K032183 · Boston Scientific Corp · DQX · Aug 1, 2003 · Cardiovascular
Device Facts
Record ID
K032183
Device Name
IQ HYDROPHILIC GUIDE WIRE
Applicant
Boston Scientific Corp
Product Code
DQX · Cardiovascular
Decision Date
Aug 1, 2003
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1330
Device Class
Class 2
Indications for Use
The IQ™ Hydrophilic Guide Wires are intended to facilitate the placement of balloon dilatation catheters or other therapeutic devices during PTCA, PTA, or other intravascular interventional procedures. The IQ™ Hydrophilic Guide Wires are not intended for use in the cerebral vasculature.
Device Story
Steerable, spring-tipped guide wire; 0.014-inch diameter; 185cm or 300cm lengths. Corewire features PTFE-coated stainless steel segment coupled to tapered Nitinol segment; distal end jacketed with Tungsten-loaded polyurethane sleeve and ICE® hydrophilic coating. Distal 3cm includes radiopaque spring coil, centering sleeve, and stainless steel ribbon. Used by physicians in clinical settings to facilitate catheter placement during intravascular interventions. Provides tactile feedback and steerability for navigation through vasculature. 185cm version compatible with AddWire™ Extension Wire. Single-use, sterile, non-pyrogenic.
Clinical Evidence
Bench testing only. Evaluated torque response, tip prolapse, tip shapeability, marker location, coating adherence, tensile strength, combined load, visual inspection, polymer peel, device compatibility, biocompatibility, and shelf-life. Results met all minimum requirements.
Indicated for patients undergoing PTCA, PTA, or other intravascular interventional procedures requiring placement/exchange of balloon dilatation catheters or therapeutic devices. Contraindicated for use in cerebral vasculature.
Regulatory Classification
Identification
A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.
Special Controls
*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}------------------------------------------------
K032183
AUG - 1 2003
Special 510(k): Device Modification
BSC IQ Hydrophilic Guide Wire
## 510(k) Summary
per 21 CFR §807.92
| Submitter's Name<br>and Address | Boston Scientific Corporation (BSC)<br>One Scimed Place<br>Maple Grove, MN 55311 | | |
|---------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|------------------|
| Contact Name and<br>Information | Anne V. Rossi<br>Specialist, Regulatory Affairs<br>Phone: 763-494-2827<br>Fax: 763-494-2222<br>e-mail: Anne.Rossi@bsci.com | | |
| Date Prepared | July 16, 2003 | | |
| Proprietary<br>Name(s) | IQ™ Hydrophilic Guide Wire | | |
| Common Name | Catheter Guide Wire | | |
| Product Code | 74DQX | | |
| Classification of<br>Device | Class II, 21 CFR Part 870.1330 | | |
| Predicate Devices | PT2™ Guide Wire | K030617 | May 21, 2003 |
| | Luge™ Guide Wire | K973945 | January 12, 1998 |
| Device<br>Description | The IQ™ Hydrophilic Guide Wires with ICE® hydrophilic coating<br>are steerable, spring-tipped guide wires. The IQ™ Hydrophilic<br>is available in a nominal diameter of 0.014 inches, Moderate<br>Support, and nominal lengths of 185 and 300 centimeters with<br>Brachial and Femoral marks at 90cm and 100cm respectively.<br>The distal three centimeters of all models are radiopaque spring<br>coils and available in either a straight shapeable or a pre-<br>formed J-Tip.<br>The IQ™ Hydrophilic corewire consists of a PTFE coated SS<br>corewire segment joined via a nickel-chromium alloy coupler to<br>a tapered Nitinol corewire segment. The distal segment of this<br>corewire is coated with an adhesive pre-coat that is jacketed<br>with a Tungsten loaded polyurethane sleeve. This polymer<br>sleeve is coated with ICE® Hydrophilic coating. A SS centering<br>sleeve attaches a SS ribbon to the distal end of the corewire.<br>The distal most portion of the corewire, the centering sleeve and<br>the SS ribbon reside within the hollow center of the 3-cm spring-<br>coil.<br>The 185-centimeter version of the IQ™ Hydrophilic is designed<br>with a proximal extension section that allows connection to the<br>AddWire™ Extension Wire. | | |
{1}------------------------------------------------
The IQ™ Hydrophilic Guide Wires are intended to facilitate the Intended Use of placement and exchange of balloon dilatation catheters or other Device therapeutic devices during PTCA or PTA or other intravascular interventional procedures. The IQ™ Hydrophilic Guide Wires are not intended for use in the cerebral vasculature. The devices are provided non-pyrogenic, sterile, and intended for one procedure only.
Technological The IQ™ Hydrophilic Guide Wires utilize similar materials and Characteristics methods of construction as the currently marketed PT2™ Guide Wire family of quide wires. The differences in construction are the addition of a spring-coil to the distal most tip and a SS ring at the ribbon/corewire joint.
Non-Clinical Test Testing and evaluation of the IQ™ Hydrophilic Guide Wires Summary included torque response, tip prolapse, tip shapeability, marker location, coating adherence/presence, tensile, combined load, visual inspection, polymer peel, device compatibility, biocompatibility, and product shelf-life. Test results verified that the IQ™ Hydrophilic Guide Wires met all of the minimum requirements and are adequate for their intended use. The IQ™ Hydrophilic Guide Wires are considered to be substantially equivalent to guide wires currently marketed by Boston Scientific based on a comparison of intended use, design and the results of in vitro testing and evaluation.
KC32183
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is an abstract symbol that resembles an eagle or a bird in flight.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG - 1 2003
Boston Scientific Corporation c/o Anne V. Rossi Specialist, Regulatory Affairs One Scimed Place Maple Grove, MN 55311
Re: K032183
> IQ™ Hydrophilic Guide Wire Regulation Number: 870.1330 Regulation Name: Catheter guide wire Regulatory Class: Class II Product Code: DQX Dated: July 16, 2003 Received: July 21, 2003
## Dear Ms. Rossi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. .
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Ms. Anne V. Rossi
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Bram D. Zuckerman, M.D.
Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
{4}------------------------------------------------
## Indications for Use Statement
| 510(k) Number<br>(if known) | K032183 |
|-----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | IQ™ Hydrophilic Guide Wire |
| Indications For<br>Use | The IQ™ Hydrophilic Guide Wires are intended to facilitate the<br>placement of balloon dilatation catheters or other therapeutic<br>devices during PTCA, PTA, or other intravascular interventional<br>procedures. The IQ™ Hydrophilic Guide Wires are not intended<br>for use in the cerebral vasculature. |
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
X
(Per 21 CFR §801.109)
Over-The-Counter Use OR
N'DeCartha
iovascular Devices
510(k) Number K032183
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.