← Product Code [GKZ](/productcode/GKZ) · K032524

# ADVIA 60 HEMATOLOGY ANALYZER (K032524)

_Bayer Healthcare, LLC · GKZ · Sep 8, 2003 · Hematology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K032524

## Device Facts

- **Applicant:** Bayer Healthcare, LLC
- **Product Code:** [GKZ](/productcode/GKZ.md)
- **Decision Date:** Sep 8, 2003
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 864.5220
- **Device Class:** Class 2
- **Review Panel:** Hematology

## Indications for Use

For in-vitro diagnostic use in the quantitative determination of blood cells and hematologic parameters in whole blood.

## Device Story

Hematology analyzer modification; software update to improve measurement linearity for WBC, RBC, HGB, HCT, and PLT parameters. Device processes blood samples to provide quantitative hematological measurements. Used in clinical laboratory settings by trained personnel. Output used by clinicians for diagnostic assessment of blood cell counts. Modification expands linearity ranges: WBC (0-100), RBC (0-8.0), HGB (0-26), HCT (0-80), PLT (0-2200/4000).

## Clinical Evidence

Bench testing only; verification and validation activities performed per design controls to confirm performance against expanded linearity specifications for WBC, RBC, HGB, HCT, and PLT parameters.

## Technological Characteristics

Hematology analyzer; software-based measurement processing; expanded linearity ranges for blood parameters. No changes to fundamental scientific technology.

## Regulatory Identification

An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

SPECIAL 510(k): Device Modification
ODE Review Memorandum

To: THE FILE
RE: DOCUMENT NUMBER K 032524

This 510(k) submission contains information/data on modifications made to the SUBMITTER'S own Class II, Class III or Class I devices requiring 510(k). The following items are present and acceptable (delete/add items as necessary):

1. The name and 510(k) number of the SUBMITTER'S previously cleared device. (For a preamendment device, a statement to this effect has been provided.)
2. Submitter's statement that the INDICATION/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use, package labeling, and, if available, advertisements or promotional materials.
3. A description of the device MODIFICATIONS, including clearly labeled diagrams, engineering drawings, photographs, user's and/or service manuals in sufficient detail to demonstrate that the FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed.

This change was for software modification and improved linearity for WBC, RBC, HGB, HCT, and PLT measurement.

4. Comparison Information (similarities and differences) to applicant's legally marketed predicate device including, labeling, intended use, physical characteristics, and performance data. The differences include changes to linearity as followings: WBC from 0.5-80 to 0-100; RBC from 0.2-7.5 to 0-8.0; HGB from 2.5-23 to 0-26; HCT from 11.6-62.4 to 0-80; PLT from 10-1000 to 0-2200 (HGB ≥ 2) and 0-4000(HGB &lt; 2).

5. A Design Control Activities Summary which includes:

a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis
b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied
c) A declaration of conformity with design controls. The declaration of conformity should include:

i) A statement signed by the individual responsible, that, as required by the risk analysis, all verification and validation activities were performed by the designated individual(s) and the results demonstrated that the predetermined acceptance criteria were met, and
ii) A statement signed by the individual responsible, that the manufacturing facility is in conformance with design control procedure requirements as specified in 21 CFR 820.30 and the records are available for review.

6. A Truthful and Accurate Statement, a 510(k) Summary or Statement and the Indications for Use Enclosure (and Class III Summary for Class III devices).

The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the particular modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared (or their preamendment) device.

Comments

(Reviewer's Signature)

(Date)

revised:3/27/98

---

**Source:** [https://fda-staging.innolitics.com/device/K032524](https://fda-staging.innolitics.com/device/K032524)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
