The PoleStar N-20 is a Magnetic Resonance Diagnostic Device intended to produce transverse, sagittal, coronal, and oblique 2D and 3D images of the extremities and selected sections of the head. The images produced by the PoleStar N-20 reflect the spatial distribution of protons (Hydrogen Nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2) and T2*. The PoleStar N-20 is intended to be used intraoperatively in a standard operating room. When interpreted by trained physicians, the MR images provide information that can be useful in determining a diagnosis.
Device Story
PoleStar N-20 is a compact, open-design intraoperative MRI system utilizing a permanent magnet. Device acquires 2D and 3D images of extremities and head sections using various pulse sequences (spin echo, gradient echo, fast spin echo, steady state free precession). Operated by clinical staff in standard operating rooms; unit is mobile, moving between operating table and storage cabinet. System transforms proton magnetic resonance signals into visual images based on T1, T2, T2*, and proton density. Output is displayed for physician interpretation to aid intraoperative clinical decision-making. Benefits include real-time or intraoperative diagnostic imaging capability without requiring patient transport to a dedicated radiology suite.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on bench testing and technical comparison of imaging parameters, magnet specifications, and system performance characteristics against the predicate device.
Technological Characteristics
Permanent magnet MRI system; 0.13T field strength. Features closed-loop water cooling for gradients. Imaging modes: 2D/3D, spin echo, gradient echo, fast spin echo. Connectivity: DICOM 3.0 interface. Computer system: Pentium 586 CPU, 64MB RAM, 4xDSP C44 TI array processor. Form factor: compact, mobile unit for intraoperative use. No cryogens used.
Indications for Use
Indicated for patients requiring diagnostic imaging of extremities and selected sections of the head. Used intraoperatively in standard operating rooms to provide MR images for physician interpretation to assist in diagnosis.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
SEP 1 2 2003
Image /page/0/Picture/2 description: The image shows the logo for Odin Medical Technologies. The logo features the word "ODIN" in bold, sans-serif font, with the words "Medical Technologies" in a smaller font size underneath. A stylized, incomplete circle or ring is positioned to the right of the word "ODIN", with a small, solid circle placed at the top of the ring, giving the impression of a planet with a ring system. The overall design is clean and modern.
# Special 510(k) Summary of Safety and Effectiveness
The Following Special 510(k) Summary of Safety and Effectiveness has been prepared pursuant to requirements for 510(k) summaries specified in 21 CFR § 807.92(a).
#### 807.92(a)(1) - Submitter Details:
| Submitter name:<br>Address:<br>Phone:<br>Fax:<br>E-mail: | Einav Shlomovitz -- Quality and Regulatory Director<br>P.O. Box 548, Yokneam Elit 20698, Israel<br>+972-4-959-1010<br>+972-4-959-1011<br>einav@odin.co.il |
|----------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Einav Shlomovitz -Quality and Regulatory Director |
| Date: | August 4, 2003 |
### 807.92(a)(2) - Device Details:
| Trade Name and Common Name: | PoleStar N-20 - Magnetic Resonance<br>Diagnostic Device |
|-----------------------------|-----------------------------------------------------------------------------------------------|
| Classification: | 21 CRF 892.1000 Magnetic Resonance<br>Diagnostic Device. |
| Class: | II<br>MRDD were reclassified by FDA from<br>Class III to Class II effective July 28,<br>1998. |
| Product Code: | LNH – Magnetic Resonance Imaging<br>System |
{1}------------------------------------------------
### 807.92(a)(3) -- Predicate Devices:
The PoleStar N-20 is comparable to Odin's PoleStar N-10
| Medical<br>Device Name | Applicant Name | 510(k)<br>Number | Classification |
|------------------------|---------------------------------------------------|------------------|-------------------|
| | PoleStar N-10 Odin Medical Thechnologies Ltd. | K010850 | l Class Il device |
Additional Substantial Equivalence Information is provided in the attached Substantial Equivalence Comparison Table.
#### 807.92(a)(4) - Device Description:
The PoleStar N-20 utilizes a permanent magnet to acquire 2D single-slice, multi slice, and 3d volume images. A wide variety of pulse sequences are provided to the operator, including spin echo, gradient echo, fast spin echo, and steady state free precession acquisitions. The PoleStar N-20 is a widely open and compact Intraoperative MRI unit intended to be used in a typical pre-existing operating room. The PoleStar N-20 can be moved within the room between procedures, from the operating table to its Magnet Storage Cabinet, thus allowing the operating room to be used for any type of surgery.
{2}------------------------------------------------
### 807.92(a)(5) - Device Intended Use:
#### The general purpose of the device as defined in 21 CFR 892.1000:
The PoleStar N-20 is a Magnetic Resonance Diagnostic Device intended to produce transverse, sagittal, coronal, and oblique 2D and 3D images of the extremities and selected sections of the head. The images produced by the PoleStar N-20 reflect the spatial distribution of protons (Hydrogen Nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2) and T2*.
- . Anatomical regions:
Diagnostic uses:
extremities and selected sections of the head.
- . Nuclei excited:
.
T1, T2, T2* and density weighted imaqing.
The PoleStar N-20 is intended to be used intraoperatively in a standard operating room. When interpreted by trained physicians, the MR images provide information that can be useful in determining a diagnosis.
H-1
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Odin Medical Technologies. The word "odin" is in bold, sans-serif font, with the "o" slightly larger than the other letters. Below "odin" is the phrase "Medical Technologies" in a smaller, italicized font. A circular design element is to the right of the word "odin".
## 807.92(a)(6) – Substantial Equivalence Comparison Table:
| Model | Odin | Odin |
|-------------------------------------------|----------------------------------------------------|----------------------------------------------------|
| parameter | PoleStar N-10 (K010850) | PoleStar N-20 |
| Clinical application | Extremities and selected<br>sections of the head | Extremities and selected<br>sections of the head |
| Magnet type | Permanent | Permanent |
| Field strength | 0.12T | 0.13T |
| 5 gauss fringe field<br>(radial/axial, m) | 1.5 | 2.2 |
| Shimming | Passive, active | Passive, active |
| Gradient subsystem | | |
| Strength mT/m | 25 | 22 |
| Rise time to 10mT/m | <1 | <1 |
| msec | | |
| Computer system | | |
| - CPU: | Pentium 586 | Pentium 586 |
| - Memory size [MB] | 64 | 64 |
| array processor | 4xDSP C44 TI | 4xDSP C44 TI |
| - Memory size [MB] | 4000 | 4000 |
| storage media | magnetic disk, floppy disk | magnetic disk, floppy disk |
| number of images stored | 5000 | 5000 |
| Imaging modes: | | |
| - single | Yes | Yes |
| - multislice | Yes | Yes |
| - volume study | Yes | Yes |
| - other | No | No |
| Reconstruction time: | | |
| - single slice, sec | <3/slice | <3/slice |
| - multislice, sec | <3/slice | <3/slice |
| - volume sec | <20/volume | <20/volume |
| Model | Odin | Odin |
| parameter | PoleStar N-10 (K010850) | PoleStar N-20 |
| Cardiac gating<br>(ECG/peripheral) | No | No |
| Respiratory gating | No | No |
| Angiography | Optional | Optional |
| Spectroscopy | No | No |
| Imaging: | | |
| - pulse sequence | Spin Echo, Fast Spin Echo,<br>Gradient Echo, 2D 3D | Spin Echo, Fast Spin Echo,<br>Gradient Echo, 2D 3D |
| - repetition time, msec | 10-5000 increments of 1 | 10-5000 increments of 1 |
| - echo time, msec | 3-150 | 3-150 |
| - inversion time, msec | N/A | N/A |
| - slice thickness, mm | 2-10 | 2-10 |
| - scan orientation | Transverse, coronal,<br>sagittal, oblique | Transverse, coronal,<br>sagittal, oblique |
| - measuring matrix | 64x64 to 256x256 steps of<br>1 in phase encoding | 64x64 to 256x256 steps of<br>1 in phase encoding |
| - display matrix | 1024x768 | 1024x768 |
| - pixel intensity | 0-4095 | 0-4095 |
| Surface coils: | | |
| - spine | No | No |
| - knee | Yes | Yes |
| - neck | No | No |
| - TMJ | No | No |
| - extremity | Yes | Yes |
| - head | Yes | Yes |
| - breast | No | No |
| - shoulder | No | No |
| - others | No | No |
| Model | Odin | |
| parameter | PoleStar N-10 (K010850) | PoleStar N-20 |
| Bore diameter or WxH, cm | 24.5x39 | 25.2x42 |
| Bore features | Open access to patient | Open access to patient |
| Cooling system type | Closed loop water cooling<br>(Gradients only) | Closed loop water cooling<br>(Gradients only). |
| Cryogen use | No | No |
| Magnet weight, kg | 330 | 400 |
| HxWxD, cm | 145x96x120 | 153x97x120 |
| Filed Of View (FOV),cm | 5-18 | 5-20 |
| Dicom 3.0 interface | Yes | Yes |
| Power requirements: | | |
| - line voltage, V | 3x208 (3 phase) | 3x208 (3 phase) |
| - Kva | 16 | 15 |
| - A/C, BTU/hr | <10000 | <10000 |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Odin Medical Technologies. The word "odin" is in bold, black letters, with the words "Medical Technologies" in a smaller font below. There is a circular design to the right of the word "odin".
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Odin Medical Technologies. The word "odin" is in large, bold, sans-serif font, with a stylized circle above the "i". Below "odin" are the words "Medical Technologies" in a smaller, sans-serif font. A partial circle is behind the word "odin".
·
{6}------------------------------------------------
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP 1 2 2003
Mr. Einav Shlomovitz Ouality and Regulatory Director Odin Medical Technologies Ltd. P.O. Box 548 Yokneam Elit, 20698 ISRAEL
Re: K032541
Trade/Device Name: PoleStar N-20 Regulatory Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: 90 LNH Dated: August 4, 2003 Received: August 18, 2003
Dear Mr. Shlomovitz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{7}------------------------------------------------
Page 2
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:
| 8xx. 1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301).594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours.
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
510(k) Number (if known):
Device Name: PoleStar N-20
Indication For Use:
The PoleStar N-20 is a Magnetic Resonance Diagnostic Device intended to produce transverse, sagittal, coronal, and oblique 2D and 3D images of the extremities and selected sections of the head. The images produced by the PoleStar N-20 reflect the spatial distribution of protons (Hydrogen Nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2) and T2*.
Anatomical regions: extremities and selected sections of
the head.
| Nuclei excited: | H-1 |
|------------------|----------------------------------|
| Diagnostic uses: | T1, T2, T2* and density weighted |
| imaging. | |
The PoleStar N-20 is intended to be used intraoperatively in a standard operating room. When interpreted by trained physicians, the MR images provide information that can be useful in determining a diagnosis.
(PLEASE DO NOT WRITE BELOW THE LINE -- CONTINUE ON ANOTHER PAGE IF NECESSARY)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
Daniel A. Bergman
(Division Sign-Off) Division of Reproductive, and Radiological Devices 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.