MICROLIFE INSTANT DIGITAL ELECTRONIC THERMOMETER, MODEL QT1JA1
K031958 · Microlife Intellectual Property GmbH · FLL · Aug 15, 2003 · General Hospital
Device Facts
Record ID
K031958
Device Name
MICROLIFE INSTANT DIGITAL ELECTRONIC THERMOMETER, MODEL QT1JA1
Applicant
Microlife Intellectual Property GmbH
Product Code
FLL · General Hospital
Decision Date
Aug 15, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2910
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Microlife Instant Digital Thermometer, Model, QT1JA1 is used for the intermittent measurement and monitoring of human body temperature, orally, rectally and under the arm. The Microlife Instant Digital Thermometer, Model, QT1JA1 is a hand-held, non-sterile, reusable clinical thermometer intended for the determination of human temperature in either predictive (Instant) mode (5- seconds orally, 7- seconds rectally and under the arm-predictive temperature), or standard mode (actual determination of temperature). The device is for the adult and pediatric population.
Device Story
Hand-held, reusable digital thermometer; measures human body temperature via oral, rectal, or axillary routes. Operates by detecting thermistor resistance changes; converts resistance to frequency via R-C oscillator circuit; frequency changes correspond to pulse number changes to calculate temperature. Offers predictive (instant) mode (5-7 seconds) and standard mode. Used by clinicians or patients for intermittent monitoring. Output displayed for provider/user to assess patient temperature status; aids clinical decision-making regarding fever management or health monitoring.
Clinical Evidence
Controlled human clinical studies conducted. Evaluated clinical bias, clinical uncertainty, and clinical repeatability per Microlife Clinical Test Protocol.
Technological Characteristics
Thermistor-based sensing; R-C oscillator circuit; hand-held form factor; reusable; non-sterile. Compliance with ASTM E1112, IEC60601-1, and IEC60601-1-2.
Indications for Use
Indicated for intermittent measurement and monitoring of human body temperature in adult and pediatric populations via oral, rectal, or axillary routes.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
{0}------------------------------------------------
## AUG 1 5 2003
### Exhibit #1
### 510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
The assigned 510(k) number is: jn 031.958
#### 1. Submitter's Identification:
Microlife Intellectual Property GmbH, Switzerland Max Schmidheiny-Strasse 201 9435 Heerbrugg / Switzerland
Date Summary Prepared: June 24, 2003
#### 2. Name of the Device:
Microlife Instant Digital Electronic Thermometer, Model QT1JA1
#### 3. Predicate Device Information:
AVITA Instant Thermometer, Model IT101, K#012136, AVITA International Corporation
#### 4. Device Description:
This predictive (instant) digital electronic thermometer enables very fast and reliable measurements. With its predicative technology, this thermometer offers very high clinical accuracy and, has been designed to provide maximum userfriendliness.
The basic principle of this thermometer is that a change of thermistor resistance, caused by changes of temperature, are converted to changes of frequency of R-C oscillator circuit. Therefore, temperature can be given by measuring the frequency of the oscillator.
For a given time period by applying to R-C oscillator circuit, changes of temperature will correspond to changes of pulse number.
{1}------------------------------------------------
#### 5. Intended Use:
The Microlife Instant Digital Thermometer, Model, QT1JA1 is used for the intermittent measurement and monitoring of human body temperature, orally, rectally and under the arm.
The Microlife Instant Digital Thermometer, Model, QT1JA1 is a hand-held, nonsterile, reusable clinical thermometer intended for the determination of human temperature in either predictive (Instant) mode (5- seconds orally, 7- seconds rectally and under the arm-predictive temperature), or standard mode (actual determination of temperature).
The device is for the adult and pediatric population.
#### 6. Comparison to Predicate Devices:
The Microlife Instant Digital Electronic Thermometer, Model QT1JA1 is similar in design and intended use to the AVITA Instant Thermometer, Model IT101, K#012136, AVITA International Corporation, differing mostly in response time, physical dimensions, a wider measurement range, display resolution and wider operating conditions.
### 7. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows:
Compliance to applicable voluntary standards includes ASTM E1112, as well as IEC60601-1 and IEC60601-1-2 requirements.
Guidance documents included the "FDA Guidance on the Content of Premarket Notification 510(K) Submissions for Clinical Electronic Thermometers.
#### 8. Discussion of Clinical Tests Performed:
Controlled human clinical studies were conducted using the Microlife Instant Digital Electronic Thermometer, Model QT1JA1. Clinical data was presented which evaluated clinical bias, clinical uncertainty and clinical repeatability per the Microlife Clinical Test Protocol outline.
#### 9. Conclusions:
The Microlife Instant Digital Electronic Thermometer, Model QT1JA1 has the same intended use and similar technological characteristics as the AVITA Instant Thermometer, Model IT101. Moreover, bench testing contained in this submission supplied demonstrates that any differences in their characteristics do not raise any new questions of safety or effectiveness. Thus, the Microlife Instant
{2}------------------------------------------------
Digital Electronic Thermometer, Model QT1JA1 is substantially equivalent to the predicate device.
:
:
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized caduceus, a symbol often associated with healthcare, with three lines forming the wings of the symbol. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG 15 2003
Microlife Intellectual Property GmbH C/O Ms. Susan D. Goldstein-Falk Official Correspondent mdi Consultants, Incorporated 55 Northern Boulevard, Suite 200 Great Neck, New York 11021
Re: K031958
Trade/Device Name: Microlife Instant Digital Thermometer, Model, QT1JA1 Regulation Number: 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: II Product Code: FLL Dated: June 24, 2003 Received: June 25, 2003
Dear Ms. Goldstein-Falk:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Ms. Goldstein- Falk
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4618. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Susan Runner, DDS, MA
Interim Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Exhibit B
Page
# 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: Microlife Instant Digital Thermometer, Model, QT1JA1
Indications For Use:
The Microlife Instant Digital Thermometer, Model, QT1JA1 is used for the intermittent measurement and monitoring of human body temperature, orally, rectally and under the arm.
The Microlife Instant Digital Thermometer, Model, QT1JA1 is a hand-held, non-sterile, reusable clinical thermometer intended for the determination of human temperature in either predictive (Instant) mode (5- seconds orally, 7seconds rectally and under the arm-predictive temperature), or standard mode (actual determination of temperature).
The device is for the adult and pediatric population.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
OR
Over-The Counter Use \$\frac{i}{}\$
(Optional Format 1-2-96)
Patricia Crescent
n of Anesthesiology, General Hospital, ion Control, Dental Devices
510(k) Number: *K031958*
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.