The ViziLite Blue Oral Exam Kit with ViziLite and The ViziLite Blue Oral Lesion ldentification and Marking System are intended for use only as adjuncts to the conventional oral mucosa examination with incandescent light. The ViziLite examination, using chemiluminescent light, is intended only as an adjunct to traditional oral examination by incandescent light to increase identification, evaluation, and monitoring of oral mucosal abnormalities in populations at increased risk for oral cancer. The ViziLite Blue Oral Lesion Identification and Marking System, as supported by clinical trials, is an adjunct or an accessory to the ViziLite examination and assists in the further work up of ViziLite-Identified white oral mucosal lesions for patients at increased risk for oral cancer.
Device Story
ViziLite Blue Oral Exam Kit is a two-part system for oral mucosal examination. First, ViziLite (OralLite) provides a diffuse chemiluminescent light source used after a pre-exam acetic acid rinse to enhance visualization of atypical or dysplastic mucosal abnormalities (acetowhite lesions) that contrast with surrounding tissue. Second, the ViziLite Blue Oral Lesion Identification and Marking System uses a metachromatic vital tissue dye (Tolonium chloride) to stain ViziLite-identified lesions. The dye allows the clinician to remove the ViziLite device and continue visualizing the lesion under standard incandescent light to assess borders, measure size, or perform a biopsy. Used in clinical settings by dentists or physicians. The device does not replace conventional screening or biopsy; it serves as an adjunct to assist clinical decision-making regarding lesion management. It is a single-use, disposable device.
Clinical Evidence
Clinical trials conducted at three university hospital sites involved 86 adult patients with a history of leukoplakia and 92 biopsied lesions. Results demonstrated that the ViziLite device improved the examiner's ability to visualize mucosal changes of leukoplakia and erythroleukoplakia compared to projected incandescent light alone. The Tolonium chloride marking system successfully stained 51% of ViziLite-identified white lesions, allowing clinicians to visualize and evaluate these lesions under incandescent light after removing the ViziLite device.
Technological Characteristics
System includes a chemiluminescent light source (ViziLite), 1% acetic acid rinse, and a Tolonium chloride (vital tissue dye) swab. The light source is a single-use, disposable, non-toxic chemical light. The marking system is a three-component swab system. No electronic components or software algorithms are utilized; the device relies on chemical interaction (acetic acid dehydration) and optical contrast enhancement.
Indications for Use
Indicated for use by trained health care providers as an adjunct to conventional oral mucosal screening for identification, evaluation, and monitoring of oral mucosal abnormalities in patients at increased risk for oral cancer. The marking system is indicated for further evaluation of ViziLite-identified white lesions. Contraindicated in pregnant or lactating women, patients with liver or renal insufficiency, patients with difficulty swallowing, or patients with known hypersensitivity to ingredients.
Regulatory Classification
Identification
A dental operating light, including the surgical headlight, is an AC-powered device intended to illuminate oral structures and operating areas.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles an abstract human figure or a bird in flight, composed of three curved lines.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Dr. Martin Lonky CEO and President The Trylon Corporation 970 West 190th Street, Suite 850 Torrance, California 90502-1037
JAN 3 1 2005
Re: K033033
Trade/Device Name: ViziLite Blue Oral Examination Product Regulation Number: 884.1720 Regulation Name: Gynecologic Laparoscope and Accessories Regulatory Class: II Product Code: MPU, EAZ Dated: September 26, 2004 Received: September 26, 2004
Dear Dr. Lonky:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
#### Page 2 - Dr. Lonky
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Chiu Lin, Ph.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# . Indications for Use
510(k) Number (if known): K033033
Device Name: ViziLite Blue Oral Exam Product
Indications for Use: The ViziLite Blue Oral Exam Product consists of the Vizilite and ViziLite Blue Oral Lesion Identification and Marking System.
The Vizilite (OralLite) is a chemiluminescent light source system indicated for use as an adjunct to conventional oral mucosal screening by trained health care providers for the identification, evaluation, and monitoring of oral mucosal abnormalities in a population at increased risk for oral cancer.
The ViziLite Blue Oral Lesion Identification and Marking-System, is a threecomponent swab system which is indicated as an adjunct to the ViziLite Test for oral mucosa lesions, for further evaluation and monitoring of lesions by providing physical marking of lesions already differentially identified with Vizil ite in a population at increased risk for oral cancer
The ViziLite Blue Oral Lesion Identification and Marking System is not being proposed for use in the initial oral mucosal examination without initial lesion identification with ViziLite. Furthermore, this ViziLite Blue Oral Lesion Identification and Marking System is not intended to be used as an indicator of lesions warranting further study, including biopsy. Whether a lesion is marked with the dye or not should not alter the clinician's clinical behavior as dictated by the results of the ViziLite examination. The marking dye, when positive, acts as a lesion marker that allows for the removal of the ViziLite device while preserving the anatomic character of the lesion.
Prescription Use V AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| | Page 1 of 1 |
|--|-------------|
|--|-------------|
| General Hospital, |
|-------------------|
| |
| K Number | KC-33033 |
|----------|----------|
|----------|----------|
{3}------------------------------------------------
## XIII. SUMMARY OF SAFETY AND EFFECTIVENESS STATEMENT
#### DEVICE DESCRIPTION:
ViziLite Blue Oral Exam Kit includes the ViziLite used as an adjunct to visual examination of the oral mucosa with incandescent light for identification of oral mucosal abnormalities and ViziLite Blue Oral Lesion Identification and Marking System to further assist with the evaluation and monitoring of ViziLiteldentified oral white mucosal abnormalities in patients at increased risk for oral cancer.
The ViziLite-Blue™ Oral Exam Kit is a visualization system that is intended as an adjunct to conventional visual examination with incandescent light of oral mucosa and employs the same components and mechanism of action as the previously cleared ViziLite Test Kit (a.k.a. ViziLite Comprehensive Exam Tray K012070 and the OralLite Test Kit, K03995 and Acetic Acid Rinse) and is indicated for the same patient population and intended use. The ViziLite examination light source and accessories, including the 1% Acetic Acid Rinse, have not been changed nor significantly modified for production of the ViziLite-Blue Oral Exam Kit.
The ViziLite Blue Oral Lesion Identification and Marking System, consists of three swab components, two swabs of 1% Acetic Acid Rinse, including a postdye decolorizer, and one swab with a metachromatic vital tissue dye, Tolonium chloride, also called Toluidine Blue. The application of the dye to ViziLite-Identified oral white mucosal lesions in a population at increased risk for oral cancer, during clinical trials was able to assist with evaluation, monitoring, and tissue sampling and helped physically mark and stain 51% of the ViziLite-Identified white lesions to allow the health care provider to visualize lesions with incandescent light.
#### INTENDED USE:
The ViziLite Blue Oral Exam Kit with ViziLite and The ViziLite Blue Oral Lesion ldentification and Marking System are intended for use only as adjuncts to the conventional oral mucosa examination with incandescent light.
The ViziLite examination, using chemiluminescent light, is intended only as an adjunct to traditional oral examination by incandescent light to increase identification, evaluation, and monitoring of oral mucosal abnormalities in populations at increased risk for oral cancer.
The ViziLite Blue Oral Lesion Identification and Marking System, as supported by clinical trials, is an adjunct or an accessory to the ViziLite examination and assists in the further work up of ViziLite-Identified white oral mucosal lesions for patients at increased risk for oral cancer.
{4}------------------------------------------------
The ViziLite Blue Oral Lesion Identification and marking System is not intended to be used as an initial screening examination for identification and evaluation of oral mucosa abnormalities, nor is it intended to replace initial conventional screening under incandescent light, ViziLite examination or biopsy. The marking system is not intended to be used as an indicator of lesions warranting further study, including biopsy. Whether a lesion is marked with the dye or not should not alter the clinician's clinical behavior as dictated by the results of the ViziLite examination. The marking dye, when positive, acts as a lesion marker that allows for the removal of the ViziLite device while preserving the anatomic character of the lesion.
#### INDICATIONS FOR VIZILITE BLUE ORAL EXAM KIT:
The ViziLite Blue Oral Exam Kit consists of the Vizilite and ViziLite Blue Oral Lesion Identification and Marking System.
The Vizilite (OralLite) is a chemiluminescent light source system indicated for use as an adjunct to conventional oral mucosal screening by trained health care providers for the identification, evaluation, and monitoring of oral mucosal abnormalities in a population at increased risk for oral cancer.
The ViziLite Blue Oral Lesion Identification and Marking System, is a threecomponent swab system which is indicated as an adjunct to the ViziLite Test for oral mucosa lesions, for further evaluation and monitoring of ViziLite-Identified white lesions in a population at increased risk for oral cancer.
The ViziLite Blue Oral Lesion Identification and Marking System is not indicated to be used as an initial screening examination for identification and evaluation of oral mucosa abnormalities, nor is it intended to replace initial conventional screening under incandescent light and any of the clinician's best clinical judgment as to patient management. In addition, the marking system is not intended to be used as an indicator or lesions warranting further study, including biopsy. Whether a lesion is marked with the dye or not should not alter the clinician's clinical behavior as dictated by the results of the ViziLite examination. The marking dye, when positive, acts as a lesion marker that allows for the removal of the ViziLite device while preserving the anatomic character of the lesion.
#### ViziLite Blue Oral Exam PROCEDURE:
The ViziLite examination consists of visualization of leukoplakic areas using a diffuse chemiluminescent light source. Following conventional oral examination with incandescent light, there is a pre-exam rinse with ViziLite Rinse, activation of the ViziLite, with placement in the ViziLite Retractor, dimming of the ambient light and re-examination with ViziLite. Under ViziLite, atypical or dysplastic mucosal
{5}------------------------------------------------
abnormalities will appear bright white, distinctly demarcated, sharply marginated areas that contrast the surrounding non involved epithelium. Any lesion identified with ViziLite is further evaluated using the adjunctive ViziLite Blue Oral Lesion Identification and Marking System (pre and post acetic acid swabs and Tolonium Chloride stain swab).
#### PRINCIPLE OF ACTION:
#### ViziLite (OralLite):
Traditionally, following application of a cytoplasmic dehydration agent such as an acetic acid solution, acetowhite or leukoplakic lesions are seen with changes in refractile properties that occur in atypical nonkeratinized squamous epithelium due to an increased nuclear cytoplasmic ratio.
Adding diffuse chemiluminescent light (ViziLite (OralLite)/Speculite) to a conventional projected incandescent light exam has been clinically shown to increase the detection of biopsy proven squamous cell dysplasia and malignancy in squomous epithelium in the lower female genital tract when compared with detection by the naked eye and detection with magnified visualization with incandescent light.
This device has preiously been cleared for use as an adjunct to a conventional examination. (K 03995, K 012070)
#### ADVERSE EVENTS:
None Known.
#### CLINICAL STUDIES:
In IRB monitored clinical trials conducted at three University Hospital sites involving 86 adult patients with a history of leukoplakia and 92 biopsied lesions. the data continued to demonstrate that the ViziLite device made it easier for the examiner to see mucosal changes of leukoplakia and erythroleukoplakia when compared with projected incandescent light (as was demonstrated in K 003995). Thus, it appears that the ViziLite is a useful adjunct to help direct the examiner to lesions that would be difficult or poorly seen with incandescent light alone, but is non-specific with regard to underlying pathology.
The addition of the ViziLite Blue Oral Lesion Identification and Marking System using Tolonium Chloride dye retention by ViziLite-Identified acetowhite lesions provided additional information to the practitioner when examining patients at high risk for oral malignancy, particularly in patients with a prior history of squamous cell cancer or dysplasia. When the Tolonium Chloride dve marked the acetowhite lesions, the health care provider was able to remove the ViziLite
{6}------------------------------------------------
device from the patient's mouth and see the lesion with incandescent illumination. This was true whether the uptake of the stain had been diffuse (the entire lesion stains with the marker-dve) or the lesion had an irreqular or stippled uptake. The provider has the ability to choose to study these lesions with the aid of loupes, to futher determine lesion characteristics, or use the now freed up hand to perform a biopsy.
The usefulness of this adjunctive approach to lesions seen with the ViziLite device is in the assisting the practitioner in his/her decision to immediately bipsy (or refer for biopsy) an acetowhite lesion seen with ViziLite or deferring a definitive biopsy in favor of following the lesion using standard clinical observation techniques.
## CONTRAINDICATIONS:
The ViziLite (OralLite) examination is an adjunct to conventional oral examination with incandescent light and should not be used without first performing a conventional examination.
The ViziLite Blue Oral Lesion Identification and Marking System (with Tolonium chloride) is not intended to be used without prior traditional oral visual examination using ambient projected light followed by ViziLite (OralLite) examination.
The ViziLite Blue Oral Lesion Identification and Marking System (with Tolonium chloride) is not intended for "initial identification" of abnormalities of the oral mucosa, but is provided to the health care professional as an adjunct to the already cleared Vizilite (OralLite) examination for the evaluation, monitoring, and sampling of "Vizilite-Identified lesions" in patients identified as high-risk for oral cancer.
The ViziLite Blue Oral Lesion Identification and Marking System (with Tolonium chloride) is contraindicated in patients with a known history or hypersensitivity to any of the ingredients or their analogs.
The ViziLite Blue Oral Lesion Identification and Marking System (with Tolonium chloride) is contraindicated in patients who are pregnant or lactating.
Due to the lack of safety data, The ViziLite Blue Oral Lesion Identification and Marking System (with Tolonium chloride) should not be used in children, patients with liver or renal insufficiency, or patients with difficulty swallowing.
The ViziLite Blue Oral Lesion Identification and Marking System (with Tolonium chloride) should be used with caution in patients who may have difficultv
{7}------------------------------------------------
following directions during the lesion staining procedure (e.g. patients with severe physical or mental disabilities).
# WARNING/PRECAUTIONS - ViziLite
#### To prevent swallowing or choking:
The health care provider should ensure that the activated ViziLite is firmly inserted into the provided ViziLite (OralLite) Retractor before placing it into the patient's mouth.
The health care provider should hold firmly onto the ViziLite Retractor while it is placed inside the patient's mouth.
## To prevent potential leakage of ViziLite chemicals into the mouth:
Insect the ViziLite (OralLite) for any evidence of chemical leakage prior to capsule activation. Discard any capsule that does not appear to be intact.
Inspect the activated ViziLite (OralLite) for potential leakage before placing it into the patient's mouth for ViziLite (OralLite).
Do not use a capsule for ViziLite (OralLite) if the capsule does not appear to be intact or functioning properly.
[Note: The chemiluminescent chemicals and the materials used in the manufacturer of the ViziLite (OralLite) have been shown to be non-toxic in animal studies should they be either swallowed or applied to the epithelial surface. Therefore, the ViziLite (OralLite) should not present a significant risk to humans when used as a light source for oral examination under usual conditions. However, oral exposure to the contents of the ViziLite (OralLite) may cause transient irritation to the mouth, throat and gastrointestinal tract. I
[Remedy for accidental exposure to chemicals: Rinse mouth immediately and dilute with 4 to 8 ounces of milk or water. Decontamination with syrup of ipecac, activated charcoal or gastric lavage is not indicated.1
After 10 minutes the chemiluminescent light begins to fade and the ability to visualize white lesions will also decrease. Therefore, perform the examination of oral tissues within 10 minutes of ViziLite (OralLite) activation.
ViziLite (OralLite) is to be used as an adjunct to conventional oral examination. ViziLite (OralLite) is not intended to be used for grading acetowhite lesions.
All activated and/or used ViziLite's (OralLite) must be discarded in a proper receptacle per the procedures of the facility.
ViziLite (OralLite) is intended as a disposal, single use patient device.
{8}------------------------------------------------
ViziLite (OralLite) is not reusable.
## WARNINGS/PRECAUTIONS - ViziLite (OralLite) Rinse
ViziLite (OralLite) is intended as a disposal, single use patient device.
ViziLite (OralLite) is not reusable.
ViziLite (OralLite) Rinse is not intended to be swallowed.
ViziLite (OralLite) Rinse is intended for oral use only. Keep out of reach of children.
Do not refrigerate ViziLite (OralLite) Rinse.
# WARNINGS/PRECAUTIONS - The ViziLite Blue Oral Lesion Identification and Marking System
No severe adverse reactions are expected when The ViziLite Blue Oral Lesion ldentification and Marking System (with Tolonium chloride) is used according to package instructions. Some patients may find the taste of the product unappealing and may gag. Patients may also experience a slight burning sensation of the oral mucosa due to the acetic acid.
Removal oral prosthesis should be removed and any associated trauma or inflammation given time to heal prior to use of the product.
Patients should be informed prior to the ViziLite Blue Oral Lesion Identification and Marking procedure that there may be a residual bluish discoloration on the vermillion border, dorsum of the tongue, and dental plaque, which usually wears off in 4-6 hours. Staining in these areas with Tolonium chloride is normal and should not be considered a positive result in the absence of clinical suspicion.
Care should be taken to protect clothing, as well as equipment and environmental surfaces from being stained. The patient should be advised of the possibility of this effect, and assured that the color change is temporary. The patient may also be instructed to contact his/her examining physician/dentist if this occurs.
Should any of the material be accidentally swallowed, the urine and and/or stools may be colored temporarily blue-green or blue, respectively. The patient should be advised of the possibility of this effect, and assured that the color change is temporary. The patient may also be instructed to contact his/her examining physician/dentist if this occurs.
Common restorative materials including porcelain, composites, and acrylics are not known to stain permanently.
{9}------------------------------------------------
# INTERACTIONS
Interactions of the ViziLite Blue Oral Lesion and Identification Marking System components with other medications have not been studied, but are unlikely.
## PREGNANCY AND LACTATION
The ViziLite Blue Oral Lesion Identification and Marking System is contraindicated for use for pregnant women. There have been no well-controlled studies in pregnant women to know the safety of use in this population.
The ViziLite Blue Oral Lesion Identification and Marking System is contraindicated for lactating women. It is not known whether Tolonium chloride is excreted in human breast milk.
## OVERDOSAGE
No adverse events have been reported in the published literature concerning the use of Tolonium chloride in the mouth for staining oral lesions. The ViziLite Blue Oral Lesion Identification and Marking System contains a Tolonium chloride dye swab with 7mg of tolonium chloride, which conservatively yields an exposure of approximately 0.1mg/kg (based on a 60kg person). Reports from published studies indicate physiological alterations are not observed at dose levels below 5 ma/kq administered intravenously. Some patients receiving 100 mg commercially available tolonium chloride orally in capsule from three times daily reported nausea. The NOAEL (No-Observable-Adverse-Effect Level) determined in toxicology studies or orally administered tolonium chloride was 20 mg/kg in rats and was 15mg/kq in rabbits. Data regarding Tolonium chloride and toluidine blue toxicity are summarized in Appendix A. Additional toxicology data for this product, beyond the data furnished in the Appendix, are available upon request.
#### Recommended Storage Conditions
Store between 15°C and 25°C (59°F - 77°F)
ViziLite (OralLite), when used in combination with conventional visual oral mucosal examination by health care providers, improved identification, evaluation, and monitoring of oral mucosal abnormalities in a population at increased risk for oral cancer.
The ViziLite-Blue Oral Lesion Identification and Marking System (with Tolonium chloride) can allow the examiner to continue to visualize oral mucosal lesions using incandescent (conventional projected)light, even after the ViziLite (OralLite) and its holder are removed from the oral cavity. This allows the examiner to measure the lesion size, observe the lesions borders, and obtain an appropriate tissue sample (biopsy) when clinically indicated.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.