← Product Code [IYO](/productcode/IYO) · K032956

# ACCUSONIC A-SCAN  MODEL, 24-4000 (K032956)

_Accutome, Inc. · IYO · Oct 10, 2003 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K032956

## Device Facts

- **Applicant:** Accutome, Inc.
- **Product Code:** [IYO](/productcode/IYO.md)
- **Decision Date:** Oct 10, 2003
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1560
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

The instrument is used for measuring the axial length, anterior chamber depth and lens thickness of the eye. It is also used for calculating the optical power of the IOL to be implanted during cataract surgery. Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Ophthalmic (A-mode).

## Device Story

AccuSonic A-Scan is a biometer using pulsed echo ultrasound to measure ocular structures; specifically axial length, anterior chamber depth, and lens thickness. Device utilizes an eye-contact probe to generate/receive ultrasound pulses; provides one-dimensional display of returning echoes with positive peaks indicating ocular structure locations. Distance between peaks is measured to calculate IOL power for cataract surgery. Used in clinical settings by eye care professionals. Output allows clinicians to determine appropriate IOL power for patient implantation, aiding surgical planning.

## Clinical Evidence

Bench testing only. No clinical data provided.

## Technological Characteristics

A-mode 10 MHz transducer; pulsed echo ultrasound principle; one-dimensional display; eye-contact probe; standalone biometer system.

## Regulatory Identification

An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Special Controls

*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- Accutome Advent AB ([K960765](/device/K960765.md))
- Teknar Ophthasonic A-Scan ([K860757](/device/K860757.md))
- Teknar Opthasonic A/P III ([K903666](/device/K903666.md))
- DGH 3000A A-Scan ([K872726](/device/K872726.md))
- DGH 4000 A-Scan/Pach ([K913067](/device/K913067.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

OCT 1 0 2003

K032956

# 510(k) SUMMARY AccuSonic A-Scan

| Applicant:                             | Accutome Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                                |
|----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address:                               | 263 Great Valley Parkway<br>Malvern, PA 19355                                                                                                                                                                                                                                                                                                                                                                                                                |
| Contact Person:                        | Jeffrey L. Wright<br>Manager, Engineering, Manufacturing & QC                                                                                                                                                                                                                                                                                                                                                                                                |
| Telephone:                             | (610) 889-0200<br>(610) 889-3233 fax                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Preparation Date:                      | August 25, 2003                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Device Name:                           | AccuSonic A-Scan                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Common Name:                           | Biometer                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Classification:                        | System, Imaging, Pulsed Echo, Ultrasonic<br>(see: 21 CFR 892.1560) Product Code: IYO Panel: 90                                                                                                                                                                                                                                                                                                                                                               |
| Legally Marketed<br>Predicate Devices: | Accutome Advent AB (K960765) previously by Mentor<br>Ophthalmics and acquired by Accutome in December of<br>1999. (However, the AccuSonic A-Scan does not function<br>as a B-Scan.); Teknar Ophthasonic A-Scan (K860757);<br>Teknar Opthasonic A/P III (K903666); DGH 3000A A-<br>Scan (K872726); DGH 4000 A-Scan/Pach (K913067)                                                                                                                             |
| Description of the<br>Device:          | The AccuSonic A-Scan device is designed as a biometer,<br>which uses pulsed echo ultrasound to measure the axial<br>length, and the location of other structures of the eye. It<br>utilizes an eye-contact probe to generate and receive the<br>ultrasound pulses, and provides a one-dimensional display<br>of returning pulse echoes, with positive peaks to indicate<br>the location of ocular structures. The distance between<br>peaks can be measured. |
| Indications for Use:                   | The instrument is used for measuring the axial length,<br>anterior chamber depth and lens thickness of the eye. It is<br>also used for calculating the optical power of the IOL to be<br>implanted during cataract surgery.                                                                                                                                                                                                                                  |

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the words "Public Health Service" in bold, black font. The words are stacked on top of each other, with "Public" on the top line, "Health" on the second line, and "Service" on the third line. The text appears to be extracted from a document or sign.

Image /page/1/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a bird in flight, composed of three stylized human profiles.

OCT 1 0 2003

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. Jeffrey L. Wright Manager, Engineering, Manufacturing & QC Accutome, Inc. 263 Great Valley Parkway MALVERN PA 19355

Re: K032956

Trade Name: AccuSonic A-Scan Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulatory Class: II Product Code: 90 IYO Dated: August 25, 2003 Received: September 22, 2003

Dear Mr. Wright:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

This determination of substantial equivalence applies to the following transducer intended for use with the AccuSonic A-Scan, as described in your premarket notification:

### Transducer Model Number

### A-mode 10 MHz

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality

{2}------------------------------------------------

systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This determination of substantial equivalence is granted on the condition that prior to shipping the first device, you submit a postclearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested in Appendix G, (enclosed) of the Center's September 30, 1997 "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers." If the special report is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded.

The special report should reference the manufacturer's 510(k) number. It should be clearly and prominently marked "ADD-TO-FILE" and should be submitted in duplicate to:

> Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850

This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801, please contact the Office of Compliance at (301) 594-4591. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".

If you have any questions regarding the content of this letter, please contact Rodrigo C. Perez at (301) 594-1212.

Sincerely yours,

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

{3}------------------------------------------------

## AccuSonic A-Scan 510(k) Notification

4.3 Indications for Use

K032956 510(k) Number (if known):_

Device Name: Accusonic A-Scan

Indications for Use:

The instrument is used for measuring the axial length, anterior chamber depth and lens thickness of the eye. It also is used for calculating the optical power of the IOL to be implanted during cataract surgery.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

OR

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use***_***_\/

Over-The Counter Use__________________________________________________________________________________________________________________________________________________________

Nancy C hogdon

(Division Sign-C Division of Reproduct and Radiological Devices 510(k) Number

Page 4 of 19

{4}------------------------------------------------

# AccuSonic A-Scan 510(k) Notification

#### Diagnostic Ultrasound Indications for Use Form

#### Fill out one form for each ultrasound system and each transducer.

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application             | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
|----------------------------------|---|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Ophthalmic                       | N |   |   |     |     |                  |                      |                              |                       |                    |
| Fetal                            |   |   |   |     |     |                  |                      |                              |                       |                    |
| Abdominal                        |   |   |   |     |     |                  |                      |                              |                       |                    |
| Intraoperative specify           |   |   |   |     |     |                  |                      |                              |                       |                    |
| Intraoperative Neurological      |   |   |   |     |     |                  |                      |                              |                       |                    |
| Pediatric                        |   |   |   |     |     |                  |                      |                              |                       |                    |
| Small Organ specify              |   |   |   |     |     |                  |                      |                              |                       |                    |
| Neonatal Cephalic                |   |   |   |     |     |                  |                      |                              |                       |                    |
| Adult Cephalic                   |   |   |   |     |     |                  |                      |                              |                       |                    |
| Cardiac                          |   |   |   |     |     |                  |                      |                              |                       |                    |
| Transesophageal                  |   |   |   |     |     |                  |                      |                              |                       |                    |
| Transrectal                      |   |   |   |     |     |                  |                      |                              |                       |                    |
| Transvaginal                     |   |   |   |     |     |                  |                      |                              |                       |                    |
| Transurethral                    |   |   |   |     |     |                  |                      |                              |                       |                    |
| Intravascular                    |   |   |   |     |     |                  |                      |                              |                       |                    |
| Peripheral Vascular              |   |   |   |     |     |                  |                      |                              |                       |                    |
| Laparoscopic                     |   |   |   |     |     |                  |                      |                              |                       |                    |
| Musculo-skeletal<br>Conventional |   |   |   |     |     |                  |                      |                              |                       |                    |
| Musculo-skeletal Superficial     |   |   |   |     |     |                  |                      |                              |                       |                    |

N= new indication: P= previously cleared by FDA: E= added under Appendix E

Additional Comments:

CONTINUE ON ANOTHER PAGE IF NEEDED.

evice Evaluation

System includes A-More ID H/A Transducer.

Nancy Hodgson

(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices

510(k) Numb

---

**Source:** [https://fda-staging.innolitics.com/device/K032956](https://fda-staging.innolitics.com/device/K032956)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
