K032201 · Perusahaan Pelindung Getah (M) Sdn Bhd · LYY · Oct 10, 2003 · General Hospital
Device Facts
Record ID
K032201
Device Name
RUBBERCARE POWDERED LATEX EXAMINATION GLOVES
Applicant
Perusahaan Pelindung Getah (M) Sdn Bhd
Product Code
LYY · General Hospital
Decision Date
Oct 10, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
The device is intended as a protective device to be worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Device Story
RubberCare powdered latex examination gloves; manufactured from natural rubber latex; USP cornstarch used as donning powder. Manufacturing process includes pre and post leaching stages to remove residual chemicals and water-soluble proteins. Used in clinical settings by healthcare professionals as a protective barrier on hands or fingers to prevent contamination between patient and examiner. Device provides physical barrier protection; protein content limited to 100 micrograms or less per gram.
Clinical Evidence
Bench testing only. Testing included physical dimensions (length, width, thickness), tensile strength and ultimate elongation (before and after aging), water extractable protein content, water leak test (AQL), and biocompatibility (skin irritation and dermal sensitization tests). All results met or exceeded ASTM D3578-01a22 standards and FDA requirements.
Technological Characteristics
Material: Natural rubber latex. Donning powder: USP cornstarch. Manufacturing: Pre and post leaching stages for protein/chemical removal. Dimensions: Length 240-245mm, palm thickness 0.14-0.16mm, finger thickness 0.16-0.20mm. Physical properties: Tensile strength 20-29 MPa (before aging), elongation 750-900% (before aging). Protein content: <= 100µg/gram. Biocompatibility: Passed skin irritation and dermal sensitization tests. Standards: ASTM D3578-01a22.
Indications for Use
Indicated for use as a disposable protective barrier worn on the examiner's hand or finger to prevent cross-contamination between patient and examiner.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image is a simple black and white image with a few small, dark spots scattered across the white background. The spots are irregularly shaped and appear to be randomly placed. The overall impression is one of sparseness and simplicity, with the dark spots providing a minimal contrast against the bright background.
K032201
# PERUSAHAAN PELINDUNG GETAH (M) SDN BHD
(Company No. 169997-P) Lot 110, Lorong Senawang 4/3, Off Jalan Senawang Empat, Senawang Industrial Estate, 70450 Seremban, Negeri Sembilan Darul Khusus, Malaysia. Fax: 606-6772780 Tel: 606-6772781
Image /page/0/Picture/5 description: The image contains two logos. The logo on the left has the letters "BVQi" stacked on top of each other, with a horizontal line underneath. Below the line, the text "ISO 9001" is printed. The logo on the right has a crown above a check mark, with the text "UKAS QUALITY MANAGEMENT" printed below.
Amended: October 2, 2003 (Original page 8 of 20)
## 510(K) SUMMARY
| Applicant: | PERUSAHAAN PELINDUNG GETAH (M) SDN BHD | | |
|---------------------|-------------------------------------------------------------------------------------------------------------------------------------|---------|--------------|
| Address | Lot 110, Lorong Senawang 4/3, Off Jalan Senawang Empat<br>Senawang Industrial Estate, 70450 Seremban,<br>Negeri Sembilan, Malaysia. | | |
| Phone No. | 60-6-6772781 | Fax No. | 60-6-6772780 |
| Contact Person | Peter Yew Nieng Choon | | |
| Date of Summary | October 2, 2003 | | |
| Device Information | | | |
| Trade Name | RUBBERCARE | | |
| Common Name | Powdered Latex Exam Gloves | | |
| Classification Name | Patient Examination Glove, powdered | | |
## Claim of Equivalence
The device is a class I latex patient examination gloves 80LYY powdered with absorbable dusting powder, that meets all the requirements of ASTM standard D 3578-01a2.
## Device Description
It is made from natural rubber latex using USP cornstarch as donning powder. The manufacturing process includes pre and post leaching stages to remove residual chemicals and water-soluble proteins. The extent of protein removal is within the limit of the ASTM D3578-01a24 recommended limits for both powder and water-soluble protein.
{1}------------------------------------------------
## PERUSAHAAN PELINDUNG GETAH (M) SDN BHD
(Company No. 169997-P) Lot 110, Lorong Senawang 4/3, Off Jalan Senawang Empat, Senawang Industrial Estate, 70450 Seremban. Negeri Sembilan Darul Khusus, Malaysia Tel: 606-6772781 Fax: 606-6772780
Image /page/1/Picture/2 description: The image contains two logos. The logo on the left has the letters "BVQi" stacked on top of each other, with the text "ISO 9001" underneath. The logo on the right has a crown above a check mark, with the text "UKAS QUALITY MANAGEMENT" underneath.
Amended: October 2, 2003 (Original page 9 of 20)
### Intended Use of Device
The device is intended as a protective device to be worn on the examiner's hand or finger to prevent contamination between patient and examiner.
### Technological Characteristics
Following is a table showing the measured parameters of the gloves (e.g. lengths, thickness, widths, physical properties, protein contents, etc) as compared to ASTM. Also data that meets FDA biocompatibility, pinhole, and any other requirements, and any other parameter on which we want to make a labeling claim.
| Parameter | ASTM Specifications | Measured Values |
|--------------------------------------|---------------------|----------------------|
| Length (sizes M, L), mm | 230 min. | 240 - 245 |
| Thickness (palm), mm | 0.08 min. | 0.14-0.16 |
| Thickness (finger), mm | 0.08 min. | 0.16-0.20 |
| Width (size M), mm | $95 \pm 10$ | 94 - 98 |
| Tensile Strength, Before Aging, Mpa | 18 min. | 20 - 29 |
| Tensile Strength, After Aging, Mpa | 14 min. | 20 - 26 |
| Ultimate Elongation, Before Aging, % | 650 min. | 750 - 900 |
| Ultimate Elongation, After Aging, % | 500 min. | 750-950 |
| Water Extractable Protein, µg/gram | 200µg/dm² | 100µg/gram and below |
| | (recommended) | |
| Water Leak Test, Before Aging, AQL | 2.5 | 1.5 and below |
| Water Leak Test, After Aging, AQL | n.a. | 2.5 and below |
| Residual Powder, mg/dm² | 10 | 9-10 |
| Skin Irritation Test | n.a | Passed* |
| Dermal Sensitization Test | n.a | Passed* |
* Please refers to attachment H (page 19)
#### Conclusions
Based on the test data given above, we certify that our gloves:
- meet or exceed the ASTM standard D3578-01a22 a.
- b. meet the FDA pinhole requirements; and
- c. meet our labeling claim on protein content.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract image of an eagle or bird-like figure, represented by three curved lines that suggest wings or feathers.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## OCT 1 0 2003
Mr. Peter Yew Nieng Choon Managing Director Perusahaan Pelindung Getah (M) SDN BHD Lot 110, Lorong Senawang 4/3, Off Jalan Senawang Empat Senawang Industrial Estate, 70450 Seremban, Negeri Sembilan, Darul Khusus, MALAYSIA
#### Re: K032201
Trade/Device Name: Rubbercare Powdered Latex Examination Gloves (containing 100 Micrograms or Less of Total Water Extractable Protein Per gram) Regulation Number: Patient Examination Glove Regulation Name: 880.6250 Regulatory Class: I Product Code: LYY Dated: October 2, 2003 Received: October 6, 2003
Dear Mr. Nieng Choon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Nieng Choon
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits vour device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4618. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours.
Chiu Lin, Ph.D.
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## PERUSAHAAN PELINDUNG GETAH (M) SDN BHD
(Company No. 169997-P) Lot 110, Lorong Senawang 4/3, Off Jalan Senawang Empat, Senawang Industrial Estate, 70450 Seremban, Negeri Sembilan Darul Khusus, Malaysia. Fax: 606-6772780 Tel: 606-6772781
Image /page/4/Picture/2 description: The image contains two logos. The logo on the left has the letters "BVQi" in a bold, sans-serif font, with a horizontal line above and below the letters. Below the logo, the text "ISO 9001" is printed. The logo on the right has a check mark above the text "UKAS QUALITY MANAGEMENT".
Amended: October 2, 2003 (Original page 3 of 20)
## INDICATIONS FOR USE
PERUSAHAAN PELINDUNG GETAH (M) SDN BHD Applicant:
K032201 510(k) No.
RubberCare Powdered Latex Examination Gloves (containing 100 Device Name: micrograms or less of total water extractable protein per gram)
#### Indications for Use
The RubberCare powdered latex examination glove is a disposable device intended for medical The Raober on on the examiner's hand or finger to prevent contamination between patient and examiner.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Sat. Pat. fx. Chan-Len
(Division Sign-Off)
Division of Anesthesiology, General Hospital,
Infection Control, Dental Devices
510(k) Number: K032201
Prescription Use (Pcr 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.