K031970 · Siemens Medical Solutions USA, Inc. · DQK · Sep 9, 2003 · Cardiovascular
Device Facts
Record ID
K031970
Device Name
INFINITY MEGACARE
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
DQK · Cardiovascular
Decision Date
Sep 9, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1425
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
INFINITY MegaCare is a software application for viewing, manual editing, printing, and archiving of digitized electrocardiograph records from Rest ECG devices, Exercise ECG devices, Ambulance ECG devices, Holter ECG devices and the Infinity Monitoring System. INFINITY MegaCare is intended to provide analysis of Rest ECG's and to provide preliminary data for editing and confirmation by a physician. Infinity MegaCare can provide a serial comparison of Rest ECG data to facilitate the review of current and previous Rest ECG's. INFINITY MegaCare is designed for network compatibility to facilitate retrieval of data and to interface with other hospital information systems though HL7 protocols. The device is intended for use in an environment where patient care is provided by Healthcare Professionals, i.e. physicians, nurses, and technicians, trained on the use of the device, who will determine when use of the device is indicated, based upon their professional assessment of the patient's medical condition. The device is intended for use with all patient populations.
Device Story
INFINITY MegaCare is a software application for managing digitized ECG records from various sources (Rest, Exercise, Ambulance, Holter, Monitoring System). It functions as a diagnostic computer system; it ingests ECG data to provide automated analysis and serial comparison of Rest ECGs. The software facilitates manual editing, printing, and archiving of records. Used in clinical environments by trained healthcare professionals (physicians, nurses, technicians), the system integrates into hospital networks via HL7 protocols. The output provides preliminary diagnostic data for physician review and confirmation, supporting clinical decision-making by enabling longitudinal review of patient ECG history. The device is not MRI compatible.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on bench testing and the use of a previously cleared ECG analysis algorithm.
Technological Characteristics
Software-based diagnostic computer system. Operates on user-provided IBM-compatible servers. Utilizes the Glasgow ECG analysis algorithm. Connectivity via HL7 protocols for hospital information system integration. Not MRI compatible.
Indications for Use
Indicated for all patient populations requiring analysis, viewing, editing, printing, or archiving of digitized ECG records (Rest, Exercise, Ambulance, Holter, or Monitoring System). Intended for use by trained healthcare professionals (physicians, nurses, technicians) to provide preliminary data for physician confirmation.
Regulatory Classification
Identification
A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.
Predicate Devices
Infinity MultiView WorkStation Enhanced with Diagnostic Statements (Rest ECG) (K980625)
{0}------------------------------------------------
#### 510(k) SUMMARY
as required per 807.92(c)
Submitters Name, Address:
Siemens Medical Solutuions Electromedical Systems Group, PCS 16 Electronics Avenue Danvers, MA 01923 Tel: (978) 907-7500 Fax: (978) 750-6879 Official Correspondent: Connie Hertel, Director Ouality Assurance & Regulatory Affairs Contact person for this submission: Penelope H. Greco Date submission was prepared: June 20, 2003
Trade Name, Common Name and Classification Name:
A. Trade Name: INFINITY MegaCare
- B. Common Name, Classification Name, Class and Regulation Number:
| Common Name | Product Code | Class | Regulation Number |
|------------------------------------|--------------|-------|-------------------|
| System, ECG Analysis | LOS | | |
| Computer, diagnostic, programmable | DQK | II | 21CFR 870.1425 |
Predicate Device Identification:
| K980625 | Infinity MultiView WorkStation Enhanced with<br>Diagnostic Statements (Rest ECG) |
|---------|----------------------------------------------------------------------------------|
| K992637 | Muse Cardiovascular Information System |
| K974420 | TraceMaster ECG Management System |
| K946281 | Burdick Eclipse 4 Electrocardiograph |
Device Description:
INFINITY MegaCare is a computer software program that allows viewing, manual editing, printing and archiving of digitized electrocardiograph records from Rest ECG devices, Exercise ECG devices, ambulance ECG devices, Holter ECG devices and the INFINITY Monitoring System.
Page 1 of 2
#### COMPANY CONFIDENTIAL
Siemens Medical Solutuions Electromedical Systems Group, PCS
16 Electronics Avenue Danvers, MA 01923
Tel: (978) 907-7500 Fax: (978) 750-6879
{1}------------------------------------------------
INFINITY MegaCare uses the Microsoft Windows 2000 server operating system, Microsoft IIS Web server, and Microsoft SOL Server 2000 relational database. The system consists of a software application, which is installed on a user provided IBM compatible server running the Microsoft Windows 2000 Server operation system. MegaCare utilizes the ECG analysis algorithm developed under the direction of Dr. Peter MacFarlane at the University of Glasgow and used for the INFINITY MVWS Rest ECG (K980625) and Burdick's Eclipse 4 Electrocardiograph (K946281).
### Intended Use:
INFINITY MegaCare is a software application for viewing, manual editing, printing, and archiving of digitized electrocardiograph records from Rest ECG devices, Exercise ECG devices, Ambulance ECG devices, Holter ECG devices and the Infinity Monitoring System.
INFINITY MegaCare is intended to provide analysis of Rest ECG's and to provide preliminary data for editing and confirmation by a physician. Infinity MegaCare can provide a serial comparison of Rest ECG data to facilitate the review of current and previous Rest ECG's.
INFINITY MegaCare is designed for network compatibility to facilitate retrieval of data and to interface with other hospital information systems though HL7 protocols.
# Assessment of non-clinical performance data for equivalence: Substantially equivalent (Section S)
Assessment of clinical performance data for equivalence: Substantially equivalent (Section T)
Biocompatability: Not applicable
Sterilization: Not applicable
Standards and Guidance: Section R
## COMPANY CONFIDENTIAL
Siemens Medical Solutions Electromedical Systems Group, PCS
16 Electronics Avenue Danvers, MA 01923
Tel: (978) 907-7500 Fax: (978) 750-6879
{2}------------------------------------------------
Public Health Service
Image /page/2/Picture/2 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a stylized symbol that resembles three overlapping human profiles or figures. The symbol is abstract and appears to be in motion or flowing.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP - 9 2003
Siemens Medical Solutions USA, Inc. Electromedical Systems Group, PCS c/o Ms. Penelope H. Greco Regulatory Submissions Manager 16 Electronics Avenue Danvers, MA 01923
### Re: K031970
Trade Name: Siemens INFINITY MegaCare Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable diagnostic computer Regulatory Class: Class II (two) Product Code: DQK Dated: June 20, 2003 Received: June 26, 2003
#### Dear Ms. Greco:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Ms. Penelope H. Greco
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
A. E. Zeller, M.D.
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# K031970
Page 1 of 1
510(k) Number (if known):
Siemens INFINITY MegaCare Device Name:
Indications for Use:
INFINITY MegaCare is a software application for viewing, manual editing, printing, and archiving of digitized electrocardiograph records from Rest ECG devices, Exercise ECG devices, Ambulance ECG devices, Holter ECG devices and the Infinity Monitoring System.
INFINITY MegaCare is intended to provide analysis of Rest ECG's and to provide preliminary data for editing and confirmation by a physician. INFINITY MegaCare can provide a serial comparison of Rest ECG data to facilitate the review of current and previous Rest ECG's.
INFINITY MegaCare is designed for network compatibility to facilitate retrieval of data and to interface with other hospital information systems though HL7 protocols.
The device is intended for use in an environment where patient care is provided by Healthcare Professionals, i.e. physicians, nurses, and technicians, trained on the use of the device, who will determine when use of the device is indicated, based upon their professional assessment of the patient's medical condition.
The device is intended for use with all patient populations.
MRI Compatibility Statement:
The INFINITY MegaCare is not compatible for use in a MRI magnetic field.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
rdiovascular Devices
**510(k) Number** K031170
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.