← Product Code [FOZ](/productcode/FOZ) · K032777

# MODIFICATION TO: NIPRO SAFE TOUCH SAFETY FISTULA NEEDLE (K032777)

_Nipro Medical Corp. · FOZ · Nov 14, 2003 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K032777

## Device Facts

- **Applicant:** Nipro Medical Corp.
- **Product Code:** [FOZ](/productcode/FOZ.md)
- **Decision Date:** Nov 14, 2003
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 880.5200
- **Device Class:** Class 2
- **Review Panel:** General Hospital

## Indications for Use

The Nipro® SafeTouch Safety Fistula Needle is intended for use as a blood access device for blood purification and for other treatments requiring an extracorporeal circuit or larger volumes of blood. Secondly, it is designed with an active sharp safety feature requiring physical action by the clinician to aid in the prevention of accidental needle stick injuries.

## Device Story

Non-implanted blood access device; consists of arterial/venous fistula adapter with flexible tubing and needle. Features active sharps safety mechanism; includes Safety Hub, Safety Stopper, and Safety Protector. Clinician operates device; upon needle removal from patient, user releases latch and slides winged cover over needle until it latches. Prevents accidental needle stick injuries. Available in fixed (stationary) and turnable (rotating) wing types; various gauges, lengths, and tubing configurations. Sterile; single-use only. Physician-ordered use.

## Clinical Evidence

Bench testing only. Biocompatibility data provided, including sterility safety, pyrogenicity, intracutaneous reactivity, systemic injection, hemolysis, and implantation testing.

## Technological Characteristics

Materials: PMMA compounds (Methyl Methacrylate Polymer and Butadiene Polymer) for Safety AVF hub; Polyvinyl chloride (PVC) tubing. Design: Fixed or turnable wing types. Safety mechanism: Active latching sharps injury prevention system. Sterile, single-use. No software or electronic components.

## Regulatory Identification

An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.

## Predicate Devices

- Nipro SafeTouch Safety Fistula Needle ([K002813](/device/K002813.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K032777
Page 1 of 2

## 510(k) Summary of Safety and Effectiveness

#### 1. Manufacturer and Contact Information:

| Manufacturer:        | Nipro (Thailand) Corporation Limited<br>10/2 Moo 8, Bangnomko, Sena<br>Ayuthaya, Thailand 13110                                                      |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|
| U. S. Distributor:   | Nipro Medical Corporation<br>3150 NW 107 Avenue<br>Miami, FL 33172                                                                                   |
| Contact Information: | Richard D. Bliss, Jr.<br>Quality Systems Engineering<br>510 Stonemont Drive<br>Weston, FL 33326<br>Telephone: (954) 385-1690<br>Fax:  (954) 385-1256 |

#### 2. Device Classification Name:

The Gastroenterology and Urology Devices Panel has classified Blood Access Device and accessories as Class II. Reference 21 CFR 876.5540.

#### 3. Intended Use:

The Nipro® SafeTouch Safety Fistula Needle is intended for use as a blood access device for blood purification and for other treatments requiring an extracorporeal circuit or larger volumes of blood. Secondly, it is designed with an active sharp safety feature requiring physical action by the clinician to aid in the prevention of accidental needle stick injuries.

#### 4. Device Description and Characteristics:

This 510(k) Summary of Safety and Effectiveness is being submitted in accordance with the requirements for the Safe Medical Device Act (SMDA) 1990.

The Nipro® SafeTouch Safety Fistula Needle includes arterial and venous fistula adapter consisting of flexible tube and needle with an active sharps safety feature (non-implanted blood access device) as described in 21 CFR 876.5540. The Nipro® SafeTouch Safety Fistula Needle was previously described in detail as part of Premarket Notification cleared by FDA under K002813 on November 21, 2000.

{1}------------------------------------------------

K032777

Page 2 of 2

The fistula needle included in this 510(k) is modified to include a change in material of the Safety AVF hub from ABS to PMMA compounds consisting of Methyl Methacrylate Polymer and Butadiene Polymer and a change in the shape of the Safety AVF stopper to enhance the activation of the safety mechanism.

The sharps injury prevention system is connected between winged needle and polyvinyl chloride tubing and consists of three parts, the Safety Hub, Safety Stopper and Safety Protector. The needle is attached to the Safety Hub and is connected to the Safety Stopper. The Safety Protector then covers the Safety Hub.

The safety feature is easily operated through the release of a latch mechanism whereby the user slides a winged cover over the needle, as it is removed from the patient. Once the needle is covered, the safety cover latches into place. The safety feature presented in this document represents a substantially equivalent version of the current brand, which previously cleared FDA in K002813.

The Nipro® SafeTouch Safety Fistula Needle includes 2 basic types of design with clamps which remain the same: fixed wing type (stationary) and turnable wing type (rotating) which is offered in 96 configurations with options within it to include needle gauge, needle length, type of needle, and tubing length availability. A table with all the device configurations is included in Section 3 of this report.

The devices are packaged sterile and are labeled for single use only. There is no ability to clean and reuse these devices. They are restricted for sale by or on the order of a physician. The results of biocompatibility data support the equivalence of the predicate device and include sterility safety, pyrogenicity, intracutaneous reactivity, systemic iniection, hemolysis, implantation testing. Please refer to Tab 9 of the Biocompatibility Section.

#### 5. Substantial Equivalence

Nipro Medical Corporation considers the modified Nipro® SafeTouch Safety Fistula Needle to be substantially equivalent to the current Nipro SafeTouch Safety Fistula Needle with regard to intended use, materials, biocompatibility, and overall performance characteristics. The labeling is equivalent to the predicate device in intended use.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three curved lines representing its body and wings. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

# NOV 1 4 2003

Nipro Medical Corp. c/o Richard D. Bliss, Jr. Quality Systems Engineering 510 Stonemont Drive WESTON FL 33326

Re: K032777

Trade/Device Name: Nipro SafeTouch Safety Fistula Needles Regulation Number: 21 CFR 876.5540 Regulation Name: Blood access device and accessories Product Code: 78 FIE Regulation Number: 21 CFR §880.5200 Regulation Name: Intravascular catheter Product Code: 80 FOZ Regulatory Class: II Dated: October 13, 2003 Received: October 15, 2003

Dear Mr. Bliss:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:

| 8xx.1xxx                         | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4654 |
| Other                            | (301) 594-4692 |

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Nipro® SafeTouch Safety Fistula Needle Device Name:

### Indications for Use:

The Nipro® SafeTouch Safety Fistula Needles is intended for use as a blood access device for blood purification and for other treatments requiring extracorporeal circuit. The Nipro SafeTouch Safety Fistula Needle aids in the prevention of accidental needle stick injuries. The compatibility of available configurations is the responsibility of the physician in charge.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation ODE)

Prescription Use (Per 21 CFR 801.109)

OR

Over-The-Counter-Use (Optional Format 1-2-96)

Nancy C Brogdon
Division Stop Off

(Division Stgn-Off) Division of Reproductive, Abdominal, and Radiological Devices

510(k) Number K032770

---

**Source:** [https://fda-staging.innolitics.com/device/K032777](https://fda-staging.innolitics.com/device/K032777)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
