TRSANSDERM IONTO SYSTEM

K032968 · Mattioli Engineering Corp. · EGJ · Dec 2, 2003 · Physical Medicine

Device Facts

Record IDK032968
Device NameTRSANSDERM IONTO SYSTEM
ApplicantMattioli Engineering Corp.
Product CodeEGJ · Physical Medicine
Decision DateDec 2, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 890.5525
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The TRANSDERM IONTO System is a powered iontophoresis drug delivery system that is indicated for the local administration of ionic drug solutions into the body for medical purposes and can be used as an alternative to injections.

Device Story

TRANSDERM IONTO System is a powered iontophoresis device for local delivery of ionic drug solutions. System components include a probe, power supply (TRANS-BASE module), and disposable drug delivery cap electrode. Device operates by generating electrical pulses via an electronic pulse generator; pulses are applied to skin through electrodes on a plastic applicator plate. Drug delivery cap electrode is adhered to patient skin and saturated with ionic solution; probe applicator head attaches to cap. Device requires 9V DC power. Used as an alternative to injections for medical purposes. Healthcare providers apply the device to patients. Output is the controlled delivery of ionic drugs into the body. Benefits include non-invasive drug administration.

Clinical Evidence

Bench testing only. Performance testing and testing to IEC 60601-1 standards were performed to demonstrate substantial equivalence.

Technological Characteristics

Powered iontophoresis device; 9V DC power supply (15W AC-DC switch mode module). Components: probe, power supply, disposable drug delivery cap electrode. Applicator uses a plastic plate with electrodes. Connectivity: mains-powered via removable cord. Sterilization method not specified. Software/algorithm details not provided.

Indications for Use

Indicated for local administration of ionic drug solutions into the body for medical purposes as an alternative to injections. Patient population not specified.

Regulatory Classification

Identification

An iontophoresis device is a device that is intended to use a direct current to introduce ions of soluble salts or other drugs into the body and induce sweating for use in the diagnosis of cystic fibrosis or for other uses if the labeling of the drug intended for use with the device bears adequate directions for the device's use with that drug. When used in the diagnosis of cystic fibrosis, the sweat is collected and its composition and weight are determined. Additionally, an iontophoresis device intended for any other purposes is a prescription device that is intended to use a current to introduce ions of drugs or non-drug solutions into the body for medical purposes other than those specified in paragraph (a) of this section, meaning that the device is not intended for use in diagnosis of cystic fibrosis, or a specific drug is not specified in the labeling of the iontophoresis device.

Special Controls

*Classification.* Class II (special controls). The device is classified as class II. The special controls for this device are:(i) The following performance testing must be conducted: (A) Testing using a drug approved for iontophoretic delivery, or a solution if identified in the labeling, to demonstrate safe use of the device as intended; (B) Testing of the ability of the device to maintain a safe pH level; and (C) If used in the ear, testing of the device to demonstrate mechanical safety. (ii) Labeling must include adequate instructions for use, including sufficient information for the health care provider to determine the device characteristics that affect delivery of the drug or solution and to select appropriate drug or solution dosing information for administration by iontophoresis. This includes the following: (A) A description and/or graphical representation of the electrical output; (B) A description of the electrode materials and pH buffer; (C) When intended for general drug delivery, language referring the user to drug labeling approved for iontophoretic delivery to determine if the drug they intend to deliver is specifically approved for use with that type of device and to obtain relevant dosing information; and (D) A detailed summary of the device-related and procedure-related complications pertinent to use of the device, and appropriate warnings and contraindications, including the following warning: *Warning:* Potential systemic adverse effects may result from use of this device. Drugs or solutions delivered with this device have the potential to reach the blood stream and cause systemic effects. Carefully read all labeling of the drug or solution used with this device to understand all potential adverse effects and to ensure appropriate dosing information. If systemic manifestations occur, refer to the drug or solution labeling for appropriate action.(iii) Appropriate analysis/testing must demonstrate electromagnetic compatibility, electrical safety, thermal safety, and mechanical safety. (iv) Appropriate software verification, validation, and hazard analysis must be performed. (v) The elements of the device that may contact the patient must be demonstrated to be biocompatible. (vi) The elements of the device that may contact the patient must be assessed for sterility, for devices labeled as sterile. (vii) Performance data must support the shelf life of the elements of the device that may be affected by aging by demonstrating continued package integrity and device functionality over the stated shelf life.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ | | DEC - 2 2003 | | | SUMMARY OF SAFETY AND EFFECTIVENESS | K032968 | |-----|--------------|------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------|---------| | I. | | | GENERAL INFORMATION | | | | | A. | Submitted By: | Mattioli Engineering Corporation<br>7918 Jones Branch Drive<br>Suite 600<br>McLean, VA 22102<br><br>Tel: 703-312-6000<br>Fax: 703-243-9139 | | | | | | Contact Person: | Ms. Linda Ferri<br>At address above | | | | | B. | Device Trade Name:<br>Common Name:<br>Classification Name: | TRANSDERM IONTO System<br>Iontophoresis Device<br>Device, Iontophoresis, Specific Uses | | | | | C. | Predicate Devices: | | | | | | | Manufacturer | Product Name | 510(k) No. | | | | | Life-Tech, Inc.<br>Life-Tech, Inc. | Microphor®<br>Iontophor® II | K913601<br>K863166 | | | II. | | DEVICE DESCRIPTION | | | | #### II. DESCRIPTION The TRANSDERM IONTO System consists of a probe, power supply/support, and a disposable drug delivery cap clectrode. Flectrical pulses are produced by an electronic pulse generator that is able to generate bursts of pulses that are applied to the skin through electrodes applied to a plastic plate (the applicator). The design of the device provides one channel output. The TRANSDERM IONTO System requires a DC supply of 9 V, 1.0 A max. This power source is supplied by, the TRANS-BASE module which is connected to mains through a removable supply cord and is equipped with a proper connector receptacle to fit the connector malc mounted on TRANSDERM IONTO probe supply cable. The TRANS-BASE is equipped with a 15 w AC-DC switch mode power module. The drug delivery cap electrode is adhered to the patient's skin and then saturated with ionic solution. Once the drug delivery cap electrode is affixed to the {1}------------------------------------------------ patient's skin, the TRANSDERM IONTO probe's applicator head can be attached to the internal side of the drug delivery cap electrode. ### III. INDICATIONS FOR USE The TRANSDERM IONTO System is a powered iontophorcsis drug delivery system that is indicated for the local administration of ionic drug solutions into the body for medical purposes and can be used as an alternative to injections. # IV. TECHNOLOGICAL COMPARISON The TRANSDERM IONTO System, Microphor® (K913601), and Iontophor® II (K863166) have similar indications for use and overall function and perform in a similar manner with respect to iontophoretic drug delivery. Performance testing supports that there is no significant difference between the TRANSDERM IONTO System and the Iontophor® II. The TRANSDERM IONTO System utilizes a plastic plate containing electrodes and an adhesive drug delivery can electrode to administer the ionic solution to the patient rather than the lead wires and adhesive clectrodes of the Microphor® and Iontophor® II. # V. TESTING Performance testing and the applicable IEC 60601-1 standards testing was performed. #### VI. CONCLUSIONS The TRANSDERM IONTO System is substantially equivalent to the legally marketed iontophoretic devices in intended use and performance. {2}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized eagle emblem, with three curved lines representing the eagle's wings and a wavy line representing its body. Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 DEC - 2 2003 Mattioli Engineering Corporation C/o Ms. Melissa Mahall Bio-Reg Associates, Inc. 11800 Baltimore Avenue, Suite 105 Beltsville, Maryland 20705 Re: K032968 > Trade/Device Name: TRANSDERM IONTO System Regulation Number: 21 CFR 890.5525 Regulation Name: Iontophoresis device Regulatory Class: III Product Code: EGJ Dated: November 20, 2003 Received: November 24, 2003 Dear Ms. Mahall: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalcnt (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act), as long as you comply with all of the Act's requirements relating to drugs labeled or promoted with the device as described below. You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into cither class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Ecderal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not {3}------------------------------------------------ limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Our substantially equivalent decision does not apply to the drugs that you will label or promote for use with your device. Therefore, you may neither label nor promote your promotor use with specific drugs, nor package drugs with your device prior to FDA having approved the drugs for iontophoretic administration. For information on the requirements for marketing new drugs, you may contact: > Director Division of Drug Labeling Compliance (HFD-310) Center for Drug Evaluation and Research Food and Drug Administration 5600 Fishers Lane Rockville, Maryland 20857 As you are aware, iontophoresis devices that are intended to use a direct current to introduce ions of soluble salts or other drugs into the body and induce sweating for use in the diagnosis of cystic fibrosis or for other uses, if the labeling of the drug intended for usc with the device bears adequate directions for the device's use with that drug, were classified into Class II. An iontophoresis device that is intended to use a direct current to introduce ions of soluble salts or other drugs into the body for medical purposes other than those specified for class II devices is classified into Class III (21 CFR 890.5525). We published our strategy for calling for premarket approval (PMA) applications in the enclosed Federal Register, dated May 6, 1994, and the enclosed memorandum, dated April 19, 1994, and the enclosed Federal Register, dated August 22, 2000. If you have any questions regarding this letter, you may contact: Kevin Lee, M.D. Food and Drug Administration Center for Devices and Radiological Health Division of General, Restorative and Neurological Devices 9200 Corporate Boulevard (HFZ-410) Rockville, Maryland 20850 (301) 594-1296 This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. {4}------------------------------------------------ Page 3 - Ms. Melissa Mahall If you desire specific advice for your device on our labeling regulation, please contact the Office of Compliance at (301) 594 - 4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll free number (800) 638-2041 or (301) 443-6597, or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, for Mark A. Wilkerson Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # INDICATIONS FOR USE K032968 510(k) Number: Device Name: TRANSDERM IONTO System Mattioli Enginecring Corporation Indications for Use: Sponsor Name: The TRANSDERM IONTO System is a powered iontophoresis drug delivery system that is indicated for the local administration of ionic drug solutions into the body for medical purposes and can be used as an alternative to injections. Prescription Use 风 (21 CFR 801 Subpart D) And/Or Over-The-Counter Use □ (21 CFR 807 Subpart C) Do Not Write Below This Line - Continue on Another Page if Needed Concurrence of CDRH, Office of Device Evaluation (ODE) Mark N Milkman Jivision Sign-C Division of General, Restorative and Neurological f 510(k) Number -
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...