HARVARD 1 SYRINGE PUMP/HARVARD 1 WITH PCA SYRINGE PUMP
K033054 · Harvard Clinical Technology · MEA · Dec 16, 2003 · General Hospital
Device Facts
Record ID
K033054
Device Name
HARVARD 1 SYRINGE PUMP/HARVARD 1 WITH PCA SYRINGE PUMP
Applicant
Harvard Clinical Technology
Product Code
MEA · General Hospital
Decision Date
Dec 16, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
Harvard 1 The Harvard 1 is designed to provide, intravenous, intra-arterial, or epidural delivery of drugs, blood or blood products or other parenteral fluids when administered by a health care professionals such as physicians and nurses. The Harvard 1 is compatible for use inside the MRI room when mounted outside the 150 gauss line with MRI systems of 1.5 tesla or less. Harvard 1 with PCA Harvard 1 with PCA is designed provide patient controlled analgesia, intravenous, intra-arterial, or epidural delivery of drugs, blood or blood products, or other parenteral fluids when administered by a health care professionals such as physicians and nurses. The Harvard 1 with PCA is compatible for use inside the MRI room when mounted outside the 150 gauss line with MRI systems of 1.5 tesla or less.
Device Story
Syringe infusion pump for delivery of parenteral fluids/drugs; accommodates 1-60 ml syringes. Features active matrix color LCD, rotary controls for operation/data entry. Microprocessor-based system: master controller, motor/sensor pump processor, system supervisor. Sensors detect syringe size, plunger capture, occlusion force; calculates infusion rate/distance. PCA version includes pendant/lock for patient-controlled analgesia. Optional PC-based 'Harvard Library' software downloads drug-specific parameters (concentrations, rates, bolus) to flash memory. Optional laser barcode reader scans syringe labels to select drug parameters, reducing input errors. Used in clinical environments, including MRI rooms (if outside 150 gauss line). Provides bidirectional RS232 communication. Alarms for occlusion, empty syringe, system faults, battery status. Benefits: precise drug delivery, reduced medication errors via library/barcode, MRI compatibility.
Clinical Evidence
Bench testing only. Performance testing conducted on prototype pumps reflecting final design configuration. Gravimetric rate accuracy and occlusion system testing performed per IEC-601-2-24. Results indicated mean error within specified device range.
Technological Characteristics
Syringe infusion pump; 1-60 ml syringe capacity. Microprocessor-controlled motor drive. Sensors: syringe size, plunger capture, occlusion force. Interface: color LCD, rotary knobs. Connectivity: RS232 serial port. Optional laser barcode reader. Flash memory for drug library. MRI compatible (1.5T or less, outside 150 gauss line). Power: AC/Battery. Sterilization: Not specified.
Indications for Use
Indicated for intravenous, intra-arterial, or epidural delivery of drugs, blood, blood products, or parenteral fluids in clinical settings. Includes patient-controlled analgesia (PCA) for pain management. Administered by healthcare professionals (physicians/nurses). Compatible with 1.5T or less MRI systems when mounted outside the 150 gauss line.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
Predicate Devices
Harvard 2 Syringe Pump
Medex Protégé
B. Braun Horizon with Dose Scan
Baxter PCA-II
Submission Summary (Full Text)
{0}------------------------------------------------
K033054
36
DEC 1 6 2003
# 510(K) SUMMARY
Harvard Clinical Technology Inc. 22 Pleasant Street South Natick MA 01760
# Contact Person:
Eric J. Flachbart Vice President, Research and Development Harvard Clinical Technology, Inc. 22 Pleasant Street South Natick, MA 01760 (508)-655-2000 (508)-655-2211 Fax eflachbart@harvardclinical.com
Date Prepared: July 23, 2003
Trade Name: Harvard 1 Syringe Pump, Harvard 1 with PCA Pump Common Name: Syringe Infusion Pump Classification Name: Infusion Pump
## Predicate Devices
Harvard 2 Syringe Pump, Medex Protégé, B. Braun Horizon with Dose Scan, Baxter PCA-II
#### Intended Use
### Harvard 1
The Harvard 1 is designed to provide, intra-arterial, or epidural delivery of drugs, blood or blood products or other parenteral fluids when administered by a health care professionals such as physicians and nurses.
The Harvard 1 is compatible for use inside the MRI room when mounted outside the 150 gauss line with MRI systems of 1.5 tesla or less.
### Harvard 1 with PCA
Harvard with PCA to provide patient controlled analgesia, intra-arterial, or epidural delivery of drugs, blood or blood products, or other parenteral fluids when administered by a health care professionals such as physicians and nurses.
The Harvard 1 with PCA is compatible for use inside the MRI room when mounted outside the 150 gauss line with MRI systems of 1.5 tesla or less.
Harvard 1 / Harvard 1 with PCA Premarket Notification - 23 July, 2003
{1}------------------------------------------------
# Device Description:
The Harvard Pump is intended for the delivery of parenteral fluids. It accommodates syringe sizes from 1 through 60 ml from multiple manufacturers.
Its user interface consists of an active matrix color LCD display with two rotary knobs for controling pump operation. The rotary Data Entry knob provides scrolling and selection of data and memu items as well as state selection. The knob is turned to scroll, and pressed for selection. The rotary Function knob controls the state (Purge, Setup, Stop, Run and Bolus) of the pump.
The pump has several microprocessors, one master which controls operation of the device, one pump processor which controls the operation of the motor and sensors specific to the syringe drive mechanism, and one supervisor which monitors the status of the system.
The pump has sensors which permit its microprocessors to determine the syringe size loaded, determine if the syringe plunger is securely captured by the pusher block assembly, measure occlusion force and calculate the proper rate and distance the plunger of the syringe must travel to cause an infusion of fluid at a given rate and volume.
The pump provides for bidirectional remote communications via an RS232 serial port.
The Harvard 1 with PCA extends the Harvard 1 functionality to include Patient Controled Analgesia capability through the addition of an optional PCA pendent and PCA cover and lock assembly. PCA capability allows the patient to request delivery of a dose of pain medication from a regimen prescribed by the physician.
The pump may be used with The Harvard Library, an optional computer program which runs on MS Windows based personal computers. The Harvard Library is capable of downloading and uploading drug information in the form of a drug library to and from the pump. After downloading, the drug library is resident in flash memory within the pump, enabling the pump to provide drug specific defaults for drug concentrations, pumping rates, bolus amounts, and bolus times.
Barcode versions of the pump provide a laser scanning barcode reader is used to scan a label placed on the syringe which selects a drug from the pump's resident drug library. This allows the pump to retrieve from its internal memory the drug's specific parameters such as concentration, infusion rate etc. In an effort to minimize user input errors.
{2}------------------------------------------------
| Warnings / Status | Purge | Purge | Purge | Purge | Purge<br>Stop | Infusion near end | | Hold time exceeded |
|-------------------------------------------|-------------------------------------------|----------------------------------------|----------------------------------------|----------------------------------------|---------------|-----------------------|--|--------------------|
| Setup | Setup | Setup | Setup | Setup | | Infusion end | | Low battery |
| Stop | Stop | Stop | Stop | Stop | | Hi-pressure-Occlusion | | Low flow from |
| Run | Run | Run | Run | Run | | System Malfunction | | |
| Bolus | Bolus | Bolus | Bolus | Bolus | | Syringe not capture | | |
| AC Power | AC Power | AC Power | AC Power | AC Power | | Plunger disengaged | | |
| Battery Charging | Battery Charging | Battery Charging | Battery Charging | Battery Charging | | Low Battery | | |
| Remote | Remote | Remote | Remote | Remote | | | | |
| Syringe will be empty<br>in 15 minutes | Syringe will be empty<br>in 15 minutes | Syringe will be empty<br>in 15 minutes | Syringe will be empty<br>in 15 minutes | Syringe will be empty<br>in 15 minutes | | | | |
| Battery will be depleted<br>in 30 minutes | Battery will be depleted<br>in 30 minutes | | | Low Battery | | | | |
.
Harvard 1 / Harvard 1 with PCA Premarket Notification - 23 July, 2003
{3}------------------------------------------------
| | Harvard 1 | Harvard with PCA | Harvard 2 | Baxter PCA II | Protégé 3010 | B. Braun<br>Horizon Lite |
|--------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|
| Alarms | Occlusion<br>Battery will be depleted<br>in 5 minutes<br>Syringe plunger not<br>captured<br>Syringe barrel<br>Pusher moved<br>Syringe Empty<br>System Fault<br>System Fault - Battery<br>Depleted | Occlusion<br>Battery will be depleted<br>in 5 minutes<br>Syringe plunger not<br>captured<br>1 hour limit reached<br>Syringe barrel<br>Pusher moved<br>Syringe Empty<br>System Fault<br>System Fault - Battery<br>Depleted | Occlusion<br>Battery will be depleted<br>in 5 minutes<br>Syringe plunger not<br>captured<br>Syringe barrel<br>Pusher moved<br>Syringe Empty<br>System Fault<br>System Fault - Battery<br>Depleted | Occlusion<br>Door open<br>Door unlocked<br>1 hour limit reached<br>Syringe plunger not<br>captured<br>Syringe empty<br>Battery depleted<br>System Fault | Infusion end<br>Hi-pressure-Occlusion<br>System Malfunction<br>Syringe not capture<br>Plunger disengaged<br>Low Battery | Air-in-line<br>Container empty<br>Door open<br>Downstream occlusion<br>container<br>System error<br>Upstream occlusion<br>Set improperly loaded |
| Inputs | Concentration<br>Rate<br>Bolus amount<br>Bolus time<br>Syringe manufacturer<br>Syringe size3<br>Patient weight<br>Dose amount<br>Dose time<br>Number of doses<br>Dose interval | Concentration<br>Rate<br>Bolus amount<br>Bolus time<br>Syringe manufacturer<br>Syringe size<br>Patient weight<br>Dose amount<br>Dose time<br>Number of doses<br>Dose interval<br>PCA dose<br>1 hour limit | Concentration<br>Rate<br>Bolus amount<br>Bolus time<br>Syringe manufacturer<br>Syringe size<br>Patient weight<br>Dose amount<br>Dose time<br>Number of doses<br>Dose interval | Concentration<br>Basal Rate<br>Bolus Amount<br>PCA Amount<br>1 Hour limit<br>PCA Delay time<br>Number of PCA does<br>Dose interval | Concentration<br>Rate<br>Bolus amount<br>Bolus time<br>Syringe manufacturer<br>Syringe size<br>Patient weight<br>Dose amount<br>Dose time<br>Number of doses<br>Dose interval | Concentration<br>Rate<br>Bolus amount<br>Bolus time<br>Patient weight<br>Dose amount<br>Dose time<br>Number of doses<br>Dose interval |
| Serial<br>Communications | Bidirectional | Bidirectional | Bidirectional | None | Bidirectional | Unknown |
| Barcode Reader | Optional laser scanning<br>reader | Optional laser scanning<br>reader | None | None | None | Laser scanning reader |
| MRI<br>Compatibility | Yes | Yes | No | No | Yes | No |
<sup>3</sup> Detected by syringe size mechanism and confirmed by user.
Harvard 1 / Harvard 1 with PCA Premarket Notification - 23 July, 2003
{4}------------------------------------------------
Performance testing using prototype pumps which functionally reflect the final design configuration was performed to substantial
equivalence claims.
Cravimetric Rate Accuracy and occlusion system
cesting per IRC-601-2-24 - Particular requirement
cesting per IEC-601-2-24 - Particular requirement
for safety of infusion pump
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
slusion test results indicate a mean error within
specified range of the device.
arvard 1 / Harvard 1 with PCA Premarket Notification - 23 July, 200
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three lines forming its body and wings. The eagle is enclosed in a circle, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the top half of the circle.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 1 6 2003
Harvard Clinical Technology Ms. Susan Gill Responsible Third Party Official Underwriters Laboratories, Incorporated 12 Laboratory Drive Research Triangle Park, North Carolina
Re: K033054
Trade/Device Name: Harvard 1 Syringe Pump/ Harvard 1 with PCA Syringe Pump Regulation Number: 880.5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: MEA, FRN Dated: December 1, 2003 Received: December 2, 2003
Dear Ms. Gill:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4618. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely vours,
Carl
Chiu Lin, Ph.D.
hiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# INDICATIONS FOR USE
Harvard Clinical Technology, Inc
510(k) Number (if known): Device Name:
Ko33054 Harvard 1 Harvard 1 with PCA Syringe Pumps
Indications For Use:
Harvard 1
The Harvard 1 is designed to provide, intravenous, intra-arterial, or epidural delivery of drugs, blood or blood products or other parenteral fluids when administered by a health care professionals such as physicians and nurses.
The Harvard 1 is compatible for use inside the MRI room when mounted outside the 150 gauss line with MRI systems of 1.5 tesla or less.
Harvard 1 with PCA
Harvard 1 with PCA is designed provide patient controlled analgesia, intravenous, intra-arterial, or epidural delivery of drugs, blood or blood products, or other parenteral fluids when administered by a health care professionals such as physicians and nurses.
The Harvard 1 with PCA is compatible for use inside the MRI room when mounted outside the 150 gauss line with MRI systems of 1.5 tesla or less.
Patrizio Curante
Division Sign-Off) (Division Sign-Olf)
Division of Anesthesiology, General Hospit Infection Control, Dental De
Infection Control:
510(k) Number: K033054
510(k) Number.
Prescription Use
(Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF
NEEDED)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.