ORTHORALIX 9200 DDE, ORTHORALIX 9200 DDE PLUS CEPH, MODELS 5105 0161WW, 5105 0181WW

K032355 · Dentsply Gendex · EHD · Sep 25, 2003 · Dental

Device Facts

Record IDK032355
Device NameORTHORALIX 9200 DDE, ORTHORALIX 9200 DDE PLUS CEPH, MODELS 5105 0161WW, 5105 0181WW
ApplicantDentsply Gendex
Product CodeEHD · Dental
Decision DateSep 25, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.1800
Device ClassClass 2

Indications for Use

The Orthoralix 9200 DDE Digital Panoramic and Cephalometric Dental X-Ray System is to be used as an extraoral source of x-rays for imaging of the dentomaxilofacial area.

Device Story

Extraoral X-ray system for dentomaxilofacial imaging. Device uses high-frequency inverter for DC power to X-ray tubehead; beam collimated via slit diaphragm. During panoramic exposure, tubehead and cassette holder rotate around patient head; patient remains stationary, positioned via bite guide and headrest. Four independent motors control radiographic projection geometry. System includes motorized wall-mounted column, overhead carriage, and optional cephalometric arm/head positioner. Operated by dental professionals in clinical settings. Microprocessor-controlled electronic unit manages technique factors and exposure. Output is radiographic image for diagnostic assessment by clinician. Benefits include standardized, accurate imaging of dental structures.

Clinical Evidence

Bench testing only. Evidence includes performance testing against product specifications, software design/performance validation, and clinical image exposure assessments to demonstrate safety and effectiveness.

Technological Characteristics

Extraoral X-ray system; high-frequency inverter; DC power supply; motorized column; microprocessor-controlled electronics; 15x30 cm flat cassette drive system; slit diaphragm collimation; optional cephalometric arm. Connectivity includes remote control box and handswitch.

Indications for Use

Indicated for extraoral X-ray imaging of the dentomaxilofacial area in patients requiring dental or maxillofacial diagnostic imaging.

Regulatory Classification

Identification

An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K032355 SEP 2 5 2003 Image /page/0/Picture/2 description: The image shows the logos for Dentsply and Gendex. The Dentsply logo is in a bold, sans-serif font, with the letters closely spaced together. Below Dentsply is the Gendex logo, which is in a thinner, sans-serif font. Both logos are in black and are stacked vertically. man mana DENTSPLY International Gendex Division 901 West Oakton Street Des Plaines, IL 60018-1884 Phone (847) 640-4800 Fax (847) 640-4970 510(k) Summary Statement for the Gendex Orthoralix 9200 DDE Digital Panoramic and Cephalometric Dental X-Ray System #### I. General Information | Submitter: | DENTSPLY International<br>Gendex Division<br>901 West Oakton St.<br>Des Plaines, IL 60018 | |---------------------------|-------------------------------------------------------------------------------------------| | Telephone: | (847) 640-4800 - Company Number<br>(847) 640-4924 - Contact Person | | Fax: | (847) 640-4970 | | Contact Person: | John R. Miller<br>Director, Quality Assurance and Regulatory Affairs | | Summary Preparation Date: | June 30, 2003 | #### II. Names | Device Name: | Orthoralix 9200 DDE Digital Panoramic and Cephalometric<br>Dental X-Ray System | |------------------------------|--------------------------------------------------------------------------------| | Primary Classification Name: | 76EHD – Unit, X-Ray, Extraoral with Timer | #### III. Predicate Devices - Gendex Orthoralix 9200 and 9200 Plus - Instrumentarium Orthoceph® OC100 D Direct Digital Cephalometric Imaging System ● - Planmeca DIMAX2 Direct Digital Imaging System - . Sirona Orthophos Plus DS {1}------------------------------------------------ ## Orthoralix 9200 DDE Summary Statement Page 2 of 3 #### IV. Product Description The DENTSPLY International, Gendex Division Orthoralix 9200 DDE Digital Panoramic and Cephalometric Dental X-Ray System is an extraoral source of x-rays for imaging of the dentomaxilofacial area. The DENTSPLY International, Gendex Division Orthoralix 9200 DDE Digital Panoramic and Cephalometric Dental X-Ray System is comprised of the following main components: - High frequency inverter which supplies a true DC output and accurate technique factors . - A microprocessor controlled user-friendly electronic control . - A motorized column to be fixed to the wall ● - Counter balanced overhead carriage with controls for patient positioning, setting and control . of technique factors and radiographic projection geometry - Cassette drive system with flat cassette for 15x30 cm film - X-ray tubehead, with DC power supply via electronic converter . - Remote control box and handswitch - . Optional cephalometric arm and head postioner - Optional utilities for transverse scanography of the jaw ("Transcan®") ● During a panoramic exposure, the x-ray tube and cassette holder moves around the patient's head. The beam from the x-ray tube is collimated by a slit diaphragm. The flat film cassette passes behind a secondary collimator which blocks the radiation scattered from the patient. Patient positioning is by means of a bite guide and a headrest. During exposure, the patient, remains still while the motorized components rotate. All movements for the panoramic radiographic projection are performed by four independent motors, which are microprocessor controlled. #### V. Indications for Use/Rationale for Substantial Equivalence The Orthoralix 9200 DDE Digital Panoramic and Cephalometric Dental X-Ray System is to be used as an extraoral source of x-rays for imaging of the dento-maxilofacial area. It shares the same indications for use, similar materials, design, operations, and functional features and therefore is substantially equivalent to the predicate devices listed in Section III of this summary. #### VI. Safety and Effectiveness Information Safety and Effectiveness is demonstrated by: - Performance testing to meet product specifications - - -Software testing to validate software design/performance - Effective clinical image exposures - - Risk analysis including risk level and solution - - Same indications for use as predicate devices. .. {2}------------------------------------------------ ### Orthoralix 9200 DDE Summary Statement Page 3 of 3 All the above steps and evaluations combine to demonstrate that the Orthoralix 9200 DDE Digital Panoramic and Cephalometric Dental X-Ray System is safe and effective when the device is used as labeled #### VII. Conclusion The DENTSPLY International, Gendex Division Orthoralix 9200 DDE Digital Panoramic and Cephalometric Dental X-Ray System was found to be substantially equivalent to the predicate devices, the Gendex Orthoralix 9200 and 9200 Plus, the Instrumentarium Orthoceph® OC100 D Direct Digital Cephalometric Imaging System, the Planmeca DIMAX2 Direct Digital Imaging System, and the Sirona Orthophos Plus DS. The DENTSPLY International, Gendex Division Orthoralix 9200 DDE Digital Panoramic and Cephalometric Dental X-Ray System shares the same indications for use, similar materials, design, operational, and functional features as the current marketed predicate devices. It has been shown to be safe and effective when used as labeled. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows a black and white logo. The logo is circular in shape and features a stylized image of an eagle in flight. The eagle is facing to the left. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES" is arranged around the top half of the circle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 SEP 2 5 2003 Mr. John R. Miller Director DENTSPLY International Gendex Division 901 West Oakton Street DES PLAINES IL 60018-1884 Re: K032355 Trade/Device Name: Orthoralix 9200 Panoramic and Cephalometric Dental X-Ray System Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral source x-ray system Regulatory Class: II Product Code: 90 EHD Dated: July 21, 2003 Received: July 30, 2003 Dear Mr. Miller: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050. {4}------------------------------------------------ This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter: | 8xx. 1xxx | (301) 594-4591 | |----------------------------------|----------------| | 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 | | 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 | | 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 | | Other | (301) 594-4692 | Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours. Nancy C. Brygdon Nancy C. B rogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ K032355 ## 510(k) Number (if known): Not Assigned Orthoralix 9200 DDE Digital Panoramic and Cephalometric Device Name: Dental X-Ray System # Indications for Use: The Orthoralix 9200 DDE Digital Panoramic and Cephalometric Dental X-Ray System is to be used as an extraoral source of x-rays for imaging of the dentomaxilofacial area. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) David R. deymm (Division Sign-Off Division of Reproductive, Abdominal and Radiological Device 510(k) Number Prescription Use (Per 21CFR 801.109) OR Over-The-Counter Use __ (Optional Format 1-2-96) Image /page/5/Picture/13 description: The image shows the number 000046 in a bold, sans-serif font. The numbers are black against a white background. The numbers are evenly spaced and aligned horizontally. The number is likely a serial number or some other type of identification number.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...