K032311 · Roche Diagnostics Corp. · CEM · Oct 17, 2003 · Clinical Chemistry
Device Facts
Record ID
K032311
Device Name
ROCHE DIAGNOSTICS OMNI S ANALYZER
Applicant
Roche Diagnostics Corp.
Product Code
CEM · Clinical Chemistry
Decision Date
Oct 17, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1600
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Roche Diagnostics Omni S Analyzer is a fully automated critical care analyzer intended to be used for the measurement of pH, PO2, PCO2, sodium, potassium, ionized calcium, chloride, hematocrit, glucose, lactate, urea/BUN, total hemoglobin, oxygen saturation, oxyhemoglobin, deoxyhemoglobin, carboxyhemoglobin and methemoglobin in samples of whole blood, serum, plasma and aqueous solutions as appropriate.
Device Story
Omni S Analyzer is a fully automated critical care analyzer; measures pH, blood gases, electrolytes, hematocrit, glucose, lactate, urea/BUN, and hemoglobin derivatives. Inputs: whole blood, serum, plasma, or aqueous solutions. Operation: utilizes blood gas, ISE, and MSS sensor modules; incorporates Braunstein technology for glucose/lactate; features enhanced sample detection and standby solution aspiration. Output: quantitative analytical results for the measured parameters. Used in clinical settings by healthcare professionals to support critical care decision-making. Benefits: provides rapid, comprehensive diagnostic data from a single platform, improving workflow efficiency via consolidated reagent packs.
Clinical Evidence
Bench testing only. The device demonstrates substantial equivalence through comparative analysis of sensor modules, sample types, and analytical performance parameters against the predicate Omni Modular and Omni C systems.
Technological Characteristics
Fully automated analyzer. Technologies: Ion Selective Electrode (ISE), enzymatic, MSS chamber sensors, Braunstein technology. Connectivity: PC-based interface, barcode reader, network of microcontrollers for data acquisition/sensor handling. Calibration: Automated 8-24 hour cycles; O2 sensor recalibrated every 30 minutes. Sample handling: Capillary, syringe, sealed containers. Standards: NCCLS EP6-P, NCCLS E5-T.
Indications for Use
Indicated for the measurement of pH, blood gases, electrolytes, hematocrit, glucose, lactate, urea/BUN, and hemoglobin derivatives in whole blood, serum, plasma, and aqueous solutions. Intended for use in critical care settings.
Regulatory Classification
Identification
A potassium test system is a device intended to measure potassium in serum, plasma, and urine. Measurements obtained by this device are used to monitor electrolyte balance in the diagnosis and treatment of diseases conditions characterized by low or high blood potassium levels.
{0}
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
DEVICE AND INSTRUMENT TEMPLATE
A. 510(k) Number: K032311
B. Analyte: Potassium, Calcium, Sodium, Chloride, Blood Gases (PCO2, PO2) and pH, Hemoglobin, Hematocrit, Oxygen Saturation, Carboxyhemoglobin, Oxyhemoglobin, Methemoglobin, Deoxyhemoglobin, Lactic Acid, Glucose, Urea Nitrogen.
C. Type of Test: Ion selective electrode – Potassium, Chloride, Sodium, Calcium, Blood gases, pH and Urea Nitrogen
Colorimetric – Glucose
Enzymatic – Lactic Acid
D. Applicant: Jennifer Tribbett
Roche Diagnostics
Patient Care Division
9115 Hague Road
P.O. Box 50457
Indianapolis, IN 46250
E. Proprietary and Established Names: Proprietary Name: The Roche OMNI S Analyzer. Established name – Fully automated Critical Care Analyzer for the measurement of pH, Blood gases, electrolytes, Hematocrit, hemoglobin, glucose, lactate, urea/BUN, total hemoglobin, Oxygen saturation, oxy - deoxy-carboxy and methemoglobin
F. Regulatory Information:
1. Regulation section: 862.1600, 862.1145, 862.1665, 862.1170, 862.1120, 862.1450, 862.1345, 862.1770, 864.5620, 864.5600, 864.7425, 864.7500
2. Classification: Class II
3. Product Code: Potassium - CEM, Calcium - JFP, Sodium - JGS, Chloride - CGZ, Blood Gases (PCO2, PO2) and pH - CHL, Hemoglobin - GKR, Hematocrit - GKF, Oxygen saturation, Carboxyhemoglobin - GHS, Oxyhemoglobin - GGZ, Methemoglobin - KHG, Deoxyhemoglobin - GKA, Lactic Acid - KHP, Glucose- CGA, Urea Nitrogen - CDS.
{1}
Page 2 of 11
4. Panel: 75, 81
## G. Intended Use:
1. Intended use(s): The Roche Diagnostics Omni S Analyzer is a fully automated critical care analyzer intended to be used for the measurement of pH, PCO2, PO2, sodium, potassium, ionized calcium, chloride, hematocrit, glucose, lactate, urea/BUN, total hemoglobin, oxygen saturation, oxyhemoglobin, deoxyhemoglobin, carboxyhemoglobin, and methemoglobin in samples of whole blood, serum, plasma, and aqueous solutions as appropriate.
2. Indication(s) for use: The Roche Diagnostics Omni S Analyzer is a fully automated critical care analyzer intended to be used for the measurement of pH, PCO2, PO2, sodium, potassium, ionized calcium, chloride, hematocrit, glucose, lactate, urea/BUN, total hemoglobin, oxygen saturation, oxyhemoglobin, deoxyhemoglobin, carboxyhemoglobin, and methemoglobin in samples of whole blood, serum, plasma, and aqueous solutions as appropriate.
3. Special condition for use statement(s): none
4. Special instrument Requirements: none
## H. Device Description: The Roche Omni S is an analyzer with an integrated AutoQC drawer option. Depending upon combination and configuration, the following parameters can be measured in whole blood, serum, plasma, acetate and bicarbonate containing dialysis solutions and QC materials.
## I. Substantial Equivalence Information:
1. Predicate device name(s): Omni Modular Analyzer and Omni C Analyzer
2. Predicate K number(s): Omni Modular Analyzer (K990092) and Omni C Analyzer (K013373)
3. Comparison with predicate:
| Similarities & Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended use | The Roche Diagnostics Omni S Analyzer is intended to be used for the measurement of | The Roche Diagnostics Omni Modular Analyzer is intended to be used for the measurement of |
{2}
Page 3 of 11
| | pH, PCO2, PO2, sodium, potassium, ionized calcium, chloride, hematocrit, glucose, lactate, urea/BUN, total hemoglobin, oxygen saturation, oxyhemoglobin, deoxyhemoglobin, carboxyhemoglobin, and methemoglobin in samples of whole blood, serum, plasma, and aqueous solutions as appropriate. | pH, PCO2, PO2, sodium, potassium, ionized calcium, chloride, hematocrit, glucose, lactate, urea/BUN, total hemoglobin, oxyhemoglobin, deoxyhemoglobin, carboxyhemoglobin, sulfhemoglobin, and methemoglobin in samples of whole blood, serum, plasma, and aqueous solutions as appropriate. |
| --- | --- | --- |
| Parameters measured | pH, PCO2, PO2, sodium, potassium, ionized calcium, chloride, hematocrit, glucose, lactate, urea/BUN, total hemoglobin, oxygen | The Roche Diagnostics Omni C Analyzer is intended to be used for the measurement of pH, PCO2, PO2, sodium, potassium, ionized calcium, chloride, hematocrit, total hemoglobin, and oxygen saturation, in samples of whole blood, serum, plasma, and aqueous solutions as appropriate. |
{3}
Page 4 of 11
| | saturation,
oxyhemoglobin,
deoxyhemoglobin,
carboxyhemoglobin, and
methemoglobin | calcium, chloride,
hematocrit,
glucose, lactate,
urea/BUN, total
hemoglobin,
oxyhemoglobin,
deoxyhemoglobin,
carboxyhemoglobin,
sulfhemoglobin,
and
methemoglobin |
| --- | --- | --- |
| Sample types | Samples of whole
blood, serum,
plasma, and
aqueous solutions
as appropriate. | The Roche
Diagnostics Omni
C Analyzer: pH,
PCO2, PO2,
sodium,
potassium, ionized
calcium, chloride,
hematocrit, total
hemoglobin, and
oxygen saturation
Same for both
predicate devices |
J. Standard/Guidance Document Referenced (if applicable): NCCLS EP6-P, NCCLS – E5-T, and Reviewer Guidance for Computer Controlled Medical Devices undergoing 510(k) review.
K. Test Principle: Ion Selective Electrode, Enzymatic, Sensor (O2 saturation).
L. Performance Characteristics (if/when applicable):
1. Analytical performance:
Precision/Reproducibility: With-in run (SWR) and Total (ST) precision were determined from 2 runs a day with 2 replicates per run for 20 days on four Roche Omni S instruments using samples of each specimen type suitable for measurement on the analyzer. All assays available on the Omni S were evaluated as appropriate for the specimen type.
{4}
Page 5 of 11
a. Linearity/assay reportable range: Linearity was established by utilizing the NCCLS Document EP6-P.
Tonometered whole blood: Whole blood was tonometered at 37°C to various levels of gravitimetrically prepared gases with CO2 and O2 concentrations certified to 0.03% absolute by the manufacturer. Expected and observed values were corrected to 760 mmHg.
Aqueous solutions: Expected values for the aqueous solutions were based on weighted samples.
NIST Standards: NIST standards are precise serums with accredited target values.
Hematocrit: Measurements for the hemofuge, which represents the gold standard for hematocrit measurements, was used as expected values for hematocrit results.
The linearity for each analyte was described as follows:
PO2 – slope 0.9904-1.0097 Intercept +/- 0.857 Correlation 0.9998
Range 0 – 800mmHg
PCO2 – slope 0.9898 – 1.0103 Intercept +/- 1.225 Correlation 0.9999
Range 4 – 200mmHg
pH – slope 0.9825 – 1.0178 Intercept +/- 0.133 Correlation 0.998
Range 6.0 – 8.0
Hct– slope 0.997 – 1.003 Intercept +/- 0.620 Correlation 0.999
Range 10 80%
Sodium – slope 0.988 – 1.012 Intercept +/- 0.365 Correlation 0.9999
Range 20 – 250mmol/L
Potassium – slope 0.960 – 1.042 Intercept +/- 0.109 Correlation 0.9999
Range 0.2 - 20 mmol/L
Ion. Calcium – slope 0.975 – 1.026 Intercept +/- 0.024 Correlation 0.9999
Range 0.1 – 4.0 mmol/L
Chloride – slope 0.959 – 1.043 Intercept +/- 2.908 Correlation 0.9999
Range 20 – 250 mmol/L
Glucose – slope 0.919 – 1.088 Intercept +/- 1.773 Correlation 0.998
Range 0.5 – 40 mmol/L
Lactate – slope 0.961 – 1.041 Intercept +/- 0.191 Correlation 0.9989
Range 0.2 – 20 mmol/L
Urea – slope 0.979 – 1.021 Intercept +/- 0.198 Correlation 0.9991
Range 0.5 – 30 mmol/L
Total Hgb – slope 0.9892 – 1.0109 Intercept +/- 0.0833 Correlation 0.9904
Range 3 - 25 g/dl
SO2 – slope 0.99999 – 1.00001 Intercept +/- 0.856 Correlation 0.9874
Range 50 – 100%
b. Traceability (controls, calibrators, or method): Tonometered whole blood, aqueous solutions, NIST standards
c. Detection limit: Detection Limits for analytes on the Omni S Analyzer are defined on the Omni S Analyzer as the linear range of each assay. The assays performed on the Omni S are measured with established normal and critical ranges for each
{5}
Page 6 of 11
analyte. These ranges fall well within the linearity of the respective assays.
d. Analytical specificity: Tonometered whole blood: Whole blood was tonometered at 37°C to various levels of gravitimetrically prepared gases with CO2 and O2 concentrations certified to 0.03% absolute by the manufacturer. Expected and observed values were corrected to 760 mmHg.
Aqueous solutions: Expected values for the aqueous solutions were based on weighted samples.
NIST Standards: NIST standards are precise serums with accredited target values.
Hematocrit: Measurements for the hemofuge, which represents the gold standard for hematocrit measurements, was used as expected values for hematocrit results.
e. Assay cut-off: The assay cut-off is established as the linearity range of the Roche Omni S for each analyte.
2. Comparison studies:
a. Method comparison with predicate device: External testing was conducted to demonstrate the correlation of the Omni S analyzer to legally marketed devices in a clinical setting, operated by personnel trained to perform and report these analyses.
Specimens analyzed were remnant from patient samples of whole blood, serum and plasma collected from routine analysis on existing instrumentation.
The Omni S has incorporated the same blood gas modules, ISE modules, total hemoglobin and Oxygen saturation, MSS chamber sensors for BUN/urea nitrogen as with the predicate devices. For lactate and glucose the Omni S utilizes Braunstein technology.
The correlations to the predicate devices are as follows:
pH, Comparison Instrument - Omni Modular, Slope/Intercept - Y = -0.063 + 1.009 * X, Bias +0.007, Corr. Coeff. [r] = 0.990, # samples = 134
pH, Comparison Instrument - Radiometer 725, Slope/Intercept - Y = -0.496 + 0.933 * X, Bias +0.003, Corr. Coeff. [r] = 0.990, # samples = 99
pO2, Comparison Instrument - Omni Modular, Slope/Intercept - Y = -0.643 + 1.031 * X, Bias +1.6, Corr. Coeff. [r] = 0.987, # samples = 136, Unit (mmHG)
{6}
Page 7 of 11
pO2, Comparison Instrument - Radiometer 725, Slope/Intercept - Y = -4.433 + 1.013 * X, Bias +6.6, Corr. Coeff. [r] = 0.996, # samples = 137, Unit (mmHG)
pCO2, Comparison Instrument - Omni C, Slope/Intercept - Y = -1.452 + 1.038 * X, Bias +1.6, Corr. Coeff. [r] = 0.987, # samples = 129, Unit (mmHG)
pCO2, Comparison Instrument - Radiometer 725, Slope/Intercept - Y = -0.301 + 1.000 * X, Bias +0.4, Corr. Coeff. [r] = 0.988, # samples = 144, Unit (mmHG)
tHb, Comparison Instrument - Radiometer 725, Slope/Intercept - Y = 0.581 + 1.083 * X, Bias +2.0, Corr. Coeff. [r] = 0.814, # samples = 96, Unit (g/dl)
Sodium, Comparison Instrument - Omni Modular, Slope/Intercept - Y = -13.193 + 1.106 * X, Bias +0.9, Corr. Coeff. [r] = 0.948, # samples = 108, Unit (mmol/L)
Sodium, Comparison Instrument - Radiometer 725, Slope/Intercept - Y = -2.143 + 1.106 * X, Bias +0.9, Corr. Coeff. [r] = 0.972, # samples = 107, Unit (mmol/L)
Potassium, Comparison Instrument - Omni Modular, Slope/Intercept - Y = -0.126 + 1.020 * X, Bias -1.4, Corr. Coeff. [r] = 0.986, # samples = 131, Unit (mmol/L)
Potassium, Comparison Instrument - Radiometer 725, Slope/Intercept - Y = -2.143 + 1.106 * X, Bias +0.9, Corr. Coeff. [r] = 0.972, # samples = 107, Unit (mmol/L)
Calcium, Comparison Instrument - Omni Modular, Slope/Intercept - Y = -0.039 + 1.024 * X, Bias -0.8, Corr. Coeff. [r] = 0.941, # samples = 131, Unit (mmol/L)
Calcium, Comparison Instrument - Radiometer 725, Slope/Intercept - Y = -0.096 + 1.073 * X, Bias -1.1, Corr. Coeff. [r] = 0.981, # samples = 98, Unit (mmol/L)
Chloride, Comparison Instrument - Omni Modular, Slope/Intercept - Y = -12.459 + 1.118 * X, Bias -0.7, Corr. Coeff. [r] = 0.960, # samples = 139, Unit (mmol/L)
{7}
Page 8 of 11
Chloride, Comparison Instrument - Radiometer 725, Slope/Intercept - Y = 17.100 + 0.800 * X, Bias -4.0, Corr. Coeff. [r] = 0.965, # samples = 98, Unit (mmol/L)
SO2, Comparison Instrument - Omni Modular, Slope/Intercept - Y = 10.066 + 0.903 * X, Bias 1.1, Corr. Coeff. [r] = 0.991, # samples = 130, Unit (%)
SO2, Comparison Instrument - Radiometer 725, Slope/Intercept - Y = -3.969 + 1.037 * X, Bias -4.0, Corr. Coeff. [r] = 0.904, # samples = 102, Unit (%)
MetHb, Comparison Instrument - Omni Modular, Deviation of mean -0.3% abs, # samples = 129
MetHb, Comparison Instrument - Radiometer 725, , Deviation of mean +0.2% abs, # samples = 131
COHb, Comparison Instrument - Omni Modular, Deviation of mean 0.7% abs, # samples = 130
COHb, Comparison Instrument - Radiometer 725, , Deviation of mean +0.1% abs, # samples = 132
Hct, Comparison Instrument - Omni Modular, Slope/Intercept - Y = 0.100 + 1.00 * X, Bias 0.4, Corr. Coeff. [r] = 0.967, # samples = 132, Unit (%)
Hct, Comparison Instrument - Radiometer 725, Slope/Intercept - Y = -0.689 + 1.040 * X, Bias 0.6, Corr. Coeff. [r] = 0.946, # samples = 141, Unit (%)
Glucose, Comparison Instrument - Omni Modular, Slope/Intercept - Y = -0.461 + 1.034 * X, Bias -3.9, Corr. Coeff. [r] = 0.938, # samples = 134, Unit (mmol/L)
Glucose, Comparison Instrument - Radiometer 725, Slope/Intercept - Y = -0.867 + 1.201 * X, Bias 5.2, Corr. Coeff. [r] = 0.986, # samples = 107, Unit (mmol/L)
Urea, Comparison Instrument - Omni Modular, Slope/Intercept - Y = -0.343 + 0.850 * X, Bias -10.8, Corr. Coeff. [r] = 0.957, # samples = 122, Unit (mmol/L)
{8}
Page 9 of 11
Urea, Comparison Instrument - Cobas Mira, Slope/Intercept - Y = -0.001 + 0.887 * X, Bias -11.1, Corr. Coeff. [r] = 0.981, # samples = 129, Unit (mmol/L)
Lactate, Comparison Instrument - Omni Modular, Slope/Intercept - Y = -0.200 + 1.000 * X, Bias -9.5, Corr. Coeff. [r] = 0.936, # samples = 136, Unit (mmol/L)
Lactate, Comparison Instrument - Cobas Mira, Slope/Intercept - Y = -0.297 + 1.074 * X, Bias -3.0, Corr. Coeff. [r] = 0.968, # samples = 137, Unit (mmol/L)
b. Matrix comparison: The Omni S Analyzer utilizes the same sample types – whole blood, serum, plasma and aqueous material (QC material and dialysis fluids) as previously cleared on the predicate device, the Omni C analyzer (k013373)
3. Clinical studies:
a. Clinical sensitivity: None stated
b. Clinical specificity: None stated
c. Other clinical supportive data (when a and b are not applicable): None
4. Clinical cut-off: None stated
5. Expected values/Reference range: A complete listing of critical values and expected values are provided in the technical manual of the Omni S. The values reflect normal or expected ranges including ranges for neonates, children and adults as needed. Additionally, explanations of typical disease states with each critical value for each assay is provided. These ranges and supportive documentation are derived from previously established literature. Specifically: "Critical Care Testing: A Quick Reference Guide" by Andrew St. John First edition 2001, Tietz "Textbook of Clinical Chemistry" third edition 1999, and "Labor und Diagnose" Lothar Thomas, fifth expanded edition 2000.
M. Instrument Name: Roche Diagnostics Omni S Analyzer
N. System Descriptions:
1. Modes of Operation: Routine, STAT
2. Software: The Roche Omni S utilizes a PC based interface to the user with a touch sensitive color matrix display and a barcode reader for input
{9}
Page 10 of 11
and display of data. Additional microcontrollers controls all operations of the Roche Omni S. Lower level tasks are done by a network of microcontrollers. The micro-controller network is responsible for data acquisition and sensor/actuator handling. The programming languages are standard for this type of task.
FDA has reviewed applicant's Hazard Analysis and software development processes for this line of product types:
Yes ☐ X or No ☐
3. Sample Identification: Samples are identified via Bar Code or manual entry.
4. Specimen Sampling and Handling: Samples are a combination of whole blood, serum, plasma and aqueous solution (QC or dialysate). Samples are introduced via capillary tubes, syringes and sealed containers.
5. Assay Types: Ion selective Electrode, Enzymatic, MSS chamber sensors for BUN/urea nitrogen, for lactate and glucose the Omni S utilizes Braunstein technology.
6. Reaction Types: ISE, Enzymatic, MSS chamber sensors for BUN/urea nitrogen as the predicate devices. For lactate and glucose the Omni S utilizes Braunstein technology.
7. Calibration: The instrument automatically calibrates every 8, 12 or 24 hours (default) established by the user. This calibration includes: wavelength calibration of the polychromator, cleaning with internal cleaning solution, Automatic conditioning of the Na+ probe, Calibration of the mixing system and 2 point calibration of all parameters. Calibration of the O2 sensor is done every 30 minutes with ambient air and a zero point solution. O2 is recalibrated after every measurement. User can calibrate any parameter by performing a user activated calibration in the instrument software. Calibrators are sold separately.
8. Quality Control: Quality Control material is sold separately and consists of the following materials: Combitrol TS+ (k032453), Auto-trol+ (AutoQC material) (k032453), Combitrol Plus B and Auto-trol Plus B (AutoQC Material) (k032453). QC target ranges are included with each lot of QC material. Each analyte tested has three levels of QC range. QC timing is established by the user via the user interface.
O. Other Supportive Instrument Performance Characteristics Data Not Covered In The "L. Performance Characteristics" Section Of The SE Determination Decision Summary. The Roche Diagnostics Omni S Analyzer performs the following assays: Potassium, Calcium, Sodium,
{10}
Page 11 of 11
Chloride, Blood Gases (PCO2, PO2) and pH, Hemoglobin, Hematocrit, Oxygen Saturation, Carboxyhemoglobin, Oxyhemoglobin, Methemoglobin, Deoxyhemoglobin, Lactic Acid, Glucose, and Urea Nitrogen. Hemoglobin, Hematocrit, Deoxyhemoglobin, Carboxyhemoglobin, Oxyhemoglobin, Methemoglobin and Deoxyhemoglobin are regulated by the Office of InVitro Diagnostics Division of Hematology and Immunology. These assays were referred to the division for consultation. The results of that consultation are included with the review.
P. Conclusion: Based upon the information provided, I recommend that the Roche Diagnostics Omni S Analyzer be found substantially equivalent to the respective predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.