← Product Code [LPN](/productcode/LPN) · K031957

# MODIFICATION TO KARATS MULTI-PURPOSE SOLUTION (K031957)

_Ciba Vision Corporation · LPN · Aug 14, 2003 · Ophthalmic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K031957

## Device Facts

- **Applicant:** Ciba Vision Corporation
- **Product Code:** [LPN](/productcode/LPN.md)
- **Decision Date:** Aug 14, 2003
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 886.5928
- **Device Class:** Class 2
- **Review Panel:** Ophthalmic

## Indications for Use

Karats Multi-Purpose Solution is indicated for cleaning, rinsing, chemical (not heat) disinfecting, protein removal and storing soft (hydrophilic) lenses (including silicone hydrogel lenses) as recommended by your eye care practitioner.

## Device Story

KARATS Multi-purpose Solution is a sterile aqueous chemical solution for contact lens care. Ingredients include sorbitol, tromethamine, pluronic F127, sodium phosphate dihydrogren, dexpanthnol, edetate disodium dihydrate, and polyhexanide (0.0001) preservative. Used by patients at home for daily maintenance of soft (hydrophilic) and silicone hydrogel contact lenses. Functions via chemical disinfection and protein removal to maintain lens hygiene and biocompatibility. Clinical benefit includes safe lens storage and cleaning, preventing ocular irritation or toxicity. Efficacy verified through microbiological stand-alone testing, XPS surface analysis, and lens compatibility studies.

## Clinical Evidence

Clinical studies previously submitted and reviewed under K021635. Bench testing includes microbiological disinfection efficacy (meeting FDA stand-alone criteria), in vitro protein removal efficacy, and cytotoxicity testing (non-cytotoxic/non-irritant). Silicone hydrogel compatibility verified via XPS surface analysis and physical property measurements, showing no significant changes to lens surface or properties compared to saline control.

## Technological Characteristics

Sterile aqueous solution. Ingredients: sorbitol, tromethamine, pluronic F127, sodium phosphate dihydrogren, dexpanthnol, edetate disodium dihydrate, polyhexanide (0.0001) preservative. Chemical disinfection principle. Compatible with soft (hydrophilic) and silicone hydrogel lenses.

## Regulatory Identification

A soft (hydrophilic) contact lens care product is a device intended for use in the cleaning, rinsing, disinfecting, lubricating/rewetting, or storing of a soft (hydrophilic) contact lens. This includes all solutions and tablets used together with soft (hydrophilic) contact lenses and heat disinfecting units intended to disinfect a soft (hydrophilic) contact lens by means of heat.

## Special Controls

*Classification.* Class II (Special Controls) Guidance Document: “Guidance for Industry Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products.”

## Predicate Devices

- KARATS Multi-purpose Solution ([K021635](/device/K021635.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K031957

# AUG 1 4 2003

#### PART IX. 510(k) SUMMARY

In response to the requirements addressed by the Safe Medical Devices Act (SMDA) of 1990, a summary follows with the safety and effectiveness information upon which the substantial equivalence determination is based.

### 510(k) SUMMARY FOR KARATS Multi-purpose Solution

#### 1. Submitter Information

CIBA Vision Corporation 11460 Johns Creek Parkway Duluth, Georgia 30097 Contact Person: Telephone No.

Steven Dowdley 678-415-3897

#### 2. Device Name

Classification Name: Proprietary Name:

Soft (hydrophilic) Contact Lens Solution KARATS Multi-purpose Solution

#### 3. Predicate Device(s)

KARATS Multi-purpose Solution

#### 4. Description of the Device

Karats Multi-Purpose Solution is a sterile aqueous solution containing sorbitol, tromethamine, pluronic F127, sodium phosphate dihydrogren, dexpanthnol, edetate disodium dihydrate and preserved with polyhexanide 0.0001

#### ട. Indications for Use

Karats Multi-Purpose Solution is indicated for cleaning, chemical (not heat) disinfecting, protein removal and storing soft (hydrophilic) lenses (including silicone hydrogel lenses) as recommended by your eye care practitioner.

#### 6. Description of Safety and Substantial Equivalence

A series of preclinical and clinical studies have been completed on this product and were previously submitted under submission K021635. The non-clinical studies were completed to demonstrate the substantial equivalence of KARATS to other currently marketed solutions. All testing was conducted in accordance with and in conformance to applicable device regulations. Results demonstrate the solution is non-toxic and biocompatible, and is comparable to other currently marketed soft contact lens solutions.

#### Silicone Hydrogel Lens Compatibility Data

A study was conducted to verify that Lotrafilcon A (silicone hydrogel) lenses are compatible with KARATS Multi-purpose Solution. The study showed there was no significant difference between KARATS and the saline control solution, with respect to optical and physical changes in the measured properties of the lenses. KARATS Multi-purpose Solution meets the guidelines set forth in FDA's May 1, 1997 Guidance for Industry, Premarket Notification 510(k) Guidance Document for Contact Lens Care products.

### Surface Analysis by XPS

X-Ray Photoelectron Spectroscopy (XPS) was used to analysis the surface of the lenses to determine if the surface coating was compromised by the solution. Following cycling in the solution, lens samples were processed and analyzed. The kinetic energy of the photoelectrons

{1}------------------------------------------------

was measured and the binding energy of the photoelectron was calculated. The results showed that there were no significant changes to the surface of the Focus NIGHT & DAY lens as analyzed bv XPS testing.

### In Vitro Cleaning Efficacy

Results of the study showed that KARATS is substantially equivalent to currently marketed products in terms of daily protein removal. This data was previously submitted and reviewed in original 510(k) submission - K021635.

### Cytotoxicity

A series of cytotoxicity studies were previously conducted to demonstrate the safety of KARATS. Results of the testing demonstrated that KARATS is non-cytotoxic and is a non-intitant. This data was previously submitted and reviewed in original 510(k) submission - K021635.

### Microbiology

A series studies were previously completed to demonstrate the microbiological efficacy of KARATS. These studies were previously submitted under 510(k) K021635. These studies demonstrate that KARATS meets the stand-alone criteria of the disinfection efficacy test of the FDA May 1, 1997 Guidance for Industry; Premarket Notification 510(k) Guidance Document for Contact Lens Care Products.

### Clinical Testing

A series of clinical studies have been conducted, submitted and reviewed in 510(k) submission K021635. Data from the clinical studies supported the substantial equivalence of KARATS.

### 7. Substantial Equivalence

The data provided in this 510(k) submission concludes that Karats Multi-Purpose Solution is substantially equivalent to Karats Multi-Purpose Solution for cleaning, rinsing, chemical (not heat) disinfecting, protein removal and storing soft (hydrophilic) lenses (including silicone hydrogel lenses) as recommended by your eye care practitioner.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. Inside the circle is an abstract image of an eagle.

Public Health Service

AUG 1 4 2003

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

CIBA Vision Corporation c/o Steven Dowdley, RAC 11460 Johns Creek Pkwy. Duluth, GA 30097

Re: K031957

Trade/Device Name: Karats Multi-Purpose Solution Regulation Number: 21 CFR 886.5928 Regulation Name: Soft (hydrophilic) Contact Lens Care Products Regulatory Class: Class II Product Code: LPN Dated: June 23, 2003 Received: July 21, 2003

Dear Mr. Dowdley:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

## Page 2 - Steven Dowdley, RAC

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4613. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

A. Réti korenthal

A. Ralph Rosenthal, M.D Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

{4}------------------------------------------------

## PART III. INDICATIONS FOR USE STATEMENT

510(k) Number: (Number to be assigned)

Device Name: KARATS Multi-purpose Solution

#### Indications for Use:

Kants Multi-Purpose Solution is indicated for cleaning, rinsing, chemical (not heat) disinfecting, protein removal and storing soft (hydrophilic) lenses (including silicone hydrogel lenses) as recommended by your eye care practitioner.

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use: 

or over-the-counter: 

(Division Sign-Off)
Division of Ophthalmia Fou

Division of Ophthalmic Ear, Nose and Throat Devises

510(k) Number K031957

---

**Source:** [https://fda-staging.innolitics.com/device/K031957](https://fda-staging.innolitics.com/device/K031957)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
