NOVELENS CONTACT LENS CLEANING ENHANCEMENT AND STORAGE ACCESSORY

K032309 · Novelens, Inc. · LRX · Dec 4, 2003 · Ophthalmic

Device Facts

Record IDK032309
Device NameNOVELENS CONTACT LENS CLEANING ENHANCEMENT AND STORAGE ACCESSORY
ApplicantNovelens, Inc.
Product CodeLRX · Ophthalmic
Decision DateDec 4, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 886.5928
Device ClassClass 2

Indications for Use

The NoveLens® Contact Lens Case is indicated for storage of hard, rigid gas permeable (silicone acrylate and fluro silicone acrylate) and soft (hydrophilic) contact lenses during chemical disinfection.

Device Story

NoveLens® Contact Lens Case is a plastic storage device for contact lenses. It features an upper chamber with two concave reservoirs for lenses and a lower chamber for collecting solution. User introduces solution into the upper chamber; gravity forces it through drainage holes and a tube into the lower chamber. Once filled, screw caps hermetically seal the system for chemical disinfection. Spent solution is expelled via external drainage holes by tipping the unit. Used by patients at home for lens storage during disinfection cycles. Device ensures proper containment of lenses and disinfection solution.

Clinical Evidence

No clinical data provided. Bench testing only; toxicology testing (Cytotoxicity, Ocular Irritation, Systemic Toxicity) performed on plastic components per ISO-10993-1 and FDA guidance for contact lens care products.

Technological Characteristics

Plastic contact lens storage case with two-chamber design (upper reservoir, lower collection chamber). Features gravity-fed drainage system and screw-cap hermetic seal. Materials verified via biocompatibility testing (ISO-10993-1).

Indications for Use

Indicated for storage of hard, rigid gas permeable (silicone acrylate and fluro silicone acrylate) and soft (hydrophilic) contact lenses during chemical disinfection.

Regulatory Classification

Identification

A soft (hydrophilic) contact lens care product is a device intended for use in the cleaning, rinsing, disinfecting, lubricating/rewetting, or storing of a soft (hydrophilic) contact lens. This includes all solutions and tablets used together with soft (hydrophilic) contact lenses and heat disinfecting units intended to disinfect a soft (hydrophilic) contact lens by means of heat.

Special Controls

*Classification.* Class II (Special Controls) Guidance Document: “Guidance for Industry Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products.”

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # K032309 510(k) Summary Statement ## SUBMITTER: ## Submitted on behalf of: | Company Name: | NoveLens®, Inc. | |------------------------|-------------------------------------------| | Address: | 17-25 Hunter Place | | | Fair Lawn, New Jersey 07410 | | Phone: | (201) 794-7999 | | Fax: | (201) 794-7958 | | CONTACT PERSON: | Richard E. Lippman, O.D., F.A.A.O. | | | Official Representative and Correspondent | | | R.P. Chiacchierini & Associates, LLC | | | 15825 Shady Grove Road | | | Suite 30 | | | Rockville, Maryland 20850 | | Phone: | (240) 683-3738 | | FAX: | (240) 683-9236 | | DATE SUMMARY PREPARED: | November 10, 2003 | | TRADE NAME: | NoveLens® Contact Lens Case | COMMON NAME: CLASSIFICATION: contact lens case Class II: LRX # ESTABLISHMENT REGISTRATION: PENDING # SUBSTANTIALLY EQUIVALENT TO: The NoveLens® Contact Lens Case is substantially equivalent to the following devices: | K013232 | Bausch & Lomb CyberCase Contact Lens Case | |---------|-------------------------------------------| | K852384 | Bausch & Lomb CyberCase Contact Lens Case | | K971618 | Alcon Contact Lens Case | | K974281 | Alcon Contact Lens Case | | K021248 | iCase: IKeeps, Inc, | ### DESCRIPTION of the DEVICE: The NoveLens® Contact Lens Case is a contact Iens storage case comprised of a plastic case that houses two chambers. The upper chamber which holds the lenses separately in individually attached concave reservoirs and a lower chamber that provides space for rinsed solution as it passes through holes from the upper chamber and through a drainage toube into . . {1}------------------------------------------------ the lower portion of the device. The NoveLens® Contact Lens Case is constructed to take solution introduced into the upper chamber in a rinse fashion and force, by gravity, solution through the drainage holes through a drainage tube and into a lower collecting chamber. Additional solution is added until both chambers are completely filled to the top. Screw caps when closed hermetically seal the system to hold disinfection solution while the lenses are stored and disinfected in accordance with directions consistent with labeled multipurpose contact lens solutions. The design provides for one-way ingress of solution and expulsion of spent solution through external drainage holes from the lower chamber upon tipping the unit on its side. ### INDICATIONS FOR USE: The NoveLens® Contact Lens Case is indicated for storage of hard, rigid gas permeable (silicone acrylate and fluro silicone acrylate) and soft (hydrophilic) contact lenses during chemical disinfection. #### ACTIONS: The NoveLens® Contact Lens Case provides for lens storage during chemical disinfection as well as storage up to the labeled time listed on the user's multipurpose disinfection solution. Directions for use accompany the NoveLens® Contact Lens Case for continued safe use of the device. # TOXICITY TESTING A battery of toxicology tests were conducted on the plastic components to verify the safety of the materials used in the device. The tests included Cytotoxicity, Ocular Irritation, and Systemic Toxicity in conformance with FDA Guidance for contact lens case products; Guidance for Industry-Premarket Notification 510(k) Guidance Document for Contact Lens Care Products- dated May 1, 1997, and ISO-10993-1. Results of the testing demonstrate that the safety and effectiveness of the NoveLens® Contact Lens Case is equivalent to the predicate devices listed above. ### EFFECTIVENESS TESTING for the NoveLens® Contact Lens Case . No further testing for case effectiveness was warranted for demonstration of substantial equivalence. ## CLINICAL INFORMATION Clinical studies were deemed unnecessary as the solutions used in the NoveLens® Contact Lens Case are already cleared for marketing as cleaning, rinsing, disinfection, and storage solution for soft, RGP, and hard contact lenses. {2}------------------------------------------------ # LABELING 、 . . . . . . . The NoveLens® Contact Lens Case is provided to the user with an outer carton as well as User Instructions. . : . 1 : : .. 1 : 199 1 11:44 : : 上一篇: . : ,在线上的 NoveLens® Inc. 17-25 Hunter Place Fair Lawn, New Jersey 07410 (201) 794-7999 FAX: (201) 794-7958 {3}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an image of a stylized eagle with three lines representing its wings. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 $$\mathtt{2EC} \text{ - 4 } \mathtt{2003}$$ NoveLens®, Inc. C/O Richard Lippman, O.D., F.A.A.O. Vice President, Ophthalmic Regulatory Medical Products R.P. Chiacchierini & Associates, LLC 15825 Shady Grove Rd., Suite 30 Rockville, MD 20850 Re: K032309 Trade/Device Name: NoveLens® Contact Lens Case Regulation Number: 21 CFR 886.5928, 21 CFR 886.5918 Regulation Name: Soft (hydrophilic) contact lens care products, Rigid gas permeable contact lens care products Regulatory Class: Class II Product Code: LRX Dated: November 10, 2003 Received: November 10, 2003 Dear Dr. Lippman: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {4}------------------------------------------------ # Page 2 - Richard Lippman, O.D., F.A.A.O. This letter will allow you to begin marketing your device as described in your Section 510/k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4613. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Sincerely yours, A. Ralph Rosenthal A. Ralph Rosenthal, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## Indication Statement 510(k) Number (if known): K032309 Device Name: NoveLens® Contact Lens Case # Indications for Use: The NoveLens® Contact Lens Case is indicated for storage of hard, rigid gas permeable (silicone acrylate and fluro silicone acrylate) and soft (hydrophilic) contact lenses during chemical disinfection. ## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Presription Use _ Over -The-Counter Use (Optional Format 1.26) (Oprtional Format 1-2-96) Karen Wabont ion Sion-Off Division of Ophthalmic Ear Nose and Throat Devi 510(k) Number K032309 11:11:11 : 11:11:11
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...