← Product Code [IPF](/productcode/IPF) · K032178

# RHYTHM TOUCH Q 2-WAY (K032178)

_Shockim Enterprises, Ltd. · IPF · Oct 15, 2003 · Physical Medicine · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K032178

## Device Facts

- **Applicant:** Shockim Enterprises, Ltd.
- **Product Code:** [IPF](/productcode/IPF.md)
- **Decision Date:** Oct 15, 2003
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 890.5850
- **Device Class:** Class 2
- **Review Panel:** Physical Medicine
- **Attributes:** Therapeutic

## Indications for Use

Electrically powered devices intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.

## Device Story

Rhythm Touch Q 2-Way is a dual-channel powered muscle stimulator. Device generates electrical stimulation signals delivered to muscles via skin electrodes and lead wires. Powered by two AAA batteries. Used to induce muscle contraction for therapeutic purposes. Operates by passing electrical currents through electrodes placed on the patient's body. Intended for clinical or home use to manage muscle spasms, atrophy, circulation, and post-surgical thrombosis prevention.

## Clinical Evidence

No clinical data provided; substantial equivalence is based on technological characteristics and intended use comparison to the predicate device.

## Technological Characteristics

Dual-channel powered muscle stimulator. Components include electronic stimulation module, lead wires, and 2" x 2" adhesive electrodes. Power source: two AAA batteries. Operates via electrical stimulation of muscles. No software algorithm or connectivity described.

## Regulatory Identification

A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.

## Predicate Devices

- BioStim Digital NMS ([K010749](/device/K010749.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## Shockim Enterprises, Ltd.

2500 Wilshire Blvd. Suite 1165 Los Angeles, CA 900057 213-365-2279

## 510(k) SUMMARY

| Submitter's name: | Shockim Enterprises, Ltd.                                |
|-------------------|----------------------------------------------------------|
| Address:          | 2500 Wilshire Blvd., Suite 1165<br>Los Angeles, CA 90057 |
| Phone:            | 213-365-2279                                             |

Name of contact person:

Greq Holland Requlatory Specialists, Inc 3722 Ave. Sausalito Irvine, CA 92606 Phone: 949-262-0411 fax: 949-552-2821 greg@regulatoryspecialists.com

Date the summary was prepared: July 15, 2003

| Name of the device:        | Rhythm Touch Q 2 - Way                                |
|----------------------------|-------------------------------------------------------|
| Trade or proprietary name: | Rhythm Touch Q 2 - Way                                |
| Common or usual name:      | Electrical muscle stimulation device                  |
| Classification name:       | Power muscle stimulator (per 21 CFR section 890.5850) |

The legally marketed device to which we are claiming equivalence [807.92(a)(3)]:

BioStim Digital NMS, manufactured by BioMedical Life Systems. The clearance number is K010749.

Description of the device:

The Rhythm Touch Q 2-Way is a dual channeled powered muscle stimulator. It electronically stimulates muscles. It comprises two main components, namely, an electronic stimulatory module which generates the required stimulation signals, and skin electrodes with lead wires.

The product is supplied with a set of single sided adhesive electrodes, an instruction manual, and a set of batteries. Power is derived from two AAA cells located in a compartment protected by a removable battery cover.

{1}------------------------------------------------

The electrodes, which are provided with this model, have 510(k) clearance. They are 2" x 2" in size. The leadwires also have 510(k) clearance.

Intended use of the device:

Electrically powered devices intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.

Indications For Use include:

- 1. Relaxation of muscle spasms
- 2. Prevention or retardation of disuse atrophy
- 3. Increasing local blood circulation
- 4. Muscle re-education
- 5. Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis
- 6. Maintaining or increasing range of motion

Summary of the technological characteristics of our device compared to the predicate device:

As was provided in both the Comparison and Standards sections, the Rhythm Touch Q 2-Way device and the BioStim device have similar technological characteristics and are equivalent.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three curved lines representing its body and wings. The eagle is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

Public Health Service

OCT 1 5 2003

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Shockim Enterprises, Ltd. C/o Mr. Greg Holland Regulatory Specialists, Inc. 3722 Avenue Sausalito Irvine, CA 92606

Re: K032178

Trade/Device Name: Rhythm Touch Q 2-Way Regulation Number: 21 CFR 890.5850 Regulation Name: Powered muscle stimulator Regulatory Class: II Product Code: IPF Dated: July 14, 2003 Received: July 17, 2003

Dear Mr. Holland:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 – Mr. Greg Holland

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Sincerely yours,

Mark N. Milbern

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

| Page | 1 | of | 1 |
|------|---|----|---|
|------|---|----|---|

| 510(k) Number (if known): | K032179 |
|---------------------------|---------|
|---------------------------|---------|

Device Name: Rhythm Touch Q

Indications For Use:

Electrically powered devices intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.

Indications For Use include:

- Relaxation of muscle spasms 1.
- Prevention or retardation of disuse atrophy 2.
- Increasing local blood circulation 3.
- 4. Muscle re-education
- Immediate post-surgical stimulation of calf muscles to prevent ട്. venous thrombosis
- Maintaining or increasing range of motion ട.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use
(Per 21 CFR 801.109)

OR

Over-The-Counter Use***_***_

(Optional Format 1-2-96)

Mark N. Millican
(Division Sign-Off)
Division of General Restorative

Division of General. Restorative and Neurological Levices

78 510(k) Number_

REGULATORY SPECIALISTS, INC.

---

**Source:** [https://fda-staging.innolitics.com/device/K032178](https://fda-staging.innolitics.com/device/K032178)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).