← Product Code [GEI](/productcode/GEI) · K032452

# MODIFICATION TO: STELLARTECH COAGULATION SYSTEM, MODELS 1100C-115 & 100C-230 (K032452)

_Stellartech Research Corp. · GEI · Aug 21, 2003 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K032452

## Device Facts

- **Applicant:** Stellartech Research Corp.
- **Product Code:** [GEI](/productcode/GEI.md)
- **Decision Date:** Aug 21, 2003
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The Stellartech Coagulation System is indicated for use in the coagulation of bleeding and non-bleeding sites in the gastrointestinal tract including but not limited to, the esophagus. Indications include Esophageal Ulcers, Mallory-Weiss tears, Arteriovenous Malformations, Angiomata, Barrett's Esophagus, Dieulafoy Lesions, and Angiodysplasia.

## Device Story

Stellartech Coagulation System is an electrosurgical unit for GI tract coagulation. System components include a generator, probe connection module, coagulation probe, optional sizing probe, and optional footswitch. The probe connects to the module, which connects to the generator. Used by clinicians in clinical settings to treat bleeding and non-bleeding lesions (e.g., ulcers, tears, malformations). The device delivers electrosurgical energy to target tissue to achieve coagulation. Benefits include management of GI bleeding and treatment of specific esophageal conditions.

## Technological Characteristics

Electrosurgical cutting and coagulation device. System comprises a generator, probe connection module, and probes. Operates via electrosurgical energy delivery. Class II device (21 CFR 878.4400).

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- Stellartech Coagulation System ([K013139](/device/K013139.md))
- Stellartech Coagulation System ([K023765](/device/K023765.md))
- Stellartech Coagulation System ([K032062](/device/K032062.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

'AUG 2 1 2003

KO32452

# 510(k) Safety Summary

### Name of Device A.

- Trade Name: Stellartech Coagulation System .
- Common Name: Electrosurgical Unit and Accessories .
- Classification Name: Device, Electrosurgical Cutting and Coagulation and Accessories . (21 CFR 878.4400)

## B. Predicate Devices

| Device                         | Premarket Notification |
|--------------------------------|------------------------|
| Stellartech Coagulation System | K013139, 12/18/01      |
| Stellartech Coagulation System | K023765, 11/29/02      |
| Stellartech Coagulation System | K032062, 07/29/03      |

## C. Device Description:

The Stellartech Coagulation System consists of the following components.

- Stellartech Coagulation Generator ●
- Stellartech Coagulation Probe Connection Module ●
- Stellartech Coagulation Probe ●
- Optional Stellartech Sizing Probe .
- Optional Stellartech Footswitch. .

The proximal end of the Stellartech Coagulation Probe connects to the Stellartech Coagulation Probe Connection Module. The proximal end of the Stellartech Coagulation Probe Connection Module cable connects to the Stellartech Coagulation Generator.

## D. Indicated Use

The Stellartech Coagulation System is indicated for use in the coagulation of bleeding and non-bleeding sites in the gastrointestinal tract including but not limited to, the esophagus. Indications include Esophageal Ulcers, Mallory-Weiss tears, Arteriovenous Malformations, Angiomata, Barrett's Esophagus, Dieulafoy Lesions, and Angiodysplasia.

### E. Technical characteristics

The technological characteristics of the Stellattech Coagulation System are substantially equivalent to those of the above listed predicate devices.

### F. Summary

By virtue of design, principles of operation, materials and intended use, the Stellartech Coagulation System is substantially equivalent to devices currently marketed in the United States.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, stacked on top of each other. The profiles are connected by a flowing line that forms the shape of a bird. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the image.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

AUG 2 1 2003

Mr. James R. Santos Senior Quality Engineer Stellartech Research Corporation 1346 Bordeaux Drive Sunnyvale, California 94089

Re: K032452

Trade/Device Name: Stellartech Coagulation System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: II Product Code: GEI Dated: August 9, 2003 Received: August 11, 2003

Dear Mr. Santos:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

Page 2 - Mr. James R. Santos

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely vours.

Sincerely yours,

Mark A. Millman

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

510(k) NUMBER (IF KNOWN):

KO32452

DEVICE NAME:

Stellartech Coagulation System

INDICATIONS FOR USE:

The Stellartech Coagulation System is indicated for use in the coagulation of bleeding and non-bleeding sites in the gastrointestinal tract including but not limited to, the esophagus. Indications include Esophageal Ulcers, Mallory-Weiss tears, Arteriovenous Malformations, Angiomata, Barrett's Esophagus, Dieulafoy Lesions, and Angiodysplasia.

Mark N Milken

(Division Sign-Off Division of General storative and Neurological L . nees

J(K) Number K03)452

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

ાર

Over-The-Counter-Use __

---

**Source:** [https://fda-staging.innolitics.com/device/K032452](https://fda-staging.innolitics.com/device/K032452)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
