The Lifecore Stage-1 Angled Abutment is intended to attach to an endosseous implant and provide support and retention for single or multi-unit (splinted) cement retained restorations in the mandible or maxilla where angle correction is required.
Device Story
Lifecore Stage-1 Angled Abutment System consists of 15° and 20° angled abutments, screws, and try-ins; used with endosseous implants to support/retain single or multi-unit cement-retained restorations. Facilitates tooth positioning in mandible/maxilla when implant angulation requires correction due to limited bone or access. Operated by clinicians in dental settings. System provides mechanical support for prosthetic restorations; aids in achieving ideal aesthetic and functional positioning. Benefits include improved restorative outcomes in anatomically challenging implant sites.
Clinical Evidence
Bench testing only. Evaluated static and fatigue loads; performed electrochemical corrosion testing of uncoupled and coupled implant/restorative alloys in artificial saliva. All results within acceptable limits.
Technological Characteristics
Materials: Titanium alloy (Ti 6Al-4V E.L.) per ASTM F136. Components: 15° and 20° angled abutments, angled abutment screws (.048" hex), and try-ins. Mechanical connection to endosseous implants. No software or electronic components.
Indications for Use
Indicated for patients requiring support and retention for single or multi-unit cement-retained dental restorations in the mandible or maxilla where implant angle correction is necessary due to bone availability or access limitations.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
Predicate Devices
Pre-Angled Post (Implant Innovations (3i))
ITI Angled Abutment (Straumann USA)
Stage-1 Abutment System (Lifecore Biomedical, Inc.)
Submission Summary (Full Text)
{0}------------------------------------------------
#### Kos 2495 Section 2 - 510(k) Summary and Certification
[As required by 21 CFR 807.92(c)]
#### 1. Submitter's Name / Contact Person
Diane Brinza Regulatory Affairs Supervisor Lifecore Biomedical, Inc. 3515 Lyman Blvd. Chaska, MN 55318 952-368-6394 Ph: Fax: 952-368-4278
#### 2. General Information
| Trade Name | The Lifecore Stage-1 Angled Abutment System consists of the following products:<br>Regular Diameter (RDS) 15° Cement on Crown (COC) Abutment Regular Diameter (RDS) 20° COC Abutment Stage-1 Angled Abutment Screw, .048" Hex, Titanium Accessories: Stage-1 Angled Abutment Try-Ins |
|--------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common / Usual Name | Angled Abutment System |
| Classification Name | Lifecore Stage-1 Angled Abutment system is a precision attachment system intended for use with a Class III medical device, thereby the Stage-1 Angled Abutment System is considered Class III (per 21 CFR 872.3640). |
| Identification of Equivalent Devices | Pre-Angled Post manufactured by Implant Innovations (3i). ITI Angled Abutment manufactured by Straumann USA. Stage-1 Abutment System manufactured by Lifecore Biomedical, Inc. |
#### 3. Device Description
The Lifecore Stage-1 Angled Abutment System is designed for use with endosseous implants to provide support and retention for single or multi-unit (splinted) cement retained restorations in the mandible or maxilla where angle correction is required. The Stage-1 Angled COC Abutment and Screws will facilitate ideal tooth positioning due to lack of available bone, access limitation, or to correct angulation of implant placement.
The system consists of 15 and 20 degree angled abutments, angled abutment screw, and angled abutment try-ins. One angled abutment is packaged together with an angled abutment screw to allow the clinician to have the components required for clinical use. The angled abutments and angled abutment screw are manufactured from titanium alloy (Ti 6Al-4V E.L.) which conforms to ASTM Standard Specification F136.
{1}------------------------------------------------
# 4. Intended Use
The Lifecore Stage-1 Angled Abutment is intended to attach to an endosseous implant and provide support and retention for single or multi-unit (splinted) cement retained restorations in the mandible or maxilla where angle correction is required.
## 5. Technological Characteristic Comparisons
The Lifecore Stage-1 Angled Abutment System is substantially equivalent to the Pre-Angled Post manufactured by Implant Innovations (3i), and ITI Angled Abutment manufactured by Straumann USA, and the connection features of the Stage-1 Abutment System, manufactured by Lifecore Biomedical, Inc. Compared to the predicate devices, the Lifecore Stage-1 Angled Abutment System is substantially equivalent in intended use, design, material, and accessories.
### 6. Summary of Studies / Nonclinical Tests
The Lifecore Stage-1 Angled Abutment System has been subjected to testing to verify design specifications. Test results indicate the Stage-1 Angled Abutment has met the design criteria for static and fatigue loads and is likely to resist corrosion. Dimensional inspections are routinely performed, Electrochemical Corrosion Evaluation of Uncoupled and Coupled Implant and Restorative Alloys has been performed to determine corrosion properties when exposed to artificial saliva and to determine the galvanic corrosion properties of the Precious Alloy/CP titanium and Precious Alloy/Ti-6Al-4V galvanic couples. All have been found to be within acceptable limits.
### 7. Substantial Equivalence Comparison
The Lifecore Stage-1 Angled Abutment System is substantially equivalent to the following products:
| Implant Innovations (3i)<br>Pre-Angled Posts | Straumann USA ITI<br>Angled Abutments | Lifecore Stage-1<br>Abutments |
|-------------------------------------------------------------------------------------------|----------------------------------------------------------------|--------------------------------------------------------------------------|
| Implant Innovations, Inc.<br>4555 Riverside Dr.<br>Palm Beach Gardens, FL<br>33410<br>USA | Straumann USA<br>1601 Trapelo Road<br>Waltham, MA 02451<br>USA | Lifecore Biomedical, Inc.<br>3515 Lyman Blvd.<br>Chaska, MN 55318<br>USA |
# 8. Conclusion (statement of equivalence)
The data submitted in this 510(k) is in support of substantial equivalency of Lifecore Stage-1 Angled Abutment System to the following commercially marketed devices:
- Implant Innovations (3i) Pre-Angled Posts .
- Straumann USA ITI Angled Abutments .
- . Lifecore Stage-1 Abutments
These current products as defined by their product literature, demonstrate the basis for the substantial equivalency relative to indications, materials, and design. The intended use of these devices is the same as the Lifecore Stage-1 Angled Abutment System.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows a logo with a stylized bird in flight inside a circle. The bird is composed of three curved lines that suggest feathers or wings. The circle is not a solid line but appears to be made up of small, broken segments, giving it a rough or textured look. The logo is black and white.
#### Public Health Service
Rockville MD 20850
SEP 1 7 2003
Ms. Diane Brinza Regulatory Affairs Supervisor Lifecore Biomedical, Incorporated 3515 Lyman Boulevard Chaska, Minnesota 55318-3051
Re: K032495
Trade/Device Name: Lifecore Stage-1 Angled Abutment System Regulation Number: 872.3640 Regulation Name: Endosseous Implant Regulatory Class: III Product Code: DZE Dated: August 11, 2003 Received: August 13, 2003
Dear Ms. Brinza:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice. labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
#### Page 2 -Ms. Brinza
Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4613. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Susen Runne
Susan Runner, DDS, MA Interim Director Division of Anesthesiology, General Hospital Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# INDICATIONS FOR USE STATEMENT
510(k) Number (if known):
Page 1 of 1
Device Name: Lifecore Stage-1 Angled Abutment System
Indications for Use:
The Lifecore Stage-1 Angled Abutment is intended to attach to an endosseous implant and provide support and retention for single or multi-unit (splinted) cement retained restorations in the mandible or maxilla where angle correction is required.
K032495 Lee Mulky fan KIR
esthesiology, General Hospital, ontrol. Dental Devices
510(k) Number ..
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.