The Trio monitor is intended for hospital use under the direct supervision of a licensed healthcare practitioner. The intended use of the monitor is to monitor physiologic parameter data on adult and pediatric patients. Physiologic data includes but is not restricted to: electrocardiogram, invasive blood pressure, heart rate (derived from ECG and SpO2) noninvasive blood pressure, pulse oximetry, respiration and temperature as summarized in the operating instructions manual. The information can be displayed, stored, trended and printed.
Device Story
Trio Monitor is a multi-parameter patient monitor for hospital use; operated by clinicians. Inputs include ECG (3/5-lead), temperature, SpO2, invasive blood pressure, non-invasive blood pressure, and respiration. Device processes these signals to provide real-time digital and waveform displays; includes built-in thermal recorder for hard copies of data, waveforms, and trends. Used in clinical environments (rolling stand, wall mount, bed rail, or tabletop). Output allows clinicians to observe patient status, track trends, and make informed clinical decisions. Benefits include continuous physiological monitoring to assist in patient care management.
Clinical Evidence
No clinical data. Evidence consists of bench testing, including requirements specification review, hardware and software testing, code design/reviews, environmental testing, safety testing, performance testing, and risk analysis. Validation confirmed performance equivalent to the predicate device.
Technological Characteristics
Multi-parameter monitor (3 or 4 trace). Parameters: ECG, Temp, SpO2, IBP, NIBP, Respiration. Form factor: rolling stand, wall mount, bed rail, or tabletop. Includes thermal recorder. Software-based processing for digital/waveform display and trend storage. Connectivity: standalone monitor.
Indications for Use
Indicated for adult and pediatric patients in hospital settings under direct supervision of licensed healthcare practitioners for monitoring ECG, invasive/noninvasive blood pressure, heart rate, pulse oximetry, respiration, and temperature. Not intended for home use or as an apnea monitor.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
Predicate Devices
GE Medical Systems Dash 2000 Pro Patient Monitor
Submission Summary (Full Text)
{0}------------------------------------------------
FEB 1 1 2004
## 510(k) Summary of Safety and Effectiveness
| Date: | July 22, 2003 |
|-----------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Patient Monitoring Division<br>Datascope Corp. |
| Contact Person: | Susan E. Mandy<br>Director, Clinical & Regulatory Affairs<br>Patient Monitoring Division<br>Datascope Corp.<br>Telephone: (201) 995-8025<br>Fax:(201) 995-8605 |
Device trade name: Trio Monitor
Common/usual name: Physiological Patient Monitor
## Classification name:
| 21 CFR 870.1100 | Blood pressure alarm |
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 21 CFR 870.1130 | Noninvasive blood pressure measurement systems |
| 21 CFR 870.1110 | Blood pressure computer |
| 21 CFR 870.2300 | Cardiac monitor (including cardiotachometer and rate alarm) |
| 21 CFR 870.2340 | Electrocardiograph |
| 21 CFR 870.2700 | Oximeter |
| 21 CFR 880.2910 | Clinical electronic thermometers |
| Predicate Device: | GE Medical Systems Dash 2000 Pro Patient Monitor |
| Device Description: | Mindray Bio-Medical Co. Ltd., previously known as Caymans Mindray<br>Medical Electronic Company of Shenzen, China, manufactures the Trio<br>monitor for Datascope Corp.<br><br>The Trio monitor is a three or four trace patient monitor that can be<br>mounted on a rolling stand, wall mount bracket, bed rail, or operated as a<br>tabletop device. The patient parameters that can be monitored on the Trio<br>are: ECG (3-lead or 5- lead selectable), Temperature, SpO2, Invasive<br>Blood Pressure, Respiration and Non-Invasive Blood Pressure. Digital<br>displays are provided for Heart Rate, Pulse Rate, Pulse Oximetry, Non-<br>Invasive Blood Pressure, Respiration Rate, Invasive Blood Pressure<br>(optional) and Temperature. Waveform displays are provided for ECG, |
{1}------------------------------------------------
| built-in thermal recorder provides hard copies of all digital data and<br>waveforms, as well as Tabular and Graphic Trend information. | |
|----------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use: | The Trio monitor is intended for hospital use under the direct supervision<br>of a licensed healthcare practitioner. The intended use of the monitor is to<br>monitor physiologic parameter data on adult and pediatric patients.<br>Physiologic data includes but is not restricted to: electrocardiogram,<br>invasive blood pressure, heart rate (derived from ECG and SpO2)<br>noninvasive blood pressure, pulse oximetry, respiration and temperature as<br>summarized in the operating instructions manual. The information can be<br>displayed, stored, trended and printed. |
| Technology: | The Trio monitor employs the same functional technology as the predicate<br>device. |
| Test Summary | The Trio monitor complies with the voluntary standards identified in<br>section six of this submission. During the development process of the Trio<br>monitor, the following activities were completed: |
| | Requirements specification review |
| | Hardware and software testing |
| | Code design and code reviews |
| | Environmental testing |
| | Safety testing |
| | Performance testing |
| | Risk Analysis |
| | Hardware and Software validation |
| Conclusion | The results of these measurements demonstrated that the Trio monitor is as<br>safe, as effective and performs as well as the predicate device. |
.
. .
:
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes forming its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 1 1 2004
Datascope Corp. c/o Ms. Susan E. Mandy Director, Clinical & Regulatory Affairs 800 MacArthur Blvd. Mahwah, NJ 07430
Re: K032338 Trade Name: Trio™ Monitor Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (including cardiotachometer and rate alarm) Regulatory Class: Class II (two) Product Code: MWI Dated: December 23, 2003 Received: December 24, 2003
Dear Ms. Mandy:
We have reviewed your Section 510(k) premarket notification of intent to market the arrived inte we nave reviewed your becamined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encrosule) to regally to regally the Medical Device Amendments, or to commerce provided in accordance with the provisions of the Federal Food. Drug.
devices that have been reclassified in accordance with the provisions of the United in AMA devices that have been reclassified in access approval of a premarket approval application (PMA). and Costience Act (Act) that do not require approvial controls provisions of the Act. The You may, therefore, market the devrees, bacjes to the morements for annual registration, listing of general controls provisions of the free, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into cither class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can may be subject to such additional controlier 21, Parts 800 to 898. In addition, FDA may be found in the Code of Peacharting your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Ms. Susan Mandy
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely vours.
Dana P. Vochner
( Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
510(k) Number (if known): K032338
Device Name:
Indications For Use:
## INDICATION FOR USE STATEMENT
The Trio™ monitor is intended for hospital use under the direct supervision of a licensed healthcare practitioner. The intended use of the monitor is to monitor physiologic parameter data on adult and pediatric patients. Physiologic data includes: electrocardiogram, invasive blood pressure, noninvasive blood pressure, pulse oximetry, heart rate (derived from ECG or SpO2), respiration and temperature as summarized in the operating instructions manual. The information can be displayed, stored, trended and printed.
The monitor is not intended for home use. The monitor is not intended to be an apnea monitor. It was not designed or validated for use as an apnea monitor.
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Dung R. Laune 1
(Division Sign-Off) Division of Cardiovascular Devices
610(k) Number_K032338
Page 1 of _ _ _ _ _
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.