← Product Code [MNH](/productcode/MNH) · K032855

# STRYKER SPINE MAPS SYSTEM (K032855)

_Howmedica Osteonics · MNH · Dec 10, 2003 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K032855

## Device Facts

- **Applicant:** Howmedica Osteonics
- **Product Code:** [MNH](/productcode/MNH.md)
- **Decision Date:** Dec 10, 2003
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3070
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Stryker Spine MAPS System is a pedicle screw system intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion the treatment of the following acute and chronic instabilities or deformities of the thoracic, iumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed provigus fusion (pseudarthrosis). The Stryker Spine MAPS System is also indicated for pedicle screw fixation for the treatment of severe spondylolisthesis (Craces 3 and 4) at the L5-S1 vertebra in sketetsly matured on receiving fusion by autogenous bone graft having implants attached to the lumber and sacral spine (13 to sacrum) with removal of the implants after the development of a solid fusion mass. The Stryker Spinc MAPS System is also a sacral/iliac screw fixation system indicated for degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (fracture and/or dislocation), spinal stenosis, deformities (scoliosis, kyphosis and/or lordosis), tumor, pseudoarthrosis and revision of failed fusion attempts. The Stryker Spine MAPS System is also intended to be used in conjunction with the OSS/Diapason Rods, Opus Spinal System Rods and the Multi-Axis Cross Comentors.

## Device Story

Stryker Spine MAPS System consists of titanium alloy screws and connectors; used for spinal immobilization and stabilization. Components available in various lengths to accommodate patient anatomy. System used in conjunction with OSS/Diapason Rods, Opus Spinal System Rods, and Multi-Axis Cross Connectors. Intended for surgical implantation by physicians in clinical settings to treat spinal instabilities or deformities. Provides mechanical support to spinal segments to facilitate fusion; implants may be removed after solid fusion mass development.

## Clinical Evidence

No clinical data. Substantial equivalence supported by mechanical testing demonstrating comparable properties to predicate devices.

## Technological Characteristics

Titanium alloy spinal fixation components (screws and connectors). Non-sterile. Mechanical design for pedicle and sacral/iliac fixation.

## Regulatory Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

## Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
 “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

## Reference Devices

- OSS/Diapason Rods
- Opus Spinal System Rods
- Multi-Axis Cross Connectors

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K032855

1.62

Stryker Spine MAPS System

DEC 1 0 2003

510(k) Premarket Notification

## 510(k) Summary of Safety and Effectiveness: Stryker Spine MAPS System

| Proprietary Name:                  | Stryker Spine MAPS System                                                                                                                                                                                |
|------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name:                       | Spinal Fixation Appliances                                                                                                                                                                               |
| Classification Name and Reference: | Pedicle Screw Spinal System, 21 CFR 888.3070                                                                                                                                                             |
| Proposed Regulatory Class:         | Class II                                                                                                                                                                                                 |
| Device Product Code:               | 87 MNH: Orthosis, Spondyloisthesis Spinal Fixation<br>87 MNI: Orthosis, Spinal, Pedicle Fixation                                                                                                         |
| For Information contact:           | Karen Ariemma<br>Regulatory Affairs Specialist<br>Howmedica Osteonics Corp.<br>59 Route 17<br>Allendale, NJ 07401-1677<br>Telephone: (201) 831-5718<br>Fax: (201) 831-6038<br>Email: karicmma@howost.com |
| Date Summary Prepared:             | December 9, 2003                                                                                                                                                                                         |

## Device Description

The Stryker Spine MAPS System is comprised of screws and connectors. The components are available in a variety of lengths in order to accommodate patient anatomy. The components are fabricated from titanium alloy. The components will be provided non-sterile.

## Intended Use

The Stryker Spine MAPS System is a pedicle screw system intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion the treatment of the following acute and chronic instabilities or deformities of the thoracic, iumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed provigus fusion (pseudarthrosis).

The Stryker Spine MAPS System is also indicated for pedicle screw fixation for the treatment of severe spondylolisthesis (Craces 3 and 4) at the L5-S1 vertebra in sketetsly matured on receiving fusion by autogenous bone graft having implants attached to the lumber and sacral spine (13 to sacrum) with removal of the implants after the development of a solid fusion mass.

The Stryker Spinc MAPS System is also a sacral/iliac screw fixation system indicated for degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (fracture and/or dislocation), spinal stenosis, deformities (scoliosis, kyphosis and/or lordosis), tumor, pseudoarthrosis and revision of failed fusion attempts.

{1}------------------------------------------------

The Stryker Spine MAPS System is also intended to be used in conjunction with the OSS/Diapason Rods, Opus Spinal System Rods and the Multi-Axis Cross Comentors.

## Substantial Equivalence

Equivalency of this device is based on similarities in intended use, materials, and design to other currently marketed spinal systems. Mcchanical testing demonstrated comparable mechanical properties to the predicate devices.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

DEC 1 0 2003

Ms. Karen Ariemma Regulatory Affairs Specialist Howmedica Osteonics Corporation 59 Route 17 Allendale, New Jersey 07401-1677

Re: K032855

Trade/Device Name: Stryker Spine MAPS System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: II Product Codes: MNH, MNI Dated: September 10, 2003 Received: September 12, 2003

Dear Ms. Ariemma:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendnents. or 10 devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 Cl·R Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Ms. Karen Ariemma

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please 11 you a the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Mah Millman

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

०32rsr '11

Page 1 of 1

510(k) Number (if known): K

Device Name: Stryker Spine MAPS System

The Stryker Spine MAPS System is a pedicle screw system intended to provide immobilization and stabilization of spinal segments in skeletally maturc patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis).

The Stryker Spinc MAPS System is also indicated for pedicle screw fixation for the treatment of severe spondylolisthesis (Grades 3 and 4) at the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the development of a solid fusion mass.

The Stryker Spine MAPS System is also a sacral/iliac screw fixation system indicated for degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (fracture and/or dislocation), spinal stenosis, deformities (scoliosis, kyphosis and/or lordosis), tumor, pseudoarthrosis and revision of failed fusion attempts.

The Stryker Spine MAPS System is also intended to be used in conjunction with the OSS/Diapason Rods, Opus Spinal System Rods and the Multi-Axis Cross Connectors.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE) Over-The-Counter Use (Per 21 CFR 801.109) Prescription Use OR (Optional Format 1-2-96)

R. Mark N. Mcllharron

al, Restorative

Number K032853

---

**Source:** [https://fda-staging.innolitics.com/device/K032855](https://fda-staging.innolitics.com/device/K032855)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).