ALOETOUCH POWDER FREE NEOPRENE SURGICAL GLOVES, STERILE, TURQUOISE COLOUR AND COATED WITH ALOE VERA
K032647 · Wrp Asia Pacific Sdn. Bhd. · KGO · Nov 7, 2003 · General, Plastic Surgery
Device Facts
Record ID
K032647
Device Name
ALOETOUCH POWDER FREE NEOPRENE SURGICAL GLOVES, STERILE, TURQUOISE COLOUR AND COATED WITH ALOE VERA
Applicant
Wrp Asia Pacific Sdn. Bhd.
Product Code
KGO · General, Plastic Surgery
Decision Date
Nov 7, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4460
Device Class
Class 1
Indications for Use
The Powder Free Neoprene Surgical Glove, Sterile, Turquoise Colour and coated with Aloe Vera is made of synthetic rubber latex intended to be worn on the hand of healthcare personnel, operating room personnel and similar personnel to prevent contamination between the healthcare or similar personnel and the patient's body, fluids, waste, or environment.
Device Story
Powder-free, sterile, turquoise-colored neoprene surgical glove coated with Aloe Vera. Designed for use by surgeons and operating room personnel to provide a protective barrier between the wearer and the patient's body, fluids, waste, or environment. Device functions as a physical barrier to prevent cross-contamination. No electronic or software components.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Material: Synthetic rubber latex (neoprene). Coating: Aloe Vera. Form factor: Sterile, powder-free surgical glove. Standards: ASTM D 3577-01aE2 (dimensions, physical properties, pinholes), ASTM D 6124-01 (powder-free), FDA 21 CFR 800.20. Biocompatibility: Primary skin irritation and dermal sensitization testing passed.
Indications for Use
Indicated for use by surgeons and/or operating room personnel to protect a surgical wound from contamination.
Regulatory Classification
Identification
A non-powdered surgeon's glove is a device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination. A non-powdered surgeon's glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for WRP Asia Pacific Sdn Bhd. The logo consists of the letters "WRP" in a stylized font, with the words "WRP Asia Pacific Sdn Bhd" written to the right of the letters. Below the company name is the number 147817V.
K032647
NOV - 7 2003
## 510(k) SUMMARY
#### 1.0 Submitter:
| Name: | WRP Asia Pacific Sdn Bhd |
|------------|-------------------------------------------------------------------------------------------------------------|
| Address: | Lot 1, Jalan 3, Kawasan Perushaan Bandar Baru Salak Tinggi,<br>43900 Sepang, Selangor Darul Ehsan, MALAYSIA |
| Phone No.: | +60 3 8706 1486 |
| Fax No.: | +60 3 8706 1557 |
Date of Summary Prepared: 25 AUG 2003
#### 2.0 Contact Person:
| Name: | Mr. Terence Lim |
|------------|-----------------|
| Phone No.: | +60 3 8706 1486 |
| Fax No.: | +60 3 8706 1557 |
#### 3.0 Name of the device:
| Trade Name: | 1. Aloetouch, and<br>2. Multiple or Customer's Trade Name |
|----------------------|-------------------------------------------------------------------------------------------|
| Device Name: | Powder Free Neoprene Surgical Glove, Sterile, Turquoise Colour and coated with Aloe Vera. |
| Common Name: | Surgical Gloves |
| Classification Name: | Surgeon's Gloves (per 21 CFR 878.4460) |
#### Identification of The Legally Marketed Device: 4.0
Class I Powder Free Synthetic rubber latex Surgeon's gloves, 79KGO, that meets all
the requirements of ASTM standard D 3577 – 01a22 Type 2 and FDA 21 CFR 800.20.
#### 5.0 Description of The Device:
The Powder Free Neoprene Surgical Glove, Sterile, Turquoise Colour and coated with Aloe Vera meets all the requirements of ASTM standard D 3577 - 01a22 and FDA 21 CFR 800.20.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for WRP Asia Pacific Sdn Bhd. The logo consists of the letters "WRP" in a bold, sans-serif font, enclosed in a black oval shape. To the right of the logo is the text "WRP Asia Pacific Sdn Bhd" in a similar font. Below the text is the number "147817 V".
#### 6.0 Intended Use of the Device:
The Powder Free Neoprene Surgical Glove, Sterile, Turquoise Colour and coated with Aloe Vera is made of synthetic rubber latex intended to be worn on the hand of healthcare personnel, operating room personnel and similar personnel to prevent contamination between the healthcare or similar personnel and the patient's body, fluids, waste, or environment.
#### Summary of The Technological Characteristics of The Device: 7.0
The Powder Free Neoprene Surgical Gloves, Sterile, Turquoise Colour and coated with Aloe Vera are summarized with the following technological characteristics compared to ASTM or equivalent standards.
| CHARACTERISTICS | STANDARDS | DEVICE<br>PERFORMANCE |
|-----------------------|------------------------------------------|-----------------------------------------|
| Dimensions | ASTM D 3577 - 01aE2 | Meets |
| Physical Properties | ASTM D 3577 - 01aE2 | Meets |
| Freedom from pinholes | ASTM D 3577 - 01aE2<br>FDA 21 CFR 800.20 | Meets |
| Powder-Free | ASTM D 6124 - 01 | Meets<br>< 2 mg/glove |
| Biocompatability | Primary Skin Irritation in<br>Rabbits | Passes<br>(Not a primary skin irritant) |
| | Dermal Sensitization | Passes<br>(Not a contact sensitizer) |
#### 8.0 Substantial Equivalent Based on Assessment of Non-Clinical Performance Data
The performance test data of the non-clinical tests that support a determination of substantial equivalence is the same as mentioned immediately above.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for WRP Asia Pacific Sdn Bhd. The logo consists of the letters "WRP" in a black oval shape, followed by the text "WRP Asia Pacific Sdn Bhd" in a bold, sans-serif font. Below the company name is the number "147817 V".
#### Substantial Equivalent Based on Assessment of Clinical Performance Data 9.0
Clinical data is not needed for gloves or for most devices cleared by the 510(k) process.
#### 10.0 Conclusion
It can be concluded that the Powder Free Neoprene Surgical Glove, Sterile, Turquoise Colour and coated with Aloe Vera will perform according to the glove performance standards referenced in section 7 above and meet ASTM standards, and FDA requirements for water leak test on pinhole AQL. Consequently, this device is substantially equivalent to currently marketed devices.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol featuring three stylized human profiles facing to the right, with flowing lines beneath them.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# NOV - 7 2003
Mr. Terence Lin Associate Manager, Quality Assurance Regulatory Affairs WRP Asia Pacific Sdn Bhd Lot 1, Jalan 3, Kawasan Perusahaan Bandar Baru Salak Tinggi 43900 Sepang, Selangor Darul Ehsan MALAYSIA
Re: K032647
Trade/Device Name: Aloetouch Powder Free Neoprene Surgical Gloves, Sterile, Turquoise Color Coated with Aloe Vera Regulation Number: 878.4460 Regulation Name: Surgeon's Glove Regulatory Class: I Product Code: KGO Dated: August 25, 2003 Received: August 27, 2003
Dear Mr. Lim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Lim
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4618. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Susakunos
fr. Shih-Li Liu, Ph.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for WRP Asia Pacific Sdn Bhd. The logo consists of the letters "WRP" in a stylized font, followed by the words "WRP Asia Pacific Sdn Bhd" in a more standard font. Below the company name is the number 147817V.
### INDICATIONS FOR USE
Applicant:
WRP Asia Pacific Sdn Bhd
510(k) Number (if known):
\$\k_{0.32647}\$
Device Name:
FREE NEOPRENE POWDER SURGICAL STERILE, GLOVES. TURQUOISE COLOUR AND COATED WITH ALOE VERA
Indications For Use:
The surgeon's glove is a device made of synthetic rubber latex intended to be worn by surgeons and/or operating room personnel to protect a surgical wound from contamination.
Pat By Lales Board Chy 18/6/03
510(k) Number:
Concurrence of CDRH, Office of Device Evaluation (ODE)
OR
Prescription Use (Per 21 CFR 801.109) Over-The-Counter _____________________________________________________________________________________________________________________________________________________________
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.