The PASSmed is a pedicle screw system intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). As a pedicle screw system PASSmed is indicated for the treatment of severe spondylolisthesis (Grades 3 and 4) of the LS-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion. It also includes hooks and a sacral plate indicated for degenerative disc disease (ddd) defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies, spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis), tumor, pseudoarthrosis and failed previous fusion.
Device Story
PASSmed is a posterior pedicle screw and spinal fixation system; includes polyaxial screws, rods (Ø6mm), sacral plates, and hooks. Used for thoracic, lumbar, and sacral spinal stabilization; provides immobilization as an adjunct to fusion. Components connect via polyaxial attachment mechanisms or clamps; sacral plates attach to sacrum with screws. Operated by surgeons in clinical/OR settings. Provides mechanical stabilization to spinal segments; facilitates fusion; intended for removal after solid fusion achieved. Benefits patients by addressing spinal instabilities and deformities.
Clinical Evidence
Bench testing only. Mechanical testing performed according to ASTM F1717 indicates the PASSmed system is as mechanically sound as other commercially available spinal fixation devices.
Technological Characteristics
System includes polyaxial screws, Ø6mm rods, sacral plates, and laminar/pedicle hooks. Polyaxial attachment mechanism used for screws and rod-plates. Materials and design are identical to predicate devices. Testing conducted per ASTM F1717.
Indications for Use
Indicated for skeletally mature patients requiring spinal immobilization/stabilization as an adjunct to fusion. Conditions include degenerative spondylolisthesis, severe spondylolisthesis (Grades 3-4, L5-S1), degenerative disc disease, trauma (fracture/dislocation), spinal stenosis, deformities (scoliosis, kyphosis, lordosis), spinal tumors, and pseudarthrosis.
Regulatory Classification
Identification
A spinal interlaminal fixation orthosis is a device intended to be implanted made of an alloy, such as stainless steel, that consists of various hooks and a posteriorly placed compression or distraction rod. The device is implanted, usually across three adjacent vertebrae, to straighten and immobilize the spine to allow bone grafts to unite and fuse the vertebrae together. The device is used primarily in the treatment of scoliosis (a lateral curvature of the spine), but it also may be used in the treatment of fracture or dislocation of the spine, grades 3 and 4 of spondylolisthesis (a dislocation of the spinal column), and lower back syndrome.
{0}------------------------------------------------
510(k) Summary
Date November 11, 2003
NOV 1 4 2003
## Submitter
Medicrea Z.I. Chef de Baic 17000 LaRochelle, FRANCE
## Contact person
J.I). Webb 1001 Oakwood Blvd Round Rock, TX 78681 512-388-0199
## Common name
Posterior pedicle screw system Hooks Sacral plate
## Classification name
Spondylolisthesis Spinal Fixation Device System per MNH 888.3070 Pedicle Screw Spinal System per MNI 888.3070 Spinal Interlaminal Fixation Orthosis per KWP 888.3050
#### Equivalent Device
The components covered by this submission are exactly the same as those cleared in K001024, K012175, K013191, K013431, K013442 and K020236. These devices were all submitted by Encore Orthopedics. All of the cleared devices were designed, manufactured and packaged by Medicrea.
#### Device Description
The PASSmed system includes pedicies include polyasial screws in a various lengths and diameters. Standard and offist height seres are available, with the offset used in cases of severe spondylolisticsis. Standard and realignment clamps are used to connect the screw and rods. Rods are Ø6mm in lengths ranging from 50mm. It also includes sacral plates and serens. The sacral plate takes the place of pedicle screw in connecting the rod to the sacrum. The place is attached to the sarum with two serews. The polyasial attachment mechanism to the same as the pedicle screws. The plates come in right and left configuration. The rod-plates are similar to the consist of a short rod segment that has enarged portions at the ends with holes to attach directly to screw with hemispherical nuts rather than connecting to the rod was a clamp. The rods attach to the hooks and can be used for single or multiple level lisations. They have the same rod attachment mechanism as the polyasial serews. The laminar hooks are inserted inferior around the lamina, pediele hooks are inserted inferior and superior around the pedicles.
## Intended Use
The PASSmed is a pediele serew system intended to provide immobilization of spinal segments in skelctally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities of the thoracic, lumbar, and sacral spine: degencive spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis).
As a pedicle screw system PASSmed is indicated for the treament of severe spondylolisthesis (Grades 3 and 4) of the L2-51 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacra spine (L 3 to sacrum) with removal of the implants after the attainment of a solid fusion.
I also includes hooks and a sacral plate indicated for degenerative disc disease (ddd) defined as back pain of discogence origin with degeneration of the disc confirmed by history and radiographic studies, trauma (i.e., fracture or dislocation), spinal stenosis, deformities (i.e., scolosis, kyphosis, and/or lordosis), tumor, pseudoat(trosis and failed previous fusion.
#### Summary Nonclinical Tests
Testing performed according to ASTM F1717 indicate that the PASSmed system is as mechanically sound as other devices commercially available.
K032094
Page 1 of 1
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is an abstract image of an eagle with its wings spread.
Public Health Service
NOV 1 4 2003
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Medicrea C/o Mr. J.D. Webb 1001 Oakwood Boulevard Round Rock, Texas 78681
Re: K032094 Trade/Device Name: PASSmed Spinal System Regulation Number: 21 CFR 888.3050, 21 CFR 888.3070 Regulation Name: Spinal interlaminal fixation orthosis, Pedicle screw spinal system Regulatory Class: II Product Code: KWP, MNH, MNI Dated: October 10, 2003 Received: October 14, 2003
Dear Mr. Webb:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
Page 2 - Mr. J.D. Webb
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
to Mark A. Millers
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
510(k) number (if known): K032094
Device Name: PASSmed Spinal System
Page 1 of 1
Indications for Use:
# PASSmed Spinal System Indications for Use
The PASSmed is a pedicle screw system intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis).
As a pedicle screw system PASSmed is indicated for the treatment of severe spondylolisthesis (Grades 3 and 4) of the LS-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion.
It also includes hooks and a sacral plate indicated for degenerative disc disease (ddd) defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies, spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis), tumor, pseudoarthrosis and failed previous fusion.
(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE
Prescription Use (per 21 CFR 801.109) OR
Over-the-Counter Use
(Optional format 1-2-96)
)
Mark N Mulkerson
(Division Sign-off) Division of General, Neurological and Restorative Devices
510(k) Number K032049
Page iii
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.