NASAL JACKS MASK

K032433 · Resmed, Ltd. · BZD · Oct 24, 2003 · Anesthesiology

Device Facts

Record IDK032433
Device NameNASAL JACKS MASK
ApplicantResmed, Ltd.
Product CodeBZD · Anesthesiology
Decision DateOct 24, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5905
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Nasal Jacks Mask is an accessory to a non-continuous ventilator (respirator), intended for single-patient multi-use for adult patients prescribed continuous positive airway pressure (CPAP) and bi-level therapy in hospital, clinical, and home environments.

Device Story

Nasal Jacks Mask is a patient interface accessory for non-continuous ventilators (CPAP/bi-level). Device consists of nasal cushions supported by headgear; connects via tubing to flow generator. Delivers positive pressure ventilation to lungs non-invasively. Used in hospital, clinical, or home environments by adult patients. Provides seal at nostrils; intended for single-patient, multi-use. Design features one frame size with three cushion sizes. Benefits include comfort, ease-of-use, and effective seal for respiratory therapy.

Clinical Evidence

Bench testing only. Performance data verified pressure-flow characteristics, functional dead space (CO2 re-breathing), physical dead space, and flow impedance. Biocompatibility testing conducted per ISO 10993 standards.

Technological Characteristics

Nasal mask interface with headgear; three nasal cushion sizes; one frame size. Materials selected for biocompatibility per ISO 10993 or predicate history. Connects to external CPAP/bi-level flow generator via tubing. Non-invasive delivery of positive pressure.

Indications for Use

Indicated for adult patients requiring non-invasive ventilatory support via CPAP or bi-level therapy in hospital, clinical, or home settings.

Regulatory Classification

Identification

A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ #### 510(K) SUMMARY 3 K032433 # 510(k) SUMMARY-Nasal Jacks Mask | Submitter Name: | ResMed Corp. | |---------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------| | Submitter Address: | 14040 Danielson Street, Poway CA 92064-6857<br>USA | | Contact Person: | David D'Cruz, VP US Clinical & Regulatory Affairs | | Phone Number: | (858) 746 2238 | | Fax Number: | (858) 746 2900 | | Date Prepared: | August 1, 2003 | | Device Trade Name: | Not available at time of submission | | Device Common Name/<br>Classification Name: | Nasal Jacks Mask | | Predicate Devices: | Primary predicate: ADAM™ Interface System. K900164<br>Secondary predicate: Mirage Nasal Mask (cleared<br>in<br>SULLIVAN® AUTOSET® Nasal CPAP System K980721) | ## Device Description: The Nasal Jacks Mask is designed for adult patients for the delivery of non-invasive ventilatory support using continuous positive airway pressure or bi-level therapy. It is intended for singlepatient, multi-use and is minimally obtrusive to the user providing a high level of comfort, ease-ofuse and seal. ## Intended Use: The Nasal Jacks Mask is an accessory to a non-continuous ventilator (respirator), intended for single-patient multi-use for adult patients prescribed continuous positive airway pressure (CPAP) and bi-level therapy in hospital, clinical, and home environments. {1}------------------------------------------------ ## Device Technological Characteristics and Comparison to Predicate Device(s): ### Overview: The Nasal Jacks Mask is supported by headgear to allow a seal with the patients nostrils via the nasal cushions, then connected via tubing to a CPAP or bi-level flow generator. Positive pressure ventilation is thus applied to the lungs in a non-invasive manner. The Nasal Jacks Mask comes in one frame size and has three nasal cushion sizes. The Nasal Jacks Mask design is substantially equivalent to predicate device(s) (ADAM™ Interface System and Mirage Nasal Mask). The Nasal Jacks Mask design has the same intended use and has the same fundamental scientific technology as its predicates. ### Performance Data: Performance testing has been carried out to verify and validate the safety and effectiveness of the Nasal Jacks Mask. The Nasal Jacks Mask was tested to determine the pressure-flow characteristic, functional dead space (CO2 re-breathing), physical dead space and flow impedance. The results of the performance data show that the mask is substantially equivalent to the predicate mask (refer section 5.2.1) ### Materials Biocompatibility Materials have been carefully selected to ensure patient safety and efficacy of the product. The materials used to create components of the Nasal Jacks Mask, which contact the skin, mucosal membrane and/or the air-path, are either predicate materials (i.e., cleared previously for the same intended use), or have been tested to the ISO 10993 standards by an independent certified laboratory. The details are referenced in section 5.4. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with three stripes representing the three branches of government. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle. OCT 2 4 2003 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 ResMed Limited C/O Mr. David D'Cruz Vice President US Clinical & Regulatory Affairs ResMed Corporation 14040 Danielson Street Poway, California 92064-6857 Re: K032433 Trade/Device Name: Nasal Jacks Mask Regulation Number: 868.5905 Regulation Name: Noncontinuous Ventilator (IPPB) Regulatory Class: II Product Code: BZD Dated: August 1, 2003 Received: August 11, 2003 Dear Mr. D'Cruz: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ # Page 2 - Mr. D'Cruz Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely vours. uses () ungel Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ : :: #### INDICATIONS FOR USE 4 510(k) Number (if known): | 1 | { | t | far | |---|---|---|-----| | C | 1 | | | | | | | | Device Name: Nasal Jacks Mask Indications for Use: The Nasal Jacks Mask is an accessory to a non-continuous ventilator (respirator), intended for single-patient multiple use for adult patients prescribed continuous positive airway pressure (CPAP) and bi-level therapy in hospital, clinical, and home environments. Selma M. Quinto (Division Sign Off) Division of Anesthesiology, General Hospital, Infection Control, Dental Devices 510(k) Number: الكم 32433 (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH; Office of Device Evaluation (ODE) | Prescription | | |----------------------|--| | Use | | | (Per 21 CFR 801.109) | | OR Over-The-Counter Use (Optional Format 1-2-96)
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...