# MARK 5 NUVO (M5C5) (K032509)

_Nidek Medical Products, Inc. · CAW · Mar 18, 2004 · Anesthesiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K032509

## Device Facts

- **Applicant:** Nidek Medical Products, Inc.
- **Product Code:** [CAW](/productcode/CAW.md)
- **Decision Date:** Mar 18, 2004
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.5440
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Therapeutic

## Indications for Use

The Mark 5 Nuvo (M5C5) Oxygen Concentrator is intended solely for medical use in oxygen therapy programs under the supervision of a physician. This device is available by prescription only and is not intended to support or sustain life.

## Device Story

The Mark 5 Nuvo (M5C5) is an AC-powered oxygen concentrator; utilizes pressure swing adsorption (PSA) to separate oxygen from ambient air; draws air via piston-type compressor; uses molecular sieve adsorbent to retain nitrogen; cycle controlled by motorized rotary 4-way valve; provides oxygen-enriched gas (90 ±3%) at 5 l/min ±10%. Used in clinical or home settings; operated by patients or caregivers under physician prescription. Includes humidifier, tubing, and nasal cannula. Output monitored by healthcare providers to adjust oxygen therapy. Benefits patients by providing supplemental oxygen for respiratory needs.

## Clinical Evidence

Bench testing only. Device performance validated against electrical safety and electromagnetic compatibility standards including EN 60601-1-2:2001, IEC 60601-1-2:2001, UL 60601-1:2003, and CAN/CSA-C22.2 No 601.1-M90. Testing conducted at Intertek Testing Services NA Inc. (Report numbers 3040276-27-1-1 and 3040276A).

## Technological Characteristics

AC-powered oxygen concentrator; PSA technology; molecular sieve adsorbent (OxySiv); piston-type compressor; motorized rotary 4-way valve; tubeaxial fan cooling; flow meter (Thorpe tube) for flow control; pressure sensor for monitoring; 10 Amp circuit breaker; steel spring compressor support; contoured cabinet design. Validated to EN 60601-1-2, IEC 60601-1-2, UL 60601-1, and CAN/CSA-C22.2 standards.

## Regulatory Identification

A portable oxygen generator is a device that is intended to release oxygen for respiratory therapy by means of either a chemical reaction or physical means (e.g., a molecular sieve).

## Predicate Devices

- Mark-5 (M5C5) Concentrator ([K883591](/device/K883591.md))
- Model 590 Oxygen Concentrator ([K895141](/device/K895141.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the text string "K032509" in a handwritten style. The text is oriented diagonally, running from the bottom left to the upper right corner of the image. The characters are written in a simple, sans-serif font, and the overall impression is that of a quick, informal notation.

MAR 1 8 2004

______________________________________________________________________________________________________________________________________________________________________________

## 510(k) Summary

| 510(k)<br>Summary             | This summary of 510(k) safety and effectiveness information is being<br>submitted in accordance with the requirements of 21 C.F.R. § 807.92.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                                |                                      |
|-------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------|--------------------------------------|
| Submitter                     | Nidek Medical Products, Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                                |                                      |
| Contact Person                | Jennifer McWilliams, Quality Assurance Manager<br>3949 Valley East Industrial Drive<br>Birmingham, AL 35217<br>Phone (205) 856-7200 ext. 215<br>Fax (205) 856-0563<br>jmcwilliams@nidekmedical.com                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                |                                      |
| Date Prepared                 | August 12, 2003                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                |                                      |
| Name                          | Mark 5 Nuvo (M5C5)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                |                                      |
| Classification<br>Names       | Oxygen Concentrator                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                |                                      |
| Device<br>Classification      | Classification: Generator, Oxygen Portable<br>Classification Panels: Anesthesiology<br>Regulation Number: 21CFR 868.5440                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                                |                                      |
| Predicate<br>Device(s)        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                |                                      |
|                               | Mark-5 (M5C5) Concentrator<br>Model 590 Oxygen Concentrator                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | K883591<br>K895141             | November 2, 1988<br>October 31, 1989 |
| Performance<br>Standards      | Performance standards have not been established by the FDA under<br>section 514 of the Federal, Food, Drug and Cosmetic Act                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                |                                      |
| Device<br>Description         | The Mark 5 Nuvo (M5C5) oxygen concentrator is an AC powered<br>device that provides a high level of inspired oxygen by separating<br>oxygen from ambient air utilizing pressure swing adsorbers (PSA).<br>Air is drawn into the device with a piston-type compressor and<br>exposed to molecular sieve adsorbent that selectively retains nitrogen<br>and other components until they are released when the pressure is<br>vented to the atmosphere. This cycle is controlled by a motorized<br>rotary 4-way valve and protected from over pressurization by the<br>compressor's pressure relief valve. The device provides a nominal<br>oxygen enriched gas concentration of 90 ±3% at a flow rate of 5<br>l/min ±10%. It is not a life-supporting, life-sustaining or sterile<br>device |                                |                                      |
| Device Description Continued  | The Mark 5 Nuvo (M5C5) oxygen concentrator is a durable, reusable, semi-portable unit weighing approximately 50 pounds [23 kg]. The device is available in both 115V and 230V models that have been designed and validated according to applicable requirements of EN 60601-1-2:2001, IEC 60601-1-2:2001, UL 60601-1:2003, and CAN/CSA-C22.2 No 601.1-M90 with A1 & A2:1999 as appropriate to the area of usage. The Mark 5 Nuvo (M5C5) oxygen concentrator is intended to be used with one of the many legally marketed humidifiers, connecting tubing and nasal cannula as prescribed. One of these devices may optionally be included with the device.                                                                                                                                 |                                |                                      |
| Indications for Use           | The Mark 5 Nuvo (M5C5) Oxygen Concentrator is intended solely for medical use in oxygen therapy programs under the supervision of a physician. This device is available by prescription only and is not intended to support or sustain life.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                                |                                      |
| Technological Characteristics | The Mark 5 Nuvo (M5C5) contains features and functions that are similar to the Mark-5 (M5C5) concentrator. The following table briefly describes the proposed changes that do not affect the intended use or alter the fundamental scientific technology of the device.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                |                                      |
|                               | Component                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Predicate Device               | Proposed Device                      |
|                               | Compressor Cooling                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Squirrel Cage Blower           | Tubeaxial Fan                        |
|                               | Molsiv Adsorbent                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | MO2                            | OxySiv                               |
|                               | Flow Control                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Selectable Fixed-Orifice Valve | Flow Meter (Thorpe Tube)             |
|                               | Outlet Pressure                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | 9 PSI                          | 7 PSI                                |
|                               | Circuit Breaker                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | none                           | 10 Amp                               |
|                               | Cabinet Dimensions                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Rectangular Design             | Contoured Design                     |
|                               | Compressor Support                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Natural Rubber Isolators       | Steel Springs                        |
|                               | Labeling                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Words                          | Symbols                              |
|                               | Pressure Measurement                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Pressure Switches              | Pressure Sensor                      |
| Nonclinical Performance       | The device was tested to applicable requirements EN 60601-1-2:2001, IEC 60601-1-2:2001, UL 60601-1:2003, and CAN/CSA-C22.2 No 601.1-M90 with A1 & A2:1999 as appropriate to the area of usage. Testing was conducted in June 2003 at Intertek Testing Services NA Inc. located at Duluth, GA. and documented in report numbers 3040276-27-1-1and 3040276A.                                                                                                                                                                                                                                                                                                                                                                                                                                |                                |                                      |
| Conclusion                    | The Mark 5 Nuvo (M5C5) is substantially equivalent to the legally marketed Mark-5 (M5C5) concentrator.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                |                                      |

.

{1}------------------------------------------------

.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features the department's name encircling a symbol. The symbol is a stylized representation of a bird-like figure, possibly an eagle, with three wing-like shapes above a wavy line, suggesting water or movement. 

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAR 1 8 2004

Ms. Jennifer McWilliams Quality Assurance Manager Nidek Medical Products, Incorporated 3949 Valley East Industrial Drive Birmingham, AL 35217

Re: K032509

Trade Name: Mark 5 Nuvo (M5C5) Regulation Number: 21 CFR 868.5440 Regulation Name: Portable Oxygen Generator Regulatory Class: II Product Code: CAW Dated: February 18, 2004 Received: February 20, 2004

Dear Ms. McWilliams:

We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your Security premation is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encrosule, to regary manetee to fithe Medical Device Amendments, or to commerce prior to May 20, 1970, the enacember with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). and Costience Act (Act) that to not require appro the general controls provisions of the Act. The
 You may, therefore, market the device, subject to the general controls pro You may, merefore, market the devrees, saloject to an all registration, listing of the general controls provisions of the Fisc labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (see aboro) als. Existing major regulations affecting your device can
may be subject to such additional controls. Existing major may be subject to such additional occurrents, Title 21, Parts 800 to 898. In addition, FDA may be found in the Code of Productions
publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that IDA's issualled of a sualies with other requirements of the Act
that FDA has made a determination that your device complies . You must that FDA has made a decemmancia in administered by other Federal agencies. You must on and lists or any Federal statures and regulations daminders but not limited to: registration and listing (21 let comply with an the Act s requirements, morealing, arenes, and manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 -- Ms. Jennifer McWilliams

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

tinger

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Item #7

## Indications for Use

510(k) Number (if known): K032509

Mark 5 Nuvo (M5C5) Device Name:

The Mark 5 Nuvo (M5C5) Oxygen Concentrator is Indications for Use: intended solely for medical use in oxygen therapy programs under the supervision of a physician. This device is available by prescription only and is not intended to support or sustain life.

Prescription Use __ × (Part 21 CFR 801 Subpart D) AND/O ಗಿ

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE: ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Sottwle

(Division Sign-Off)
Division of Anesthesiology, General Hospital,
Infection Control, Dental Devices
510(k) Number: K032509

Page __ of ____

ber: 5037

---

**Source:** [https://fda-staging.innolitics.com/device/K032509](https://fda-staging.innolitics.com/device/K032509)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).