K032351 · BioHorizons Implant Systems, Inc. · DZE · Oct 21, 2003 · Dental
Device Facts
Record ID
K032351
Device Name
THE MAESTRO SYSTEM 3.0MM DIAMETER IMPLANT
Applicant
BioHorizons Implant Systems, Inc.
Product Code
DZE · Dental
Decision Date
Oct 21, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3640
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The BioHorizons 3.0mm diameter implant mav be used: (1) as an artificial root structure for single tooth replacement of mandibular central and lateral incisors and maxillary lateral incisors. The implant may be immediately restored with a temporary prosthesis that is not in functional occlusion. (2) when splinted together as an artificial root structure for multiple tooth replacement of mandibular incisors. The implants may be restored after a period of time or placed in immediate function. (3) for denture stabilization using multiple implants in the anterior mandible and maxilla. The implants may be restored after a period of time or placed in immediate function.
Device Story
Maestro System 3.0mm Diameter Implant; screw-type endosseous dental implant. Functions as artificial root structure for tooth replacement or denture stabilization. Machined titanium alloy (ASTM F 136) with resorbable blast media (RBM) surface treatment (hydroxylapatite per ASTM F 1185). Available in 12mm, 15mm, 18mm lengths. Used by dental professionals in clinical settings. Implant placed into jawbone; restored with temporary or permanent prosthesis depending on clinical application (single tooth, splinted multiple teeth, or denture stabilization). Provides mechanical foundation for prosthetic restoration; benefits patients by restoring dental function and aesthetics.
Clinical Evidence
Bench testing only. No clinical data provided. Substantial equivalence established through design, material, and intended use comparisons to legally marketed predicate devices.
Technological Characteristics
Material: Titanium-6Aluminum-4Vanadium ELI alloy (ASTM F 136). Surface: Resorbable blast media (RBM) with hydroxylapatite (ASTM F 1185). Form: Screw-type endosseous implant, 3.0mm diameter, 12-18mm lengths. Sterilization: Radiation (ANSI/AAMI/ISO 11137).
Indications for Use
Indicated for single tooth replacement of mandibular central/lateral incisors and maxillary lateral incisors; multiple tooth replacement of mandibular incisors (when splinted); and denture stabilization in the anterior mandible and maxilla.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{0}------------------------------------------------
# OCT 2 1 2003
BioHorizons Implant Systems, Inc.
The Maestro System™ 3.0mm Diameter Implant 510(k) Notification, July 25, 2003 - revised October 16, 2003
### 510(k) summary
| 21 CFR 807.92 | |
|-------------------|-----------------------------------|
| Date: | July 25, 2003 |
| Official Contact: | Winston Greer, Director, QA & RA |
| Manufacturer: | BioHorizons Implant Systems, Inc. |
| | One Perimeter Park South |
| | Suite 230 South |
| | Birmingham, AL 35243 |
| | Phone: (205) 967-7880 |
| | Fax: (205) 870-0304 |
#### Proprietary Name
The Maestro System™ Maximus 3.0mm Diameter Implant
#### Common Name
Screw-type Dental Implant
#### Classification Name
Endosseous implants, surgical components, and prosthetic attachments
#### Predicate Devices
Predicate devices are:
- 1. The Maestro System™, a screw-type endosseous implant manufactured and distributed by BioHorizons Implant Systems, Inc. Authorization to legally market the oredicate implant system has been documented under 510(k) numbers K960026. K964330, K972313, K010458, K020133, K020645, K022795 and K030463.
- 2. The Altiva Natural Tooth Replacement (NTR) System 3.3mm diameter implant, a one-piece screw-type endosseous implant for long-term applications manufactured and distributed by Altiva Corporation and cleared for market under 510(k) number K992512.
- 3. The Sendax Mini Dental Implant (MDI) 1.8mm diameter implant, a one-piece screwtype endosseous implant for transitional and long-term applications manufactured and distributed by Intec Corporation and cleared for market under 510(k) numbers K972351, K990983 and K023067.
### Device Description
The BioHorizons Maestro System 3.0mm diameter dental implant is a machined titanium, screw-form implant supplied in lengths of 12mm, 15mm and 18mm. Implant raw material is titanium alloy as specified in ASTM F 136, Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy (UNS R56401) for Surgical Implant Applications.
The device is further processed by treating the surface with resorbable blast media (RBM). The product is packaged using materials known in the industry to be appropriate for medical device packaging and is provided with a minimum sterility assurance level of 10 °, validated in compliance to ANSI/AAMI/ISO 11137, Sterlization of healthcare products - Requirements for validation and routine control - Radiation Sterilization.
{1}------------------------------------------------
The Maestro System™ 3.0mm Diameter Implant 510(k) Notification, July 25, 2003 - revised October 16, 2003
The Maestro System™ 3.0mm Diameter Implant is a comprehensive system containing implants and surgical components. The BioHorizons 3.0mm diameter implant may be used: (1) as an artificial root structure for single tooth replacement of mandibular central and lateral incisors and maxillary lateral incisors. The implant may be immediately restored with a temporary prosthesis that is not in functional occlusion; (2) when splinted together as an artificial root structure for multiple tooth replacement of mandibular incisors. The implants may be restored after a period of time or placed in immediate function; (3) for denture stabilization using multiple implants in the anterior mandible and maxilla. The implants may be restored after a period of time or placed in immediate function.
All BioHorizons implants referenced in this submission are 3.0mm in diameter with D3 thread form and surface treated using resorbable blast media (hydroxylapatite conforming to ASTM F 1185, Standard Specification for Composition of Ceramic Hydroxylapatite for Surgical Implants). The following table provides a summary of the proposed implant catalog item or reference numbers by length.
| Lengths (mm) | Catalog<br>REF Numbers |
|--------------|------------------------|
| 12 | 3012D3 |
| 15 | 3015D3 |
| 18 | 3018D3 |
### Intended Use
The BioHorizons 3.0mm diameter implant mav be used:
- (1) as an artificial root structure for single tooth replacement of mandibular central and lateral incisors and maxillary lateral incisors. The implant may be immediately restored with a temporary prosthesis that is not in functional occlusion.
- (2) when splinted together as an artificial root structure for multiple tooth replacement of mandibular incisors. The implants may be restored after a period of time or placed in immediate function.
- (3) for denture stabilization using multiple implants in the anterior mandible and maxilla. The implants may be restored after a period of time or placed in immediate function.
## Technological Characteristics
The fundamental scientific technology of the device is identical or very similar to the referenced predicate devices. All materials, suppliers, processing, packaging and sterilization methods remain the same as for the predicate Maestro System devices. The BioHorizons 3.0mm diameter implants are substantially equivalent to all features of the predicate devices which could affect safety or effectiveness because of the similarities in design, material and intended use.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three curved lines representing its wings. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# OCT 2 1 2003
Mr. Weston D. Greer Director Assurance & Regulatory Affairs Biohorizons Implant System, Incorporated One Perimeter Park South Suite 230 South Birmingham, Alabama 35243
Re: K032351
Trade/Device Name: Biohorizons the Maestro System™ 3.0mm Diameter Implant Regulation Number: 872.3640 Regulation Name: Endosseous Implant Regulatory Class: III Product Code: DZE Dated: July 25, 2003 Received: July 31, 2003
Dear Mr. Greer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
#### Page 2 - Mr. Greer
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits vour device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4613. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours.
Patricia Cucarito Flor
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
BioHorizons Implant Systems, Inc.
The Maestro System™ 3.0mm Diameter 510(k) Notification, July 25, 2003; revised October 16, 2003
Page 1 of 1
510(k) Number: _______________________________________________________________________________________________________________________________________________________________
Device Name: BioHorizons The Maestro System™ 3.0mm Diameter Implant
Indications for Use:
The BioHorizons Maestro System 3.0 mm Diameter Implant may be used
- (1) as an artificial root structure for single tooth replacement of mandibular central and lateral incisors and maxillary lateral incisors. The implant may be immediately restored with a temporary prosthesis that is not in functional occlusion.
- (2) when splinted together as an artificial root structure for multiple tooth replacement of mandibular incisors. The implants may be restored after a period of time or placed in immediate function.
- (3) for denture stabilization using multiple implants in the anterior mandible and maxilla. The implants may be restored after a period of time or placed in immediate function.
Concurrence of CDRH, Office of Device Evaluation (ODE)
Susan Runser
510(k) Number. K032235
Prescription Use (per 21 CFR 801.109) OR
Over-the-Counter Use
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.