The BRM2 Brain Monitor is an Electroencephalograph as per 21 CFR §882.1400 (a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head). The BRM2 Brain Monitor is intended to monitor the state of the brain by acquisition of electroencephalogram (EEG) signals, in the intensive care unit, operating room, and for clinical research.
Device Story
BRM2 Brain Monitor is a two-channel EEG system used to measure and record brain electrical activity. System components include skin electrodes (Neonatal Sensor Set), sensor lead, acquisition unit, isolation unit, cables, monitor, and roll-pole. The Neonatal Sensor Set features two signal electrode pairs and a reference electrode pre-wired to a single connector. Device is used in intensive care units, operating rooms, and clinical research environments. Healthcare providers apply electrodes to the patient's head to acquire EEG signals. Output is displayed on the monitor to assist clinicians in assessing brain state. The device facilitates continuous monitoring, with the sensor set validated for up to 24 hours of use.
Clinical Evidence
Clinical testing was conducted to demonstrate the effective duration of use for the Neonatal Sensor Set. The study established acceptable performance in general use and confirmed the sensor set is effective for the 24-hour maximum recommended usage period.
Technological Characteristics
Two-channel EEG monitor. Components: skin electrodes, sensor lead, acquisition unit, isolation unit, cables, monitor. Neonatal Sensor Set uses pre-wired signal/reference electrodes. Standards: IEC 60601-1 (safety), UL 2601-1 (USA safety), IEC 60601-2-26 (EEG), AAMI EC12 (disposable ECG electrodes).
Indications for Use
Indicated for monitoring brain state via EEG signal acquisition in intensive care units, operating rooms, and clinical research settings. Patient population includes neonates (via Neonatal Sensor Set).
Regulatory Classification
Identification
An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image is a logo for Brainz Instruments Limited. The logo consists of a black square with a white waveform inside, above the word "BRAINZ" in large, bold, sans-serif font. Below the word "BRAINZ" is the text "INSTRUMENTS LIMITED" in a smaller, sans-serif font.
8 September, 2003
OCT 1 0 2003
Brainz Instruments Limited
25 Carbine Road, Mt Wellington PO Box 51-078, Pakuranga Auckland, New Zealand
KO33010
Telephone: +64-9 978 8888 +64-9 978 8889 Facsimile:
www.brainzinstruments.com
### 510(k) Summary of Safety and Effectiveness Information
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR §807.92:
(a)(1) Refer to information above and concluding this summary.
(a)(2) Name of the Device
BRM2 Brain Monitor Model No. / Name:
- Electroencephalograph Classification Name: Electrochecphanigraphi
Neurology Devices, 21 CFR §882.1400, Class II, OLT, ONACTLY
## (a)(3) Identification of Legally Marketed Devices
| K030489 | BRM2 Brain Monitor | Brainz Instruments Ltd |
|---------|---------------------------------------|--------------------------------|
| K960732 | HydroSpot EEG Electrode Model 1501 | Physiometrix Inc |
| K911529 | Bunny Electrode / Stealth Electrode | Lead-Lok Inc |
| K000206 | PALS Neonatal Pediatric ECG Electrode | Axelgaard Manufacturing Co Ltd |
#### (a)(4) Description of the Device
The BRM2 Brain Monitor is a two-channel EEG device, consisting of skin electrodes, Sensor Lead, Acquisition Unit, Isolation Unit, Serial and Power Cables, Monitor, and Roll-Pole. All components other than the skin electrodes and Sensor Lead are the same as described in 510(k) K030489 for the BRM2 Brain Monitor.
Changes to the BRM2 Brain Monitor consist of using the Neonatal Sensor Set for the EEG skin electrodes, which has two signal electrode pairs and a reference electrode all pre-wired to a single connector. The Sensor Lead has been modified to add a compatible connector for the Neonatal Sensor Set. Products used for skin preparation at the EEG electrode application sites have been made available in a convenience kit.
#### (a)(5) Statement of the Intended Use
The BRM2 Brain Monitor is an Electroencephalograph as per 21 CFR §882.1400. It is used to measure and record the electrical activity of a patient's brain, obtained by placing electrodes on the head. Refer to the Indications for Use statement for further information.
1/2
{1}------------------------------------------------
K033010 2/2
# 510(k) Summary continued - BRM2 Brain Monitor
# (a)(6) Technological Characteristics Summary
The technological characteristics of the BRM2 Brain Monitor are equivalent to the predicate devices listed above. The EEG monitor system is the same as described in 510(k) K030489 for the BRM2 Brain Monitor. The Neonatal Sensor Set is equivalent in terms of materials and construction to the electrode predicate devices.
# (b)(1) Discussion of the Non-Clinical Tests
Non-clinical testing of the BRM2 Brain Monitor has been carried out to cover the changes to the system, including verification of electrical safety and performance, biocompatibility, and product shelf life.
The BRM2 Brain Monitor and Neonatal Sensor Set meet the requirements of IEC 60601-1 general safety international standard. They meet relevant USA deviations of the UL 2601-1 standard for general safety, and particular requirements of the IEC 60601-2-26 standard for electroencephalographs.
The Neonatal Sensor Set meets the requirements of the AAMI EC12 standard for disposable ECG electrodes, and FDA / IEC requirements for safety of electrode lead wires.
# (b)(2) Discussion of the Clinical Tests
Clinical testing was carried out to demonstrate the effective duration of use for the Neonatal Sensor Set. This established acceptable performance in general use of the electrodes and in particular over the 24-hour maximum recommended usage period.
# (b)(3) Conclusions Demonstrating Safety, Effectiveness and Performance
The testing carried out for the BRM2 Brain Monitor and Neonatal Sensor Set indicates that they meet design and performance requirements. The subject device continues to meet the requirements of IEC and UL medical electrical equipment standards for safety, and the IEC particular standard for electroencephalographs. The Neonatal Sensor Set meets appropriate performance requirements for ECG electrodes, and effectiveness was demonstrated in a clinical evaluation.
This information indicates that the BRM2 Brain Monitor and Neonatal Sensor Set are equivalent to the predicate devices in terms of safety, effectiveness and performance.
te: 8 Sept 2003
signed:
Chris Mander Regulatory & Quality Manager Brainz Instruments Ltd
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" written around the perimeter. Inside the circle is an image of an eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans. The eagle is stylized and drawn in black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Brainz Instruments Ltd. c/o Ms. Denise L. Klinker Underwriters Laboratories, Inc. 1655 Scott Boulevard Santa Clara, California 95050-4169
APR - 9 2012
Re: K033010
Trade/Device Name: Model BRM2 Brain Monitor with Neonatal Sensor Set Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: II Product Code: OLT, OMC, GXY Dated (Date on orig SE ltr): September 25, 2003 Received (Date on orig SE ltr): September 26, 2003
Dear Ms. Klinker:
This letter corrects our substantially equivalent letter of October 10, 2003.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Ms. Denise L. Klinker
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Kesia Alexander
Image /page/3/Picture/8 description: The image shows a close-up of a handwritten signature on a document. The signature is stylized and illegible, with looping strokes and curves. To the right of the signature are three lines of text, also illegible, written in a formal font. The text appears to be part of a legal or official document.
Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear. Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image is a logo for Brainz Instruments Limited. The logo features a black square with a white waveform inside, resembling brainwaves. Below the square, the word "BRAINZ" is printed in large, bold, sans-serif font, followed by "INSTRUMENTS LIMITED" in a smaller font size.
8 September, 2003
#### Brainz Instruments Limited
25 Carbine Road, Mt Wellington PO Box 51-078, Pakuranga Auckland, New Zealand
Telephone: +64-9 978 8888 Facsimile: +64-9 978 8889
www.brainzinstruments.com
[510(k)] Number:
# Brainz Instruments Ltd - BRM2 Brain Monitor
#### PREMARKET NOTIFICATION 510(k) INDICATIONS FOR USE STATEMENT
The Brainz Instruments Ltd BRM2 Brain Monitor is an Electroencephalograph as per 21 CFR §882.1400 (a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head).
The BRM2 Brain Monitor is intended to monitor the state of the brain by acquisition of electroencephalogram (EEG) signals, in the intensive care unit, operating room, and for clinical research.
Concurrence of CDRH, Office of Device Evaluation (ODE)
Miriam C. Provost
Division of General, Restorative and Neurological Devices
510(k) Number K033010
Prescription Use (Per 21 CFR §801.109)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.