regulatory Information:
1. Regulation section: 21 CFR 866.3305
2. Classification: III
3. Product Code: MXJ
4. Panel: 83
H. Intended Use:
1. Intended use and Indications for Use: The Trinity Biotech Captia™ Herpes Simplex Virus (HSV) 1 Type Specific IgG kit is an Enzyme-linked Immunosorbent Assay (ELISA) intended for qualitatively detecting the presence or absence of human IgG class antibodies to HSV-1 in human sera. In conjunction with the Trinity Biotech Captia™ Herpes Simplex Virus (HSV) 2 Type Specific IgG kit, the test is indicated for testing sexually active adults or expectant mothers for aiding in the presumptive diagnosis of HSV infection.
2. Special condition for use statement(s): The performance of this assay has not been established for use in a pediatric population, for neonatal screening, for testing of immunocompromised patients, or for use with automated equipment.
3. Special instrument Requirements: N/A
I. Device Description: Enzyme-Linked Immunosorbent Assays (ELISA) rely on the ability of biological materials (i.e., antigens) to adsorb to plastic surfaces such as polystyrene (solid phase). When antigens bound to the solid phase are brought into contact with a patient's serum, antigen specific antibody, if present, will bind to the
{1}
Page 2 of 14
antigen on the solid phase forming antigen-antibody complexes. Excess antibody is removed by washing. This is followed by the addition of goat anti-human IgG conjugated with horseradish peroxidase, which then binds to the antibody-antigen complexes. The excess conjugate is removed by washing, followed by the addition of chromogen/substrate, Tetramethylbenzidine (TMB). If specific antibody to the antigen is present in the patient's serum, a blue color develops. When the enzymatic reaction is stopped with 1N H₂SO₄, the contents of the wells turn yellow. The color, which is indicative of the concentration of antibody in the serum, can be read on a suitable spectrophotometer or ELISA microwell plate reader.
## J. Substantial Equivalence Information:
1. Predicate device name(s): Focus Technologies HerpeSelect® 1 ELISA IgG
2. Predicate K number(s): K021429
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| 1. Both utilize the type-specific immunoglobulins g (HSV gG1 IgG). | Captia™ HSV 1 Type-Specific IgG | Focus Technologies HerpeSelect® 1 ELISA IgG |
| 2. Both use recombinant antigens. | | |
| 3. Both did cross-reactivity testing on only taxonomically related viruses | | |
| 4. Both were adequately tested with sexually active adults and expectant mothers | | |
| 5. Both are intended for presumptive diagnosis. | | |
| 6. Both are limited with testing on pediatric, neonatal or immunocompromised populations | | |
| 7. Both are compared to Western blot. | | |
| 8. Both incubate serum for 30 minutes | | |
| 9. The substrate is TMB and the stop is H₂SO₄ for both. | | |
| 10. Both use a high and low positive and negative control along with a cutoff calibrator. | | |
14
{2}
Page 3 of 14
| 11. The index/ISR values for interpretation of results are the same for both (adjusted per lot number for each calibrator). | | |
| --- | --- | --- |
| Differences | | |
| Item | Device | Predicate |
| 1. Trinity did not compare their device to automated methodology where Focus did. | Captia™ HSV 1 Type-Specific IgG | Focus Technologies HerpeSelect® 1 ELISA IgG |
| 2. Trinity incubates substrate for 15 minutes and Focus for 10 minutes | | |
| 3. Trinity recommends washing 3X and Focus 5X. | | |
| 4. Trinity uses air as a blank and Focus uses the sample diluent as such. | | |
| 5. Both assays are to be read at 450 nm but Trinity also incorporates a dual wavelength of 650 nm. | | |
J. Standard/Guidance Document Referenced (if applicable): N/A
K. Test Principle: HSV 1 and HSV 2 have approximately 50% sequence homology and show considerable cross-reactivity. The Trinity Biotech HSV 1 Type Specific IgG ELISA uses a recombinant glycoprotein g, which is type specific for HSV 1. This allows for a rapid and less expensive sero-diagnosis of HSV 1 infection than virus isolation techniques (9, 10, 11, 12).
The Trinity Biotech HSV 1 Type Specific IgG kit utilizes the ELISA technology where a purified recombinant HSV 1 antigen is bound to the wells of a microplate. A peroxidase coupled anti-human IgG conjugate is used as the detection system. When rHSV1 antigens bound to the solid phase are brought into contact with a patient's serum, antigen specific antibody, if present, will bind to the antigen on the solid phase forming antigen-antibody complexes. Excess antibody is removed by washing. This is followed by the addition of goat anti-human IgG conjugated with horseradish peroxidase, which then binds to the antibody-antigen complexes. The excess conjugate is removed by washing, followed by the addition of chromogen/substrate, Tetramethylbenzidine (TMB). If specific antibody to the antigen is present in the patient's serum, a blue color develops. When the enzymatic reaction is stopped with 1N H₂SO₄, the contents of the wells turn yellow. The color, which is indicative of the concentration of antibody in the serum, can be read on a suitable spectrophotometer or ELISA microwell plate reader.
15
{3}
Page 4 of 14
# M. Performance Characteristics (if/when applicable):
## 1. Analytical performance:
### a. Precision/Reproducibility:
Table 1
HSV 1 Type Specific IgG Intra and Inter Assay Precision
Study Site 1
| Serum ID | Assay 1 (n=10) | | | Assay 2 (n=10) | | | Assay 3 (n=10) | | | Inter Assay (n=30) | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | X | S.D. | CV% | X | S.D. | CV% | X | S.D. | CV% | X | S.D. | CV% |
| 1 | 2.84 | 0.109 | 4% | 2.71 | 0.239 | 9% | 2.81 | 0.152 | 5% | 2.79 | 0.179 | 6% |
| 2 | 5.28 | 0.113 | 2% | 5.08 | 0.280 | 6% | 5.36 | 0.146 | 3% | 5.24 | 0.222 | 4% |
| 3 | 3.54 | 0.077 | 2% | 3.40 | 0.073 | 2% | 3.62 | 0.096 | 3% | 3.52 | 0.122 | 3% |
| 4 | 2.27 | 0.117 | 5% | 2.11 | 0.107 | 5% | 2.23 | 0.067 | 3% | 2.20 | 0.118 | 5% |
| 5 | 0.15 | 0.039 | 26% | 0.12 | 0.027 | 23% | 0.12 | 0.013 | 11% | 0.13 | 0.032 | 24% |
| 6 | 0.09 | 0.036 | 42% | 0.05 | 0.015 | 32% | 0.03 | 0.007 | 21% | 0.06 | 0.031 | 55% |
Table 2
HSV 1 Type Specific IgG Intra and Inter Assay Precision
Study Site 2
| Serum ID | Assay 1 (n=10) | | | Assay 2 (n=10) | | | Assay 3 (n=10) | | | Inter Assay (n=30) | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | X | S.D. | CV% | X | S.D. | CV% | X | S.D. | CV% | X | S.D. | CV% |
| 1 | 3.40 | 0.109 | 3% | 3.32 | 0.196 | 6% | 2.75 | 0.118 | 4% | 3.16 | 0.141 | 4% |
| 2 | 6.44 | 0.261 | 4% | 6.41 | 0.261 | 4% | 6.30 | 0.337 | 5% | 6.39 | 0.286 | 4% |
| 3 | 4.22 | 0.127 | 3% | 4.20 | 0.136 | 3% | 4.01 | 0.098 | 2% | 4.16 | 0.121 | 3% |
| 4 | 2.70 | 0.091 | 3% | 2.68 | 0.097 | 4% | 2.67 | 0.081 | 3% | 2.68 | 0.090 | 3% |
| 5 | 0.20 | 0.014 | 7% | 0.19 | 0.011 | 6% | 0.18 | 0.009 | 5% | 0.19 | 0.011 | 6% |
| 6 | 0.24 | 0.032 | 13% | 0.23 | 0.030 | 13% | 0.23 | 0.031 | 13% | 0.24 | 0.025 | 13% |
Table 3
HSV 1 Type Specific IgG Intra and Inter Assay Precision
Study Site 3
| Serum ID | Assay 1 (n=10) | | | Assay 2 (n=10) | | | Assay 3 (n=10) | | | Inter Assay (n=30) | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | X | S.D. | CV% | X | S.D. | CV% | X | S.D. | CV% | X | S.D. | CV% |
| 1 | 2.82 | 0.107 | 4% | 2.75 | 0.101 | 3% | 3.05 | 0.141 | 4% | 2.87 | 0.172 | 6% |
| 2 | 5.31 | 0.103 | 2% | 5.07 | 0.237 | 4% | 5.46 | 0.127 | 2% | 5.28 | 0.228 | 4% |
| 3 | 3.47 | 0.063 | 2% | 3.25 | 0.186 | 5% | 3.51 | 0.207 | 6% | 3.41 | 0.196 | 6% |
| 4 | 2.08 | 0.066 | 3% | 2.08 | 0.069 | 3% | 2.12 | 0.056 | 2% | 2.09 | 0.064 | 3% |
| 5 | 0.19 | 0.040 | 2% | 0.22 | 0.020 | 9% | 0.22 | 0.006 | 3% | 0.21 | 0.029 | 14% |
| 6 | 0.17 | 0.013 | 8% | 0.17 | 0.007 | 4% | 0.16 | 0.010 | 6% | 0.16 | 0.011 | 7% |
Table 4
HSV 1 Type Specific IgG Inter Site Precision
| Sample ID | Site 1 | Site 2 | Site 3 | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | X | X | X | X | S.D. | C.V. |
| 1 | 2.79 | 3.16 | 2.87 | 2.94 | 0.195 | 7% |
| 2 | 5.24 | 6.39 | 5.28 | 5.64 | 0.653 | 12% |
| 3 | 3.52 | 4.16 | 3.41 | 3.70 | 0.405 | 11% |
| 4 | 2.20 | 2.68 | 2.09 | 2.32 | 0.314 | 14% |
| 5 | 0.13 | 0.19 | 0.21 | 0.18 | 0.042 | 24% |
| 6 | 0.06 | 0.24 | 0.16 | 0.15 | 0.090 | 59% |
X = Mean ISR Value
S.D. = Standard Deviation
C.V. = Coefficient of Variation
16
{4}
Page 5 of 14
b. Linearity/assay reportable range: N/A
c. Traceability, Stability, Expected values (controls, calibrators, or method):
A CDC serum panel was obtained and tested by Trinity Biotech. The results of this testing are presented as a means to convey further information on the performance of this assay with a masked, characterized serum panel. This does not imply an endorsement of the assay by the CDC.
The panel consisted of 58.0% HSV 1 positive and 42.0% HSV 1 negative specimens. The Trinity Biotech HSV 1 Type Specific IgG ELISA demonstrated 96.0% total agreement with the CDC results. Of the results obtained by Trinity Biotech, there was 93.1% agreement with all HSV 1 positive specimens (This includes sera that are positive for HSV 1 only and sera that are positive for both HSV 1 and HSV 2). There was 100.0% agreement with the specimens that were HSV 1 positive only and 100.0% agreement with the specimens that were HSV negative for both.
d. Type Specificity with HSV 2 Western Blot Positives
An outside investigator at a Pacific Northwest University assessed the type specificity using HSV 2 Western Blot positive and HSV 1 Western Blot negative sera from the above described populations (n = 56): expectant mothers, sexually active adults, low prevalence persons, and HSV 1 culture positives. Of 56 HSV 2 Western Blot positive and HSV 2 Western Blot negative samples, Trinity ELISA was 54 negative and 2 positive.
Type Specificity with HSV 2 Western Blot Positives (n = 56)
| Characteristic | % (EL/WB) | 95% Confidence Interval (CI) |
| --- | --- | --- |
| Type-specificity relative to WB | 96.4% (54/56) | 87.7-99.6% |
| Type cross-reactivity relative to WB | 3.6% (2/56) | 0.43-12.3% |
e. Detection limit: N/A
f. Analytical specificity: Cross-Reactivity with Taxonomically Related Viruses
A study was performed by the manufacturer to determine the cross-reactivity of the Trinity Biotech HSV 1 Type Specific IgG ELISA test with 31 sera containing IgG antibody to taxonomically similar viruses including Cytomegalovirus (CMV), Varicella-Zoster Virus (VZV), and Epstein-Barr Virus (EBV). Of the 31 sera, 5 tested positive for CMV IgG by ELISA, 25 tested positive for VZV IgG by ELISA, and 30 tested positive for EBV IgG by ELISA. All 31 sera were negative by the Trinity Biotech HSV 1 Type Specific IgG ELISA indicating that antibodies to these viruses do not cross-react with the Trinity Biotech HSV 1 Type Specific IgG ELISA. Consequently, because adequate cross-reactivity testing was not performed the sponsor has added a Warning in the Cross-reactivity section of the PI to state, “Because minimal cross-reactivity performance testing
17
{5}
6 of 14
was
| Characteristic | % (TBU ELISA / Alt. ELISA) | 95% Confidence Interval (CI) |
| --- | --- | --- |
| Percent Positive Agreement | 93.88 % (92 / 98) | 87.2 – 97.7% |
| Percent Negative Agreement | 97.06 % (99 / 102) | 91.6 – 99.4 % |
| Percent Agreement | 95.50 % (191 / 200) | 91.6 – 97.9 % |
done, each laboratory should consider performing testing on taxonomically related viruses and viruses which could cause a syndrome similar to HSV such as HPV and gonorrhea. The levels tested should exceed ISR values of 3.10."
g. Assay cut-off: Because this is a qualitative assay, low level performance testing was done as reproducibility. See above, M.1.a. In addition, 48 HSV Type 1 negative sera were assayed by the HSV 1 IgG Type Specific ELISA test. The mean and standard deviation of the optical density readings for the sera was 0.155 and 0.073 respectively. The positive threshold for the assay was determined by adding the mean and 2.5 standard deviations (0.155 + 2.5 (0.073) = 0.338). A positive serum was titrated to give a constant ratio of the threshold value to obtain a calibrator sera. On all subsequent assays, this sera was run and the assay calibrated by multiplying the O.D. value for the calibrator by the ratio to the cut off to obtain the cut off O.D. This value was then divided into the O.D. for the patient sera to obtain an index value. By definition, the cut off index is equal to 1.00. To account for inherent variation in the immunoassay, values of 0.91 – 1.09 were considered equivocal. Therefore values ≤ 0.90 are considered negative and the values ≥ 1.10 are considered positive. Analytical validation was performed using levels 15% above and 30% below the cut off.
2. Comparison studies:
a. Method comparison with predicate device:
% Agreement Positive and % Agreement Negative to Alternate HSV 1 Type Specific IgG ELISA
An outside investigator at a Pacific Northwest University assessed the % agreement positive and % agreement negative of the Trinity Biotech Captia™ HSV 1 Type Specific IgG kit and an alternate HSV 1 type specific IgG ELISA test with 200 prospective, unselected, sequentially submitted specimens.
| Prospectively Collected, Sequential Sera | | Alternate HSV 1 Type Specific IgG | | |
| --- | --- | --- | --- | --- |
| | | + | - | E |
| TrinityBiotechCaptia
HSV 1 Type Specific | + | 92 | 3 | 0 |
| | - | 6 | 99 | 0 |
| | E | 0 | 0 | 0 |
b. Matrix comparison: N/A since this is intended for serum only.
7. Clinical studies:
a. Clinical sensitivity:
i) % Agreement Positive and % Agreement Negative with Expectant Mothers (n = 210†)
18
{6}
Page 7 of 14
An outside investigator assessed the % agreement positive and % agreement negative with consented, coded, unselected, banked and masked sera from expectant mothers (n = 210). The reference method was an HSV 1 Western Blot (WB) from a Pacific Northwest university. Of 155 WB positives, Trinity ELISA was 136 positive, 18 negative and 1 equivocal. Of 55 WB negatives, Trinity ELISA was 54 negative and 1 positive.
| Characteristic | % (EL/WB)* | 95% Confidence Interval (CI) |
| --- | --- | --- |
| % agreement positive to WB | 88.31% (136/154) | 83.2-93.4%‡ |
| % agreement negative to WB | 98.18% (54/55) | 90.3-100.0% |
| | + | |
* Excludes one ELISA equivocal
† The word “% agreement” refers to comparing this assay’s results with those of a similar assay. No attempt was made to correlate the assay results to disease presence or absence. No judgment can be made on the similar assay’s accuracy in predicting disease.
‡ 95% CI calculated using the normal approximate method.
ii) % Agreement Positive and % Agreement Negative with Sexually Active Adults (n = 198)†
An outside investigator assessed the % agreement positive and % agreement negative with consented, unselected and masked sera from sexually active adults over the age of 14 (n = 198). The reference method was an HSV 1 Western Blot (WB) from a Pacific Northwest university. Of 116 WB positives, Trinity ELISA was 102 positive and 14 negative. Of 80 WB negatives, Trinity ELISA was 80 negative.
| Characteristic | % (EL/WB)* | 95% Confidence Interval (CI) |
| --- | --- | --- |
| % agreement positive to WB | 87.93% (102/116) | 82.0-93.9%‡ |
| % agreement negative to WB | 100.00% (80/80) | 95.5-100.0% |
Excludes two atypical Western Blots.
† The word “% agreement” refers to comparing this assay’s results with those of a similar assay. No attempt was made to correlate the assay results to disease presence or absence. No judgment can be made on the similar assay’s accuracy in predicting disease.
‡ 95% CI calculated using the normal approximate method.
iii) % Agreement Positive and % Agreement Negative with a Low Prevalence Population (n = 184)†
An outside investigator assessed the % agreement positive and % agreement negative with unselected, banked and masked sera from a low prevalence population (n = 184). The reference method was an HSV 1 Western Blot (WB) from a Pacific Northwest university. Of 131 WB negatives, Trinity ELISA was 128 negative, 1 positive and 2 equivocal. Of 53 WB positives, Trinity ELISA was 42 positive, 8 negative and 3 equivocal.
| Characteristic | % (EL/WB)* | 95% Confidence Interval (CI) |
| --- | --- | --- |
| % agreement positive to WB | 84.00% (42/50) | 70.9-92.8% |
| % agreement negative to WB | 99.22% (128/129) | 95.8-100.0% |
(9)
{7}
Page 8 of 14
* Excludes five ELISA equivocals.
† The word "% agreement" refers to comparing this assay's results with those of a similar assay. No attempt was made to correlate the assay results to disease presence or absence. No judgment can be made on the similar assay's accuracy in predicting disease.
iv) % Agreement Positive with Culture Positives†
An outside investigator assessed the % agreement positive using unselected, retrospective and masked sera from patients that were at least six weeks but not more than one year post clinical presentation and culture HSV 1 positive (n = 53). Reference methods included culture (infection) and an HSV 1 Western Blot (WB) (antibody) from a Pacific Northwest university. Of 53 culture positives: 1) Trinity ELISA was 37 positive, 12 negative and 4 equivocal and, 2) WB was 44 positive and 9 negative. Of 44 WB positives: Trinity ELISA was 36 positive, 6 negative, and 2 equivocal.
% Agreement Positive with Culture Positives (n = 53)†
| Characteristic | % (EL/WB or Culture) | 95% Confidence Interval (CI) |
| --- | --- | --- |
| % agreement positive to culture | 75.51% (37/49)* | 61.1-86.7% |
| % agreement positive to WB | 85.71% (36/42)** | 71.5-94.6% |
Excludes four ELISA equivocals.
** Excludes two ELISA equivocals.
† The word "% agreement" refers to comparing this assay's results with culture, considered the gold standard.
b. Clinical specificity: Specificity testing was performed with only taxonomically related viruses and a limitation added to the product insert.
4. Clinical cut-off: There was only an analytical cutoff determination performed.
5. Expected values/Reference range: Pending receipt of appropriate information.
N. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
20
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.