K032242 · Omi Manufacturing Pty., Ltd. · GES · Sep 23, 2003 · General, Plastic Surgery
Device Facts
Record ID
K032242
Device Name
PERSONNA PLUS SAFETY SCALPEL SYSTEM
Applicant
Omi Manufacturing Pty., Ltd.
Product Code
GES · General, Plastic Surgery
Decision Date
Sep 23, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4800
Device Class
Class 1
Attributes
3rd-Party Reviewed
Indications for Use
The PersonnaPlus Safety Scalpel System is a surgical safety scalpel intended for use as a cutting device during surgical, pathology and minor medical procedures. It is intended to aid in the protection against accidental injuries during loading, passing, and disposal.
Device Story
PersonnaPlus Safety Scalpel System is a manual surgical instrument featuring a retractable plastic shroud to protect against sharps injuries. Available as a totally disposable unit with a plastic handle or a 'reposable' unit with a stainless steel handle and replaceable blade. Used in surgical, pathology, and minor medical settings by clinicians. The device functions as a standard scalpel with an integrated safety mechanism to shield the blade during loading, passing, and disposal. The safety feature is designed to reduce the risk of accidental injury to healthcare personnel. Substantial equivalence is supported by a simulated user study conducted per FDA guidance for sharps injury prevention features.
Clinical Evidence
No clinical data. Evidence consists of a simulated user study conducted in accordance with the FDA guidance document 'Supplementary Guidance on Premarket Notifications for Medical Devices with Sharps Injury Prevention Features'.
Technological Characteristics
Manual surgical scalpel with retractable plastic shroud. Materials: stainless steel blades, plastic or stainless steel handles. Supplied sterile. Conforms to applicable ISO standards. No electronic components or software.
Indications for Use
Indicated for use as a cutting device during surgical, pathology, and minor medical procedures for patients requiring surgical intervention.
Regulatory Classification
Identification
A manual surgical instrument for general use is a nonpowered, hand-held, or hand manipulated device, either reusable or disposable, intended to be used in various general surgical procedures. The device includes the applicator, clip applier, biopsy brush, manual dermabrasion brush, scrub brush, cannula, ligature carrier, chisel, clamp, contractor, curette, cutter, dissector, elevator, skin graft expander, file, forceps, gouge, instrument guide, needle guide, hammer, hemostat, amputation hook, ligature passing and knot-tying instrument, knife, mallet, disposable or reusable aspiration and injection needle, disposable or reusable suturing needle, osteotome, pliers, rasp, retainer, retractor, saw, scalpel blade, scalpel handle, one-piece scalpel, snare, spatula, stapler, disposable or reusable stripper, stylet, suturing apparatus for the stomach and intestine, measuring tape, and calipers. A surgical instrument that has specialized uses in a specific medical specialty is classified in separate regulations in parts 868 through 892 of this chapter.
{0}------------------------------------------------
K032242
## EXHIBIT 2 OMI Manufacturing Pty Ltd Unit 1, 12 Booran Drive Slacks Creek Queensland 4127 Australia Phone : + 61- 7 3209 3099 Fax: + 61- 7 3209 4765 Contact: Mr. Bruce L. Kiehne, Director/Secretary 510(k) Summary
- 1. Identification of the Device: Proprietary-Trade Name: PersonnaPlus Safety Scalpel System also known as OMI Safe Scalpel Classification Name: Blade: GES, Handle: GDZ Common/Usual Name: Scalpel blade, scalpel handle This device is Class I but with sharps injury protection feature.
- 2. Equivalent legally marketed device: K923170 Los Alamos Retractable Knife and Personna Safety Scalpel" K924503
- 3. Indications for Use (intended use) The PersonnaPlus Safety Scalpel System is a surgical safety scalpel intended for use as a cutting device during surgical, pathology and minor medical procedures. It is intended to aid in the protection against accidental injuries during loading, passing, and disposal.,
- 4. Description of the Device: The device is a combination scalpel blade and handle incorporating a sharps injury protection feature. There are two product options -(1) a totally disposable unit and (2) a metal handle with disposable blade and guard - our "reposable" product. The totally disposable unit uses a plastic handle while the reposable product has a stainless steel handle.
{1}------------------------------------------------
| Device<br>Characteristics | K923170 Los Alamos<br>Retractable Knife | PersonnaPlus Medical Safety Scalpel<br>and Blades |
|---------------------------|-----------------------------------------|--------------------------------------------------------------------|
| Indications for use | General surgical use | SAME |
| Safety feature | Retractable blade | Blade guard retracts |
| Single use or reusable | Single use only, disposable. | Reusable with change of blade and<br>re-sterilization of handle |
| Blade size(s) | 10, 11, 15 | Handle sizes 3, 4<br>Blade sizes: 10,11,15,15C, 20, 21, 22, 23, 24 |
| Blade material | Stainless steel | SAME |
| Handle material | Plastic | Plastic or Stainless steel |
| Supplied | Sterile | SAME |
- 5. Safety and Effectiveness, comparison to predicate device:
- 6. In all respects, the PersonnaPlus Safety Scalpel System also known as OMI Safe Scalpel are substantially equivalent to one or more scalpels currently marketed in the USA. The scalpels are constructed of identical materials and conform to applicable ISO standards. An added safety feature of a retractable plastic shroud around the blade enhances safety. A successful simulated user study was conducted according to FDA guidance document Supplementary Guidance on Premarket Notifications for Medical Devices with Sharps Injury Prevention Features; Guidance for Industry and FDA Document Issued on: December 31, 2002
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo is circular in shape, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. In the center of the circle is a stylized image of a human figure, with three wavy lines representing the head, body, and legs.
Public Health Service
SEP 2 3 2003
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Omi Manufacturing Pty, Limited C/O Mr. George Kreimer Responsible Third Party Official Underwriters Laboratories, Incorporated 333 Pfingsten Road Northbrook, Illinois 60062
Re: K032242
Trade/Device Name: Personna Plus Safety Scalpel System Regulation Number: 878.4800 Regulation Name: Manual Surgical Instrument for General Use Regulatory Class: I Product Code: GES, GDZ Dated: September 4, 2003 Received: September 5, 2003
Dear Mr. Kreimer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 -Mr. Kreimer
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4618. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Susan Runner
Susan Runner, DDS, MA Interim Director Division of Anesthesiology, General Hospital Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## k) Indications for Use
510(k) Number
Device Name: PersonnaPlus Safety Scalpel System also known as OMI Safe Scalpel..
Indications for Use: The PersonnaPlus Safety Scalpel System is a surgical safety scalpel intended for use as a cutting device during surgical, pathology and minor medical procedures. It is intended to aid in the protection against accidental injuries during loading, passing, and disposal.
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
Over the Counter Use (Per 21 CFR 801.109)
Patricio Cussenite
OR
10(k) Number: K032242
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.