← Product Code [FPA](/productcode/FPA) · K032767

# LILY VOLUME METERIC ADMINISTRATION SET, MODELS IB-2111 AND CIB-120ES (K032767)

_Lily Medical Supplies Co., Ltd. · FPA · Jun 3, 2004 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K032767

## Device Facts

- **Applicant:** Lily Medical Supplies Co., Ltd.
- **Product Code:** [FPA](/productcode/FPA.md)
- **Decision Date:** Jun 3, 2004
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5440
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic

## Indications for Use

The LILY Volume Meteric Administration Set is designed to administer fluids from a container to a patient's vascular system through a catheter inserted into a vein.

## Device Story

The LILY Volume Meteric Administration Set is a single-use, sterile, non-toxic, non-pyrogenic infusion set. It facilitates the delivery of fluids from a fluid container to a patient's vascular system through an indwelling intravenous catheter. The device is intended for use in clinical settings by healthcare professionals to manage fluid therapy. It functions as a passive conduit for fluid administration.

## Clinical Evidence

Bench testing only. The device underwent biocompatibility, sterility, and performance testing as per the FDA Guidance on Premarket Notification for Intravascular Administration Sets.

## Technological Characteristics

Single-use, sterile, non-toxic, non-pyrogenic infusion set. Designed for intravascular fluid administration. No electronic components, software, or energy sources.

## Regulatory Identification

An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.

## Special Controls

*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.

## Predicate Devices

- TUTA Healthcare Burette - In Line (150 ml) ([K023595](/device/K023595.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

JUN 0 3 2004

K03.2767

### 510(K) Summary of Safety and Effectiveness II.

# Information

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

(Pcr 21 CFR 807.92)

- LILY Medical Supplies Co., Ltd. 1. Submitter: No.28-2, Shun Chan Tsan, Chu-Nan, Miao-Li Hsien, Taiwan, 350, ROC

| Contact Person: | Dr. Ke-Min Jen         |
|-----------------|------------------------|
|                 | Official Correspondent |
|                 | 886-3-5208829 (Tel)    |
|                 | 886-3-5209783 (Fax)    |

- Date Prepared: 2003/8/30 2.
- 3. Device Name:

| -Proprietary Name:    | LILY Volume Meteric Administration Set, CIB-100ES,<br>CIB-120ES, CIB-150ES |
|-----------------------|----------------------------------------------------------------------------|
| -Common Name:         | Infusion Set                                                               |
| -Classification Name: | Set, Administration, Intravascular                                         |
| -Product code:        | FPA,                                                                       |
| -Regulation:          | 880.5440                                                                   |

#### 4. Predicate Device:

TUTA Healthcare Burette - In Line ( 150 ml )(K023595)

#### న్, Device Description:

The LILY Volume Meteric Administration Set, CIB-100ES, CHB-120ES, CIB-150ES, is a single-use, non-toxic, non-pyrogenic and sterile infusion set to administer fluids from a container to a patient's vascular system through a catheter inserted into a vein.

#### 6. Intended Use:

The LILY Volume Meteric Administration Set is designed to administer fluids from a container to a patient's vascular system through a catheter inserted into a vein.

## 7. Substantial Equivalence (SF) and Safety and Effectiveness Information:

A claim of substantial equivalence is made to TUTA Healthcare Burette - In Line ( 150 mL)(K023595). The LILY VOLUME METERIC ADMINISTRATION SET and the TUTA Healthcare Burette - In Line ( 150 mL) are designed to administer fluids from a container to a patient's vascular system through a catheter inserted into a vein.

{1}------------------------------------------------

## 8. Performance Testing:

Bench, biocompatibility, sterility are employed upon submission of this 510(K) premarket notification according to the _Guidance on Premarket Notification for Intravascular Administration Sets document provided by CDRH/ FDA.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUN 0 3 2004

LILY Medical Supplies Company, Limited C/O Dr. Ke-Min-Jen Official Correspondent Roc Chinese-European Industrial Research Society No. 58, Fu-Chiun Street Hsin-Chu City, CHINA 300 (TAIWAN)

Re: K032767

K052767
Trade/Device Name: Lily Volume Meteric Administration Set Regulation Number: 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: FPA Dated: March 15, 2004 Received: March 22, 2004

Dear Dr. Jen:

We have reviewed your Section 510(k) premarket notification of intent to market the device We nave reviewed your bootion by (1) (1) +) + + + + + = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = relerenced above and have determined are to tegally marketed predicate devices marketed in increations for use stated in the chelosary to 10g.
interstate commerce prior to May 28, 1976, the enactment date of the Medical Device interstate collinered prior to thay 20, 1978, and establed in accordance with the forcements of Amendinents, of to devroes that have o Act (Act) that do not require approval of a premarket the Federal Food, Drug, and Cosment + to (1 co) .............................................................................................................................. approval appreation (1 Mrx). - 1 ou 1.0), - 1 ou 1.0, 1 on 1.0 min of the Act include Controls provisions of the rec. "The genting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III If your device is classified (see above) into controls. Existing major regulations affecting (PMA), it may be subject to Such adam 6 Federal Regulations, Title 21, Parts 800 to 898. In the Collections your device can oc found in the Sode of received.
addition, FDA may publish further announcements concerning your device in the Federal Register.

Legend: = Road =

{3}------------------------------------------------

Page 2 - Dr. Jen

Please be advised that FDA's issuance of a substantial equivalence determination does not Please be advised that FDA STISSuance of a substantial vice complies with other requirements mean that FDA has made a decemmance inter Jour Jour Federal agencies.
of the Act or any Federal statutes and regulations administered by segistering of the Act of ally receital statutes and regulations, but not limited to: registration You must colliply with an the Net 3 requirements and 801); good manufacturing practice and listing (21 CFR Fart 607), laooling (21 CFR Pat 820), and if 1
requirements as set forth in the quality systems (QS) regulation (21 CFR Part 620), and if requirements as set forth in the quality systems (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) This letter will anow you to oegil mationing , bostantial equivalence of your device to a premarket notification. The I DA Imaling of substantial on for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), ri you desire specific advice for your de recomments of the regulation prease contact une Office or Ochiphanes and (200). " (21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the may obtain of Small Manufacturers, International and Consumer Assistance at its toll-free Division (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Chiu-Ling, Ph.D.

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

K032767

### Indications for Use Statement IV.

Applicant: I I ILY MEDICAL SUPPLIES CO., LTD.

510(k) Number: ___K032767

Device Name: LILY VOLUME METERIC ADMINISTRATION SET,

CIB-100ES, CIB-120ES, CIB-150ES

Indications for Use:

.

The LILY Volume Meteric Administration Set is designed to administer fluids from a container to a patient's vascular system through a catheter inserted into a vein.

Prescription Use

AND/OR

Over-The-Counter Use

(Part 21 CFR 801 Subpart D)

(21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off

---

**Source:** [https://fda-staging.innolitics.com/device/K032767](https://fda-staging.innolitics.com/device/K032767)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com