DIMENSION TOTAL TRIIODOTHYRONINE CALIBRATOR (RC414)

K032697 · Dade Behring, Inc. · JIS · Nov 7, 2003 · Clinical Chemistry

Device Facts

Record IDK032697
Device NameDIMENSION TOTAL TRIIODOTHYRONINE CALIBRATOR (RC414)
ApplicantDade Behring, Inc.
Product CodeJIS · Clinical Chemistry
Decision DateNov 7, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1150
Device ClassClass 2

Indications for Use

The T3 Calibrator is intended for use in the calibration of Total Triiodothyronine (T3) method on the Dimension® clinical chemistry system with the Heterogeneous Immunoassay Module.

Device Story

The Dimension® Total Triiodothyronine (T3) Calibrator (RC414) is a liquid, multi-level calibrator kit consisting of 10 vials (two each at five levels). Level 1 contains stripped human serum; levels 2-5 contain L-triiodothyronine in a stripped human serum base. The device is used in a clinical laboratory setting to calibrate the T3 assay on the Dimension® clinical chemistry system equipped with a Heterogeneous Immunoassay Module. By establishing a calibration curve, the device enables the system to accurately quantify T3 concentrations in patient samples, supporting clinical diagnostic decision-making regarding thyroid function.

Clinical Evidence

bench testing only

Technological Characteristics

Liquid calibrator kit; 5 levels (0, 1, 2, 4, 6.5 ng/mL); matrix: stripped human serum base; analyte: L-triiodothyronine. Designed for use with Dimension® clinical chemistry system Heterogeneous Immunoassay Module.

Indications for Use

Indicated for use as a calibrator for the Total Triiodothyronine (T3) assay on the Dimension® clinical chemistry system with the Heterogeneous Immunoassay Module.

Regulatory Classification

Identification

A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)

Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Submission Summary (Full Text)

{0} 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY DEVICE ONLY TEMPLATE A. 510(k) Number: K032697 B. Analyte: Triiodothyronine (T3) C. Type of Test: N/A D. Applicant: Dade Behring Inc. E. Proprietary and Established Names: Dade Behring Dimension® Total Triiodothyronine (T3) Calibrator (RC414) F. Regulatory Information: 1. Regulation section: §862.1150, Calibrator, Primary 2. Classification: Class II 3. Product Code: JIS 4. Panel: 75 (Chemistry) G. Intended Use: 1. Indication(s) for use: The T3 Calibrator is intended for use in the calibration of Total Triiodothyronine (T3) method on the Dimension® clinical chemistry system with the Heterogeneous Immunoassay Module. 2. Special condition for use statement(s): None 3. Special instrument Requirements: Dade Behring Dimension® clinical chemistry system with the Heterogeneous Immunoassay Module. H. Device Description: The Dimension® Total Triiodothyronine (T3) Calibrator (RC414) is a liquid product. The kit consists of 10 vials, two each at levels 1 through 5. Level 1 vials contain 2 mL of stripped human serum. Vials for levels 2 through 5, contain 1 mL with concentrations of L-Triiodothyronine in a stripped human serum base. {1} Page 2 of 3 # I. Substantial Equivalence Information: 1. Predicate device name(s): Opus Total T3 Calibrator 2. Predicate K number(s): K953160 3. Comparison with predicate: | | Dimension® T3 Calibrator | Opus Total T3 Calibrator | | --- | --- | --- | | Intended Use | Calibrator | Calibrator | | Analyte | T3 | T3 | | Matrix | Stripped human serum base | Stripped human serum base | | Form | Liquid | Liquid | | Volume | 2 mL per vial @ level 1, 1 mL per vial @ levels 2-5 | 2 mL per vial @ level 1, 1 mL per vial @ levels 2-6 | | Levels | 5 levels @ 0, 1, 2, 4, 6.5 ng/mL | 6 levels @ 0, 0.5, 1, 2, 4, 6 ng/mL | # J. Standard/Guidance Document Referenced (if applicable): FDA guidance “Abbreviated 510(k) submissions for In Vitro Diagnostic Calibrators” # K. Test Principle: N/A # L. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: N/A b. Linearity/assay reportable range: N/A c. Traceability (controls, calibrators, or method): The assigned values are referenced to USP L-Triiodothyronine (USP Catalog # 36800). Users of this product will be notified if there is a change in the assigned values. Systeme International d'Unites Conversion factor: ng/mL x 1.536 = nmol/L d. Detection limit: N/A e. Analytical specificity: N/A {2} f. Assay cut-off: N/A 2. Comparison studies: a. Method comparison with predicate device: N/A b. Matrix comparison: N/A 3. Clinical studies: a. Clinical sensitivity: N/A b. Clinical specificity: N/A c. Other clinical supportive data (when $a$ and $b$ are not applicable): 4. Clinical cut-off: N/A 5. Expected values/Reference range: The following table shows the calibrator values obtained on the Dimension® Clinical Chemistry System with Heterogeneous Immunoassay Module. Please note the assignment is absolute, there is no range: | | Assigned Value | Units | Assigned Value | SI Units | | --- | --- | --- | --- | --- | | Level 1 | 0 | ng/mL | 0 | nmol/L | | Level 2 | 1 | ng/mL | 1.5 | nmol/L | | Level 3 | 2 | ng/mL | 3.1 | nmol/L | | Level 4 | 4 | ng/mL | 6.1 | nmol/L | | Level 5 | 6.5 | ng/mL | 9.9 | nmol/L | # M. Conclusion: Based on the information provided, I recommend that the Dade Behring Dimension® Total Triiodothyronine (T3) Calibrator is substantially equivalent to the Opus Total T3 Calibrator (K953160) predicate device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...