← Product Code [DXN](/productcode/DXN) · K032927

# FULL AUTO ARM DIGITAL BLOOD PRESSURE MONITOR (K032927)

_Sein Electronics Co., Ltd. · DXN · May 7, 2004 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K032927

## Device Facts

- **Applicant:** Sein Electronics Co., Ltd.
- **Product Code:** [DXN](/productcode/DXN.md)
- **Decision Date:** May 7, 2004
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1130
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

SE-7070 is intended to measure systolic and diastolic pressure and pulse rate of adults in a home care environment using arm cuff and oscillometric method of measurement.

## Device Story

The SE-7070 is an automated, non-invasive blood pressure monitor for home use. It utilizes an arm cuff and the oscillometric method to measure systolic/diastolic pressure and pulse rate. The device features automatic inflation and deflation, with a fuzzy logic function that dynamically adjusts initial inflation pressure based on the patient's blood pressure. The user initiates the measurement via a power button; the device processes the signal and displays results (systolic, diastolic, pulse, date, time) on an LCD screen. It includes a memory function for the 58 most recent readings. The device is operated by the patient without specialized training. It provides objective data to assist in monitoring blood pressure trends, aiding in the management of hypertensive, normotensive, or hypotensive conditions.

## Clinical Evidence

No clinical data provided. Substantial equivalence is based on technological characteristics and performance comparisons to the predicate device.

## Technological Characteristics

Non-invasive oscillometric blood pressure monitor. Features automatic cuff inflation/deflation. Includes fuzzy logic for inflation pressure control. LCD display for systolic, diastolic, pulse, date, and time. Memory capacity for 58 readings. Powered by internal electronics. Home-use form factor.

## Regulatory Identification

A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.

## Predicate Devices

- Full Auto Blood Pressure Monitor, Model SE-7700H ([K003282](/device/K003282.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

MAY - 7 2004

K032927
p1/2

# 510(k) Summary As Required by 21 section 807.92 ( c )

- 1-Submitter Name: Sein Electronics Co., Ltd
2-Address: Room #506, Youcheon Factopia, #196, Anyang 7-dong, Manan-Gu, Anyang-city, Kyunggi-do, Korea

- 3-Phone: (82) 31-467-2188
- 4-Fax: (82) 31-467-2107

5-Contact Person: Won-Ky Kim

- 6-Date summary prepared: September 9", 2003
- 7-Device Trade or Proprietary Name: Full Auto Arm Digital Blood Pressure Monitor, Model SE-7070
- 8-Device Common or usual name: Blood pressure monitor

9-Device Classification Name: Non Invasive blood pressure measuring system

- 10-Substantial Equivalency is claimed against the following device: Full Auto Blood Pressure Monitor, Model SE-7700H, manufactured by Sein Electronics Co., Ltd.
### 11-Description of the Device:

SE-7070 is intended to measure systolic and diastolic pressure and pulse rate of adults in a home care environment using arm cuff and oscillometric method of measurement.

There are no contraindications; the subject device may be employed in the care of normotensive, hypertensive, or hypotensive patents.

The user interface panel has a power switch, a mode switch, a memory switch and a liquid crystal display ("LCD") for displaying the systolic blood pressure, diastolic blood pressure, pulse rate, date and time. This device has the memory function that permits memory and display of the 58 most recent measurement results.

The device measures blood pressure through the use of a automatic inflating cuff. Pressurization is automatically governed. The device has fuzzy logic function that establishes automatically initial inflation pressure according to blood pressures of patients. The cuff automatically deflates during blood pressure measurement.

The patient is responsible for applying the cuff, for initiating the measurements sequence by pressuring the "Power" button, and for recording results. The patient cannot alter bleed-down rate.

All system functions are preprogrammed. The user is cautioned in the instruction manual against attempting any programming or other modification.

PAGE 10

{1}------------------------------------------------

No special training beyond basic ability to follow instruction is required. Since the products are designed for home use, detailed instructions on avoidance of practices that adversely affect the accuracy of measurements are included in the instruction manual.

#### 12-Intended use of the device:

SE-7070 is intended to measure systolic and diastolic pressure and pulse rate of adults in a home care environment using arm cuff and oscillometric method of measurement.

#### 13-Safety and Effectiveness of the device:

This device is safe and effective as the other predicate device cited above. This is better expressed in the tabulated comparison (Paragraph 14 below)

#### 14-Summary comparing technological characteristics with other predicate device:

Please find below a tabulated comparison supporting that this device is substantially equivalent to other medical devices in commercial distribution. Also, Equivalency overview chart path is attached. Indeed, this device is SIMILAR to the predicate device.

| FDA file reference number                        | 510k # K003282                                                                                |
|--------------------------------------------------|-----------------------------------------------------------------------------------------------|
| Attachments inside notification submission file  | 510k summary print out                                                                        |
| <b>TECHNOLOGICAL CHARACTERISTICS</b>             | <b>Comparison result</b>                                                                      |
| Indications for use                              | Identical, except that the end<br>user is the individual at home,<br>not medical practitioner |
| Target population                                | Identical                                                                                     |
| Design                                           | Similar                                                                                       |
| Materials                                        | Similar                                                                                       |
| Performance                                      | Similar                                                                                       |
| Sterility                                        | Similar                                                                                       |
| Biocompatibility                                 | Similar                                                                                       |
| Mechanical safety                                | Similar                                                                                       |
| Chemical safety                                  | Not Applicable                                                                                |
| Anatomical sites                                 | Identical                                                                                     |
| Human factors                                    | Similar                                                                                       |
| Energy used and/or delivered                     | Similar                                                                                       |
| Compatibility with environment and other devices | Similar                                                                                       |
| Where used                                       | Identical                                                                                     |
| Standards met                                    | Similar                                                                                       |
| Electrical safety                                | Similar                                                                                       |
| Thermal safety                                   | Similar                                                                                       |
| Radiation safety                                 | Similar                                                                                       |

REFER TO MAIN SUBMISSION FOR COMPLETE DETAILS.

{2}------------------------------------------------

Image /page/2/Picture/12 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of an eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAY - 7 2004

Sein Electronics Co., Ltd. c/o Jay Mansour Mansour Consulting LLC 1308 Morningside Park Drive Alpharetta, GA 30022

Re: K032927

Trade Name: Full Auto Arm Digital Blood Pressure Monitor, Model SE-7070 Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: II (two) Product Code: DXN Dated: April 04, 2004 Received: April 12, 2004

Dear Mr. Mansour:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA is issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or 

{3}------------------------------------------------

Page 2 - Mr. Jay Mansour

any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4648. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Meftaxmiya

Bram D. Zuckerman Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## Indications for Use

510(k) Number (if known): K032927

Device Name: FULL AUTO ARM DIGITAL BLOOD PRESSURE MONITOR, MODEL SE-7070

Indications For Use:

This device is an over the counter device, and is indicated for use to measure systolic and diastolic pressure and pulse rate of adults by the individual, in a home care environment, using arm cuff and oscillometric method of measurement

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Me Jak Mayer
sign-on
Joe Davisson

810(k) Number K032927

Page 1 of ____________________________________________________________________________________________________________________________________________________________________

---

**Source:** [https://fda-staging.innolitics.com/device/K032927](https://fda-staging.innolitics.com/device/K032927)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
